System Care and Maintenance LOGIQ E9 User Guide 3-33 Direction 5180302-100 Rev. 1 System Care and Maintenance Inspecting the System Examine the following on a monthly basis: • Connectors on cables for any mechanical defects. • Entire length of electrical and power cables for cuts or abrasions. • Equipment for loose or missing hardware. • Control panel and keyboard for defects. • Casters for proper locking operation. CAUTIONTo avoid electrical shock hazard, do not remove panels or covers from console. This servicing must be performed by qualified service personnel. Failure to do so could cause serious injury. Electrical HazardIf any defects are observed or malfunctions occur, do not operate the equipment but inform a qualified service person. Contact a Service Representative for information.
After the Exam is Over 3-34 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Weekly Maintenance The system requires weekly care and maintenance to function safely and properly. Clean the following: • LCD • Operator control panel • B/W Printer • CD/DVD-R Multi Drive Failure to perform required maintenance may result in unnecessary service calls. System Cabinet To clean the system cabinet: 1. Moisten a soft, non-abrasive folded cloth with a mild, general purpose, non-abrasive soap and water solution. 2. Moisten a soft, non-abrasive folded cloth. 3. Wipe down the top, front, back, and both sides of the system cabinet. NOTE: Do not spray any liquid directly into the unit. Touch Panel For the Touch Panel, use a non-ammonia (isopropyl and water) lens cleaner. These are available at most computer outlet stores. DO NOT use Windex, Screen-Clean, etc. as these contain ammonia, which will remove the anti-glare coating on the Touch Panel.
System Care and Maintenance LOGIQ E9 User Guide 3-35 Direction 5180302-100 Rev. 1 Operator Controls ONLY use the following cleaners on the LOGIQ E9 Operator Panel: • Palmolive Dishwashing Liquid (manufactured by Colgate- Palmolive) • Sani Wipes Alcohol-free (manufactured by Micorgen Inc.) • T-Spray II (manufactured by Pharmaceutical Innovations, Inc.) DO NOT USE: • Original T-Spray (T-Spray I) • SaniCloth Plus (red cap); Use Sani-Cloth HB (green cap) instead. • Any cleaning/disinfecting solution containing Alcohol. To clean the operator control panel: 1. Moisten a soft, non-abrasive folded cloth with a mild, general purpose, non-abrasive soap and water solution. 2. Wipe down operator control panel. 3. Use a cotton swab to clean around keys or controls. Use a toothpick to remove solids from between keys and controls. NOTE: When cleaning the operator control panel, make sure not to spill or spray any liquid on the controls, into the system cabinet, or in the probe connection receptacle. NOTE: In case of SARS, use bleach or Cidex in a normal diluted form for cleaning/disinfecting the operator panel. NOTE: DO NOT use T-spray or Sani Wipes on the control panel.
After the Exam is Over 3-36 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Other Maintenance Manually Moving the Operator Panel Up/Down The LOGIQ E9 is equipped with an automatic Operator Panel up/down, left/right, forward/backward control. But if the system is powered down when you need to lower the Operator Panel, you can use the manual release handle located at the back of the unit to manually move the Operator Panel up/down. To move the Operator Panel up/down using the manual control, 1. Locate the manual release handle which extends through the air vent at the back of the LOGIQ E9. 2. Move the manual release handle to the left with one hand while pushing up/down on the Operator Panel to position it.
System Care and Maintenance LOGIQ E9 User Guide 3-37 Direction 5180302-100 Rev. 1 Cleaning the Air Filters Clean the systems air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned quarterly (once every three months). LocatingThe LOGIQ E9 has two air filters which need to be cleaned. The top air filter is located on the back of the LOGIQ E9 below the power cord and the bottom air filter is located underneath the LOGIQ E9. Figure 3-11. Location of Air Filters CAUTIONBe sure to lock the wheels before cleaning the air filters to avoid injury by any unexpected movement of the system.
After the Exam is Over 3-38 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Cleaning the Air Filters (continued) CleaningTo clean the top air filter, 1. Gently remove the filter cover. Pull the cover out and away from the LOGIQ E9. Set it down. 2. Remove the foam filter. 3. Clean the foam filter with soap and water. Let dry. 4. Replace the filter. 5. Replace the filter cover. Position it back into place on the back of the LOGIQ E9. The bottom filter is held in place with magnets. To clean the bottom air filter, 1. You need to kneel down on the floor to access the bottom filter. Locate the piece of cloth extending from the air filter on the bottom of the LOGIQ E9. 2. You need two hands to remove this filter. One hand is used to pull on the piece of cloth that extends from the filter and the other hand is used to wedge a finger just under a corner of the filter. In this step, firmly take hold of the piece of cloth that extends from the filter. 3. Position your other hand at a corner of the filter and wedge a finger between the filter and the bottom of the LOGIQ E9. 4. Pull down on the filter with the finger that is wedged between the filter and the LOGIQ E9 while pulling on the piece of cloth that extends from the filter at the same time to remove the filter. 5. Clean this filter with soap and water. Let dry. 6. Reposition the filter at the bottom of the LOGIQ E9.
