Page 231
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual217 FC092326-03 2D Area measurements 1. Generate the 2D image. 2. Press FREEZE to stop the cineloop. Alternative: Press CALIPER and AREA assignable. 3. Press MEASURE on the Control Panel. 4. Select Area (trace) in the Measurement Menu (see Figure 6-2). 5. Trackball the cursor to the start point of the measurement. See the Status bar to get the next step to perform.6. Press SELECT to anchor the start point of the measurement. 7. Trace...
Page 232
Cardiac Measurement and Analysis 218Vivid7/Vivid7 PRO Users Manual FC092326-03 2D Volume measurements The measurements described in this section enable volume measurement in a defined zone. The measurements tool generates results by two methods: For measurement formulae, refer to the Reference Manual.• Method of Disk (displayed as Vmod in the Measurement result table), known as Simpsons method. • Area/Length method (displayed as Va -l in the Measurement result table). To perform a volume...
Page 233
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual219 FC092326-03 3. Press MEASURE on the Control Panel. See the Status bar to get the next step to perform.4. Press the assignable POINT to select the depth measurement function. 5. Trackball the cursor to the position to measure. The current distance from the probe is displayed in the Measurement result table and is instantaneously updated when moving the cursor. The measurements displayed on the 2D image and the cor- responding...
Page 234
Cardiac Measurement and Analysis 220Vivid7/Vivid7 PRO Users Manual FC092326-03 M-Mode Measurements In M-Mode, the user can perform distance and time measurements. This measurement package has also the following pre-defined measurement studies: • LA/Ao • LV • RV M-Mode Length measurements 1. Generate the M-Mode image. 2. Press FREEZE to stop the cineloop. Alternative: Press CALIPER on the con- trol panel and press CALIPER assignable. 3. Press MEASURE on the Control Panel. 4. Select caliper in the...
Page 235
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual221 FC092326-03 Alternative: Press CALIPER on the con- trol panel and press the AO/LA assign- able. 3. Press MEASURE on the Control Panel. 4. Select Ao/LA in the Measurement Menu. 5. Trackball the cursor along the time axis to the required point to start measurement of Aorta root diameter. See the Status bar to get the next step to perform.6. Press SELECT. The starting point for the measurement is anchored. 7. Trackball to the end point...
Page 236
Cardiac Measurement and Analysis 222Vivid7/Vivid7 PRO Users Manual FC092326-03 8. Press SELECT. The IVSd measurement end point is anchored and the value is displayed in the Measurement result table. The end point of the IVSd is also the start point for the LVIDd. 1. Trackball to the end point of the LVIDd measurement. 2. Press SELECT. The LVIDd measurement end point is anchored and the value is displayed in the Measurement result table. The end point of the LVIDd is also the start point for the...
Page 237
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual223 FC092326-03 The RVIDs value is displayed in the Measurement result table.
Page 238
Cardiac Measurement and Analysis 224Vivid7/Vivid7 PRO Users Manual FC092326-03 Doppler Measurements The following measurements may be calculated on Doppler mode spectra: For measurement formulae, refer to the Reference Manual.• Maximum (peak) and mean velocity • Maximum and mean pressure gradient • Pressure half-time (PHT) • Velocity time integral (VTI) • Mitral valve area (MVA), derived from PHT Velocity and Pressure point measurements 1. Generate the spectrum to be measured. 2. Press FREEZE to...
Page 239
Cardiac Measurement and Analysis Vivid7/Vivid7 PRO Users Manual225 FC092326-03 • Time difference (dT) between anchored points position The measurement display on the spec- trum and the corre- sponding result are numbered.9. To assign a label to the measurement, see page 227. 10. Repeat steps 5 through 8 to make additional measurements. Manual Doppler trace measurements Adjust Compress and reject controls to optimize the Doppler signal.1. Generate the spectrum to be measured. 2. Press FREEZE to...
Page 240
Cardiac Measurement and Analysis 226Vivid7/Vivid7 PRO Users Manual FC092326-03 7. Trackball to the end trace position. 8. Press SELECT to anchor the end point of the trace. The trace is automatically generated and the following measurements are displayed in the Measurement result table: • Maximum and mean Velocities • Maximum and mean pressures • Env. Ti • Velocity time integral (VTI) 9. Trackball the cursor to the next heart beat. 10. Press SELECT to anchor the next heart beat starting point....