Page 381
Probes Vivid7/Vivid7 PRO Users Manual367 FC092326-03 Disconnecting the probe To disconnect probes: 1. Rotate the lock handle counter-clockwise to the horizontal position to unlock the connector. The probes that are not connected to the unit should be stored in their stor- age case.2. Remove the connector from the port. 3. Ensure that the probe head is clean before placing the probe in its storage case. CAUTIONMake sure that the probe and application names displayed on the screen correspond to the...
Page 382
Probes 368Vivid7/Vivid7 PRO Users Manual FC092326-03 Care and Maintenance This section covers: • Planned maintenance • Probe inspection • Probe cleaning • Probe disinfection Planned maintenance It is recommended to keep a maintenance log and note all probe malfunctions. Follow the maintenance schedule below to ensure optimum operation and safety: After each use: • Inspect the probe • Clean the probe • If required disinfect the probe Before each use: • Inspect the probe CAUTIONImproper handling can...
Page 383
Probes Vivid7/Vivid7 PRO Users Manual369 FC092326-03 Inspecting the probe After each use: 1. Inspect the lens, the probe housing and the cable (Figure 10-4. 2. Look for damage that might allow liquid into the probe. Before each use: 1. Inspect the lens, the probe housing and the cable (Figure 10-4). 2. Look for damage that might allow liquid into the probe. 3. Test the functionality of the probe. Figure 10-4: Probe parts CAUTIONIf any damage is found, DO NOT use the probe until it has been...
Page 384
Probes 370Vivid7/Vivid7 PRO Users Manual FC092326-03 Cleaning and disinfecting probes Cleaning probes Cleaning procedure 1. Disconnect the probe from the unit. 2. Remove the coupling gel by wiping the probe lens with a soft cloth. 3. Wipe the probe and cable with a soft cloth moisten in a warm soap and water solution (
Page 385
Probes Vivid7/Vivid7 PRO Users Manual371 FC092326-03 endocavity (TV,TR,TE) probes after contact with mucosal membrane. High-level disinfection procedure Follow the manu- facturers instruc- tions for storage, use and disposal of the disinfection so- lution.1. Prepare the germicide solution according to the manufacturers instructions. 2. Place the cleaned dried probe in contact with the germicide for the time duration specified by the manufacturer. 3. Rinse the part of the probe which was in contact...
Page 386
Probes 372Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 10-5: Probe immersion levels 1. Fluid level 2. Contact face with patient environment
Page 387
Probes Vivid7/Vivid7 PRO Users Manual373 FC092326-03 Probe safety This section includes information on hazards to both the user and the equipment, as follow: • Electrical hazards • Mechanical hazards • Biological hazards Electrical hazards Probes are driven by electricity, which can injure the patient or user when exposed to contact with conductive solution. Mechanical hazards Take precaution to avoid mechanical hazards. WARNINGDo not immerse the probe into any liquid beyond the level shown in Figure...
Page 388
Probes 374Vivid7/Vivid7 PRO Users Manual FC092326-03 Biological hazards To minimize disease transmission, legally marketed and sterile pyrogen-free sheaths should be used for each probe recommended for intra-cavity procedures. Adequate cleaning and disinfection are essential to prevent disease transmission. It is the responsibility of the user to verify and maintain the effectiveness of the infection control procedures in use. CAUTIONTransesophageal probes require a special handling. Refer to the...
Page 389
Probes Vivid7/Vivid7 PRO Users Manual375 FC092326-03 Biopsy The Vivid 7/Vivid 7 PRO supports biopsy capability for the 3.5C, 10L and 12L probes. The biopsy option is intended for use by a duly licensed physician who has received the appropriate training in biopsy techniques as dictated by current relevant practices, as well as in proper operation of the Vivid 7/ Vivid 7 PRO ultrasound unit. Precaution concerning the use of biopsy procedures WARNINGDo not freeze the image during a biopsy procedure....
Page 390
Probes 376Vivid7/Vivid7 PRO Users Manual FC092326-03 Preparing the Biopsy guide attachment The 3.5C, 10L and 12L probes have an optional biopsy kit specific for each probe. The biopsy kit consists of: • One reusable non-sterile bracket • Five disposable sterile Ultra-Pro II TM Needle guide kits (Civco Medical Instruments Co, Inc.) consisting of: • Two sets with needle inserts covering gauge size 14 through 23 (2.1mm to 0.6mm) • One sterile sheath • Two rubber bands • gel • One reusable needle...