LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 10Application Context Name 1.2.840.10008.3.1.1.1 The Maximum Length PDU negotiation is included in all association establishment requests. The maximum length PDU for an association initiated by LOGIQ 7 is: Maximum Length PDU 32768 The SOP Class Extended Negotiation is not supported. The user information Items sent by this product are: · Maximum PDU Length · Implementation UID · Implementation Version Name 2.3.1.1.2 Number of Associations The LOGIQ 7 AE will initiate multiple DICOM associations. 2.3.1.1.3 Asynchronous Nature Asynchronous mode is not supported. All operations will be performed synchronously. 2.3.1.1.4 Implementation Identifying Information The Implementation UID for this DICOM Implementation is: LOGIQ 7 Implementation UID 1.2.840.113619.6.115 The Implementation Version Name for this DICOM Implementation is: LOGIQ 7 Implementation Version Name LOGIQ 7_4 Note: The Implementation Version Name may change in the future without modification of this document. 2.3.1.2 Association Initiation Policy The LOGIQ 7 AE attempts to establish a new association with a remote device due to six Real-World Activities: · Exam save initiated by the operator for images and reports and sending request for Storage Commitment. · Verification, which verifies application level communication between peer DICOM AE’s for service purposes. · Worklist initiated by the operator for receiving worklist information. · Image Store/End Exam sending messages to Modality Performed Procedure Step. · Print initiated by the operator for a specific image or group of images. · Exam Query/Retrieve initiated by the operator for receiving examination information and selecting examination to retrieve.
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 112.3.1.2.1 Real-World Activity A (‘Exam save’ Operation) 2.3.1.2.1.1 Associated Real-World Activity Upon a request by the operator (manual or automatic), images will be sent to a DICOM Storage SCP. 2.3.1.2.1.2 Proposed Presentation Context Tables The Proposed Presentation Context Table depends on compression according to the following table: Presentation Context Table – Proposed Abstract Syntax Transfer Syntax Role Extended Negotiation Name UID Name List UID List Presentation Context Table: Compression set to None Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Image Storage (retired) 1.2.840.10008.5.1.4.1.1.6 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Multi-frame Image Storage (retired) 1.2.840.10008.5.1.4.1.1.3 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Presentation Context Table: Compression set to RLE Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 Run Length Encoding, RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.5 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Run Length Encoding, RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.5 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Run Length Encoding, RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.5 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Image Storage (retired) 1.2.840.10008.5.1.4.1.1.6 Run Length Encoding, RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.5 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Ultrasound Multi-frame Image 1.2.840.10008.5.1.4.1.1.3 Run Length Encoding, 1.2.840.10008.1.2.5 SCU None
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 12Storage (retired) RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 Presentation Context Table: Compression set to JPEG Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Ultrasound Multi-frame Image Storage (retired) 1.2.840.10008.5.1.4.1.1.3 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Presentation Context Table for Structured Reports Comprehensive Structured Report 1.2.840.10008.5.1.4.1.1.88.33 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.1.2.1 SOP Specific DICOM Conformance Statement for all Storage SOP Classes This operation also sends a Storage Commitment Request, with the following proposed presentation context. The result from the SCP is expected on another association for the Storage Commitment result. Presentation Context Table Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Storage Commitment Push Model SOP Class 1.2.840.10008.1.20.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None For these SOP class all status codes with status Refused or Error are treated as failures and terminate the association and operation. On a failure, the request will be put in a holding queue for the user to manually retry the request. All status codes with status Warning or Success are treated as successes. 2.3.1.2.2 Real-World Activity B (‘Echo’ Operation) 2.3.1.2.2.1 Associated Real-World Activity The user may initiate a DICOM Verification Request in the Config screen.
