Probes LOGIQ E9 User Guide 1-21 Direction 5180302-100 Rev. 1 Probes Introduction Only use approved probes. DO NOT plug any LOGIQ 9 probes into the LOGIQ E9 system. Use the left-most probe port for the 3CRF and S1-5 probes; use the 3 right probe ports for any “___-D” probe. Figure 1-13. Probe Connectors 1. 3CRF and S1-5 Probe Connector 2. _____-D Probe Connector Refer to the Probes chapter for more information.
Page 32
Getting Started 1-22 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Connecting the Probe Probes can be connected at any time, regardless of whether the console is powered on or off. To ensure that the ports are not active, place the system in the image freeze condition. To connect a probe: 1. Place the probes carrying case on a stable surface and open the case. 2. Carefully remove the probe and unwrap the probe cord. 3. DO NOT allow the probe head to hang free. Impact to the probe head could...
Page 33
Probes LOGIQ E9 User Guide 1-23 Direction 5180302-100 Rev. 1 Cable Handling Take the following precautions with probe cables: • Keep free from wheels. Use the cable hooks located below the operator panel. • Do not bend the cable acutely • Avoid crossing cables between probes. Activating the Probe To activate the probe, select the appropriate probe from the probe indicators on the Touch Panel. Figure 1-14. Probe Indicators 1. Probe Indicators The probes default settings for the mode and selected...
Page 34
Getting Started 1-24 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Disconnecting the Probe Probes can be disconnected at any time. However, the probe should not be active when disconnecting the probe. 1. Ensure the probe is deactivated. Deactivate by selecting another probe or pressing Freeze. 2. Move the probe locking handle to the left. 3. Pull the probe and connector straight out of the probe port. 4. Carefully slide the probe and connector away from the probe port and around the right side...
Page 35
Probes LOGIQ E9 User Guide 1-25 Direction 5180302-100 Rev. 1 LOGIQ E9 Applications The LOGIQ E9 probes are ONLY approved for use for the applications listed in the table below. Table 1-4: Probe Indications for Use Probe Application9L -D 11 L- D ML 6- 15- D M4 S- D S1 -5 C1 -5- DM6 C- D 3C RF IC 5- 9 DRN A5 -9- DRA B2 -5- DRI C5 -9- DRS P6- 16- D Abdomen X X X X X X Small Parts X X X X Periph. Vasc./ VascularXX X X X X Obstetrics X X X X X X X Gynecology X X X X X X Pediatrics X X X X X X Neonatal X...
Page 36
Getting Started 1-26 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 LOGIQ E9 Features Table 1-5: Probe Features Probe Application9L -D 11 L- D ML 6- 15- D M4 S- D S1 -5 C1 -5- DM6 C- D 3C RF IC 5- 9- DRN A5 -9- DRA B2 -5- DRI C5 -9- DRS P6- 16- D Coded HarmonicsXX XXX XXXXX X X X SRI-HD XX XXX XXXXX X X X Coded ContrastXX XXXX X LOGIQ View X X X X X X X X X X X X X Virtual ConvexXX X X X X Easy 3D X X X X X X X X X Advanced 3D X X X X X X X X X Tru3D X X X X 4DXXXX Raw Data X X X X X X X X X X X...
Page 37
Beginning an Exam LOGIQ E9 User Guide 1-27 Direction 5180302-100 Rev. 1 Beginning an Exam Scanning a New Patient 1. Press Patient. 2. Press New Patient on the Patient menu. 3. If there are images on the clipboard, a pop-up menu appears. Specify whether you want to store images permanently, delete images, or go to active images. 4. Choose the exam category. 5. Verify the dataflow. NOTE: DO NOT use the removable media Dataflows on the New Patient menu. NOTE: The system can display a warning...
Page 38
Getting Started 1-28 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Scanning a New Patient (continued) 10. Store the raw data to the clipboard. To store the still image, press Freeze and run the cineloop using the Trackball. Select the frame and press P1 (or the assigned Printer key). To store the cineloop, press Freeze and run the cineloop using the Trackball. Select the start/end frame and run the selected loop. Press P1 (or the assigned Print key). 11. When you have completed the study, press...
Page 39
Beginning an Exam LOGIQ E9 User Guide 1-29 Direction 5180302-100 Rev. 1 Patient Screen Figure 1-15. Patient Screen (Example: Category OB) 1. Image Management 2. Function Selection 3. EZBackup/EZMove 4. Dataflow Selection 5. Exit 6. Patient Information 7. Category Selection 8. Exam Information 9. Patient View 10. Scan Assistant Program
Page 40
Getting Started 1-30 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Patient Screen (continued) Enter Patient Data with the alphanumeric keyboard. To navigate through the Patient Entry menu, use the Tab key or Trackball and Set to move and fix the cursor. The Patient Screen details are: 1. Image Management • Patient–Provides a search and creation of patient. (currently selected) • Image History–Provides a list of images per exam for the currently selected patient. • Active Images–Provides preview...