5 Alarms and Troubleshooting 1006-0939-0005-155-15 Breathing system problems (no alarm) Symptom Problem Solution(s) Gas scavenging flow is too low. Suction supply problem. Use a different suction supply. Filter blockage. Active systems have a flow indicator to show this.Replace the filter. Refer to “Disassemble the scavenging path” in Section2, Cleaning and Sterilization. The bellows fills when the Bag/ Vent switch is set to Bag or the bag fills when the switch is set to Vent.Leak through Bag/Vent switch. Causes include dirt or damage to the o-ring.Refer to “Manifold maintenance” in Section 4, User Maintenance. The ventilator does not read the position of the Bag/Vent switch. Use manual ventilation, if necessary.Ventilator or absorber malfunction.Ventilate manually. Ask a qualified service representative to repair the system. The ventilator does not correctly identify the breathing circuit module.Broken identification tabs. Remove the module and look at the tabs on the rear. If there is damage, use a different module. Dirt on the optical sensors that read the tabs.Use a DAMP alcohol prep to gently clean. Squeeze dry the prep before you use it. Ventilator malfunction. Ask a qualified service representative to repair the system. APL valve does not operate correctly.APL valve problem. Replace APL Valve seal and diaphragm - Refer to User Maintenance. Bellows falls > 100 mL/min during bellows tests.Leak in the breathing system.Check/clean/reposition the pressure relief valve. If the problem persists, replace the pressure relief valve, bellows base, or bellows assembly.
Aestiva 5-161006-0939-000 5-16 Electrical problems (power failure, etc.) ww w w WW W W AA A A RR R R NN N N III I NN N N GG G G If a circuit breaker opens frequently, do not use the system. Have an approved service representative repair the system Symptom Problem Solution Mains indicator is not on. The electrical power cable is not connected.Connect the power cable. The inlet circuit breaker (toggle switch) is open.Close the circuit breaker (Figure 5-1). The power cable is damaged.Replace the power cable. The electrical socket the power cable connects to has no power.Use a different electrical socket. An internal fuse is open. Have a service representative repair the system. The electrical outlets do not have power but the mains indicator is on.The main outlet circuit breaker is open.Close the circuit breaker. One electrical outlet does not have power.The outlet circuit breaker is open.Close the circuit breaker. A circuit breaker opens frequently.Equipment connected to the outlet(s) uses more current than the circuit breaker rating (Figure 5-1).Use a different power supply for some of the equipment. The equipment connected to the outlet has a short.Do not use the equipment until it is repaired. Tec 6 has no power. Not plugged into outlet. Connect power cable.
5 Alarms and Troubleshooting 1006-0939-0005-175-17 1. Outlet CIrcuit Breakers* 2. Master Circuit Breaker (all outlets)* 3. Inlet Circuit Breaker* 4. Ventilator Fuse 5. Mains Indicator *Labels show ratings. Figure 5-1 • Circuit breakers, fuse, and the mains indicator 4 Open (No Power) Closed (OK) 123 Out (No Power)In (OK) AA.96.109 AA.96.108 AA.96.117 5 AB.29.007
Aestiva 5-181006-0939-000 5-18 Pneumatic problems wCC C C AA A A UU U U TT T T III I OO O O NN N N ::: : No repair should ever be attempted by anyone not having experience in the repair of devices of this nature. Symptom Problem Solution High-pressure leak Controls are not set correctly. Set the system switch to Standby and the auxiliary flowmeter to Off. Equipment connected to pneumatic outlet.Disconnect the equipment. High-pressure leak Incorrect cylinder connection (cylinder yokes).Make sure that there is only one cylinder gasket, the gasket is in good condition, and the T handle is tight. Incorrect cylinder connection (DIN connection.)Make sure the nut is tight. Low-pressure leak with a vaporizer on.The vaporizer is not correctly installed. Correctly install the vaporizer. The vaporizer fill spout is loose (fill port type vaporizer).Tighten the fill spout. Vaporizer port o-rings (external) are damaged or not installed.Install new o-rings. A vaporizer malfunction (the leak stops if you use a different vaporizer in the same position).Send the vaporizer to a Datex-Ohmeda Service Center. A port valve malfunction (the leak continues if you use a different vaporizer in the same manifold position). Have an approved service person repair the vaporizer manifold.
