Page 41
Changing a programme Provided the Start button has not yet been pressed,a wrongly selected programme may be changed as de - scribed below (if it has been pressed, see Cancelling a programme): ^Turn the programme selector to the required programme. ^Select additional function DRYING 0again if required (see Selecting an additional function). ^Press the Start button6. Selecting an additional function The following additional function can be selected: ^0DRYING may be selected as an additional function as...
Page 42
To cancel a programme Once a programme is running it should only be cancelled in extreme cases, e.g. when articles in the machine rattle or jostle against each other and have to be rearranged. See notes in next sec - tion regarding the control of epidemics. ^Turn the programme selector to the fposition (the programme is can - celled after approx. 2 secs.). ^Open the doora Caution. Water and items in the ma - chine may be hot. Danger of burning or scalding. ^Arrange the load securely, (wear protective...
Page 43
For future servicing work, please document any alterations to the stan - dard factory settings. Enter any alterations in the field Value set:_________. General notes: –Switch positions that have not been allocated a function are indicated in the middle part of the display by a bar (-). –The programming level is called up via the STOP positionf. – If any changes have been made to programmes or dispensing systems these must be documented in a log book kept with the machine in accor- dance with the Medical...
Page 44
1.1.2 Venting the dispensing system: a) After setting the dosage concentra - tion: ^Turn the programme selector to theA position. do1 appears in the display. ^Pressß. The dispensing system will be vented automatically. 0 will appear in the dis - play when venting is finished. Having changed a value or all values: ^Press the6button. SP appears. ^Press6again. The change has been recorded. b) If only venting is required (e.g. be- cause the dispensing system has been allowed to empty by mistake) ^Select...
Page 45
1.2 DOS 3 (Neutralising agent) 1.2.1 Setting the dosage concentra - tion: Set the dosage concentration accord - ing to the neutralising agent manufac - turer’s recommendations. Example: If the manufacturer recom - mends a concentration level of 0.3 % set the value to 0.30. ^Turn the programme selector to the CHEM 60°C-5’position. The dosage concentration set will appear as a % in the display: 0.05 - 0.50 = 0.05 % - 0.5 % dosage concentration. ^Pressßas many times as neces- sary or hold it pressed in...
Page 46
1.3 DOS 4 (chemical disinfecting agent) via the optional DOS-Module G 60 or C 60 . 1.3.1 Setting the dosage concentra - tion: ^Set the dosage selector on the DOS-Module to position 10 (applies to DOS Module C 60 only). Set the concentration level according to the disinfecting agent manufacturer’s recommendations. Example: If the manufacturer recom - mends a concentration level of 1 % set the value to 1.00. ^Turn the programme selector to the varioTD ANposition. The dosage concentration set will appear as...
Page 47
2. Altering the wash tempera - ture and/or the temperature holding time in the "Main wash" (except with the SPECIAL (AN) 93°C10’ programme) and "Final rinse" programme stages Please make a note of any changes to the temperature and / or the tem - perature holding time in the appropri - ate section of the programme chart . The programme chart shows the factory settings for the Main wash temperature together with temperature holding time and the Final rinse temperature with its temperature...
Page 48
^Pressßor hold pressed in, until the required value appears in the display. Final rinse temperature: ^Turn the programme selector to the STOP positionf. ^Press0as often as necessary or hold it pressed in until E25 (pro - gramming level 25 = Final rinse 2) appears in the display. Then: ^Turn the programme selector back to the programme which needs to be changed. The actual value appears in the display field after°C. ^Pressßor hold pressed in, until the required value appears in the display. Note: When...
Page 49
3. Audible tone (Buzzer) This special function allows you to programme a buzzer (lasting up to a maximum of 30 seconds) for the follow - ing functions: –Programme end = Constant buzzer –Fault warning = Buzzer at one sec - ond intervals –Removing a test item from the final rinse water (this function has to be programmed by a Miele service en - gineer) = 3 x short beeps followed by 2 sec - ond pause, then 3 short beeps etc. ^Turn the programme selector to the STOP positionf. ^Switch off the machine. ^Press...
Page 50
4. Re-setting all parameter changes back to the factory settings. ^Turn the programme selector to the STOP positionf. ^Switch off the machine. ^Press and holdßand6at the same time, and whilst doing so, switch the machine on with theI-0 button. The current programme status P... will appear in the display and thep/ 6indicator light will come on. ^Press0as often as necessary or hold it pressed in until E33 (pro- gramming level 33) appears in the display. ^Turn the programme selector to any switch position...