Measurement and Analysis LOGIQ E9 User Guide 2-17 Direction 5180302-100 Rev. 1 Circumference and area (trace) measurement (continued) Open TraceTo trace the circumference of a portion of the anatomy and calculate its length: 1. Press Measure. 2. Press the appropriate Trackball key to select Open trace; a trace caliper displays. NOTE: For open trace, Area (A) is not displayed in the result window. 3. To position the trace caliper at the start point, move the Trackball. 4. To fix the trace start point, press Set. The trace caliper changes to an active caliper. 5. To trace the measurement area, move the Trackball around the anatomy. A dotted line shows the traced area. 6. To complete the measurement, press Set. The system displays the circumference and the length in the Results Window. HINTSBefore you complete the trace measurement: • To erase the line (bit by bit) back from its current point, move the Trackball or adjust the Ellipse control counterclockwise. • To erase the dotted line but not the trace caliper, press Clear once. • To clear the trace caliper and the current data measured, press Clear twice.
Performing an Exam 2-18 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Circumference and area (spline trace) measurement To trace the circumference of a portion of the anatomy and calculate its area: 1. Press Measure. 2. Press the appropriate Trackball key to select Spline Trace; a Spline trace caliper displays. 3. To position the first caliper at the start point, move the Trackball. 4. To fix the trace start point, press Set. The first caliper turns yellow. The second caliper appears at the same position as the first caliper and is green. NOTE: When pressing the Clear key once, the second caliper disappears and the first caliper is activated. If Clear is pressed again, the first caliper disappears and the Spline trace is cancelled. 5. To position the second caliper, move the Trackball and press Set. The third caliper appears at the same position. NOTE: The Clear key functionality is the same as noted in the previous step. The spline trace requires at least three points to draw the trace. Continue setting the points of the trace until the desired points are set. 6. Press Set again after the last caliper is fixed to finalize the spline trace. All points are removed from the line and the spline trace turns yellow. NOTE: Pressing Set twice finishes the trace measurement. If Clear is pressed twice when more than 3 points exist on the trace, all points are removed and the first caliper again displays. NOTE: Spline trace is not available through the factory default. The system defaults to trace. To enable spline trace, modify the Measure Key Sequence preset found in Utility -> Measure -> Advanced preset menu.
Measurement and Analysis LOGIQ E9 User Guide 2-19 Direction 5180302-100 Rev. 1 Circumference and area (spline trace) measurement (continued) Edit the spline trace1. Select Cursor Select. The spline trace changes to green and all points appear on the trace as yellow. A pick-caliper appears on the center of the image and the message “Edit spline trace” displays at the bottom of the screen. NOTE: The pick-caliper is used to select and move the trace points. Figure 2-1. Edit spline trace NOTE: Pressing Clear at this time removes all points and the trace graphic. 2. Move the pick-caliper to the desired point and press Set. The point is activated and turns green. 3. Move the point to the desired position and press Set. The point is fixed and turns yellow. The pick-caliper appears on the center of the image. NOTE: The spline trace is updated at run time. NOTE: To remove a point, press Clear while moving the point. The trace turns green and the remaining points continue to be shown as yellow. If there are less than three points, the spline trace is removed. 4. Press Set again. All points are removed from the trace and the trace is shown as yellow.
Performing an Exam 2-20 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 IMT Measurement You can measure the average of the intima media thickness for use as the index of arterial sclerosis. NOTE: The IMT measurement is available with linear probes only. NOTE: Due to the physical properties of ultrasound imaging, the posterior IMT measurement is generally more accurate than the anterior IMT measurement. 1. Before you measure the IMT, add the IMT measurement to the Carotid folder via the Measurement & Analysis screen (by selecting one of the three types of IMT measurements under Add Measurement in the M&A screen). • IMT: Three vertical lines are parallel. Place the start point on the line and place the end point anywhere. • IMT2: Each vertical line can be rotated with the Ellipse control. You must place the start and end points on the line. • 5mm Scale: The horizontal line can be rotated with the Ellipse control. A maximum of 20 distance values which produce one average value can be taken. The number of distance values is specified when adding the measurement in the M&A screen. IMT and IMT2 have three kinds of measurements: • IMT --+/IMT2 --+: Measure from right to left. • IMT -+-/IMT -+-: First measure at the center, then right and left are last. • IMT+--/IMT2 --+: Measure from left to right. 2. Scan the carotid artery, then press Freeze.
