GE Logiq 9 User Manual
Have a look at the manual GE Logiq 9 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Optimizing the Image LOGIQ E9 User Guide 2-7 Direction 5180302-100 Rev. 1 Sample Volume Length (SV Length)Yes Sizes the sample volume gate. Packet Size Yes Controls the number of samples gathered for a single color flow vector. Map Compress No When you increase the value, high velocity elements in the map are compressed so that the map darkens. When you decrease the value, low velocity elements in the map are compressed so that the map lightens. The effect is visible in the color bar. CF/PDI Vertical Size No You can set the default CF/PDI ROI vertical size. CF/PDI Width No You can set the default CF/PDI ROI width. CF/PDI Center DepthNo You can set the default CF/PDI center depth. CF/PDI Frequency (MHz)No You can set the default CF/PDI Frequency (MHz). CF/PDI Auto FrequencyNo You can set the default CF/PDI Auto Frequency. CF/PDI Focus Depth (%)No You can set the default CF/PDI center depth. Power Doppler Imaging (PDI)No Power Doppler Imaging (PDI) is a color flow mapping technique used to map the strength of the Doppler signal coming from the flow rather than the frequency shift of the signal. Using this technique, the ultrasound system plots color flow based on the number of reflectors that are moving, regardless of their velocity. PDI does not map velocity, therefore it is not subject to aliasing. Table 2-3: Color Flow Mode Controls (Continued) Control Possible BioeffectDescription/ Benefit
Performing an Exam 2-8 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Doppler Mode Controls Table 2-4: Doppler Mode Controls Control Possible BioeffectDescription/ Benefit Update Yes Toggles between simultaneous and update presentation while viewing the timeline. SV Length (Trackball)Yes Moves the sample volume gate on the B-Modes Doppler Mode cursor. The gate is positioned over a specific position within the vessel. Doppler sample volume length (SV Length)Yes Sizes the sample volume gate. Scale (Velocity Scale)Yes Adjusts the velocity scale to accommodate faster/slower blood flow velocities. Velocity scale determines pulse repetition frequency. If the sample volume gate range exceeds single gate Scale capability, the system automatically switches to high PRF mode. Multiple gates appear, and HPRF is indicated on the display. Angle Correct No Estimates the flow velocity in a direction at an angle to the Doppler vector by computing the angle between the Doppler vector and the flow to be measured. NOTE: When the Doppler Mode Cursor and angle correct indicator are aligned (the angle is O), you cannot see the angle correct indicator. Quick Angle No Quickly adjusts the angle by 60 degrees. Wall Filter No Insulates the Doppler signal from excessive noise caused from vessel movement. Baseline No Adjusts the baseline to accommodate faster or slower blood flows to eliminate aliasing. Mode Cursor No Displays the Doppler Mode cursor on the B-Mode image. Steer and Fine Steer Yes You can slant the ROI of the Color Flow linear image left or right to get more information without moving the probe. The angle steer function only applies to linear probes. Volume No Controls audio output. Invert No Vertically inverts the spectral trace without affecting the baseline position. Dynamic Range (Compression)No Dynamic range controls how echo intensities are converted to shades of gray, thereby increasing the range of contrast you can adjust. Trace Method (Spectral Trace)No Traces the average mean and peak velocities in realtime or frozen images. Cycles to Average No The average value over a number of cycles (from 1-5). Trace Sensitivity No Adjust the trace to follow the waveform for signal strength.
Optimizing the Image LOGIQ E9 User Guide 2-9 Direction 5180302-100 Rev. 1 Trace Direction No Specifies trace direction. Cursor Moving No Cursor Moving lets you ‘walk’ Doppler through a vessel while the Doppler gate is moving. Display Format No Changes the horizontal/vertical layout between B-Mode and M- Mode, or timeline only. Modify Auto Calcs No Activates the menu to select which calculations are automatically calculated. Auto Calcs No Activates the calculation automatically which you select in the Modify Auto Calculation when the system is in a state of freeze or live. Table 2-4: Doppler Mode Controls (Continued) Control Possible BioeffectDescription/ Benefit
Performing an Exam 2-10 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Other Controls Zoom Write zooming an image changes the frame rate which tends to change thermal indices. The position of the focal zones may also change which may cause the peak intensity to occur at a different location in the acoustic field. As a result, the MI (TI) may change. To zoom an image, adjust Zoom. A reference image appears in the upper, left-hand section of the display. To exit zoom, adjust Zoom until the reference zoom image is removed or press B-Mode. Freezing an image To freeze an image, 1. Press Freeze. The Freeze key backlight turns green. If you are in a mixed mode, both screen formats stop immediately. Deactivating Freeze restarts both modes and places a black bar on the trace to indicate the time discontinuity. To reactivate the image, 1. Press Freeze again. NOTE: Deactivating Freeze erases all measurements and calculations from the display (but not from the worksheet). Use the Trackball to start CINE after pressing Freeze. Activating CINE To activate CINE, 1. Press Freeze. 2. Move the Trackball. Acoustic Output HazardObserve the output display for possible effects.
