Page 171

Stress Echo
Vivid7/Vivid7 PRO Users Manual157
FC092326-03
Running Continuous capture
1. Do all your pre-stress acquisitions in the Cardiac 
application.
The application Ex-
ercise should be 
used in order to get 
maximum continu-
ous capture buffer.2. Press PROTOCOL to enter the stress echo mode.
The Protocol screen is displayed (see Figure 4-1, 
page 149).
3. Press Te m p l a t e.
The template list is displayed.
4. Select the template Exercise.
The Exercise proto-
col template is auto-
matically...

Page 172

Stress Echo
158Vivid7/Vivid7 PRO Users Manual
FC092326-03
Continuous capture with additional image acquisition
If the buffer is filled up before all the image acquisitions are 
done, additional loops can be stored in the clipboard before 
doing image assignment to the views:
1. Perform Continuous capture as described above (steps 1 to 
10).
2. • If the buffer is not filled up: press More and Protocol 
Pause. Live scanning is activated.
• If the buffer is filled up: press Select later in the 
Continuous...

Page 173

Stress Echo
Vivid7/Vivid7 PRO Users Manual159
FC092326-03
The Protocol screen is displayed.
5. Press on the Continuous capture icon on the lower left 
corner of the Protocol screen.
The Continuous capture selection screen is displayed.
6. Assign the cineloops to the views (see page 160).
7. Press Done when finished.
8. Press Delete to discard the loop
OR
Press Store all to keep the entire loop.
The normal procedure is to discard the loop. The loop is 
very big and will take a lot of disk space.
9....

Page 174

Stress Echo
160Vivid7/Vivid7 PRO Users Manual
FC092326-03
Assigning and storing the loops
The cineloops captured in the buffer are assigned to the stress 
protocol views and stored from the Continuous capture 
selection screen (see Figure 4-6).
 Figure 4-6: The Continuous capture selection screen
Assigning a cineloop to a view
1. Trackball to the desired loop in order to assign it to a 
particular view of the stress template.
The frame of the loop is highlighted.
2. Press 
SELECT.
1. Rotate CHANGE PAGE...

Page 175

Stress Echo
Vivid7/Vivid7 PRO Users Manual161
FC092326-03
A pop-up menu is displayed with the view names of the 
template (see Figure 4-7).
The views that are 
already assigned are 
tick marked (see 
Figure 4-7).3. Trackball to the required view name.
4. Press 
SELECT.
The name of the view is displayed above the timers in the 
loop window.
5. Repeat steps Figure 1 through Figure 4 to assign loops to 
the other views of the level.
6. Press the assignable 
DONE when completed.
A dialogue window is...

Page 176

Stress Echo
162Vivid7/Vivid7 PRO Users Manual
FC092326-03
 Figure 4-7: Loop assignment in Continuous capture
1. Assigned loop
2. Highlighted loop
3. Views pop-up menu
4. Highlighted views
5. Already assigned 
view 

Page 177

Stress Echo
Vivid7/Vivid7 PRO Users Manual163
FC092326-03
Analysis
Analysis consists of viewing previously saved loops and 
assigning scores to each cardiac segment, in order to quantify 
the function of the muscle, or wall motion.
Depending on the protocol configuration, the analysis stage 
can be started automatically after completion of the stress test 
or it can be started manually. In this case, the usual procedure 
consists of sequentially opening all image groups (if defined) 
and perform scoring...

Page 178

Stress Echo
164Vivid7/Vivid7 PRO Users Manual
FC092326-03
 Figure 4-8: Image selection from the Protocol screen
Manual selection of images in the Analysis screen
1. When currently in protocol analysis in the Stress analysis 
quad screen (Figure 4-9), hold down 
SHIFT while performing 
steps 2 to 4.
2. Trackball to the first image to select in the Template matrix.
3. Press 
SELECT.
The selected loop is displayed in the Stress analyze 
screen and the next window in the quad screen is 
automatically...

Page 179

Stress Echo
Vivid7/Vivid7 PRO Users Manual165
FC092326-03
4. Repeat steps 2 and 3 to select other images.
Alternative: Double 
click on the last se-
lected image to open 
images.5. Press ANALYZE to open images in the Analyze screen (see 
page 165).
Scoring acquired loops
1. After image selection (see page 163), press ANALYZE.
The Stress Echo Analysis screen is displayed (see 
Figure 4-9).
 Figure 4-9: The stress echo analysis screen (Quad screen)
To edit a score, se-
lect it and choose a 
new score.2....

Page 180

Stress Echo
166Vivid7/Vivid7 PRO Users Manual
FC092326-03
3. Trackball to a score.
Alternative: Press 
the arrow heads at 
the bottom of the 
scoring diagram 
(see Figure 4-9)4. Press SELECT.
The score is displayed in the relevant segment area in the 
diagram (see Figure 4-10).
5. Repeat steps 1 through 3 to score relevant segments.
6. Rotate the assignable 
CHANGE PAGE to display next group 
of images.
7. Repeat steps 1 through 3 to score relevant segments on the 
new loops. 