Page 51
Optimizing the Image LOGIQ E9 User Guide 2-7 Direction 5180302-100 Rev. 1 Sample Volume Length (SV Length)Yes Sizes the sample volume gate. Packet Size Yes Controls the number of samples gathered for a single color flow vector. Map Compress No When you increase the value, high velocity elements in the map are compressed so that the map darkens. When you decrease the value, low velocity elements in the map are compressed so that the map lightens. The effect is visible in the color bar. CF/PDI...
Page 52
Performing an Exam 2-8 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Doppler Mode Controls Table 2-4: Doppler Mode Controls Control Possible BioeffectDescription/ Benefit Update Yes Toggles between simultaneous and update presentation while viewing the timeline. SV Length (Trackball)Yes Moves the sample volume gate on the B-Modes Doppler Mode cursor. The gate is positioned over a specific position within the vessel. Doppler sample volume length (SV Length)Yes Sizes the sample volume gate....
Page 53
Optimizing the Image LOGIQ E9 User Guide 2-9 Direction 5180302-100 Rev. 1 Trace Direction No Specifies trace direction. Cursor Moving No Cursor Moving lets you ‘walk’ Doppler through a vessel while the Doppler gate is moving. Display Format No Changes the horizontal/vertical layout between B-Mode and M- Mode, or timeline only. Modify Auto Calcs No Activates the menu to select which calculations are automatically calculated. Auto Calcs No Activates the calculation automatically which you select in...
Page 54
Performing an Exam 2-10 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Other Controls Zoom Write zooming an image changes the frame rate which tends to change thermal indices. The position of the focal zones may also change which may cause the peak intensity to occur at a different location in the acoustic field. As a result, the MI (TI) may change. To zoom an image, adjust Zoom. A reference image appears in the upper, left-hand section of the display. To exit zoom, adjust Zoom until the...
Page 55
Optimizing the Image LOGIQ E9 User Guide 2-11 Direction 5180302-100 Rev. 1 Body Patterns To activate body patterns, press the Body Pattern/Ellipse control. A maximum of 6 body pattern packages are displayed on the Touch Panel according to exam category and preset. Select the desired body pattern on the Touch Panel. The selected body pattern is displayed on the monitor. Press the Move Pattern control on the Touch Panel to reposition the body pattern with the Trackball and Set controls. A probe mark...
Page 56
Performing an Exam 2-12 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Annotating an Image Pressing the Comment key or any keys on the alphanumeric keyboard initiates the comment mode. This assigns the trackball function to controlling the cursor and displays the comment library on the Touch Panel menu. In comment mode, text can be added by using the comment library or by typing from the alphanumeric keyboard. After activating the comment mode, a vertical bar type cursor appears on the screen. Use...
Page 57
Measurement and Analysis LOGIQ E9 User Guide 2-13 Direction 5180302-100 Rev. 1 Measurement and Analysis B-Mode Measurements Two basic measurements can be made in B-Mode. • Distance • Circumference and Area • Ellipse Method • Trace Method • Spline Method NOTE: The following instructions assume that you first scan the patient and then press Freeze.
Page 58
Performing an Exam 2-14 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Distance measurement To make a distance measurement: 1. Press Measure once; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points, if preset...
Page 59
Measurement and Analysis LOGIQ E9 User Guide 2-15 Direction 5180302-100 Rev. 1 Circumference and area (ellipse) measurement You can use an ellipse to measure circumference and area. To measure with an ellipse: 1. Press Measure once; an active caliper displays. 2. To position the active caliper, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second caliper, move the Trackball. 5. Adjust the Ellipse...
Page 60
Performing an Exam 2-16 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Circumference and area (trace) measurement Tr a c eTo trace the circumference of a portion of the anatomy and calculate its area: 1. Press Measure. 2. Press the appropriate Trackball key to select Trace; a trace caliper displays. 3. To position the trace caliper at the start point, move the Trackball. 4. To fix the trace start point, press Set. The trace caliper changes to an active caliper. 5. To trace the measurement...