Home > GE > Medical Equipment > GE Logiq 9 User Manual

GE Logiq 9 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 9 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							System Care and Maintenance
    LOGIQ E9 User Guide 3-33
    Direction 5180302-100 Rev. 1
    System Care and Maintenance
    Inspecting the System
    Examine the following on a monthly basis:
    •  Connectors on cables for any mechanical defects.
    •  Entire length of electrical and power cables for cuts or 
    abrasions.
    •  Equipment for loose or missing hardware.
    •  Control panel and keyboard for defects.
    •  Casters for proper locking operation. 
    CAUTIONTo avoid electrical shock hazard, do not remove panels or 
    covers from console.  This servicing must be performed by 
    qualified service personnel.  Failure to do so could cause 
    serious injury.
    Electrical
    HazardIf any defects are observed or malfunctions occur, do not 
    operate the equipment but inform a qualified service person.  
    Contact a Service Representative for information. 
    						
    							After the Exam is Over
    3-34 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Weekly Maintenance
    The system requires weekly care and maintenance to function 
    safely and properly.  Clean the following:
    • LCD
    •  Operator control panel
    • B/W Printer
    •  CD/DVD-R Multi Drive
    Failure to perform required maintenance may result in 
    unnecessary service calls.
    System Cabinet
    To clean the system cabinet:
    1.  Moisten a soft, non-abrasive folded cloth with a mild, 
    general purpose, non-abrasive soap and water solution.
    2.  Moisten a soft, non-abrasive folded cloth.
    3.  Wipe down the top, front, back, and both sides of the system 
    cabinet.
    NOTE:  Do not spray any liquid directly into the unit.
    Touch Panel
    For the Touch Panel, use a non-ammonia (isopropyl and water) 
    lens cleaner. These are available at most computer outlet 
    stores. DO NOT use Windex, Screen-Clean, etc. as these 
    contain ammonia, which will remove the anti-glare coating on 
    the Touch Panel.  
    						
    							System Care and Maintenance
    LOGIQ E9 User Guide 3-35
    Direction 5180302-100 Rev. 1
    Operator Controls
    ONLY use the following cleaners on the LOGIQ E9 Operator 
    Panel: 
    •  Palmolive Dishwashing Liquid (manufactured by Colgate-
    Palmolive)
    •  Sani Wipes Alcohol-free (manufactured by Micorgen Inc.)
    •  T-Spray II (manufactured by Pharmaceutical Innovations, 
    Inc.)
    DO NOT USE: 
    •  Original T-Spray (T-Spray I)
    •  SaniCloth Plus (red cap); Use Sani-Cloth HB (green cap) 
    instead. 
    •  Any cleaning/disinfecting solution containing Alcohol. 
    To clean the operator control panel:
    1.  Moisten a soft, non-abrasive folded cloth with a mild, 
    general purpose, non-abrasive soap and water solution.
    2.  Wipe down operator control panel.
    3.  Use a cotton swab to clean around keys or controls.  Use a 
    toothpick to remove solids from between keys and controls.
    NOTE:  When cleaning the operator control panel, make sure not to spill 
    or spray any liquid on the controls, into the system cabinet, or in 
    the probe connection receptacle.
    NOTE:  In case of SARS, use bleach or Cidex in a normal diluted form 
    for cleaning/disinfecting the operator panel. 
    NOTE:  DO NOT use T-spray or Sani Wipes on the control panel.  
    						
    							After the Exam is Over
    3-36 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Other Maintenance
    Manually Moving the Operator Panel Up/Down
    The LOGIQ E9 is equipped with an automatic Operator Panel 
    up/down, left/right, forward/backward control. But if the system 
    is powered down when you need to lower the Operator Panel, 
    you can use the manual release handle located at the back of 
    the unit to manually move the Operator Panel up/down. 
    To move the Operator Panel up/down using the manual control, 
    1.  Locate the manual release handle which extends through 
    the air vent at the back of the LOGIQ E9.  
    2.  Move the manual release handle to the left with one hand 
    while pushing up/down on the Operator Panel to position it.  
    						
