    							Optimizing the Image
    LOGIQ E9 User Guide 2-7
    Direction 5180302-100 Rev. 1
    Sample Volume 
    Length (SV Length)Yes Sizes the sample volume gate.
    Packet Size Yes Controls the number of samples gathered for a single color flow 
    vector. 
    Map Compress No When you increase the value, high velocity elements in the map are 
    compressed so that the map darkens. When you decrease the 
    value, low velocity elements in the map are compressed so that the 
    map lightens. The effect is visible in the color bar. 
    CF/PDI Vertical Size No You can set the default CF/PDI ROI vertical size.
    CF/PDI Width No You can set the default CF/PDI ROI width.
    CF/PDI Center 
    DepthNo You can set the default CF/PDI center depth. 
    CF/PDI Frequency 
    (MHz)No You can set the default CF/PDI Frequency (MHz).
    CF/PDI Auto 
    FrequencyNo You can set the default CF/PDI Auto Frequency.
    CF/PDI Focus Depth 
    (%)No You can set the default CF/PDI center depth. 
    Power Doppler 
    Imaging (PDI)No Power Doppler Imaging (PDI) is a color flow mapping technique 
    used to map the strength of the Doppler signal coming from the flow 
    rather than the frequency shift of the signal. Using this technique, the 
    ultrasound system plots color flow based on the number of reflectors 
    that are moving, regardless of their velocity. PDI does not map 
    velocity, therefore it is not subject to aliasing. 
    Table 2-3:  Color Flow Mode Controls (Continued)
    Control
    Possible
    BioeffectDescription/
    Benefit 
    						
    							Performing an Exam
    2-8 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Doppler Mode Controls
    Table 2-4:  Doppler Mode Controls
    Control
    Possible
    BioeffectDescription/
    Benefit
    Update Yes Toggles between simultaneous and update presentation while 
    viewing the timeline. 
    SV Length 
    (Trackball)Yes Moves the sample volume gate on the B-Modes Doppler Mode 
    cursor. The gate is positioned over a specific position within the 
    vessel. 
    Doppler sample 
    volume length (SV 
    Length)Yes Sizes the sample volume gate.
    Scale (Velocity 
    Scale)Yes Adjusts the velocity scale to accommodate faster/slower blood flow 
    velocities. Velocity scale determines pulse repetition frequency.
    If the sample volume gate range exceeds single gate Scale 
    capability, the system automatically switches to high PRF mode. 
    Multiple gates appear, and HPRF is indicated on the display. 
    Angle Correct No Estimates the flow velocity in a direction at an angle to the Doppler 
    vector by computing the angle between the Doppler vector and the 
    flow to be measured. 
    NOTE: When the Doppler Mode Cursor and angle correct indicator 
    are aligned (the angle is O), you cannot see the angle correct 
    indicator.
    Quick Angle No Quickly adjusts the angle by 60 degrees. 
    Wall Filter No Insulates the Doppler signal from excessive noise caused from 
    vessel movement. 
    Baseline No Adjusts the baseline to accommodate faster or slower blood flows to 
    eliminate aliasing. 
    Mode Cursor No Displays the Doppler Mode cursor on the B-Mode image. 
    Steer and Fine Steer Yes You can slant the ROI of the Color Flow linear image left or right to 
    get more information without moving the probe. The angle steer 
    function only applies to linear probes.
    Volume No Controls audio output.
    Invert No Vertically inverts the spectral trace without affecting the baseline 
    position. 
    Dynamic Range 
    (Compression)No Dynamic range controls how echo intensities are converted to 
    shades of gray, thereby increasing the range of contrast you can 
    adjust. 
    Trace Method 
    (Spectral Trace)No Traces the average mean and peak velocities in realtime or frozen 
    images.
    Cycles to Average No The average value over a number of cycles (from 1-5).
    Trace Sensitivity No Adjust the trace to follow the waveform for signal strength. 
    						
    							Optimizing the Image
    LOGIQ E9 User Guide 2-9
    Direction 5180302-100 Rev. 1
    Trace Direction No Specifies trace direction. 
    Cursor Moving No Cursor Moving lets you ‘walk’ Doppler through a vessel while the 
    Doppler gate is moving. 
    Display Format No Changes the horizontal/vertical layout between B-Mode and M-
    Mode, or timeline only. 
    Modify Auto Calcs No Activates the menu to select which calculations are automatically 
    calculated. 
    Auto  Calcs No Activates the calculation automatically which you select in the Modify 
    Auto Calculation when the system is in a state of freeze or live. 
    Table 2-4:  Doppler Mode Controls (Continued)
    Control
    Possible
    BioeffectDescription/
    Benefit 
    						
    							Performing an Exam
    2-10 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Other Controls
    Zoom
    Write zooming an image changes the frame rate which tends to 
    change thermal indices. The position of the focal zones may 
    also change which may cause the peak intensity to occur at a 
    different location in the acoustic field. As a result, the MI (TI) 
    may change.
    To zoom an image, adjust Zoom. A reference image appears in 
    the upper, left-hand section of the display. 
    To exit zoom, adjust Zoom until the reference zoom image is 
    removed or press B-Mode. 
    Freezing an image
    To freeze an image, 
    1. Press Freeze. The Freeze key backlight turns green. 
    If you are in a mixed mode, both screen formats stop 
    immediately. Deactivating Freeze restarts both modes and 
    places a black bar on the trace to indicate the time discontinuity. 
    To reactivate the image, 
    1. Press Freeze again. 
    NOTE:  Deactivating Freeze erases all measurements and calculations 
    from the display (but not from the worksheet). 
     Use the Trackball to start CINE after pressing Freeze. 
    Activating CINE
    To activate CINE,
    1. Press Freeze.
    2.  Move the Trackball.
    Acoustic
    Output
    HazardObserve the output display for possible effects.  
    						
