GE Aestiva 5 Service Manual
Have a look at the manual GE Aestiva 5 Service Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
7 Theory of Operation and Specifications 1006-0939-0007-177-17 Pneumatic specifications Gas suppliesPipeline gases: O2, Air, N2O Cylinder gases: CO 2, Heliox, O2, N2O, Air (maximum: 2 cylinders of each gas; 4 cylinders total on narrow models; 5 cylinders total on the wide model. Cylinder connections: Pin indexed (all gases); nut and gland DIN 477 (O 2, N2O, Air); large cylinder kit available for O 2 and N2O. Primary regulator output pressure: • With a rear panel label: The primary regulator is set to pressure less than 345 kPa. The primary regulator is set to pressure less than 414 kPa. •Without a rear panel label: approximately 587- 690 kPa Pressure relief valve: approximately 883 kPa. Pipeline connections (filtered): DISS-Male; DISS-Female; DIN 13252; AS 4059 (Australian); F90-116 (French Air Liquide); CEN (QPREN 737-6); or NIST (ISO 5359). All fittings available for O 2, Air, and N 2O. Pressure displays: Color coded gauges Pneumatic outlet: O 2 DISS or Mini Schrader connections Common gas outlet reliefValve limits fresh gas to 138 kPa at the flush flow. < 345 kPa < 414 kPa 034.5 69.0 103.5 138 172.5 207 241.5 276 Outlet Pressure (kPa) 130 120 110 100 90 80 70 60 50 40 30 20 10 Flow (L/min)DISS Power Outlet ISO Power Outlet Inlet pressure constant at 241.5 kPa
Aestiva 7-181006-0939-000 7-18 Flow specifications Flow rates Minimum O 2 flow: 25 to 75 mL/min Accuracy At 20°C with gas supply pressures at 345 kPa and an outlet pressure of 101.3 kPa (absolute) flowmeter accuracy agrees with VDE 3513 Part 3, Accuracy Class 2.5 or better. Different breathing circuit pressures, barometric pressures or temperatures change the accuracy. With some conditions, these changes can be larger than the tolerances. Flush flow 35-50 L/min O 2 supply failure alarm and shutoffGas Flow Range Scale (One Flow Tube)Scale (Two Flow Tubes) O 20.05-15 L/min ------------------ 0.05 -0.95 L/min 1-15 L/min N 2O 0-10 L/min1------------------ 0.05 -0.95 L/min 1-10 L/min Air 0-15 L/min 1-15 L/min 0.05 -0.95 L/min 1-15 L/min He 0-15 L/min 0.5-15 L/min ------------------ CO 20-0.5 L/min 0.05-0.5 L/min ------------------ 1.The link system sets the nominal O2 flow at 25% of the total O2 and N2O flow. O2 Pressure O 2 supply failure alarm193 to 221 kPa N 2O shutoff3.5 kPa CO 2 or Heliox shutoff69 kPa
7 Theory of Operation and Specifications 1006-0939-0007-197-19 Electrical power Supply voltage 100-120 or 220-240 Vac ± 10% at 50 or 60 Hz Electrical circuit breakers Inlet circuit breakers: Master circuit breaker-all outlets: Outlet circuit breakers: Circuit Breaker for Tec 6 Outlet (limited by master circuit breaker) System leakage current limit - do not exceed UL and CSA rated systems (USA and Canada)
Aestiva 7-201006-0939-000 7-20 Power cord Electrical Fuses T6.3L/250V Battery InformationA sealed lead acid battery supplies battery backup for the Aestiva. The Aestiva is not a portable unit. Batteries are used as back up power in case of a power failure. Thus the battery is in a float charge state most of the time. Batteries meet the following: 1. Capacity to operate for 30 minutes. 2. Unit functions to specifications through the transition to battery power. 3. Long float charge life. 4. Battery pack is internally fused - in line replaceable. 5. Battery terminals and connecting wires are protected against short circuits. Only Datex-Ohmeda service representatives are to replace the battery. Batteries must be disposed of in accordance with applicable regulatory requirements in effect at the time and place of disposal. Electro-magnetic Compatibility Environment Suitable for use in the EM environment described in EN 60601-1-2 Immunity Levels The Aestiva complies with the requirements of EN 60601-1-2 (Electromagnetic Compatibility - Requirements and tests). The following basic EMC standards were applied to verify conformance. Emissions CISPR 11 Group 1 (EN 55011) Immunity IEC 801-2, 8 kV air, 3 kV contact IEC 801-3, 3 V/m IEC 801-4, 2 kV power line IEC 801-5, 2 kV line to earth, 1 kV line to lineLength 5 meters Voltage rating 125 to 264Vac at 10A, by country and/or translation Current Carrying Capacity 10A Type Three conductor medical grade power supply cord by country and/or translation
7 Theory of Operation and Specifications 1006-0939-0007-217-21 Physical specifications All specifications are approximate values and can change without notice. Weight2 vaporizer system: 136 kg 3 vaporizer system: 154 kg Dimensions2 vaporizer system Height: 135.8 cm Width: 75.0 cm Depth: 83.0 cm Shelves: Number: 0, 1, or 2 Size: 47.5 x 35 cm or 67.5 x 35 cm Usable height (2 shelves) 1: 26 or 36 cm Weight limit: 23 kg per shelf 3 vaporizer system Height: 135.8 cm Width: 93.0 cm Depth: 83.0 cm Shelves: Number: 0, 1, or 2 Size bottom shelf: 87.5 x 35 cm or 67.5 x 35 cm Size top shelf: 47.5 x 35 cm or 67.5 x 35 cm Usable height (2 shelves) 2: 26 or 36 cm Weight Limit: 23 kg per shelf Casters12.5 cm with brakes on the front casters DrawersOptional: 14.5 x 38.5 x 26.0 cm Standard drawer: 10.5 x 38.5 x 26 cm Ventilator display7.6 x 15.2 cm 1.Flat or 10° angle shelves have the same usable height.2.Flat or 10° angle shelves have the same usable height.
