    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     40Attribute Name Tag Type Attribute Description Contrast/Bolus Volume (0018,1041) 3 Not used Contrast/Bolus Start Time (0018,1042) 3 Not used Contrast/Bolus Stop Time (0018,1043) 3 Not used Contrast/Bolus Total Dose (0018,1044) 3 Not used Contrast Flow Rate(s) (0018,1046) 3 Not used Contrast Flow Duration(s) (0018,1047) 3 Not used Contrast/Bolus Ingredient (0018,1048) 3 Not used Contrast/Bolus Ingredient Concentration (0018,1049) 3 Not used 4.5.5.4 Palette Color Lookup Table Module 
    This section specifies the attributes that describe the Lookup table data for images with Palette Color 
    photometric interpretation.  
    TABLE 4.5-9  
    PALETTE COLOR LOOKUP MODULE 
    Attribute Name Tag Type  Attribute Description Red Palette Color Lookup Table Descriptor (0028,1101) 1C Only used when reading Palette images. Green Palette Color Lookup Table Descriptor (0028,1102) 1C Only used when reading Palette images. Blue Palette Color Lookup Table Descriptor (0028,1103) 1C Only used when reading Palette images. Palette Color Lookup Table UID (0028,1199) 3 Not used Red Palette Color Lookup Table Data (0028,1201) 1C Only used when reading Palette images. Green Palette Color Lookup Table Data (0028,1202) 1C Only used when reading Palette images. Blue Palette Color Lookup Table Data (0028,1203) 1C Only used when reading Palette images. Segmented Red Palette Color Lookup Table 
    Data (0028,1221) 1C Not used Segmented Green Palette Color Lookup Table 
    Data (0028,1222) 1C Not used Segmented Red Palette Color Lookup Table 
    Data (0028,1223) 1C Not used  
    4.5.5.5 VOI LUT Module 
    This section specifies the attributes that identify and describe the VOI LUT Module  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     41TABLE 4.5-10  
    VOI LUT MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description VOI LUT Sequence (0028,3010) 3 Not used >LUT Descriptor (0028,3002) 3 Not used >LUT Explanation (0028,3003) 3 Not used >LUT Data (0028,3006)  3 Not used Window Center (0028,1050)  3 Value set to 127 if Photometric Interpretation has 
    value MONOCHROME2. Window Width (0028,1051)  3 Value set to 256 if Photometric Interpretation has 
    value MONOCHROME2. Window Center & Width Explanation (0028,1055)  3 Not used  
    4.5.6 General Modules 
    The SOP Common Module is mandatory for all DICOM IODs. 
    4.5.6.1 SOP Common Module 
    This section defines the attributes that are required for proper functioning and identification of the 
    associated SOP Instances. They do not specify any semantics about the Real-World Object represented 
    by the IOD. 
    TABLE 4.5-11  
    SOP COMMON MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description SOP Class UID (0008,0016) 1 Set to 
    “1.2.840.10008.5.1.4.1.1.3.1” 
    “1.2.840.10008.5.1.4.1.1.3” 
    “1.2.840.10008.5.1.4.1.1.6.1” 
    “1.2.840.10008.5.1.4.1.1.6” or 
    “1.2.840.10008.5.1.4.1.1.7”  SOP Instance UID (0008,0018) 1 Uniquely generated by the equipment Specific Character Set (0008,0005) 1C Set to ISO IR 100 or ISO 2022 IR 87 if extended 
    characters are used. Image Read: images using other 
    extended character set than ISO IR 100 or ISO 2022 
    IR 87 are rejected. Instance Creation Date (0008,0012) 3 Not used Instance Creation Time (0008,0013) 3 Not used Instance Creator UID (0008,0014) 3 Not used Instance Number (0020,0013) 3 Not used 4.5.7 US Modules 
    This Section describes US Series, Equipment, and Image Modules. These Modules contain attributes 
    that are specific to US Image IOD. 
