Page 121
Scanning Modes Vivid7/Vivid7 PRO Users Manual107 FC092326-03 Color Mode controls Color Mode assignable controls Horizontal sweep (Color M-Mode only) Adjusts the horizontal refresh rate of the M-Mode area of the display. Scale Adjusts the repetition rate of the Doppler pulses transmitted to acquire the data for color flow mapping. The Scale (Nyquist limit) should be adjusted so that no aliasing occurs, while still having good resolution of velocities. The Nyquist limit should be somewhat above the...
Page 122
Scanning Modes 108Vivid7/Vivid7 PRO Users Manual FC092326-03 Cineloop (in Freeze, Color 2D mode only) Starts cineloop acquisition. Color-Mode Soft menu controls Sample volume Adjusts the size of the color flow Doppler sampling area. Lower setting gives better flow resolution while a higher setting increases sensitivity and helps to locate turbulent flows. Tissue priority Emphasize either the color of the color mode or the greyscale tissue detail of the 2D image. Tissue priority is available in both...
Page 123
Scanning Modes Vivid7/Vivid7 PRO Users Manual109 FC092326-03 Trackball controls ROI (Region Of Interest) size When the trackball command Size is selected (see also ’Trackball operation’ on page 59), the height and width of the color area (or ROI) is adjusted from the trackball. ROI (Region Of Interest) position When the trackball command Pos (position) is selected (see also ’Trackball operation’ on page 59), the position of the color area (or ROI) is adjusted with the trackball.
Page 124
Scanning Modes 110Vivid7/Vivid7 PRO Users Manual FC092326-03 Using Color Mode Color 2D 1. From an optimized 2D image press COLOR. 2. Use the trackball to position the ROI frame over the area to be examined. The assignable con- trols of the trackball are displayed in the trackball status bar in the bottom right corner of the screen.3. Press TRACKBALL. The instruction Size should be highlighted in the trackball status bar. If not, press TRACKBALL again to select Size. 4. Use the trackball to adjust...
Page 125
Scanning Modes Vivid7/Vivid7 PRO Users Manual111 FC092326-03 Optimizing Color Mode Refer to page 400 about creating pre- sets.The use of preset gives optimum performance with minimum adjustment. If necessary, the following controls can be adjusted to further optimize the Color Mode display: • Adjust the Active mode gain to set the gain in the color flow area. The scale value may affect FPS, Low Ve- locity Reject, and Sample Volume.• Adjust Scale to the highest setting that provides adequate flow...
Page 126
Scanning Modes 112Vivid7/Vivid7 PRO Users Manual FC092326-03 PW and CW Doppler PW and CW Doppler over view Figure 3-8: The PW/CW Doppler Mode screen 1. Sample volume (PW only) 2. Angle correction marker 3. Velocity scale 4. Low velocity reject 5. Nyquist velocity 6. Doppler baseline 7. Frequency scale (optional, see page 398) 8. Status window: 9. Soft menu Controls marked with R are also available in freeze and cine replay.
Page 127
Scanning Modes Vivid7/Vivid7 PRO Users Manual113 FC092326-03 Figure 3-9: The PW/CW controls on the front panel 1. Assignable keys: • Horizontal sweep R • Scale • Baseline R • LVR • Invert R • LPRF (in PW mode) • Color maps R MORE • Quick angle • Angle correction 2. Zoom: • Display zoom • HR zoom 3. Img Size: • Small/large 2D display • Top/bottom display • Side by side display 4. Soft menu: • Sample Volume (PW) • Compress R • Reject R • Frequency • Frame rate • Angle Corr. R • Power 5. Freeze 6. CW 7....
Page 128
Scanning Modes 114Vivid7/Vivid7 PRO Users Manual FC092326-03 PW and CW Doppler controls PW and CW Doppler assignable controls Horizontal sweep Adjusts the horizontal refresh rate of the Doppler area of the display. Horizontal sweep is available in live and cine replay. Scale Adjusts the velocity scale to accommodate faster/slower blood flow velocities. Velocity scale determines pulse repetition frequency. Baseline Enables the Doppler baseline to be shifted up and down. The default Doppler baseline is...
Page 129
Scanning Modes Vivid7/Vivid7 PRO Users Manual115 FC092326-03 raised beyond a certain limit, more than one Doppler gate is displayed on the screen. Color maps Displays a drop down menu of different Doppler colorization maps.Use the trackball to select the desired map and press SELECT to activate the map. Quick angle and Angle correction In non-cardiac ap- plications, Angle correction is con- trolled from the Trackball.Enables correction of the Doppler velocity scale by defining the angle between the...
Page 130
Scanning Modes 116Vivid7/Vivid7 PRO Users Manual FC092326-03 Frame rate Adjusts the frame rate. The relative setting of the frame rate is displayed in the status window (under 2D). Power When power is re- duced, it reduces the signal-to-noise ratio, so that the im- age may become noisier.Controls the amount of acoustic power applied in all modes. When power is set to maximum, it is equal to or less than the maximum acoustic power permitted by the FDA. The Thermal Index (TI) and the Mechanical...