Page 261
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual247 FC092326-03 Deletion of a trace The user can delete all traces at once or one at a time. To delete all traces 1. If necessary, press TRACKBALL until the QA trackball assignment is selected. 2. Press UPDATE MENU in the trackball area on the control panel. The System menu is displayed. 3. Trackball to Delete all traces. 4. Press SELECT. To delete one specific trace 1. If necessary, press TRACKBALL until the QA trackball assignment is selected. 2....
Page 262
Quantitative Analysis 248Vivid7/Vivid7 PRO Users Manual FC092326-03 Frame disabling Frame disabling excludes the actual frame from the cineloop display. Frame disabling is available only with contrast data. There are two ways to disable a frame. Disabling frames from the frame marker To re-enable a frame: Press SELECT on the cor- responding frame marker. 1. Trackball to the frame marker of the frame to disable beneath the Analysis window (see Figure 7-2). 2. Press SELECT to disable the frame. The...
Page 263
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual249 FC092326-03 3. Press SELECT. The current frame is disabled and the corresponding frame marker is displayed in red. Re-enabling all frames 1. Trackball the cursor to the Frame marker axis. 2. Press UPDATE MENU in the trackball area on the control panel. The System menu is displayed. 3. Trackball to Enable all frames. 4. Press SELECT. All previously disabled frames are re-enabled. Figure 7-2: Frame disabling 1. Analysis window 2. Frame marker...
Page 264
Quantitative Analysis 250Vivid7/Vivid7 PRO Users Manual FC092326-03 Optimizing sample area The sample area can be reshaped and labelled. Reshaping a sample area There are two ways of modifying a sample area: either from the Update menu or by selecting the SAMPLE SHAPE assignable. Reshaping a sample area from the Update menu 1. If necessary, press TRACKBALL until the QA trackball assignment is selected. 2. Trackball to the sample area to reshape. 3. Press UPDATE MENU in the trackball area on the...
Page 265
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual251 FC092326-03 Press Set as default to return to the Quantitative analysis screen and keep the settings as default. Reshaping a sample area from the assignables This procedure allows to reshape either a free sample area or a specific anchored sample area providing that the QA cursor is pointing at the actual sample area. 1. Press the SAMPLE SHAPE assignable. The Sample shape assignable controls are displayed. 2. Adjust the size and angle of the...
Page 266
Quantitative Analysis 252Vivid7/Vivid7 PRO Users Manual FC092326-03 Optimizing the trace display Optimizing the Y-axis Auto-scaling The system can be configured to display the full unit range or a range according to the maximum and minimum values of the displayed trace(s) (auto-scaling function). In addition, the auto- scaling function can be set to be live update (updates while the sample area is moved) or delayed (updated when the sample area is anchored). Setting the auto-scaling function 1. If...
Page 267
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual253 FC092326-03 Ver tical units Applicable with contrast data only.When analyzing contrast data, the Y-axis can be set to display either logarithmic scale (dB) or linear, acoustical units (AU) for both tissue intensity (2D) or Angio intensity data. Selecting the Y-axis unit 1. If necessary, press TRACKBALL until the QA trackball assignment is selected. 2. Trackball to the Analysis window. 3. Press UPDATE MENU in the trackball area on the control...
Page 268
Quantitative Analysis 254Vivid7/Vivid7 PRO Users Manual FC092326-03 The System menu is displayed. 4. Select Smoothing. The Smoothing menu is displayed. 5. Select a smoothing filter. The trace display is updated.
Page 269
Quantitative Analysis Vivid7/Vivid7 PRO Users Manual255 FC092326-03 Switching modes or traces The user can toggle between TVI, Tissue Tracking, Strain rate or Strain modes to access to the mode specific controls (soft menu and assignable) or display alternative traces from within a selected mode. To switch mode 1. Press MORE. 2. Select the desired mode (TVI, Tissue Tracking, Strain rate or Strain. The Soft menu and assignables are updated accordingly. To switch trace 1. If necessary, press TRACKBALL...
Page 270
Quantitative Analysis 256Vivid7/Vivid7 PRO Users Manual FC092326-03 Cur ve fitting analysis Curve fitting analysis is used to estimate local myocardial perfusion rate using ultrasound contrast agents. The analysis is based on two algorithms: • Wash-in curve fitting: find and estimate local perfusion rate using contrast agent. Exponential wash-in is described by the function: y(t)=A[1-e -kt]+B, where: • A (dB or AU) is the intensity from the contrast agent. Note that A+B = contrast + tissue = plateau...