    4.5.7.1 US Region Calibration Module 
    US Region Calibration Module is used to describe multiple regions. Note: if a multi-frame image has 
    been acquired with different calibration, the US Region Calibration Module will not be used.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     42TABLE 4.5-12  
    US REGION CALIBRATION MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Sequence of Ultrasound Regions (0018,6011) 1 Object contains 0 or 1 regions >Region Location Min x
    0 (0018,6018) 1 Value is 0 >Region Location Min y
    0 (0018,601A) 1 Value is 0 >Region Location Max x
    1 (0018,601C) 1 Value is image width-1. >Region Location Max y
    1 (0018,601E) 1 Value is image height-1 >Physical Units X Direction (0018,6024) 1 Enumerated Values supported: 
     0003H cm 
     0004H seconds >Physical Units Y Direction (0018,6026) 1 Enumerated Values supported: 
     0003H cm 
     0004H seconds 
     0007H cm/sec >Physical Delta X (0018,602C) 1 Varies with scanning mode >Physical Delta Y (0018,602E) 1 Varies with scanning mode >Reference Pixel x
    0 (0018,6020) 3 Varies with scanning mode >Reference Pixel y
    0 (0018,6022) 3 Varies with scanning mode >Ref. Pixel Physical Value X (0018,6028) 3 Varies with scanning mode >Ref. Pixel Physical Value Y (0018,602A) 3 Varies with scanning mode >Region Spatial Format (0018,6012) 1 Supported. The spatial organization of the data within 
    the region. >Region Data Type (0018,6014) 1 Supported. The type of data within the region. >Region Flags (0018,6016) 1 Bit 0: 0 = Opaque 
    Bit 1: 0 = Not Protected because there may be other 
    regions within the image  
    Bit 2: 0 = Velocity >Pixel Component Organization (0018,6044) 1C Pixel component calibration data does not exist for 
    any region >Pixel Component Mask (0018,6046) 1C Not used >Pixel Component Range Start (0018,6048) 1C Not used >Pixel Component Range Stop (0018,604A) 1C Not used >Pixel Component Physical Units (0018,604C) 1C Not used >Pixel Component Data Type (0018,604E) 1C Not used >Number of Table Break Points (0018,6050) 1C Not used >Table of X Break Points (0018,6052) 1C Not used >Table of Y Break Points (0018,6054) 1C Not used >Number of Table Entries (0018,6056) 1C Not used >Table of Pixel Values (0018,6058) 1C Not used >Table of Parameter Values (0018,605A) 1C Not used >Transducer Frequency (0018,6030) 3 Supported >Pulse Repetition Frequency (0018,6032) 3 Supported >Doppler Correction Angle (0018,6034) 3 Not used >Steering Angle (0018,6036) 3 Not used >Doppler Sample Volume X Position (0018,6038) 3 Not used  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     43Attribute Name Tag Type Attribute Description >Doppler Sample Volume Y Position (0018,603A) 3 Not used >TM-Line Position x
    0 (0018,603C) 3 Not used >TM-Line Position y
    0 (0018,603E) 3 Not used >TM-Line Position x
    1 (0018,6040) 3 Not used >TM-Line Position y
    1 (0018,6042) 3 Not used  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     444.5.7.2 US Image Module 
    This section specifies the attributes that describe ultrasound images. 
    TABLE 4.5-13  
    US IMAGE MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Samples Per Pixel (0028,0002) 1 Value of ‘1’ if Photometric Interpretation element 
    value has value 
     ‘MONOCHROME2’  
    Value of ‘3’ when Photometric Interpretation element 
    value has value  
     ‘RGB’ or 
     YBR_FULL or  
      YBR_FULL_422 Photometric Interpretation (0028,0004) 1 Defined Values used: 
     “MONOCHROME2”, 
     “RGB” , 
     “YBR_FULL” or 
     “YBR_FULL_422”  Bits Allocated (0028,0100) 1 Value always = 0008H Bits Stored (0028,0101) 1 Value always = 0008H High Bit (0028,0102) 1 Value always = 0007H Planar Configuration (0028,0006) 1 Enumerated value 0000H, color-by-pixel, if 
    Photometric Interpretation element value has value 
    ‘RGB’ or if image is JPEG compressed. 
