Page 311
Archiving Vivid7/Vivid7 PRO Users Manual297 FC092326-03 Figure 8-18: The Image list screen 1. Examination 2. Examination date and archive location 3. Selected image 4. Preview of selected image 5. Missing images
Page 312
Archiving 298Vivid7/Vivid7 PRO Users Manual FC092326-03 Connectivity This section describes the communication and connection options for the Vivid 7/Vivid 7 PRO ultrasound unit with other devices in the hospital information system. This section covers the procedures for optimal data management from a Vivid 7/ Vivid 7 PRO in the following scenarios: • A stand-alone Vivid 7/Vivid 7 PRO (page 301). • A Vivid 7/Vivid 7 PRO and one or several EchoPAC PC workstations in a sneaker net environment (page...
Page 313
Archiving Vivid7/Vivid7 PRO Users Manual299 FC092326-03 Figure 8-19: Examples of dataflows A default dataflow can be selected from the sublevel Dataflow in the subgroup Connectivity of the configuration management package (see page 423and following pages). A: LocalArchive-Int.HD dataflow: The local database is used for patient archiving. Images are stored to internal harddrive. B: LocalArchive-MOD dataflow: The local database is used for patient archiving. Images are stored to a MOD C:...
Page 314
Archiving 300Vivid7/Vivid7 PRO Users Manual FC092326-03 The dataflow is otherwise selected from the Search/Create Patient window (’Creating a new Patient record or starting an examination from an existing patient record’ on page 49). Refer to Chapter 12, ’Presets and System setup’ on page 391 for a complete list and description of supported dataflows.
Page 315
Archiving Vivid7/Vivid7 PRO Users Manual301 FC092326-03 Stand-alone scanner In this scenario images will most likely be reviewed from VCR tape. If digital images are stored, they should be saved directly on a MOD. Figure 8-20: Stan-alone scanner with LocalArchive-MOD dataflow Dataflow configuration 1. Select the LocalArchive-MOD dataflow as default dataflow in the sublevel Dataflow in the subgroup Connectivity of the Configuration management package (see page 424 and following pages). In this...
Page 316
Archiving 302Vivid7/Vivid7 PRO Users Manual FC092326-03 ’Backup procedure’ on page 329. Images There is no backup function for the images in this scenario (no creation of a safety copy). The image MOD is the recommended media for long-term storage of images.
Page 317
Archiving Vivid7/Vivid7 PRO Users Manual303 FC092326-03 A scanner and EchoPAC PC in a sneaker net environment In this scenario the EchoPAC PC (one or several) is used for review of studies acquired on one or more Vivid 7/Vivid 7 PRO without being connected in a network. Images can be stored on the scanner’s internal harddrive (recommended) or on a dedicated MOD. Images stored on the inter nal harddrive Figure 8-21: A scanner and EchoPAC PC in a sneaker net with images stored on the scanner’s...
Page 318
Archiving 304Vivid7/Vivid7 PRO Users Manual FC092326-03 patient records/examinations’ on page 315. Export from LocalArchive-Int.HD to Removable MOD Archive. To free space in the internal harddrive on the scanner, check the command Delete selected patients after copy in the Export patient window (see Figure 8-28, page 317). Make sure that the option Copy images is checked. 2. Import the data from the Export/import MOD to EchoPAC PC internal harddrive using the Import function as described in the...
Page 319
Archiving Vivid7/Vivid7 PRO Users Manual305 FC092326-03 Images stored on dedicated MO disk Figure 8-22: A scanner and EchoPAC PC in a sneaker net with images stored on a removable MOD In this configuration the images are stored directly on a dedicated MOD and imported to the EchoPAC PC’s internal harddrive. Scanner’s dataflow configuration 1. Select LocalArchive-MOD dataflow in the sublevel Dataflow in the subgroup Connectivity of the Configuration management packages default dataflow (see page 424...
Page 320
Archiving 306Vivid7/Vivid7 PRO Users Manual FC092326-03 Make sure that the option Copy images is unchecked. 2. Import the demographics, measurements and report data from the Export/import MOD to EchoPAC PC internal harddrive using the Import function as described in the workstation user manual. Import from Removable MOD Archive to LocalArchive- Int.HD. Review on the EchoPAC PC workstation 1. After completion of the previous step, insert the Image MOD in the EchoPAC PC MO drive. 2. Select the...