    							System Care and Maintenance
    LOGIQ E9 User Guide 3-37
    Direction 5180302-100 Rev. 1
    Cleaning the Air Filters
    Clean the systems air filters to ensure that a clogged filter does 
    not cause the system to overheat and reduce system 
    performance and reliability. It is recommended the filters be 
    cleaned quarterly (once every three months).
    LocatingThe LOGIQ E9 has two air filters which need to be cleaned. The 
    top air filter is located on the back of the LOGIQ E9 below the 
    power cord and the bottom air filter is located underneath the 
    LOGIQ E9. 
    Figure 3-11. Location of Air Filters
    CAUTIONBe sure to lock the wheels before cleaning the air filters to 
    avoid injury by any unexpected movement of the system. 
    						
    							After the Exam is Over
    3-38 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Cleaning the Air Filters (continued)
    CleaningTo clean the top air filter, 
    1.  Gently remove the filter cover. Pull the cover out and away 
    from the LOGIQ E9. Set it down. 
    2.  Remove the foam filter. 
    3.  Clean the foam filter with soap and water. Let dry. 
    4.  Replace the filter. 
    5.  Replace the filter cover. Position it back into place on the 
    back of the LOGIQ E9. 
    The bottom filter is held in place with magnets. To clean the 
    bottom air filter, 
    1.  You need to kneel down on the floor to access the bottom 
    filter. Locate the piece of cloth extending from the air filter on 
    the bottom of the LOGIQ E9. 
    2.  You need two hands to remove this filter. One hand is used 
    to pull on the piece of cloth that extends from the filter and 
    the other hand is used to wedge a finger just under a corner 
    of the filter. In this step, firmly take hold of the piece of cloth 
    that extends from the filter. 
    3.  Position your other hand at a corner of the filter and wedge 
    a finger between the filter and the bottom of the LOGIQ E9. 
    4.  Pull down on the filter with the finger that is wedged 
    between the filter and the LOGIQ E9 while pulling on the 
    piece of cloth that extends from the filter at the same time to 
    remove the filter. 
    5.  Clean this filter with soap and water. Let dry. 
    6.  Reposition the filter at the bottom of the LOGIQ E9.  
    						
    							System Care and Maintenance
    LOGIQ E9 User Guide 3-39
    Direction 5180302-100 Rev. 1
    Supplies/Accessories
    Not all features or products described in this document may be 
    available or cleared for sale in all markets.
    Contact the distributor, GE affiliate or sales representative for 
    approved peripherals. For HCATs, contact your sales person. 
    For 2million/5million number part numbers, these are service 
    replacement part numbers that may be either new or 
    refurbished. To order these, contact CARES in the US, or call 
    service in Europe and Asia. 
    The following supplies/accessories have been verified to be 
    compatible with the system:
    Peripherals
    CAUTIONDO NOT connect any probes or accessories without approval 
    by GE.
    Table 3-3:  Peripherals and Accessories
    AccessoryPart Number / HCAT
    Sony B/W Printer Model UP-D897 5171608
    Sony Color Printer Model UP-D23MD 2401985 / H4903JS
    Sony Color Printer Model UP-D55 5116559
    Canon PIXMA MP610 Report Printer 5322258 / H4908RP
    Canon PIXMA MP610 Report Printer USB Cable 5315370 
    						
    							After the Exam is Over
    3-40 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Console
    Table 3-4:  Console Accessories
    AccessoryPart Number / HCAT
    Protective Cover H4908DC
    Volume Navigation Stand 5268740 / H4908NS
    Volume Navigation Bracket Starter Kit H4908NK
    LOGIQ E9 Technical Course H4908TC
    Swedish Keyboard H4908SY
    Norwegian Keyboard H4908NY
    Greek Keyboard H4908GY
    Russian Keyboard H4908RY
    Power Cord - North American  5148381 / H4000ZA
    Power Cord - Argentina 5148381-9 / H4000ZH
    Power Cord - Europe 5148381-3 / H4000ZB
    Power Cord - UK-Ireland 5148381-4 / H4000ZD
    Power Cord - Switzerland 5148381-7 / H4000ZL
    Power Cord - Denmark 5148381-8 / H4000ZE
    Power Cord - Israel 5148381-10 / H4000ZJ
    Power Cord - Japan 5148381-2 / H4000ZK
    Power Cord - China 5148381-5 / H4000ZC
    Power Cord - Australia 5148381-6 / H4000ZG
    Power Cord - India 5148381-11 / H4000ZF 
    						
