Home > GE > Medical Equipment > GE Aestiva 5 Service Manual

GE Aestiva 5 Service Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Aestiva 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							1 Introduction
    1006-0939-0001-5
    1-5
    134°CAutoclavable
    RBag position/ manual ventilationrMechanical ventilation
    uOpen drain (remove liquid)UClose drain
    qInspiratory flowQExpiratory flow
    tO2 sensor connection End case
    The primary regulator is set to pressure 
    less than 345 kPa.The primary regulator is set to 
    pressure less than 414 kPa.
    European Union Representative 
    						
    							Aestiva 
    1-61006-0939-000
    Maintenance summary and schedule
    These schedules show the minimum frequency. You will have to service the 
    equipment more frequently if you use it:
    •In unusual conditions (dirty gas supplies, high temperature, high humidity, 
    etc.).
    •More frequently than normal.
    Operator maintenanceExamine all components and do the maintenance procedures more frequently 
    if necessary.
    Minimum 
    Frequency
    Maintenance
    Daily • Clean the external surfaces.
    •21% O
    2 calibration (circuit O2 sensor).
    •Flow sensor calibration
    Two weeks • Drain the vaporizers and discard the agent. This is not 
    necessary for Tec 6 vaporizers.
    Monthly • 100% O
    2 calibration (cIrcuit O2 sensor).
    •Put Krytox (or a lubricant approved for use with 100% 
    O
    2) on all tee handle threads.
    During cleaning 
    and setup•Inspect the parts for damage. Replace or repair as 
    necessary
    Annually • Replace the external o-rings on the vaporizer ports.
    As necessary • Install new cylinder gaskets on cylinder yokes.
    •Replace the absorbent in the absorber.
    •Drain the breathing system.
    •Drain the breathing circuit module.
    1
    •Drain the overflow trap on the optional suction 
    regulator.
    •Replace the circuit O
    2 sensor.
    •Replace the disposable flow sensors (plastic) 
    2
    •Replace the autoclavable flow sensors (metal)3.
    •Replace the receiver filter (active gas scavenging only).
    1. This is included in the preoperative test procedure.
    2. Under typical use the sensor meets specifications for 3 months 
    3. Under typical use the sensor meets specifications for 1 year. 
    						
    							1 Introduction
    1006-0939-0001-7
    1-7
    Datex-Ohmeda approved
    service
    Minimum
    Frequency Maintenance
    6 months Have an approved service person do the service tests 
    and scheduled service maintenance. 
    						
    							1006-0939-0002-1
    2-1
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    Obey applicable safety precautions:
    •During the cleaning of the liquid collecting tray, avoid skin or eye contact with 
    the contents of the absorber. In the event of skin or eye contact, immediately 
    rinse the affected area with water and seek medical assistance.
    •Read the material data sheet for each cleaning agent.
    •Read the operation and maintenance manual for all sterilization equipment.
    •Wear gloves and safety glasses. A damaged O
    2 sensor can leak and cause 
    burns (contains potassium hydroxide).
    •Do not breathe the fumes.
    wCC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    To prevent damage:
    • Refer to the manufacturer’s data if you have questions about a cleaning agent.
    •Do not use organic, halogenated, or petroleum based solvents, anesthetic 
    agents, glass cleaners, acetone, or other harsh cleaning agents.
    •Do not use abrasive cleaning agents (such as steel wool, silver polish or 
    cleanser).
    •Keep all electronic parts away from liquids.
    •Do not permit liquid to go into the equipment housings.
    •Do not soak synthetic rubber parts for more than 15 minutes. Swelling or faster 
    aging can occur.
    •Only autoclave parts that are marked 134°C.
    In this sectionSummary   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
    Clean and sterilize   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
    Disassemble the patient path  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2-7
    Canister disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2-10
    Disassemble the scavenging path   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2-12
    How to clean and disinfect the flow sensors   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2-13
    How to clean and sterilize the optional CO
    2 bypass assembly  . . . . . . . . . . . . . . . . .  2-15
    2 Cleaning and Sterilization 
    						
    							Aestiva
    2-21006-0939-000
    2-2
    Summary
    Patient pathThe parts in Figure 2-1 send exhaled gas back to the patient. They may require 
    more frequent cleaning/sterilization than parts in Figure 2-2. Refer to your 
    hospital’s infection control policy.
    Figure 2-1 • These parts return exhaled gas to the patient
    Autoclave (134°C) or
    If plastic, refer to cleaning/disinfection
    Wipe with a damp cloth. wash (mild detergent pH 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-3
    2-3
    Scavenging path
    Figure 2-2 • These parts do not send gas back to the patient
    Autoclave (134°C) or
     
    wash (mild detergent pH 
    						
    							Aestiva
    2-41006-0939-000
    2-4
    Clean and sterilize
    The Disassembly part of this section tells you how to remove parts for 
    cleaning.
    To wash (by hand or
    machine)
    Use a mild detergent (pH 
    						
    							2 Cleaning and Sterilization
    1006-0939-0002-5
    2-5
     Autoclave
    Autoclave at 134°C. Inspect the parts for deterioration. The User Maintenance 
    section tells you how to do this. 
    Patient path
    Scavenging path
    Upside down
    AB.23.087
    AB.23.086 AB.23.083
    Metal Only
      
    						
    							Aestiva
    2-61006-0939-000
    2-6
    Special requirements
    •To clean the circuit O2 sensor, wipe it with a damp cloth. Do not put the sensor 
    in liquid.
    •To clean/disinfect metal/or plastic flow sensors, use the flow sensor 
    cleaning procedure. Do not get the connectors wet.
    •Disassemble the bellows before you wash it. If not, it will take a very long time 
    to dry. Hang the bellows upside down to dry.
    •Assemble the bellows before you autoclave. Autoclave the bellows upside 
    down.
    wWW W W
    AA A A
    RR R R
    NN N N
    III I
    NN N N
    GG G G
    Do not use talc, zinc stearate, calcium carbonate, corn starch or 
    equivalent materials to prevent tackiness. These materials can go 
    into the patient’s lungs and airways and cause irritation or injury.
    wCC C C
    AA A A
    UU U U
    TT T T
    III I
    OO O O
    NN N N
    Do not put the circuit O2 sensor or flow sensor connector in liquid. 
    wDo not autoclave the Circuit O2 sensor or the plastic flow sensors. 
    wDo not clean the interior surfaces of the flow sensors. Use a damp 
    cloth
     on external surfaces only. 
    AssembleThe Setup section tells you how to assemble the breathing system. 
    Inspect all parts for deterioration. Replace them if necessary. 
    The Preoperative tests in Part 1 (the first half of this manual), tell you how to 
    test the system for correct operation.
    AB.23.088
    * Hang the bellows upside down
     (extended) to dry. If not, the
    convolutions can stick together.*  
    						
    All GE manuals Comments (0)

    Related Manuals for GE Aestiva 5 Service Manual