Home > GE > Medical Equipment > GE Logiq 7 User Manual

GE Logiq 7 User Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 7 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     30 
    3.4 AUGMENTED AND PRIVATE APPLICATION PROFILES 
    LOGIQ 7 creates Secondary Capture Image Objects in addition to the objects defined in the application 
    profiles. 
    3.5 EXTENSIONS, SPECIALIZATIONS, PRIVATIZATIONS OF SOP CLASSES AND TRANSFER 
    SYNTAXES 
    If so configured, the product will send ultrasound raw data information in private data elements 
    designated by the Private Creator element: 
    Element Name  Tag VR VM Description Private Creator 7FE1,00xx LO 1 
    GEMS_Ultrasound_MovieGroup_001  
    This means that all private tags starting with 7FE1,xx will belong to the 
    GEMS_Ultrasound_MovieGroup_001. 
    If so configured, the product will send preview image in private data elements designated by the Private 
    Creator element: 
    Element Name  Tag VR VM Description Private Creator 6003,00xx LO 1 
    GEMS_Ultrasound_ImageGroup_001  
    This means that all private tags starting with 6003,xx will belong to the 
    GEMS_Ultrasound_ImageGroup_001. 
    3.6 CONFIGURATION 
    The following parameters are configurable: 
    · Location of DICOMDIR 
    · Read or Read/Write 
    · Enable/disable raw data 
    · Frame rate reduction 
    · Enable/disable multi-frame 
    · Compression selections 
     
