GE Vivid 7 User Manual
Here you can view all the pages of manual GE Vivid 7 User Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 371
Probes Vivid7/Vivid7 PRO Users Manual357 FC092326-03 Linear Array probes 7S2D mode M-Mode Color Flow CW Doppler PW Doppler• Cardiology • Coronary • PediatricFrequency: Foot print:3.0 - 8.0 MHz 15 x 21 mm 10S2D mode M-Mode Color Flow CW Doppler PW Doppler• Cardiology • Coronary • PediatricFrequency: Foot print:4.0 - 12.0 MHz 10 x 14 mm Probe Mode Intended use Technical data Probe Mode Intended use Technical data 7L 2D mode M-Mode Color Flow CW Doppler PW Doppler• Peripheral vascular • Small...
Page 372
Probes 358Vivid7/Vivid7 PRO Users Manual FC092326-03 12L 2D mode M-Mode Color Flow CW Doppler PW Doppler• Small parts • peripheral vascularFrequency: Foot print:5.0 - 11.5 MHz 17 x 49 mm M12L a)2D mode M-Mode Color Flow CW Doppler PW Doppler• Carotid • SuperficialFrequency: Foot print:5.0 - 13.0 MHz 10 x 45 mm i8L2D mode M-Mode Color Flow CW Doppler PW Doppler• Cardiac intraoperativ eFrequency: Foot print:5.0 - 8.0 MHz 16 x 44 mm i13L2D mode M-Mode Color Flow CW Doppler PW Doppler• Cardiac...
Page 373
Probes Vivid7/Vivid7 PRO Users Manual359 FC092326-03 Cur ved Array (Convex) probes Probe Mode Intended use Technical data 3.5C2D mode M-Mode Color Flow CW Doppler PW Doppler• Abdomen • Fetal Heart • Obstetrics • Pelvic • RenalFrequency: Foot print: FOV:2.4 - 5.0 MHz 18 x 64 mm 58 degrees 5C2D mode M-Mode Color Flow CW Doppler PW Doppler• Abdomen • Fetal Heart • Obstetrics • Pelvic • RenalFrequency: Foot print: FOV:3.6 - 8.0 MHz 17 x 58 mm 73 degrees M7C2D mode M-Mode Color Flow CW Doppler PW Doppler•...
Page 374
Probes 360Vivid7/Vivid7 PRO Users Manual FC092326-03 Doppler probe Multiplane Transesophageal Phased Array probe Probe Mode Intended use Technical data 2D (P2D)CW Doppler • Cardiology Frequency: 2.0 MHz 6D (P6D)CW Doppler • Carotid Frequency: 6.0 MHz Probe Mode Intended use Technical data 6T2D mode M-Mode Color Flow CW Doppler PW Doppler• Transesopha geal CardiologyFrequency: 2.9 - 6.7 MHz 7T2D mode M-Mode Color Flow CW Doppler PW Doppler• Transesopha geal CardiologyFrequency: 3.0 - 8.0 MHz
Page 375
Probes Vivid7/Vivid7 PRO Users Manual361 FC092326-03 9T2D mode M-Mode Color Flow CW Doppler PW Doppler• Transesopha geal pediatric CardiologyFrequency: 3.3 - 10.0 MHz 5T PAMPTE2D mode M-Mode Color Flow CW Doppler PW Doppler• Transesopha geal CardiologyFrequency: 2.9 - 6.7 MHz 8T2D mode M-Mode Color Flow CW Doppler PW Doppler• Transesopha geal pediatric CardiologyFrequency: 3.3 - 8.0 MHz Probe Mode Intended use Technical data
Page 376
Probes 362Vivid7/Vivid7 PRO Users Manual FC092326-03 Probe orientation Some probes are provided with a green light (LED) orientation marking near their head (see Figure 10-1). Probes which do not have a LED have an indentation (notch) for orientation on the probe housing. This LED, or notch, corresponds with the V mark on the scanning screen. The V mark indicates the orientation of the probe to the scan. Figure 10-1: Orientation marking on probe and on screen Probe labelling Each probe is labelled...
Page 377
Probes Vivid7/Vivid7 PRO Users Manual363 FC092326-03 The probe name displayed on both the probe housing and the connector can be read when the probe is connected. Figure 10-2: Probe labelling (examples) 1. CE mark 2. Probe name M12L MODEL2250695SERIAL
Page 378
Probes 364Vivid7/Vivid7 PRO Users Manual FC092326-03 Probe Integration This section covers: • Connecting the probe • Activating the probe • Disconnecting the probe Connecting the probe Probes can be connected at any time, whether the unit is on or off. To connect a probe: 1. Hold the probe connector vertically with the cable pointing upward. 2. Turn the connector locking handle to the horizontal position. 3. Align the connector with the probe port and carefully push into place. 4. Rotate the...
Page 379
Probes Vivid7/Vivid7 PRO Users Manual365 FC092326-03 Connecting the PAMPTE or the 6Tv probe The PAMPTE and the 6Tv probes are equipped with a Vivid Five connector. To connect the these probes on the Vivid 7/ Vivid 7 PRO, an adaptor must be used (see Figure 10-3). To connect the PAMPTE or 6Tv probe: 1. Turn the probe connector locking handle to the horizontal position. 2. Align the connector with the adaptor port and carefully push into place. 3. Rotate the probe connector locking handle to the full...
Page 380
Probes 366Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 10-3: The PAMPTE probe adaptor Activating the probe When a probe is connected to the unit it is automatically detected. To select a probe and an application: To change applica- tion without chang- ing the current probe, press APPL. on the control pan- el. 1. Press PROBE on the control panel. A list of the connected probes will pop up. 2. Trackball to the desired probe. An application menu for the desired probe is then listed. 3. Trackball...