Page 221

Contrast Imaging
Vivid7/Vivid7 PRO Users Manual207
FC092326-03
Running RTMC
The RTMC Contrast application works only with the M3S and 
3S probes.
To select RTMC ap-
plication without 
changing the cur-
rent probe, press 
APPL. on the control 
panel.
1. Press PROBE on the control panel.
A list of the connected probes is displayed.
2. Trackball to the desired probe supporting the RTMC 
application.
The Application menu for the selected probe is listed.
3. Trackball to Real Time MC application.
4. Press...

Page 222

Contrast Imaging
208Vivid7/Vivid7 PRO Users Manual
FC092326-03
Optimizing RTMC
RTMC is by default scanning in low power mode to avoid 
destruction of slow moving contrast in the myocardium. This 
setting is required for real time contrast imaging, but is sub-
optimal for tissue imaging.
•  To obtain good baseline images, the power may be 
increased prior to contrast injection, but must be reset to its 
default value before contrast injection or infusion to avoid 
destruction of the contrast agent.
•  If...

Page 223

Cardiac Measurement and Analysis
Vivid7/Vivid7 PRO Users Manual209
FC092326-03
Chapter 6 
Cardiac Measurement and 
Analysis
This chapter includes the following information:
• Introduction  ................................................................................... ..  210
• The Assign and Measure modality  .............................................. ..  211
• Starting the Assign and Measure modality ............................... 211
• Entering a study and performing measurements...

Page 224

Cardiac Measurement and Analysis
210Vivid7/Vivid7 PRO Users Manual
FC092326-03
Introduction
The Vivid 7/Vivid 7 PRO Ultrasound unit provides functionality 
for two measurement conventions:
A study is a set of 
related measure-
ments, or measure-
ments that are 
logically grouped 
together. The mea-
surements in a 
study are some-
times used in a for-
mula to calculate 
new parameters 
(e.g. biplane volume 
with EF, SV and 
CO).• Assign and Measure (Measure Protocols): the user 
selects a study consisting...

Page 225

Cardiac Measurement and Analysis
Vivid7/Vivid7 PRO Users Manual211
FC092326-03
The Assign and Measure modality
In this measurement modality, the user selects a study 
consisting in a set of related pre-labelled measurements.
Star ting the Assign and Measure 
modality
1. Press MEASURE on the control panel.
The Measurement Menu is displayed in the Parameters 
window (see example Figure 6-3).
The trackball cursor is in the scanning window, ready for 
starting measurement.
In Freeze mode, the 
UPDATE MENU...

Page 226

Cardiac Measurement and Analysis
212Vivid7/Vivid7 PRO Users Manual
FC092326-03
 Figure 6-2: Example of a measurement study
1. Active application
2. Study
3. Selected study
4. Opened study
5. Measurements 
related to the area 
study for the cardiac 
application 

Page 227

Cardiac Measurement and Analysis
Vivid7/Vivid7 PRO Users Manual213
FC092326-03
Entering a study and perfor ming 
measurements
1. Press MEASURE on the front panel to enter the Assign and 
Measure modality.
The Measurement menu with a list of studies is displayed 
in the Parameters window (see example Figure 6-3).
When entering the Measurement mode for the first time, 
the Caliper tool is selected by default. When re-entering 
the Measurement mode, the first measurement in the 
actual study that has not...

Page 228

Cardiac Measurement and Analysis
214Vivid7/Vivid7 PRO Users Manual
FC092326-03
 Figure 6-3: Display of a performed measurement (example)
1. Performed 
measurement
2. Next measurement 
is automatically 
selected 

Page 229

Cardiac Measurement and Analysis
Vivid7/Vivid7 PRO Users Manual215
FC092326-03
Measure and Assign modality
In this measurement modality, the user performs a 
measurement and assign a label.
Star ting the Measure and Assign 
modality
Alternative: Press 
CALIPER on the con-
trol panel and select 
the measurement 
tool from the as-
signables.
1. Press MEASURE on the control panel.
The Measurement Menu is displayed in the Parameters 
window (see example Figure 6-4).
2. If not selected, trackball to the...

Page 230

Cardiac Measurement and Analysis
216Vivid7/Vivid7 PRO Users Manual
FC092326-03
Perfor ming measurements
2D Measurements
2D Length measurements
1. Generate the 2D image.
2. Press 
FREEZE to stop the cineloop.
Alternative: Press 
CALIPER on the con-
trol panel and press 
CALIPER assignable.
3. Press MEASURE on the Control Panel.
4. Select Caliper in the Measurement Menu (see Figure 6-2).
5. Trackball the cursor to the start point of the measurement.
6. Press 
SELECT to anchor the start point of the...