Page 161
Stress Echo Vivid7/Vivid7 PRO Users Manual147 FC092326-03 Chapter 4 Stress Echo This chapter includes the following information: • Introduction ................................................................................... .. 148 • Selection of a stress test protocol template .............................. .. 149 • Image acquisition .......................................................................... .. 150 • Starting acquisition...
Page 162
Stress Echo 148Vivid7/Vivid7 PRO Users Manual FC092326-03 Introduction The Vivid 7/Vivid 7 PRO ultrasound unit provides an integrated stress echo package, with the ability to perform image acquisition, review, image optimization, and wall segment scoring and reporting for a complete, efficient stress echo examination. The stress package provides protocol templates for exercise, as well as, pharmacological stress examinations. In addition to preset factory protocol templates, templates can be...
Page 163
Stress Echo Vivid7/Vivid7 PRO Users Manual149 FC092326-03 Selection of a stress test protocol template 1. Press PROTOCOL to enter the stress echo mode. The Protocol screen is displayed (see Figure 4-1) showing the default stress protocol for the current probe. To create or edit a template see page 177.2. •To use the current template: Turn freeze off to initiate scanning. •To use another template: Press the assignable TEMPLATE. The template list is displayed. 3. Trackball to the desired template. 4....
Page 164
Stress Echo 150Vivid7/Vivid7 PRO Users Manual FC092326-03 Image acquisition Images are acquired in a pre-defined order, according to the selected template. The highlighted cell (green) of the matrix, displayed in the Clipboard window indicates which view is currently being acquired (see Figure 4-2). The names of both the view and the level for the current cell is displayed on the top corner of the image area and under the template matrix. Figure 4-2: The stress mode acquisition screen 1. Current...
Page 165
Stress Echo Vivid7/Vivid7 PRO Users Manual151 FC092326-03 Star ting acquisition To use the Timer, see page 154.1. Turn freeze off to initiate scanning. 2. Perform a scan that conforms with the view that is highlighted in the template matrix on the Clipboard window. Smart Stress is turned on by de- fault in factory tem- plates. If the selected template has the option Smart Stress turned on (see page 180), a subset of the image acquisition settings for each view in the baseline level will be stored...
Page 166
Stress Echo 152Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 4-3: Display of the Reference loop during acquisition Selecting a view during acquisition A fixed protocol is provided for scanning, based on the selected template. The system automatically highlights the next view to be acquired in the template matrix, as images are stored. However, the order of scanning may be changed manually as follow: Manual selection of a view during acquisition 1. Use the arrow keys on the alphanumeric keyboard...
Page 167
Stress Echo Vivid7/Vivid7 PRO Users Manual153 FC092326-03 Replacing an acquired image 1. Use the arrow keys on the alphanumeric keyboard to highlight the cell that represents the view that is to be replaced. The selected cell in the template matrix is highlighted in red, indicating non-default position. 2. Turn freeze off to initiate scanning. 3. Scan and save the selected loop as explained in the previous section. After storage the system automatically highlights the next available view to be...
Page 168
Stress Echo 154Vivid7/Vivid7 PRO Users Manual FC092326-03 Blood Pressure and hear t rate data The blood pressure (systolic and diastolic) and heart rate can be manually entered during acquisition (or during Wall motion analysis). 1. Press MEASURE. The Measurement menu is displayed. 2. Select Wall Motion. The systolic and diastolic blood pressure and the heart rate for the active cell stress-level are displayed on the button over the scoring diagrams (see Figure 4-9, page 165). By pressing this...
Page 169
Stress Echo Vivid7/Vivid7 PRO Users Manual155 FC092326-03 Continuous capture mode Continuous capture mode enables the user to perform acquisition continuously for all views at any level depending on the selected template configuration. Continuous capture consists of temporary saving images acquired in a storage buffer. To enable best possible use of the limited storage buffer capacity, a Pause/Capture mode is provided, as opposed to the normal Freeze/Scan mode. The Pause mode enables scanning and...
Page 170
Stress Echo 156Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 4-5: The buffer bar in Continuous capture Controlling the capture process When entering a stress level with Continuous capture enabled, the unit is automatically set in Pause mode. 1. Press IMAGE SRORE or 2D FREEZE to start image capture. “Capture” is displayed in the buffer bar, the gauge starts filling and the percentage of filled memory buffer increases (see Figure 4-5, page 156). 2. Press IMAGE SRORE or 2D FREEZE again to stop...