    Enumerated value 0001H, color-by-plane if image is 
    RLE compressed. Pixel Representation (0028,0103) 1 Always 0000H = unsigned integer. Frame Increment Pointer (0028,0009) 1C Export: Is set to Frame Time (0018,1063) or Frame 
    Time Vector (0018,1065) if the image is multiframe 
    IOD, Not used if the image is a single frame IOD. Image Type (0008,0008) 2 The first two values contain “ORIGINAL/PRIMARY” or 
    “DERIVED/ PRIMARY”. 
    Value is the description of the type of application per 
    DICOM PS3.3-2001, section C.8.5.6.1.1 
    Value 4 is a description of the mode. 
    Values 5 and 6 may be used for private data. Lossy Image Compression (0028,2110) 1C Set to 01 if image is compressed using JPEG Baseline 
    compression. Ultrasound Color Data Present (0028,0014) 3 Supported Referenced Overlay Sequence (0008,1130) 3 Not used >Referenced SOP Class UID (0008,1150) 1C Not used >Referenced SOP Instance UID (0008,1155) 1C Not used Referenced Curve Sequence (0008,1145) 3 Not used >Referenced SOP Class UID (0008,1150) 1C Not used >Referenced SOP Instance UID (0008,1155) 1C Not used Stage Name (0008,2120) 3 Name of stage of stress test. Sent if image is acquired 
    in a stress test. The name is defined in the User 
    Interface.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     45Attribute Name Tag Type Attribute Description Stage Number (0008,2122) 3 Number of stage, starting at one. Sent if image is 
    acquired in a stress test. View Name (0008,2127) 3 Name of view of stress test. Sent if image is acquired 
    in a stress test. The name is defined in the User 
    Interface. View Number (0008,2128) 3 Number of view, starting at one. Sent if image is 
    acquired in a stress test. Number of Event Timers (0008,2129) 3 Not used Event Elapsed Time(s) (0008,2130) 3 Not used Event Timer Name(s) (0008,2132) 3 Not used Anatomic Region Sequence (0008,2218) 3 Not used >Include ‘Code Sequence Macro’    >Anatomic Region Modifier Sequence (0008,2220) 3 Not used >>Include ‘Code Sequence Macro’    Primary Anatomic Structure Sequence (0008,2228) 3 Not used >Include ‘Code Sequence Macro’    >Primary Anatomic Structure Modifier 
    Sequence (0008,2230) 3 Not used >>Include ‘Code Sequence Macro’    Transducer Position Sequence (0008,2240) 3 Not used >Include ‘Code Sequence Macro’    >Transducer Position Modifier Sequence (0008,2242) 3 Not used >>Include ‘Code Sequence Macro’    Transducer Orientation Sequence (0008,2244) 3 Not used >Include ‘Code Sequence Macro’    >Transducer Orientation Modifier Sequence (0008,2246) 3 Not used >>Include ‘Code Sequence Macro’     
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     46Attribute Name Tag Type Attribute Description Trigger Time (0018,1060) 3 Not used Nominal Interval (0018,1062) 3 Not used Beat Rejection Flag (0018,1080) 3 Not used Low R-R Value  (0018,1081) 3 Not used High R-R Value  (0018,1082) 3 Not used Heart Rate  (0018,1088) 3 Set to heart rate Output Power (0018,5000) 3 Not used Transducer Data (0018,5010) 3 Not used Transducer Type (0018,6031) 3 Not used Focus Depth (0018,5012) 3 Not used Preprocessing Function (0018,5020) 3 Not used Mechanical Index (0018,5022) 3 Not used Bone Thermal Index, (0018,5024) 3 Not used Cranial Thermal Index (0018,5026) 3 Not used Soft Tissue Thermal Index (0018,5027) 3 Not used Soft Tissue-focus Thermal Index (0018,5028) 3 Not used Soft Tissue-surface Thermal Index (0018,5029) 3 Not used Depth of Scan Field (0018,5050) 3 Not used Image Transformation Matrix (0018,5210) 3 Not used Image Translation Vector (0018,5212) 3 Not used Overlay Subtype (60xx,0045) 3 Not used  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     47 5. ULTRASOUND MULTIFRAME (US MF) INFORMATION OBJECT 
    IMPLEMENTATION 
    5.1 INTRODUCTION 
    This section specifies the use of the DICOM US Multi-frame Image IOD to represent the information 
    included in US images produced by this implementation. Corresponding attributes are conveyed using 
    the module construct. The contents of this section are: 
    5.2 - IOD Implementation 
    5.3 - IOD Entity-Relationship Model 
    5.4 - IOD Module Table 
    5.5 - IOD Module Definition 
    5.2 US MF IOD IMPLEMENTATION 
    This section defines the implementation of US Multi-Frame image information object. 
