    							2 Cleaning and Sterilization
    1006-0939-0002-7
    2-7
    Disassemble the patient path
    Step 1
    Open the access panel.
    Step 2
    Push up on the latch located under the 
    flow sensor module. 
    Step 3
    Remove the flow sensor module. You 
    will feel some resistance. Continue to 
    pull.
    Step 4
    Disconnect the cable and remove the 
    O
    2 sensor.
    AA.96p.009
    AA.96p.010
    AA.96p.011
    AA.96p013, 014 
    						
    							Aestiva
    2-81006-0939-000
    2-8
    Step 5
    Completely loosen the thumbscrew on 
    the breathing circuit module. 
    Step 6
    Remove the module. You will feel some 
    resistance. Continue to pull. 
    Step 7
    Push the metal button to the first stop. 
    Pull out the bag arm.
    Step 8
    Push the metal button to the 
    second stop and open the top panel.
    AA.96p016
    AA.96p017
    AA.96p018
    AA.96p020 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-9
    2-9
    Autoclave assemblies marked 134°C. Refer to the Section “Clean and 
    sterilize” for complete instructions. To assemble the circuit refer to the Setup 
    section.
    Step 9
    Push the release button and remove 
    the bellows assembly (pull up).
    Step 10
    Remove the main manifold.
    AA.96p021, 23
    AA.96p024 
    						
    							Aestiva
    2-101006-0939-000
    2-10
    Canister disassembly
    wCC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    To prevent damage, pull the release handle forward. Then turn the handle 
    Step 1
    Pull the release forward.
    Step 2
    Turn the release clockwise.
    Step 3
    Pull down and out on the canisters.
    wZ
    AA.96p026
    AA.96p028
    AA.96p029 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-11
    2-11
    Continue with the next section or go to "Clean and sterilize". To assemble, 
    refer to the Setup section.
    Step 4
    Push in the buttons on each side of the 
    top dish.
    Step 5
    Lift up the drain dish.
    Remove the top dish and drain dish.
    AA.96p053
    AA.96p054 
    						
    							Aestiva
    2-121006-0939-000
    2-12
    Disassemble the scavenging path
    These parts send exhaled gas to the disposal system, not the patient. 
    Exhalation valve
    Complete the basic disassembly 
    procedure.
    Push in the latches and pull out the 
    exhalation valve.
    Scavenging reservoir
    Loosen the knob and remove the rear 
    cover.
    Pull the reservoir up and to the rear.
    Twist and pull down to remove the tube.
    AB.23.089
    AB.23.090 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-13
    2-13
    How to clean and disinfect the flow sensors
    ww w w
    CC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    Do not autoclave plastic flow sensors. 
    ww w w
    Do not use high-pressure gas, or brushes to clean the flow sensors.
    ww w w
    Do not use cleaning solvents that are not approved for use with 
    Polycarbonates (e.g. CIDEX Plus).
     
    CIDEX sterilizationBoth Datex-Ohmeda and the manufacturer of CIDEX (Johnson & Johnson) 
    have tested this procedure.
    •CIDEX must be 14 day mixture, with activator vial REF REORDER # 2245
    •One liter of this solution cleans four (4) flow sensors
    Procedure (Figure 2-3)1. Remove the flow sensor module from the absorber. Refer to “Disassemble 
    the patient path” in this section.
    2. Remove the flow sensors from the module.
    •Push in the latch.
    •Pull off the cover.
    •Remove the flow sensors.
    3. Submerge the flow sensor and tubes in activated CIDEX solution. Keep the 
    connector dry.
    4. Keep the solution in the tubes for the sterilization period.
    5. Submerge the flow sensor and tubes in distilled water. Again, do not get the 
    connector wet.
    6. Rinse as indicated in CIDEX instructions. 
    7. Do steps 5 and 6 again to remove all CIDEX. 
    8. COMPLETELY dry the flow sensor and the tubes before you use the sensor. 
    Use a dry syringe, or connect vacuum or pressure to remove all liquid from the 
    sensor (sensor, tubes, and connector):
    •Minimum time: 1 min
    •Maximum vacuum: 30 in Hg
    •Maximum flow: 10 L/min flow
    •Maximum pressure: 345 kPa (50 psi). 
    						
