6 Illustrated Parts 1006-0939-000 6-76-7 Exhalation valve Figure 6-3 • Ventilator exhalation valve 1 2 3 4 5 6 AB.23.021 7 Item Description Stock Number 1Exhalation valve cover 1503-3583-000 2Lock ring 1503-3588-000 3Seat exhalation valve 1503-3584-000 4Diaphragm assy 1503-8121-000 5Base, exhalation valve 1503-3585-000 6Blank, pop-off valve 1406-3240-000 7Tube exhalation valve 1406-3566-000
Page 102
Aestiva 6-81006-0939-000 6-8 Figure 6-4 • Breathing circuit modules Item Description CircleBain- Mapleson D 1Assembly (items 2-12) 1406-8102-000 1406-8211-000 2M6-1x23 Thumb Screw 1406-3304-000 1406-3304-000 3Seal lens, flow sensor 1406-3395-000 1406-3395-000 4Module Body -------- 1406-3467-000 5Seal module cover 1406-3317-000 1406-3463-000 6Split Ring 1406-3319-00 1406-3319-000 7Back plate 1406-3391-000 1406-3391-000 8O-Ring Check Valve 1406-3397-000 Not Used 9Valve Seat Check Valve 1406-3396-000 Not...
Page 103
6 Illustrated Parts 1006-0939-000 6-96-9 Bellows Figure 6-5 • Bellows parts Item Description Stock Number 1 Bellows housing 1500-3117-000 2Bellows 1500-3378-000 3Rim 1500-3351-000 4Pressure relief valve assy 1500-3377-000 5Latch, rim 1500-3352-000 6Seal, base 1500-3359-000 7Base assembly, bellows 1406-8106-000 1 2 3 4 5 6 7 AB.23.055
Page 104
Aestiva 6-101006-0939-000 6-10 Test tools and system parts Description Stock Number Vaporizer port o-rings, external (Qty 6) 1102-3016-000 Cylinder gasket 0210-5022-300 Handle for yoke tee 0219-3372-600 Yoke plug 0206-3040-542 DIN O 2 plug (cylinder connection) 1202-7146-000 Positive low-pressure leak test device (BSI) 1001-8975-000 Positive low-pressure leak test device (ISO) 1001-8976-000 Negative low-pressure leak test device 0309-1318-800 Test lung 0219-7210-300 Krytox 1001-3854-000 Touch-up paint,...
Page 105
1006-0939-0007-1 7-1 7 Theory of Operation and Specifications In this sectionVentilator theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2 Pneumatic circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9 Suction regulator (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12 O 2 flowmeter (optional). . . . . . . . . . . . . . . . . . . . . ....
Page 106
Aestiva 7-21006-0939-000 7-2 Ventilator theory GeneralThe ventilator pneumatics are at the rear of the breathing system. A precision valve controls gas flow to the patient. During inspiration, this gas flow closes the exhalation valve and pushes the bellows down. During expiration, a small flow pressurizes the exhalation diaphragm to supply PEEP pressure. Volume and pressure measurements come from flow sensors in the flow sensor module. Two tubes from each sensor connect to a transducer that...
Page 107
7 Theory of Operation and Specifications 1006-0939-0007-37-3 ModesThe system has four modes of mechanical ventilation: •Volume control mode •Pressure control mode •Synchronized intermittent volume (SIMV) mode (optional) •Pressure support ventilation (PSVPro) mode (optional) Use the main menu to set the mode. Figure 7-1 • Ventilation Setup Menu Volume control mode Figure 7-2 • Volume control diagram Volume Control supplies a set tidal volume. The ventilator calculates a flow based on the set tidal...
Page 108
Aestiva 7-41006-0939-000 7-4 Pressure control mode Figure 7-3 • Pressure control diagram Pressure control supplies a constant set pressure during inspiration. The ventilator calculates the inspiratory time from the frequency and I:E ratio settings. A high initial flow pressurizes the circuit to the set inspiratory pressure. The flow then decreases to maintain the set pressure (Pinspired). Pressure sensors in the ventilator measure patient airway pressure. The ventilator automatically adjusts the flow...
Page 109
7 Theory of Operation and Specifications 1006-0939-0007-57-5 At the specified time interval, the ventilator will wait for the next inspiratory effort from the patient. The sensitivity of this effort is adjusted using the flow trigger level. When the ventilator senses the beginning of inspiration it synchronously delivers a volume breath using the set tidal volume, and inspiratory time that is set on the ven- tilator. If the patient fails to make an inspiratory effort during the trigger window time...
Page 110
Aestiva 7-61006-0939-000 7-6 An apnea backup mode is provided in the event the patient stops breathing. When setting the backup mode the user adjusts the inspiratory pressure (Pinsp), respiratory rate and the inspiratory time (Tinsp). As long as the patient triggers the ventilator within the set apnea delay time, the patient will get pressure-supported breaths and the ventilator will not deliver machine breaths. The apnea delay time can be set from 10 to 30 seconds with the default set at 30...