    5.3 US MF ENTITY-RELATIONSHIP MODEL 
    The Entity-Relationship diagram for the US MF Image interoperability schema is shown in Illustration 
    5.3-1. In this figure, the following diagrammatic convention is established to represent the information 
    organization: 
    · each entity is represented by a rectangular box 
    · each relationship is represented by a diamond shaped box. 
    · the fact that a relationship exists between two entities is depicted by lines connecting the corresponding 
    entity boxes to the relationship boxes. 
    The relationships are fully defined with the maximum number of possible entities in the relationship 
    shown. In other words, the relationship between Series and Image can have up to n Images per Series, 
    but the Patient to Study relationship has 1 Study for each Patient (a Patient can have more than one 
    Study on the system, however each Study will contain all of the information pertaining to that Patient).  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     48 ILLUSTRATION 5.3-1  
    US MULTIFRAME IMAGE ENTITY RELATIONSHIP DIAGRAM Patient Study is 
    the subject of contains creates Equipment US MF 
    Image contains  
    Series 0,n 1 1 1,n 1,n 1 1,n 1  
    5.3.1 Entity Descriptions 
    Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the 
    entities contained within the US Multi-Frame Information Object. 
    5.3.2 LOGIQ 7 Mapping of DICOM entities 
    TABLE 5.3-1  
    MAPPING OF DICOM ENTITIES TO LOGIQ 7 ENTITIES 
    DICOM LOGIQ 7 Entity Patient Patient Study Exam Series Exam Image Image Curve Not used   
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     49 5.4 IOD MODULE TABLE 
    Within an entity of the DICOM US Multi-Frame IOD, attributes are grouped into related set of 
    attributes. A set of related attributes is termed a module. A module facilitates the understanding of the 
    semantics concerning the attributes and how the attributes are related with each other. A module 
    grouping does not infer any encoding of information into datasets. 
    Table 5.4-1 identifies the defined modules within the entities, which comprise the DICOM US Multi-
    Frame IOD. Modules are identified by Module Name. 
    See DICOM Part 3 for a complete definition of the entities, modules, and attributes. 
     
    TABLE 5.4-1 
     US MULTI-FRAME IOD MODULES 
    Entity Name Module Name Reference Patient Patient  4.5.1.1 Study General Study 4.5.2.1  Patient Study 4.5.2.2 Series General Series 4.5.3.1 Frame of Reference Frame of Reference Not used  US Frame of Reference Not used Equipment General Equipment  4.5.4.1 Image General Image 4.5.5.1  Image Pixel  4.5.5.2  Contrast/Bolus 4.5.5.3  Cine 5.5.1.1  Multi-frame 5.5.1.2  Palette Color Lookup Table 4.5.5.4  US Region Calibration 4.5.7.1  US Image 4.5.7.2  Overlay Plane  Not used  VOI LUT 4.5.5.5  SOP Common 4.5.6.1 Curve Not used  
    5.5 INFORMATION MODULE DEFINITIONS 
    Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the 
    entities and modules contained within the US Multi-Frame Information Object. 
    The following modules are included to convey Enumerated Values, Defined Terms, and Optional 
    Attributes supported. Type 1 & Type 2 Attributes are also included for completeness and to define what 
    values they may take and where these values are obtained. It should be noted that they are the same ones 
    as defined in the DICOM Standard Part 3 (Information Object Definitions).  
    						