Measurement and Analysis LOGIQ E9 User Guide 2-21 Direction 5180302-100 Rev. 1 IMT Measurement (continued) 3. Press Measure, then select IMT1, IMT2 or 5mm scale. An active caliper displays. Figure 2-2. IMT caliper (Example) 4. Use the Trackball to move the caliper and the Ellipse control to adjust the angle. Press Set to fix the caliper. NOTE: The interval of the vertical line for IMT1 and IMT2 is 1cm and for the 5mm scale is 5mm. 5. Measure the thickness of three points for IMT1 and IMT2. OR Measure the thickness of the specified number of points for the 5mm scale. NOTE: The caliper moves to the next point automatically.
Performing an Exam 2-22 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 IMT Measurement (continued) 6. After you complete the measurement, the system calculates the average automatically. Figure 2-3. IMT Measurement
Measurement and Analysis LOGIQ E9 User Guide 2-23 Direction 5180302-100 Rev. 1Plaque Score Tool1. Before you measure, add Plaque Score to the Carotid folder at the Utility -> Measure -> Measurement & Analysis screen (by selecting 2D Plaque Score under Add Measurement in the M&A screen). Label the parameters you want measured: • Distance value (up to 20) • Sum (greater or equal to 1.1 mm) • Count (greater or equal to 1.1 mm) • Maximum value of each area • Average value • Average value of each area 2. Scan the carotid artery and press Freeze. Display dual images to measure across split screen images. 3. Press Measure and select Plaque Score Tool. An active caliper, one horizontal line and five vertical lines display. The interval of the vertical lines are 1.5 cm. Figure 2-4. Plaque score example 4. Use the Trackball to move the caliper and the Ellipse control to adjust the angle. Press Set to fix the caliper.
Performing an Exam 2-24 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Plaque Score Tool (continued)5. Measure the thickness (up to twenty times) on any place as necessary. Figure 2-5. Plaque Score Example 2 6. The system displays the measurement result. Figure 2-6. Plaque Score Example 3 NOTE: Only calculates the value larger than 1.1mm. 7. Double click the Set key to finish the measurement.
Measurement and Analysis LOGIQ E9 User Guide 2-25 Direction 5180302-100 Rev. 1 Doppler Mode Measurements Four basic measurements can be made in Doppler Mode. • Velocity • TAMAX and TAMEAN (Manual or Auto Trace) • Two Velocities with the Time Interval and Acceleration between them • Time Interval • Volume Flow NOTE: The following instructions assume that you do the following: 1. In the B-Mode part of the display, scan the anatomy you want to measure. 2. Go to the Doppler Mode part of the display. 3. Press Freeze. Ve l o c i t y To measure velocity: 1. Press Measure; an active caliper with a vertical dotted line displays. 2. To position the caliper at the desired measurement point, move the Trackball. 3. To complete the measurement, press Set. The system displays the velocity measurement in the Results Window.
Performing an Exam 2-26 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 TAMAX and TAMEAN Manual TraceThe value measured depends upon the Vol Flow Method preset. The two selections available are: Peak (TAMAX) and Mean (TAMEAN). To do a manual trace of TAMAX or TAMEAN: 1. Press Measure twice; a trace caliper displays. 2. To position the caliper at the trace start point, move the Trackball. 3. To fix the start point, press Set. 4. To trace the maximum values of the desired portion of the spectrum, move the Trackball. NOTE: To edit the trace line, move the Trackball. 5. To complete the measurement, press Set. The system displays the measurement values in the Results Window. Auto TraceThe value measured depends upon the Vol Flow Method preset. The selection available is the Peak (TAMAX). To auto trace TAMAX: 1. Press Measure twice; an active caliper with a vertical dotted line displays. 2. To position the caliper at the trace start point in the Doppler spectrum, move the Trackball. 3. To fix the start point, press Set. 4. To position the vertical caliper at the end point, move the Trackball. 5. To complete the measurement, press Set. The system automatically fixes both calipers and traces the maximum value between the two points. The system displays this value in the Results Window. NOTE: When you set the Auto Trace for Both (above and below), the system picks up the maximum power of the signal, NOT the maximum velocity. If the maximum velocity is not the maximum power, the system may not trace accurately. If you want to use maximum velocity, select either Above or Below.