GE Logiq 9 User Manual
Here you can view all the pages of manual GE Logiq 9 User Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 21
Console Overview LOGIQ E9 User Guide 1-11 Direction 5180302-100 Rev. 1 Monitor Display Figure 1-7. Monitor Display Tour 1. Institution/Hospital Name, Date, Time, Operator Identification. 2. Patient Name, Patient Identification. 3. Power Output Readout 4. GE Symbol: Probe Orientation Marker. 5. Worksheet/Direct Report 6. Measurement Summary Window. 7. Gray/Color Bar. 8. Image. 9. Measurement Calipers. 10. Measurement Results Window. 11. Image Preview. 12. Image Clipboard. 13. Probe Identifier....
Page 22
Getting Started 1-12 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Brightness Adjusting the monitors brightness is one of the most important factors for proper image quality. If these controls are set incorrectly, the Gain, TGC, Dynamic Range and even Power Output may have to be changed more often than necessary to compensate. The proper setup displays a complete gray scale. The lowest level of black should just disappear into the background and the highest white should be bright, but not...
Page 23
Moving the System LOGIQ E9 User Guide 1-13 Direction 5180302-100 Rev. 1 Moving the System Before moving the system 1. Press the Power On/Off switch to power off the system. See ‘Power Off’ on page 1-19 for more information. 2. Unplug the power cord. 3. All cables from off-board peripheral devices (external Color Digital/Report printer, etc.) must be disconnected from the console. 4. Ensure that no loose items are left on the console. 5. Wind the power cable around the cable hook below the rear...
Page 24
Getting Started 1-14 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Before moving the system (continued) 8. Store sufficient gel and other essential accessories in the provided space. 9. Adjust the LCD monitor and control panel to their lowest positions by using the up/down switch on the front of the operator panel. Make sure the operator panel is locked in place. 10. Unlock the wheels. CAUTIONDO NOT place probes into the front pocket storage area when moving the system. This is not a storage...
Page 25
Moving the System LOGIQ E9 User Guide 1-15 Direction 5180302-100 Rev. 1 When moving the system 1. Always use the rear handle grips to move the system. 2. Take extra care when moving the system long distances and on inclines. Ask for help if necessary. Avoid ramps that are steeper than ten degrees to avoid tipping over the system. NOTE: Wheel chair ramps are usually less than five degrees. Utilize additional care and personnel when moving on a steep incline (>5 degrees) or loading the system into...
Page 26
Getting Started 1-16 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Operator Panel Movement Controls The system Operator Panel can be freely moved in all directions. The vertical displacement of the Operator Panel is motor driven. The control buttons are located adjacent to the handles. Figure 1-9. Operator Panel Movement Controls To move the operator panel sideways, press and hold down the left button to release and move the Operator Panel sideways until it is in the desired position. To park...
Page 27
Moving the System LOGIQ E9 User Guide 1-17 Direction 5180302-100 Rev. 1 Setting the front wheels lock The pedal is located between the front wheels of the unit and enables the user to control the movement of the wheels. Figure 1-10. Front Pedal Table 1-3: Front Pedal Functions Pedal positionFunction 1. Left—Swivel Lock Swivel Lock. To move the system, press the Swivel Lock to set the wheels in a forward-to-back position. First move the system in the direction you will be going, then press the...
Page 28
Getting Started 1-18 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 System Start-Up Power On Login Personal IDs and associated passwords can be preset on the LOGIQ E9. If the Use Auto Logon preset is blank, you are prompted to login. Figure 1-11. Operator Login Window 1. Operator: Select the Operator. 2. Password: Enter Operator’s password (optional). 3. Select type of Logon or Cancel. • OK: Standard logon • Cancel: Cancel logon CAUTIONPress the Power On/Off switch to turn the power on. The...
Page 29
System Start-Up LOGIQ E9 User Guide 1-19 Direction 5180302-100 Rev. 1 Power Off To power off the system: 1. When you shutdown the system, enter the scan screen and lightly press the Power On/Off switch at the front of the system once. The System-Exit window is displayed. NOTE: DO NOT press and hold down the Power On/Off switch to shutdown the system. Instead, lightly press the Power On/ Off switch and select Shutdown. 2. Using the Trackball, select Shutdown. The shutdown process takes a few seconds...
Page 30
Getting Started 1-20 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Circuit breaker The Circuit Breaker is located at the rear panel of the system. When set to On, the Circuit Break supplies main power to all internal systems. When set to Off, the Circuit Breaker removes main power from all internal systems. The circuit breaker automatically shuts off power to the system in case of a power overload. If a power overload occurs: 1. Turn off all peripheral devices. 2. Reactivate the Circuit...