1006-0939-000 6-1 6-1 6 Illustrated Parts In this sectionBreathing System Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2 Main manifold . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7 Bellows . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9 Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Aestiva 6-21006-0939-000 6-2 Breathing system parts Top level Item Description (Figure 6-1) Stock Number 1Flow sensor module (cover & holder) 1406-8208-000 a. Cover only 1406-3401-000 b. Holder only 1406-3400-000 2Flow sensor (plastic) 1503-3856-000 Flow sensor (metal - autoclavable) 1503-3244-000 3O 2 sensor, cell 6050-0004-110 O-ring 1406-3466-000 4Breathing circuit module, circle 1406-8102-000 Breathing circuit module, Mapleson/Bain 1406-8211-000 5Bag arm long (adjustable) 1406-3350-000 Bag arm short (not adjustable) 1406-3380-000 6Main manifold assembly 1406-8209-000 7Autoclavable Bellows Assembly 1406-8105-000 8Blank, pop-off valve 1406-3240-000 9Exhalation valve assembly 1503-8114-000 10Rear column cover 1406-3455-000 11Gas scavenging (4 options) -------- Active gas scavenging assembly, 6.35 mm (ANSI, Not European Community)1406-8216-000 Active gas scavenging assembly, 30 mm threaded outlet1406-8214-000 Passive gas scavenging assembly, 30 mm ISO taper1406-8215-000 Passive gas scavenging assembly, 25 mm taper (DEU) 1406-8217-000 12AGSS nylon filter, 225 µ1406-3521-000
6 Illustrated Parts 1006-0939-000 6-36-3 13Upper dish (2 options) ------ Upper dish absorber plastic (seal not included) 1406-3413-000 CO 2 bypass assembly metal (seal not included)1406-3500-000 14Upper seal 1406-3414-000 15Tube corrugated 210 mm 1400-3009-000 16Elbow transfer tube 1406-3576-000 17Transfer tube 1406-3575-000 18Drain dish assembly 1406-8218-000 19Plug, sight glass-drain 0229-2080-100 O-ring, plug 0210-0594-300 20 Canister 0229-3015-800 ---Not Shown Upper seal, upper dish, corrugated tube, drain dish, canisters 1406-8207-000 Barb cap, passive scavenging 1406-3542-000 Item Description (Figure 6-1) Stock Number
Aestiva 6-41006-0939-000 6-4 Figure 6-1 • Breathing system top assemblies 1b 2 1a6 5 3 47 9 108 11 12* 13 14 15 16 17 18 20 19 * Active scavenging systems only AB.23.095
6 Illustrated Parts 1006-0939-000 6-56-5 Main manifold Figure 6-2 • Main manifold 5 Top Bottom 1 2 2 2 3 2b 2a 4 6 7 89 10 2c11 AB.23.096 AB.23.071
Aestiva 6-61006-0939-000 6-6 Item Description (Figure 6-2) Stock Number 1Seal, main manifold, APL/AGSS 1406-3316-000 2Seal, bag/vent assy (a, b, c) a. O-ring (Qty 2) 1406-3278-000 b. Poppet (Qty 2) 1406-3279-000 c. Seal only 1406-3314-000 3Plate, top, main manifold 1406-3300-000 4Plate, bottom main manifold 1406-3303-000 5Seal, Main Manifold, Absorber 1406-3315-000 6Body manifold, main 1406-3301-000 7APL diaphragm assembly 1406-3331-000 8Manifold APL/AGSS 1406-3302-000 9Split ring (Qty 3) 1406-3319-000 10Thumbscrew, M6x43 mm w/10 mm Head 1406-3306-000 11M6 x28.5 mm Thumbscrew (Qty 2) 1406-3305-000