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 13Associations will be released upon the receipt of each C-ECHO confirmation. In the event that the SCP does not respond for some reason, the operation will time out and LOGIQ 7 will close the association. 2.3.1.2.2.2 Proposed Presentation Context Table Presentation Context Table - Proposed Abstract Syntax Transfer Syntax Role Name UID Name List UID List Extended Negotiation Verification SOP Class 1.2.840.10008.1.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.3 Real-World Activity C (‘Worklist Query’ Operation) 2.3.1.2.3.1 Associated Real-World Activity The user may initiate a DICOM Worklist Query in Search screen, which will send a C-FIND-RQ to the Worklist SCP. Associations will be released upon the receipt of C-FIND-RSP confirmation. 2.3.1.2.3.2 Proposed Presentation Context Tables Presentation Context Table – Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Modality Worklist Information Model - FIND 1.2.840.10008.5.1.4.31 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.3.2.1 SOP Specific DICOM Conformance Statement for Worklist SOP Classes The LOGIQ 7 includes matching keys in the Modality Worklist quesries as described in Section 9.5. All status codes with status Refused or Error are treated as failures and terminate the association and operation. On a failure, the user will be informed and the last successful query will be used as Worklist. All status codes with status Warning or Success are treated as successes. Service Status Status Code Further Meaning Application Behaviour When receiving Status Codes Related Fields Processed if received Refused A700 Out of resources Terminate the association and operation (0000,0902) 0122 SOP Class not Supported Terminate the association and operation (0000,0902)
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 14Failed A900 Identifier does not match SOP Class Terminate the association and operation (0000, 0901) (0000, 0902) Cxxx Unable to process Terminate the association and operation (0000,0901) (0000,0902) Success 0000 Matching is complete – No final identifier is supplied None Pending FF00 Matches are continuing – Current Match is supplied and any Optional Keys were supported in the same manner Receiving process of the matches continues. Identifier FF01 Matches are continuing – Warning that one or more Optional Keys were not supported for existence for this Identifier Receiving process of the matches continues without any warnings or errors Identifier 2.3.1.2.4 Real-World Activity D (‘Image Store/End exam’ Operation) 2.3.1.2.4.1 Associated Real-World Activity The Modality Performed Procedure Step messages are sent when the first image is acquired for the start of an exam and when the exam is ended (for the case where there are no images, the N-CREATE is sent when the exam is ended). For an exam with saved images or results, the N-SET will be sent with status COMPLETED. For an exam without saved images or results, the N-SET will be sent with status DISCONTINUED. 2.3.1.2.4.2 Proposed Presentation Context Table Presentation Context Table – Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Modality Performed Procedure Step SOP Class 1.2.840.10008.3.1.2.3.3 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.4.2.1 SOP Specific DICOM Conformance Statement for Modality Performed Procedure Step Class LOGIQ 7 includes attributes in the Modality Performed Procedure Step N-CREATE as described in Section 10.2.1
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 15LOGIQ 7 includes attributes in the Modality Performed Procedure StepN-SET as described in Section 10.2.1 The mapping from Worklist attributes is described in Section 9.5.LOGIQ 7 sends N-SET after the exam is ended. The N-SET will include all acquired images’ UIDs and the status of COMPLETED or DISCONTINUED. For this SOP class, all status codes with status Refused or Error are treated as failures and terminate the association and operation. All status codes with status Warning or Success are treated as successes. 2.3.1.2.5 Real-World Activity E (‘Image Print’ Operation) 2.3.1.2.5.1 Associated Real-World Activity Upon a request by the operator, print jobs will be sent to a DICOM Print SCP. If an error occurs during the transmission, the current association is released and a new association initiated. The maximum number of retries is configurable. 2.3.1.2.5.2 Proposed Presentation Context Tables The following table is used: Presentation Context Table - Proposed Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Basic Color Print Management Meta SOP Class 1.2.840.10008.5.1.1.18 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.5.2.1 SOP Specific DICOM Conformance Statement for all Print Management SOP Classes All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status codes with status Warning or Success are treated as successes. 2.3.1.2.6 Real-World Activity F (‘Exam Query/Retrieve’ Operation) 2.3.1.2.6.1 Associated Real-World Activity The user may initiate a DICOM Exam Query in Search screen, which will send a C-FIND-RQ to the Query/Retrieve SCP. Associations will be released upon the receipt of C-FIND-RSP confirmation. The user may then select an examination to be retrieved, using the C-MOVE-RQ command to the Query/Retrieve SCP. The result from the SCP is expected on another association for the retrieved examinations. 2.3.1.2.6.2 Proposed Presentation Context Tables Presentation Context Table – Proposed Abstract Syntax Transfer Syntax Role Extended
