Contrast Imaging Vivid7/Vivid7 PRO Users Manual207 FC092326-03 Running RTMC The RTMC Contrast application works only with the M3S and 3S probes. To select RTMC ap- plication without changing the cur- rent probe, press APPL. on the control panel. 1. Press PROBE on the control panel. A list of the connected probes is displayed. 2. Trackball to the desired probe supporting the RTMC application. The Application menu for the selected probe is listed. 3. Trackball to Real Time MC application. 4. Press...
Page 222
Contrast Imaging 208Vivid7/Vivid7 PRO Users Manual FC092326-03 Optimizing RTMC RTMC is by default scanning in low power mode to avoid destruction of slow moving contrast in the myocardium. This setting is required for real time contrast imaging, but is sub- optimal for tissue imaging. • To obtain good baseline images, the power may be increased prior to contrast injection, but must be reset to its default value before contrast injection or infusion to avoid destruction of the contrast agent. • If...
Page 223
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual209 FC092326-03 Chapter 6 Cardiac Measurement and Analysis This chapter includes the following information: • Introduction ................................................................................... .. 210 • The Assign and Measure modality .............................................. .. 211 • Starting the Assign and Measure modality ............................... 211 • Entering a study and performing measurements...
Page 224
Cardiac Measurement and Analysis 210Vivid7/Vivid7 PRO Users Manual FC092326-03 Introduction The Vivid 7/Vivid 7 PRO Ultrasound unit provides functionality for two measurement conventions: A study is a set of related measure- ments, or measure- ments that are logically grouped together. The mea- surements in a study are some- times used in a for- mula to calculate new parameters (e.g. biplane volume with EF, SV and CO).• Assign and Measure (Measure Protocols): the user selects a study consisting...
Page 225
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual211 FC092326-03 The Assign and Measure modality In this measurement modality, the user selects a study consisting in a set of related pre-labelled measurements. Star ting the Assign and Measure modality 1. Press MEASURE on the control panel. The Measurement Menu is displayed in the Parameters window (see example Figure 6-3). The trackball cursor is in the scanning window, ready for starting measurement. In Freeze mode, the UPDATE MENU...
Page 226
Cardiac Measurement and Analysis 212Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 6-2: Example of a measurement study 1. Active application 2. Study 3. Selected study 4. Opened study 5. Measurements related to the area study for the cardiac application
Page 227
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual213 FC092326-03 Entering a study and perfor ming measurements 1. Press MEASURE on the front panel to enter the Assign and Measure modality. The Measurement menu with a list of studies is displayed in the Parameters window (see example Figure 6-3). When entering the Measurement mode for the first time, the Caliper tool is selected by default. When re-entering the Measurement mode, the first measurement in the actual study that has not...
Page 228
Cardiac Measurement and Analysis 214Vivid7/Vivid7 PRO Users Manual FC092326-03 Figure 6-3: Display of a performed measurement (example) 1. Performed measurement 2. Next measurement is automatically selected
Page 229
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual215 FC092326-03 Measure and Assign modality In this measurement modality, the user performs a measurement and assign a label. Star ting the Measure and Assign modality Alternative: Press CALIPER on the con- trol panel and select the measurement tool from the as- signables. 1. Press MEASURE on the control panel. The Measurement Menu is displayed in the Parameters window (see example Figure 6-4). 2. If not selected, trackball to the...
Page 230
Cardiac Measurement and Analysis 216Vivid7/Vivid7 PRO Users Manual FC092326-03 Perfor ming measurements 2D Measurements 2D Length measurements 1. Generate the 2D image. 2. Press FREEZE to stop the cineloop. Alternative: Press CALIPER on the con- trol panel and press CALIPER assignable. 3. Press MEASURE on the Control Panel. 4. Select Caliper in the Measurement Menu (see Figure 6-2). 5. Trackball the cursor to the start point of the measurement. 6. Press SELECT to anchor the start point of the...