Optimizing the Image LOGIQ E9 User Guide 2-11 Direction 5180302-100 Rev. 1 Body Patterns To activate body patterns, press the Body Pattern/Ellipse control. A maximum of 6 body pattern packages are displayed on the Touch Panel according to exam category and preset. Select the desired body pattern on the Touch Panel. The selected body pattern is displayed on the monitor. Press the Move Pattern control on the Touch Panel to reposition the body pattern with the Trackball and Set controls. A probe mark is associated with the body patterns and illustrates the probe position on the body pattern. This marker can be placed with the Trackball and rotated with the Ellipse control. The probe mark type is selectable by rotating the Probe Type control on the Touch Panel. There are different choices available with one being a blank selection. To select the active side in dual B-Mode, use the Active Side rotary control at the bottom of the Touch Panel. Press the Body Pattern/Ellipse control to erase the body pattern. The pattern is erased and the system exits the body pattern mode. Press Set on the keyboard or Scan on the Touch Panel to exit without erasing the body pattern.
Performing an Exam 2-12 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Annotating an Image Pressing the Comment key or any keys on the alphanumeric keyboard initiates the comment mode. This assigns the trackball function to controlling the cursor and displays the comment library on the Touch Panel menu. In comment mode, text can be added by using the comment library or by typing from the alphanumeric keyboard. After activating the comment mode, a vertical bar type cursor appears on the screen. Use the Trackball to move the cursor. To delete comments by character, press the Backspace key. To delete all comments and arrow marks, press the Clear key twice immediately after entering the comment mode. To move by words or by text group, press the Tab key. Arrow pointers can be used by activating the F2 (Arrow) key on the keyboard. When the pointer comes up, it is a GREEN color, indicating it is active and can be moved.
Measurement and Analysis LOGIQ E9 User Guide 2-13 Direction 5180302-100 Rev. 1 Measurement and Analysis B-Mode Measurements Two basic measurements can be made in B-Mode. • Distance • Circumference and Area • Ellipse Method • Trace Method • Spline Method NOTE: The following instructions assume that you first scan the patient and then press Freeze.
Performing an Exam 2-14 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Distance measurement To make a distance measurement: 1. Press Measure once; an active caliper displays. 2. To position the active caliper at the start point, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points, if preset accordingly. 5. To complete the measurement, press Set. The system displays the distance value in the Results Window. The following hints can help you to perform distance measurements: HINTS• Before you complete a measurement: • To toggle between active calipers, press Measure. • To erase the second caliper and the current data measured and start the measurement again, press Clear once. • After you complete the measurement: • To rotate through and activate previously fixed calipers, adjust Cursor Select. • To erase all data that has been measured to this point, but not data entered onto worksheets, press Clear.
Measurement and Analysis LOGIQ E9 User Guide 2-15 Direction 5180302-100 Rev. 1 Circumference and area (ellipse) measurement You can use an ellipse to measure circumference and area. To measure with an ellipse: 1. Press Measure once; an active caliper displays. 2. To position the active caliper, move the Trackball. 3. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. 4. To position the second caliper, move the Trackball. 5. Adjust the Ellipse control; an ellipse with an initial circle shape displays. 6. To position the ellipse and to size the measured axes (move the calipers), move the Trackball. 7. To increase the size, adjust the Ellipse control in a clockwise direction. To decrease the size, adjust the Ellipse control in a counterclockwise direction. 8. To toggle between active calipers, press the appropriate Trackball key. 9. To complete the measurement, press Set. The system displays the circumference and area in the Results Window. HINTSBefore you complete the ellipse measurement: • To erase the ellipse and the current data measured, press Clear once. The original caliper is displayed to restart the measurement. • To exit the measurement function without completing the measurement, press Clear a second time.
Performing an Exam 2-16 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Circumference and area (trace) measurement Tr a c eTo trace the circumference of a portion of the anatomy and calculate its area: 1. Press Measure. 2. Press the appropriate Trackball key to select Trace; a trace caliper displays. 3. To position the trace caliper at the start point, move the Trackball. 4. To fix the trace start point, press Set. The trace caliper changes to an active caliper. 5. To trace the measurement area, move the Trackball around the anatomy. A dotted line shows the traced area. 6. To complete the measurement, press Set. The system displays the circumference and the area in the Results Window.