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 16Name UID Name List UID List Negotiation Study Root Query/Retrieve Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Study Root Query/Retrieve Information Model - MOVE 1.2.840.10008.5.1.4.1.2.2.2 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None 2.3.1.2.6.2.1 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model - FIND SOP Classes Only a single information model, Study Root, is supported. All queries are initiated at the highest level of the information model (the STUDY level), and then for each response received, recursively repeated at the next lower levels (the SERIES and then IMAGE levels). CANCEL requests can be issued during the queries via graphical user interface. All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status codes with status Warning or Success are treated as successes. LOGIQ 7 will only support hierarchical query. STUDY ROOT REQUEST IDENTIFIER FOR QUERY Name Tag Types of Matching Filtering is Supported STUDY Level Study Date (0008,0020) *, U,R Yes Study Time (0008,0030) S,*, U,R - Accession Number (0008,0050) *, U Yes Patients Name (0010,0010) *, U Yes Patient ID (0010,0020) *, U Yes Study ID (0020,0010) S,*, U - Study Instance UID (0020,000D) UNIQUE - Modalities in Study (0008,0061) S,*, U - Referring Physicians Name (0008,0090) S,*, U - Study Description (0008,1030) S,*, U - Procedure Code Sequence (0008,1032) U - Name of Physician(s) Reading Study (0008,1060) S,*, U - Admitting Diagnoses Description (0008,1080) S,*, U - Referenced Study Sequence (0008,1110) U - Referenced Patient Sequence (0008,1120) U - Patients Birth Date (0010,0030) S,*, U,R - Patients Birth Time (0010,0032) S,*, U,R - Patients Sex (0010,0040) S,*, U - Other Patient IDs (0010,1000) S,*, U - Other Patient Names (0010,1001) S,*, U - Patient’s Age (0010,1010) S,*, U - Patient’s Size (0010,1020) S,*, U - Patient’s Weight (0010,1030) S,*, U - Ethnic Group (0010,2160) S,*, U - Occupation (0010,2180) S,*, U - Additional Patient History (0010,21B0) S,*, U -
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 17Patient Comments (0010,4000) S,*, U - Other Study Numbers (0020,1070) S,*, U - Number of Patient Related Studies (0020,1200) S,*,U - Number of Patient Related Series (0020,1202) S,*,U - Number of Patient Related Instances (0020,1204) S,*,U - Number of Study Related Series (0020,1206) S,*,U - Number of Study Related Instances (0020,1208) S,*,U - Interpretation Author (4008,010C) S,*,U - Series Level Modality (0008,0060) S,*,U - Series Number (0020,0011) S,*,U - Series Instance UID (0020,000E) UNIQUE - Number of Series Related Instances (0020,1209) S,*,U - Series Date (0008,0021) S,*,U,R - Series Time (0008,0031) S,*,U,R - Performing Physicians’ Name (0008,1050) S,*,U - Protocol Name (0018,1030) S,*,U - Series Description (0008,103E) S,*,U - Operator’s Name (0008,1070) S,*,U - Institutional Department Name (0008,1040) S,*,U - Software Versions (0018,1020) S,*,U - Performed Procedure Step Start Date (0040,0244) S,*,U,R - Performed Procedure Step Start Time (0040,0245) S,*,U,R - Image Level Instance Number (0020,0013) S,*,U - SOP Instance UID (0008,0018) UNIQUE - Contrast/Bolus Agent (0018,0010) S,*,U - Types of Matching: - Single Value matching (S) - Universal Matching (U) - Wild Card Matching (*) - Range of date, Range of Time (R) The types of Matching supported by the C-FIND SCU are: ‘S’ indicates the identifier attribute uses Single Value Matching, an ‘R’ indicates Range Matching, a “*” indicates wildcard matching, a ‘U’ indicates Universal Matching, and ‘UNIQUE’ indicates that this is the Unique Key for that query level, in which case Universal Matching or Single Value Matching is used depending on the query level. “Filtering is supported” means that matching strings can be controlled from the Search screen at Patient - Data Transfer. All other matching fields can be configured in Config screen to be either enabled, enabled with a matching string value or disabled. The constant value will be used as entered by user. 2.3.1.2.6.2.2 SOP Specific DICOM Conformance Statement for Study Root Query/Retrieve Information Model - MOVE SOP Classes All status codes with status Refused or Error are treated as failures and terminate the association and operation. All status codes with status Warning or Success are treated as successes.
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 182.3.1.3 Association Acceptance Policy The AE accepts an association when it receives a Verification Request from another network device or a store request from a Q/R SCP or a Storage Commitment result from a Storage Commitment SCP. 2.3.1.3.1 Real-World Activity B – (‘Echo’ operation) 2.3.1.3.1.1 Associated Real-World Activity An incoming Verification Request will cause the AE to accept the association and respond with a Verification Response. 2.3.1.3.1.2 Accepted Presentation Context Table Presentation Context Table - Accepted Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation Verification SOP Class 1.2.840.10008.1.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCP None 2.3.1.3.1.2.1 SOP Specific DICOM Conformance Statement for Verify SOP Class The AE provides standard conformance to the Verification SOP Class as an SCP. The port number used is configured on Config screen, default is 104. 2.3.1.3.1.3 Presentation Context Acceptance Criterion No criterion. 2.3.1.3.1.4 Transfer Syntax Selection Policies The selected transfer syntax is based on the proposed transfer syntax list. The priority order is Explicit VR Little Endian, Explicit VR Big Endian and Implicit VR Little Endian. 2.3.1.3.2 Real-World Activity F (‘Exam Query/Retrieve’ Operation) 2.3.1.3.2.1 Associated Real-World Activity If the user has initiated a retrieve by a C-MOVE-RQ, the AE will accept associations for C-STORE- RQs. The images will be stored locally. 2.3.1.3.2.2 Accepted Presentation Context Table Presentation Context Table - Accepted Abstract Syntax Transfer Syntax Role Extended Name UID Name List UID List Negotiation
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 19Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None Ultrasound Image Storage (retired) 1.2.840.10008.5.1.4.1.1.6 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None Ultrasound Multi-frame Image Storage (retired) 1.2.840.10008.5.1.4.1.1.3 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None 2.3.1.3.2.2.1 SOP Specific DICOM Conformance Statement for Storage SOP Classes The AE provides standard conformance to the Storage SOP Classes as an SCP. The port number used is configured in Config screen, default is 104. 2.3.1.3.2.3 Presentation Context Acceptance Criterion No criterion.