Contrast Imaging Vivid7/Vivid7 PRO Users Manual197 FC092326-03 MC Contrast controls MC Contrast assignables controls Horizontal sweep Adjust the refresh rate of the ECG. ECG Trig 1 Specifies the delay (ms) from R-wave to the triggered frame. ECG Trig 2 Specifies the delay (ms) from R-wave to the second frame. Tr i g 2 must be greater than Tr i g 1 for dual triggering to be active. ECG Trig Interval Controls the number of cardiac cycles between triggered images. Simultaneous Enables simultaneous...
Page 212
Contrast Imaging 198Vivid7/Vivid7 PRO Users Manual FC092326-03 MC Contrast Soft menu controls (2D Angio mode) Power Controls the amount of acoustic power applied to the transmitted pulse. Color threshold Too high a Power level will destroy the contrast agent.Controls the level of greyscale intensity that is used as a threshold for color overlay. PRF (Pulse Repetition Frequency) Adjusts the repetition rate of the Doppler pulses transmitted to acquire the data for color mapping. Lower frame rates...
Page 213
Contrast Imaging Vivid7/Vivid7 PRO Users Manual199 FC092326-03 ROI (Region Of Interest) position When the trackball command Pos (position) is selected (see also ’Trackball operation’ on page 59), the position of the color area is adjusted by rolling the trackball.
Page 214
Contrast Imaging 200Vivid7/Vivid7 PRO Users Manual FC092326-03 Running MC Contrast The MC Contrast application works only with the M3S, 3S and 5S probes. To select MC Con- trast application without changing the current probe, press APPL. on the control panel. 1. Press PROBE on the control panel. A list of the connected probes is displayed. 2. Trackball to the desired probe supporting the MC Contrast application. The Application menu for the selected probe is listed. 3. Trackball to LV Contrast...
Page 215
Contrast Imaging Vivid7/Vivid7 PRO Users Manual201 FC092326-03 See also page 62 about cineloop oper- ation.21. Select Left marker and Right marker frames to define the cineloop to transfer. 22. Press IMAGE STORE to save the contrast acquisition. Optimizing MC Contrast • A good baseline image is the key for successful myocardial contrast imaging. Using Coded Harmonic Angio, no colors should be found in the myocardium at baseline. • The ECG triggering position should be adjusted to avoid wall...
Page 216
Contrast Imaging 202Vivid7/Vivid7 PRO Users Manual FC092326-03 Real-Time Myocardial Contrast Imaging Real-Time Myocardial Contrast (RTMC) is intended for visualization of myocardial perfusion by use of ultrasound contrast agents. As opposed to MC Contrast, RTMC performs myocardial contrast imaging in a non-triggered mode. WARNINGThis application is intended for clinical research only. Diagnosis must not be based on results achieved by contrast analysis alone.
Page 217
Contrast Imaging Vivid7/Vivid7 PRO Users Manual203 FC092326-03 RTMC over view Figure 5-5: The RTMC Contrast acquisition screen 1. Sample volume 2. Status window: 3. Soft menu Controls marked with R are also available in freeze and cine replay.
Page 218
Contrast Imaging 204Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 5-6: The RTMC controls on the front panel 1. Application 2. Probe 3. Gain R 4. Assignable keys: • Frequency • Focus • Flash • Flash frames • Flash gain • Simultaneous • Angio • Color maps R • Cineloop (in Freeze) • Flash power • 2D width • ECG trig 5. Soft menu • Power • Frame rate • Color threshold • PRF • Low velocity reject • Sample volume • Radial average • Lateral average • DDP R • 6. Trackball Controls marked with R are also...
Page 219
Contrast Imaging Vivid7/Vivid7 PRO Users Manual205 FC092326-03 RTMC controls RTMC Contrast assignables controls Frequency Enables the adjustment of the probes operating frequency. A higher frequency gives better resolution. Focus Changes the location of the focal point. A triangular marker on the depth scale along the image sector indicates the position of the focal point. Flash Enables the transmission of a pulse with the maximum allowed power over a given time. The purpose of the Flash function is...
Page 220
Contrast Imaging 206Vivid7/Vivid7 PRO Users Manual FC092326-03 Flash power Controls the power (amplitude) of the Flash. The maximum allowed power is default. RTMC Soft menu controls Power Controls the amount of acoustic power applied to the transmitted pulse. Too high a Power level will destroy the contrast agent.Color threshold Controls the level of greyscale intensity that is used as a threshold for color overlay. PRF (Pulse Repetition Frequency) Adjusts the repetition rate of the Doppler pulses...