    							Aestiva
    2-141006-0939-000
    2-14
    Figure 2-3 • Steps 2-8
    Step 2
    Steps 3 and 4
    Steps 5, 6 and 7
    Step 8 
    Dry for > 1 min with these precautions:
    • Maximum vacuum 30 in Hg
    • Maximum flow 10 L/min
    • Maximum pressure 345 kPa (50 psi) 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-15
    2-15
    HH H H
    oo o o
    ww w w
        
    ttt t
    oo o o
        
    cc c c
    llll
    ee e e
    aa a a
    nn n n
        
    aa a a
    nn n n
    dd d d
        
    ss s s
    ttt t
    ee e e
    rrr r
    iiii
    llll
    iiii
    zz z z
    ee e e
        
    ttt t
    hh h h
    ee e e
        
    oo o o
    pp p p
    ttt t
    iiii
    oo o o
    nn n n
    aa a a
    llll
        
    CC C C
    OO O O22 2 2    
    bb b b
    yy y y
    pp p p
    aa a a
    ss s s
    ss s s
        
    aa a a
    ss s s
    ss s s
    ee e e
    mm m m
    bb b b
    llll
    yy y y
    These instructions assume that the system has the optional CO2 bypass 
    feature installed.
    SS S S
    ttt t
    ee e e
    pp p p
        
    11 1 1
    Remove the absorber canisters. (Also described in “Canister disassembly” in this section.)
    a. Pull the canister release handle forward.
    wCC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    To prevent damage, pull the canister release handle forward before turning the 
    handle.
    b. Turn the release handle clockwise.
    c. Lift out the canisters.
    Note: The canisters should drop under their own weight when released. If they do not, clean and lubricate the pins in 
    the drain dish and sockets in the base.
    SS S S
    ttt t
    ee e e
    pp p p
        
    22 2 2
    Release the CO2 bypass assembly by 
    pushing both buttons (one on each side 
    of the assembly).
    SS S S
    ttt t
    ee e e
    pp p p
        
    33 3 3
    Pull down on the CO2 bypass assembly 
    (a) to remove it and lift the drain dish 
    (b) up.
    a
    b 
    						
    							Aestiva
    2-161006-0939-000
    2-16
    SS S S
    ttt t
    ee e e
    pp p p
        
    44 4 4
    Clean and sterilize the bypass 
    assembly:
    a. Immerse the assembly into a 
    solution of mild detergent and 
    water.
    b. Agitate the assembly in the 
    solution while repeatedly 
    actuating and releasing the 
    plate.
    c. Rinse the assembly by 
    immersing it in clean water and 
    agitating it again with the plate 
    actuated and released.
    d. Autoclave the assembly at 
    134°C.
    e. Cool about 40 minutes at room 
    temperature.
    Note: Optionally, the bypass assembly can be divided into two smaller sections for easier cleaning by releasing the 
    top clamp (a) and pulling the hose from the assembly. When reassembling, reattach the hose to the bottom part of the 
    assembly. Secure the hose with the clamp and pinch the clamp one notch past finger-tight with pliers.
    SS S S
    ttt t
    ee e e
    pp p p
        
    55 5 5
    Reinstall the bypass assembly:
    a. Reinstall the drain dish.
    b. Insert the bypass assembly.
    c. Push back and adjust to locate 
    the access holes for the top 
    ports.
    d. Push the buttons on each side of 
    the assembly and push the 
    assembly up; both buttons 
    should snap into place.
    e. Pull down on the assembly to 
    make sure it is locked into place.
    a
    ab
    d
    c 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Aestiva 5 Service Manual