System Care and Maintenance LOGIQ E9 User Guide 3-39 Direction 5180302-100 Rev. 1 Supplies/Accessories Not all features or products described in this document may be available or cleared for sale in all markets. Contact the distributor, GE affiliate or sales representative for approved peripherals. For HCATs, contact your sales person. For 2million/5million number part numbers, these are service replacement part numbers that may be either new or refurbished. To order these, contact CARES in the US, or call service in Europe and Asia. The following supplies/accessories have been verified to be compatible with the system: Peripherals CAUTIONDO NOT connect any probes or accessories without approval by GE. Table 3-3: Peripherals and Accessories AccessoryPart Number / HCAT Sony B/W Printer Model UP-D897 5171608 Sony Color Printer Model UP-D23MD 2401985 / H4903JS Sony Color Printer Model UP-D55 5116559 Canon PIXMA MP610 Report Printer 5322258 / H4908RP Canon PIXMA MP610 Report Printer USB Cable 5315370
After the Exam is Over 3-40 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Console Table 3-4: Console Accessories AccessoryPart Number / HCAT Protective Cover H4908DC Volume Navigation Stand 5268740 / H4908NS Volume Navigation Bracket Starter Kit H4908NK LOGIQ E9 Technical Course H4908TC Swedish Keyboard H4908SY Norwegian Keyboard H4908NY Greek Keyboard H4908GY Russian Keyboard H4908RY Power Cord - North American 5148381 / H4000ZA Power Cord - Argentina 5148381-9 / H4000ZH Power Cord - Europe 5148381-3 / H4000ZB Power Cord - UK-Ireland 5148381-4 / H4000ZD Power Cord - Switzerland 5148381-7 / H4000ZL Power Cord - Denmark 5148381-8 / H4000ZE Power Cord - Israel 5148381-10 / H4000ZJ Power Cord - Japan 5148381-2 / H4000ZK Power Cord - China 5148381-5 / H4000ZC Power Cord - Australia 5148381-6 / H4000ZG Power Cord - India 5148381-11 / H4000ZF
System Care and Maintenance LOGIQ E9 User Guide 3-41 Direction 5180302-100 Rev. 1 Probes Table 3-5: Probes and Accessories Probe HCAT Global P/N Japan P/N Biopsy Guide Biopsy Guide HCAT V Nav Capable V Nav Bracket HCAT 3CRF Micro ConvexH40442LP 5196216 5214820Single Angle, Disposable with a Reusable BracketH40442LR No C1-5 ConvexH40452LE 5304539 5314506Multi Angle, Disposable with a Reusable BracketH40432LE Yes H40432LF M6C-D Matrix Array ConvexH40432LM 5182312 5189066Multi Angle, Disposable with a Reusable BracketE8385RF No IC5-9-D Micro Convex IntracavitaryH40442LK 5212417 5189066Single Angle, Disposable with a Disposable or Reusable BracketE8385MJ H40412LNNo S1-5 SectorH4908SC 5269878 Not available at time of publicationMulti Angle, Disposable with a Reusable BracketH4908SD Yes H4908ND M4S-D SectorH40442LN 5224357 5316841Not Available No 9L-D LinearH40442LM 5212849 5251786Multi Angle, Disposable with a Reusable BracketH4906BK Yes H4908NB 11L-D LinearH40432LN 5176908 5245419Multi Angle, Disposable with a Reusable BracketH40432LC No ML6-15-D Matrix Array LinearH40452LG 5271060 5314503Multi Angle, Disposable with a Reusable BracketH40432LK Yes H40432LK RAB2-5-D 4D Convex Vo l u m eH48651MN KTZ157037 5189544Single Angle, Disposable with a Reusable Bracket Single Angle, ReusableH4601AE H48621WNo
After the Exam is Over 3-42 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Options RIC5-9-D 4D Convex Vo l u m e IntracavitaryH48651MS KTZ157043 5189803Single Angle, Reusable H46721R No RNA5-9-D 4D Convex Vo l u m eH48651MY KTZ156994 5260751Single Angle, Disposable with a Reusable Bracket Single Angle, ReusableH46701AF H468651DGNo RSP6-16-D 4D Linear Vo l u m eH48651MR KTZ157046 5189162Single Angle, Disposable with a Reusable Bracket Single Angle, ReusableH47001AD H46721WNo Table 3-6: Options AccessoryPart Number / HCAT LOGIQWorks H4908LW Tru3DH4908TR DICOMH4908DM Report Writer H4908RW Realtime 4D H4908D 4D with VCI H4908MD VOCAL II (Volume Calculation) H4908HL Contrast Imaging (Not available in the USA) H4908C Volume Contrast Imaging (VCI) (Not available in the USA) H4908HM Volume Navigation H4908N Scan Assistant H4908SA Tomographic Ultrasound Imaging H4908TM DVRH4908DR Table 3-5: Probes and Accessories (Continued) Probe HCAT Global P/N Japan P/N Biopsy Guide Biopsy Guide HCAT V Nav Capable V Nav Bracket HCAT