Aestiva 7-221006-0939-000 7-22 Environmental requirements Temperature Humidity Altitude Operation 10 to 40 °C, (Oxygen cell operates to specifications at 10 to 40 °C) Storage -20 to 70 °C Oxygen cell storage is -5 to 50°C, 10 to 95% Rh, 500 to 800 mmHg Operation 15 to 95% Rh, non-condensing Storage 10 to 100% Rh, include condensing Operation 500 to 800 mmHg (3565 to -440 meters) Storage 375 to 800 mmHg (5860 to -440 meters) Compensation range 525 to 795 mmHg (3,000 to -100 meters)
7 Theory of Operation and Specifications 1006-0939-0007-237-23 Ventilation operating specifications Pneumatics Fresh gas compensation Pressure Volume Gas Source Anesthesia System Gas Composition Medical Air or O 2 Nominal Supply Pressure 350 kPa Pressure Range at Inlet 240 to 700 kPa Peak Gas Flow 120 L/m at 240 kPa, 0.75 seconds Continuous Gas Flow 80 L/m at 240 kPa Flow valve range 1 to 120 L/min at 240kPa. Flow Compensation Range 200 mL/min to 15 L/min Gas Composition O 2, N2O, N2 Air, Heliox, CO2 Anesthetic Agents Patient airway pressure range -20 to +120 cmH2O +/-2 cmH2O High pressure alarm set range 12 to 100 cmH 2O, 1 cm increment Sustained pressure alarm range6 to 30 cmH 2O, 1 cm increment Display range -20 to 120 cmH 2O Tidal volume display range 0 to 9999 mL, 1 mL resolution Setting range 20 to 1500 mL Minute volume 0.0 to 99.9 liters, 0.1 liter resolution Breath rate 4 to 100 bpm (breaths per minute), 1 bpm resolution Volume sensor type Variable flow orifice
Aestiva 7-241006-0939-000 7-24 Oxygen Display range 8 to 100% O2 Display resolution 1% increments Sensor type Galvanic fuel cell Measurement range 0 to 100% O 2 Measurement accuracy Better than ± 3% of full scale Cell response time 35 seconds 1 Low O 2 alarm range 18% to 99% High O 2 alarm setting 21% to 99% or Off Note: Low O 2 limit may not be set above the high O 2 limit, nor may the high O2 limit be set below the low O 2 limit. Expected cell life Four months of shelf life (23 °C room air) and one year of normal operation. 1.Response time of cell and adapter as measured using the test method described in ISO 7767 (1988-12-15), clause 50.9
7 Theory of Operation and Specifications 1006-0939-0007-257-25 Aestiva ventilator accuracy data The following accuracy data are based on patient conditions and settings described in ASTM F1101. The ventilator is assumed to be operating in volume mode (Heliox Off). For the following to be true, the ventilator is operating with 100 percent oxygen in the breathing system; or, it is connected to an anesthesia gas analyzer. If the ventilator is operating without being connected to an anesthesia gas analyzer, additional errors are described in the gas composition chart, Figure 8-9. NN N N oo o o ttt t ee e e ::: : Gas composition errors may be in addition to the above normalized accuracy. When adding errors, positive errors can have the effect of nulling out negative errors. NN N N oo o o ttt t ee e e ::: : Use of anesthetic agent could affect the errors by approximately -0.95%/% volume agent in normal mode and roughly -2.5%/% volume agent in Heliox mode. Volume delivery accuracy > 210 mL tidal volume - accuracy better than 7% < 210 mL tidal volume - accuracy better than 15 mL < 60 mL tidal volume - accuracy better than 10 mL Volume monitoring accuracy > 210 mL tidal volume - accuracy better than 9% < 210 mL tidal volume - accuracy better than 18 mL < 60 mL tidal volume - accuracy better than 10 mL Inspiratory pressure delivery accuracygreater of ±10% or ±3 cmH 2O PEEP delivery accuracy±1.5 cmH 2O Pressure monitoring accuracygreater of ±5% or ±2 cmH 2O
Aestiva 7-261006-0939-000 7-26 Figure 7-9 • Gas Composition Related Errors (both modes) Heliox modeThe effect on volume delivery and monitoring accuracies of Heliox mixtures is within +8% to -15% when operating the Aestiva Ventilator in the Heliox mode. Open loop mode volume delivery accuracy: ±10% of the set value under the following conditions: 1. Accuracy evaluated at the ventilator outlet. 2. This is not a normal operating mode. The ventilator is operating in an irregular condition.