    							System Care and Maintenance
    LOGIQ E9 User Guide 3-41
    Direction 5180302-100 Rev. 1
    Probes
    Table 3-5:  Probes and Accessories
    Probe
    HCAT
    Global P/N
    Japan P/N
    Biopsy Guide
    Biopsy 
    Guide 
    HCAT
    V Nav 
    Capable
    V Nav 
    Bracket 
    HCAT
    3CRF
    Micro ConvexH40442LP
    5196216
    5214820Single Angle, 
    Disposable with a 
    Reusable BracketH40442LR No
    C1-5
    ConvexH40452LE
    5304539
    5314506Multi Angle, Disposable 
    with a Reusable 
    BracketH40432LE Yes H40432LF
    M6C-D
    Matrix Array 
    ConvexH40432LM
    5182312
    5189066Multi Angle, Disposable 
    with a Reusable 
    BracketE8385RF No
    IC5-9-D
    Micro Convex 
    IntracavitaryH40442LK
    5212417
    5189066Single Angle, 
    Disposable with a 
    Disposable or 
    Reusable BracketE8385MJ
    H40412LNNo
    S1-5
    SectorH4908SC
    5269878
    Not available 
    at time of 
    publicationMulti Angle, Disposable 
    with a Reusable 
    BracketH4908SD Yes H4908ND
    M4S-D
    SectorH40442LN
    5224357
    5316841Not Available No
    9L-D
    LinearH40442LM
    5212849
    5251786Multi Angle, Disposable 
    with a Reusable 
    BracketH4906BK Yes H4908NB
    11L-D
    LinearH40432LN
    5176908
    5245419Multi Angle, Disposable 
    with a Reusable 
    BracketH40432LC No
    ML6-15-D
    Matrix Array 
    LinearH40452LG
    5271060
    5314503Multi Angle, Disposable 
    with a Reusable 
    BracketH40432LK Yes H40432LK
    RAB2-5-D
    4D Convex 
    Vo l u m eH48651MN
    KTZ157037
    5189544Single Angle, 
    Disposable with a 
    Reusable Bracket
    Single Angle, ReusableH4601AE
    H48621WNo 
    						
    							After the Exam is Over
    3-42 LOGIQ E9 User Guide
    Direction 5180302-100 Rev. 1
    Options
    RIC5-9-D
    4D Convex 
    Vo l u m e  
    IntracavitaryH48651MS
    KTZ157043
    5189803Single Angle, Reusable H46721R No
    RNA5-9-D
    4D Convex 
    Vo l u m eH48651MY
    KTZ156994
    5260751Single Angle, 
    Disposable with a 
    Reusable Bracket
    Single Angle, ReusableH46701AF
    H468651DGNo
    RSP6-16-D
    4D Linear 
    Vo l u m eH48651MR
    KTZ157046
    5189162Single Angle, 
    Disposable with a 
    Reusable Bracket
    Single Angle, ReusableH47001AD
    H46721WNo
    Table 3-6:  Options
    AccessoryPart Number / HCAT
    LOGIQWorks H4908LW
    Tru3DH4908TR
    DICOMH4908DM
    Report Writer H4908RW
    Realtime 4D H4908D
    4D with VCI H4908MD
    VOCAL II (Volume Calculation) H4908HL
    Contrast Imaging (Not available in the USA) H4908C
    Volume Contrast Imaging (VCI) (Not available in the USA) H4908HM
    Volume Navigation H4908N
    Scan Assistant H4908SA
    Tomographic Ultrasound Imaging H4908TM
    DVRH4908DR
    Table 3-5:  Probes and Accessories (Continued)
    Probe
    HCAT
    Global P/N
    Japan P/N
    Biopsy Guide
    Biopsy 
    Guide 
    HCAT
    V Nav 
    Capable
    V Nav 
    Bracket 
    HCAT 
    						
    All GE manuals Comments (0)