Hitachi F7000 Instruction Manual
Here you can view all the pages of manual Hitachi F7000 Instruction Manual. The Hitachi manuals for Fluorescence Spectrophotometer are available online for free. You can easily download all the documents as PDF.
Page 201
6.1 6 - 1 6. 3-D TIME SCAN 6.1 Flow of Operation This section explains the parameter setting and operating method for 3- D time scan. Measurement proceeds along the following flow. Fig. 6-1 Operational Flow (3-D time scan) Click button and set your analytical conditions. • Scan range • Data mode • Scan speed, etc. Click button. • Sample name • Auto File • File name, etc. Set a sample. Click...
Page 202
6.2 Creating an Analysis Method 6 - 2 6.2 Creating an Analysis Method Set the analytical conditions (method) prior to measurement. (1) Click the button on the Measurement toolbar. The window in Fig. 6-2 appears. (Or select the Method command from the Edit menu.) Fig. 6-2 Analysis Method (2) On the Analysis Method window, analytical conditions for the 3-D time scan mode can be set. The following tabs are provided on this window. These tabs are...
Page 203
6.2.1 6 - 3 The parameters shown in Fig. 6-3 are provided at the General tab. For the settable range of each parameter, refer to Appendix G. Fig. 6-3 General Tab 1 Measurement (mode) A measurement mode is selectable. Here, select “3-D Time scan.” For Wavelength scan, Time scan, Photometry and 3-D Scan, refer to Sections 2, 3, 4 and 5, respectively. 2 Operator Enter an operator name. 3 Instrument The name of the connected instrument is indicated. 4 Sampling -...
Page 204
6.2 Creating an Analysis Method 6 - 4 6 Load Click this button to read out saved analytical conditions. Upon clicking, the file opening window appears. So, select an analytical condition file. 7 Save When you click this button after setting the parameters, the analytical conditions can be saved by overwriting. In case of a new analysis method, clicking this button displays the “Save As” window. You must enter a file name for saving in this window. 8 Save As Click this...
Page 205
6.2.2 6 - 5 The window shown below is based on that 3-D Time scan is specified for Measurement at the General tab. When you click the Instrument tab, the window in Fig. 6-4 appears. For the settable range of each parameter, refer to Appendix G. Fig. 6-4 Instrument Tab 1 Scan mode Select a scan mode here. In case of Synchronous, the abscissa of the graph stands for excitation wavelength. The unit indication of the abscissa becomes ‘EX/EM (nm).’ 2 Data mode Select...
Page 206
6.2 Creating an Analysis Method 6 - 6 4 EM WL For the Excitation scan mode, enter the fixed wavelength on the emission side . For the Synchronous scan mode, enter the start wavelength on the emission side. 5 EX Start WL This is the start wavelength on the excitation side in wavelength scan. Enter the shorter wavelength side of measuring wavelength range. 6 EX End WL This is the end wavelength on the excitation side in wavelength scan. Enter the longer wavelength side of...
Page 207
6.2.2 6 - 7 15 Response When “Auto” is selected, response will be set automatically depending on scan speed. Select “Auto” usually. 16 Corrected spectra A function for determining the spectrum inherent to a sample by correcting the photometer wavelength characteristic using the saved instrument parameters, following measurement with the instrument parameters for photometer control. If there are no instrument parameters for the wavelength range to be measured, only the wavelength...
Page 208
6.2 Creating an Analysis Method 6 - 8 When selecting the Monitor tab, the window in Fig. 6-5 appears. For the settable range of each parameter, refer to Appendix G. Fig. 6-5 Monitor Tab Assign the following parameters. 1 Y-Axis Max (maximum) Enter a maximum value for Y-axis on the monitor window. 2 Y-Axis Min (minimum) Enter a minimum value for Y-axis on the monitor window. 3 Contour interval Function for setting intervals of drawing contour lines. 4...
Page 209
6.2.3 6 - 9 NOTE: When the remaining memory capacity of PC in operation is very small (system resource is less than 20%), the data processing window will not open even if this item is turned ON. Either close the presently opened data processing window or restart the FL Solutions program. 5 Print report after data acquisition Select whether or not to print a report after sample measurement. When selected (check mark is applied), printing will be made automatically after measurement....
Page 210
6.2 Creating an Analysis Method 6 - 10 On selection of the Processing tab, the window in Fig. 6-6 will be displayed. For the settable range of each parameter, refer to Appendix G. Fig. 6-6 Processing Tab 1 Peak Finding: Integration method Select an integration method for the peaks to be displayed in the data processing window. 2 Threshold Set a detection limit for the photometric value axis of peak and valley. Peaks or valleys below the setting will not be detected....
All Hitachi manuals