    							Optimizing the Image
    LOGIQ E9 User Guide 2-11
    Direction 5180302-100 Rev. 1
    Body Patterns
    To activate body patterns, press the Body Pattern/Ellipse 
    control. A maximum of 6 body pattern packages are displayed 
    on the Touch Panel according to exam category and preset. 
    Select the desired body pattern on the Touch Panel.  The 
    selected body pattern is displayed on the monitor.
    Press the Move Pattern control on the Touch Panel to 
    reposition the body pattern with the Trackball and Set controls.
    A probe mark is associated with the body patterns and illustrates 
    the probe position on the body pattern.  This marker can be 
    placed with the Trackball and rotated with the Ellipse control.
    The probe mark type is selectable by rotating the Probe Type 
    control on the Touch Panel. There are different choices available 
    with one being a blank selection.
    To select the active side in dual B-Mode, use the Active Side 
    rotary control at the bottom of the Touch Panel.
    Press the Body Pattern/Ellipse control to erase the body 
    pattern. The pattern is erased and the system exits the body 
    pattern mode.
    Press Set on the keyboard or Scan on the Touch Panel to exit 
    without erasing the body pattern. 
    						
    							Performing an Exam
    2-12 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Annotating an Image
    Pressing the Comment key or any keys on the alphanumeric 
    keyboard initiates the comment mode. This assigns the trackball 
    function to controlling the cursor and displays the comment 
    library on the Touch Panel menu.
    In comment mode, text can be added by using the comment 
    library or by typing from the alphanumeric keyboard.
    After activating the comment mode, a vertical bar type cursor 
    appears on the screen. Use the Trackball to move the cursor.
    To delete comments by character, press the Backspace key.
    To delete all comments and arrow marks, press the Clear key 
    twice immediately after entering the comment mode. 
    To move by words or by text group, press the Tab key.
    Arrow pointers can be used by activating the F2 (Arrow) key on 
    the keyboard. When the pointer comes up, it is a GREEN color, 
    indicating it is active and can be moved. 
    						
    							Measurement and Analysis
    LOGIQ E9 User Guide 2-13
    Direction 5180302-100 Rev. 1
    Measurement and Analysis
    B-Mode Measurements
    Two basic measurements can be made in B-Mode.
    • Distance
    •  Circumference and Area
    • Ellipse Method
    • Trace Method
    • Spline Method
    NOTE:  The following instructions assume that you first scan the 
    patient and then press Freeze. 
    						
    							Performing an Exam
    2-14 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Distance measurement
    To make a distance measurement:
    1. Press Measure once; an active caliper displays.
    2.  To position the active caliper at the start point, move the 
    Trackball.
    3.  To fix the start point, press Set.
    The system fixes the first caliper and displays a second 
    active caliper.
    4.  To position the second active caliper at the end point, move 
    the Trackball.
    A dotted line connects the measurement points, if preset 
    accordingly.
    5.  To complete the measurement, press Set.
    The system displays the distance value in the Results 
    Window.
    The following hints can help you to perform distance 
    measurements: 
    HINTS• Before you complete a measurement:
    •  To toggle between active calipers, press Measure.
    •  To erase the second caliper and the current data 
    measured and start the measurement again, press 
    Clear once.
    • After
     you complete the measurement:
    •  To rotate through and activate previously fixed calipers, 
    adjust Cursor Select.
    •  To erase all data that has been measured to this point, 
    but not data entered onto worksheets, press Clear. 
    						
    							Measurement and Analysis
    LOGIQ E9 User Guide 2-15
    Direction 5180302-100 Rev. 1
    Circumference and area (ellipse) measurement
    You can use an ellipse to measure circumference and area. To 
    measure with an ellipse:
    1. Press Measure once; an active caliper displays.
    2.  To position the active caliper, move the Trackball.
    3.  To fix the start point, press Set. The system fixes the first 
    caliper and displays a second active caliper.
    4.  To position the second caliper, move the Trackball.
    5. Adjust the Ellipse control; an ellipse with an initial circle 
    shape displays.
    6.  To position the ellipse and to size the measured axes (move 
    the calipers), move the Trackball.
    7.  To increase the size, adjust the Ellipse control in a 
    clockwise direction. To decrease the size, adjust the Ellipse 
    control in a counterclockwise direction.
    8.  To toggle between active calipers, press the appropriate 
    Trackball key.
    9.  To complete the measurement, press Set. The system 
    displays the circumference and area in the Results Window.
    HINTSBefore you complete the ellipse measurement:
    •  To erase the ellipse and the current data measured, press 
    Clear once. The original caliper is displayed to restart the 
    measurement.
    •  To exit the measurement function without completing the 
    measurement, press Clear a second time. 
    						
    							Performing an Exam
    2-16 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Circumference and area (trace) measurement
    Tr a c eTo trace the circumference of a portion of the anatomy and 
    calculate its area: 
    1. Press Measure.
    2. Press the appropriate Trackball key to select Trace; a trace 
    caliper displays.
    3.  To position the trace caliper at the start point, move the 
    Trackball.
    4.  To fix the trace start point, press Set. The trace caliper 
    changes to an active caliper. 
    5.  To trace the measurement area, move the Trackball around 
    the anatomy. A dotted line shows the traced area.
    6.  To complete the measurement, press Set. The system 
    displays the circumference and the area in the Results 
    Window. 
    						