GE Aestiva 5 Service Manual
Have a look at the manual GE Aestiva 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
![](/img/blank.gif)
4 User Maintenance 1006-0939-0004-114-11 Do the steps in the opposite order to assemble the module. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 4 Push in the tabs and remove the top cover. Step 5 Remove the seals. Turn and pull to remove the check valves. AA.96p.043 AA.96p.044 AA.96p.045
![](/img/blank.gif)
Aestiva 4-121006-0939-000 4-12 Bellows maintenance Step 1 Remove the bellows from the breathing system. Step 2 Turn the housing counter-clockwise and lift. Step 3 Remove the bottom edge of the bellows from the rim. AA.96p.023 AA.96p.051 AA.96p.052
![](/img/blank.gif)
4 User Maintenance 1006-0939-0004-134-13 wWW W W AA A A RR R R NN N N III I NN N N GG G G Do not disassemble the pressure relief valve. This can damage the seat or diaphragm and cause injury to the patient. Step 4 Remove the disk from the bellows. Step 5 Remove the ring from inside the top of the bellows. Step 6 Push the latch toward the center and remove the rim. Step 7 Remove the pressure relief assembly. AA.96p.031 AA.96p.032 AA.96p.033 AA.96p.034
![](/img/blank.gif)
Aestiva 4-141006-0939-000 4-14 Do these steps in the opposite order to assemble the bellows. If you see a dust-like powder on the housing or the bellows, apply a thin layer of KRYTOX lubricant to the ribs of the bellows housing. Make sure the lubricant is applied smoothly and there are no lumps. Make sure that: •The arrow on the seal points up. •You hear a double click when you install the rim. •The rim is locked in position. •The inner ring is correctly installed inside the top of the bellows. •Only the bottom ring of the bellows fits over the rim. •The housing is locked in position. You cannot lift it off. Step 8 Push the latch towards the center and remove the locking tabs. Step 9 Remove the seal. AA.96p.035 AA.96p.036
![](/img/blank.gif)
4 User Maintenance 1006-0939-0004-154-15 Bellows tests wWW W W AA A A RR R R NN N N III I NN N N GG G G Objects in the breathing system can stop gas flow to the patient. This can cause injury or death: •Do not use a test plug that is small enough to fall into the breathing system. •Make sure that there are no test plugs or other objects caught in the breathing system. wWW W W AA A A RR R R NN N N III I NN N N GG G G The bellows assembly test does not replace the preoperative tests. Always complete the tests in the section Preoperative Tests before you use the system with a patient. This test makes sure that all components are correctly assembled. It is not an alternative to a complete system checkout. If the bellows operates correctly, install it in the system. If there is a problem, disassemble the bellows. Look for and replace damaged parts. Step 1 Hold the bellows assembly vertical and close the ports (A and B). A B
![](/img/blank.gif)
Aestiva 4-161006-0939-000 4-16 Step 2 Invert the bellows. They must not fall more than 100 mL/min. If it does: •The ports are not tightly sealed. •The bellows is incorrectly installed. •The seal inside the bellows is not correctly installed (with its groove pointed up). •Parts are damaged. Step 3 Remove the plugs from the ports. Permit the bellows to fully extend. Step 4 Close port C. AA.96p.124 C
![](/img/blank.gif)
4 User Maintenance 1006-0939-0004-174-17 If the result for all the bellows tests was “passed,” install it in the system. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual. Step 5 The bellows must not fall more than 100 mL/min. If it does: •The port is not tightly sealed. •The bellows or the pressure relief valve is not correctly installed. •Parts are damaged.
![](/img/blank.gif)
Aestiva 4-181006-0939-000 4-18 O2 sensor calibration - 21% O2 Note: The 21% O2 calibration must be done before the 100% O2 calibration. During O2 calibration the screen replaces O 2 data with - - -. Mechanical ventilation must be off to perform calibration. This procedure takes three minutes or less. Step 1 Push the menu key. Step 2 Turn the knob to select Setup/Calibration (highlight). Step 3 Push the knob to show the next screen. AB.29.013 AB29.002 AB.90.036 AB.90.037 AB.90.091
![](/img/blank.gif)
4 User Maintenance 1006-0939-0004-194-19 The screen shows “Calibrating. . .”, followed by the result (“Complete” or “Failure”). Step 4 Turn, then push the knob to select O 2 Sensor Cal. Step 5 Select 21%. Then, push the knob. Step 6 Complete the steps shown on the screen. • Remove the O 2 sensor from the circuit. •Do not twist or stress the cable. •Make sure the cable is connected. Step 7 Select Start Cal. Then, push the knob. AB29.002 AB.29.012AB.90.048 AB.90.049 AB.48.080
![](/img/blank.gif)
Aestiva 4-201006-0939-000 4-20 If the calibration fails: •perform the calibration again. •do a 100% O 2 sensor calibration. If this passes, repeat the 21% O2 calibration. If the calibration passes, install the O 2 sensor. If necessary, do the 100% O2 calibration. Before you use the system, complete the preoperative test procedure. Refer to "Appendix - Preoperative Tests" section of Part 1 of the Operation Manual.