    3.7 SUPPORT OF EXTENDED CHARACTER SETS 
     
    LOGIQ 7 will support the ISO IR 100 (ISO 8859-1:1987 Latin alphabet N 1. supplementary set) and 
    ISO 2022 IR 87 (JIS X 0208: Kanji) as extended character set. Any incoming SOP instance that is 
    encoded using another extended character set will not be read.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     314. ULTRASOUND (US) INFORMATION OBJECT IMPLEMENTATION 
    4.1 INTRODUCTION 
    This section specifies the use of the DICOM US Image IOD to represent the information included in US 
    images produced by this implementation. Corresponding attributes are conveyed using the module 
    construct. The contents of this section are: 
    4.2 - IOD Implementation 
    4.3 - IOD Entity-Relationship Model 
    4.4 - IOD Module Table 
    4.5 - IOD Module Definition 
    In this section, supported means that tag is sent with value. 
    4.2 US IOD IMPLEMENTATION 
    This section defines the implementation of US image information object. 
    4.3 US ENTITY-RELATIONSHIP MODEL 
    The Entity-Relationship diagram for the US Image interoperability schema is shown in Illustration 4.3-
    1. In this figure, the following diagrammatic convention is established to represent the information 
    organization: 
    · each entity is represented by a rectangular box 
    · each relationship is represented by a diamond shaped box. 
    · the fact that a relationship exists between two entities is depicted by lines connecting the corresponding 
    entity boxes to the relationship boxes. 
    The relationships are fully defined with the maximum number of possible entities in the relationship 
    shown. In other words, the relationship between Series and Image can have up to n Images per Series, 
    but the Patient to Study relationship has 1 Study for each Patient (a Patient can have more than one 
    Study on the system, however each Study will contain all of the information pertaining to that Patient).  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     32ILLUSTRATION 4.3-1  
    US IMAGE ENTITY RELATIONSHIP DIAGRAM Patient Study is 
    the subject of contains creates Equipment US Image contains  
    Series 
    0,n 1 1 1,n 1,n 1 1,n 1  
    4.3.1 Entity Descriptions 
    Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the 
    entities contained within the US Information Object. 
    4.3.2 LOGIQ 7 Mapping of DICOM Entities 
    TABLE 4.3-1  
    MAPPING OF DICOM ENTITIES TO LOGIQ 7 ENTITIES 
    DICOM LOGIQ 7 Entity Patient Patient Study Exam Series Exam Image Image Curve Not used   
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     334.4 IOD MODULE TABLE 
    Within an entity of the DICOM US IOD, attributes are grouped into related set of attributes. A set of 
    related attributes is termed a module. A module facilitates the understanding of the semantics 
    concerning the attributes and how the attributes are related with each other. A module grouping does not 
    infer any encoding of information into data sets. 
    Table 4.4-1 identifies the defined modules within the entities, which comprise the DICOM US IOD. 
    Modules are identified by Module Name. 
    See DICOM Part 3 for a complete definition of the entities, modules, and attributes. 
    Only the single frame US Image IOD is described here. 
    TABLE 4.4-1 
     US IMAGE IOD MODULES 
    Entity Name Module Name Reference Patient Patient  4.5.1.1 General Study 4.5.2.1 Study 
    Patient Study 4.5.2.2 Series General Series 4.5.3.1 Frame of Reference Not used Frame of Reference 
    US Frame of Reference Not used Equipment General Equipment  4.5.4.1 General Image 4.5.5.1 Image Pixel  4.5.5.2 Contrast/Bolus 4.5.5.3 Palette Color Lookup Table 4.5.5.4 US Region Calibration 4.5.7.1 US Image 4.5.7.2 Overlay Plane  Not used VOI LUT 4.5.5.5 Image 
    SOP Common 4.5.6.1 Curve Not used  
    4.5 INFORMATION MODULE DEFINITIONS 
    Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the 
    entities and modules contained within the US Information Object. 
    The following modules are included to convey Enumerated Values, Defined Terms, and Optional 
    Attributes supported. Type 1 & Type 2 Attributes are also included for completeness and to define what 
    values they may take and where these values are obtained. It should be noted that they are the same ones 
    as defined in the DICOM Standard Part 3 (Information Object Definitions). 
    4.5.1 Common Patient Entity Modules 
    4.5.1.1 Patient Module 
    This section specifies the Attributes of the Patient that describe and identify the Patient who is the 
    subject of a diagnostic Study. This Module contains Attributes of the patient that are needed for 
    diagnostic interpretation of the Image and are common for all studies performed on the patient.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     34TABLE 4.5-1 
     PATIENT MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Patients Name (0010,0010) 2 May be entered from User Interface.  
    Taken from worklist if it is there. Patient ID (0010,0020) 2 May be entered from User Interface.  
    Taken from worklist if it is there. Patients Birth Date (0010,0030) 2 May be entered from User Interface. 
    Taken from worklist if it is there. Patients Sex (0010,0040) 2 May be entered from User Interface. 
    Taken from worklist if it is there. Referenced Patient Sequence (0008,1120) 3 Taken from worklist if it is there >Referenced SOP Class UID (0008,1150) 1C Taken from worklist if it is there.  >Referenced SOP Instance UID (0008,1155) 1C Taken from worklist if it is there Patients Birth Time (0010,0032) 3 Taken from worklist if it is there. Other Patient Ids (0010,1000) 3 Taken from worklist if it is there. Other Patient Names (0010,1001) 3 Not used Ethnic Group (0010,2160) 3 Taken from worklist if it is there. Patient Comments (0010,4000) 3 Taken from worklist if it is there.  
    4.5.2 Common Study Entity Modules 
    The following Study IE Modules are common to all Composite Image IODs, which reference the Study 
    IE. These modules contain Attributes of the patient and study that are needed for diagnostic 
    interpretation of the image. 
    4.5.2.1  General Study Module 
    This section specifies the attributes that describe and identify the Study performed upon the Patient. 
    TABLE 4.5-2  
    GENERAL STUDY MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Study Instance UID (0020,000D) 1 Uniquely generated by the equipment.  
    Taken from worklist if it is there. Study Date (0008,0020) 2 Is set to examination date Study Time (0008,0030) 2 Is set to examination time Referring Physicians Name (0008,0090) 2 May be entered from User Interface. 
    Taken from worklist if it is there. Study ID (0020,0010) 2  
    Taken from worklist if it is there (from Requested 
    Procedure Id) Accession Number (0008,0050) 2 May be entered from User Interface. 
    Taken from worklist if it is there. Study Description (0008,1030) 3 Taken from worklist if it is there (from Requested 
    Procedure Description). Physician(s) of Record (0008,1048) 3 Taken from worklist if it is there (from Names of 
    Intended Recipients of Result)  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     35Name of Physician(s) Reading Study (0008,1060) 3 Not used Referenced Study Sequence (0008,1110) 3 Taken from worklist if it is there. (Not used in SR 
    Documents.) >Referenced SOP Class UID (0008,1150) 1C Taken from worklist if it is there. (Not used in SR 
    Documents.) >Referenced SOP Instance UID (0008,1155) 1C Taken from worklist if it is there. (Not used in SR 
    Documents.)  
    4.5.2.2 Patient Study Module 
    This section defines attributes that provide information about the Patient at the time that the Study was 
    performed. 
    TABLE 4.5-3  
    PATIENT STUDY MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Admitting Diagnoses Description (0008,1080) 3 Not used Patients Age (0010,1010) 3 Not used Patients Size (0010,1020) 3 May be entered from User Interface.  
    Taken from worklist if it is there. Patients Weight (0010,1030) 3 May be entered from User Interface.  
    Taken from worklist if it is there. Occupation (0010,2180) 3 Not used Additional Patient’s History (0010,21B0) 3 Taken from worklist if it is there.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     364.5.3  Common Series Entity Modules 
    The following Series IE Modules are common to all Composite Image IODs, which reference the Series 
    IE. 
    4.5.3.1 General Series Module 
    This section specifies the attributes that identify and describe general information about the Series 
    within a Study. 
    TABLE 4.5-4  
    GENERAL SERIES MODULE ATTRIBUTES 
    Attribute Name Tag Type  Attribute Description Modality (0008,0060) 1 Defined Term  “US” used. 
    When reading SC all modalities are accepted. Series Instance UID (0020,000E) 1 Uniquely generated by the equipment Series Number (0020,0011) 2 Internal number which is incremented for each new 
    exam within a study. Laterality (0020,0060) 2C Not used Series Date (0008,0021) 3 Is set to Series date Series Time (0008,0031) 3 Is set to Series time Performing Physicians’ Name (0008,1050) 3 May be entered from User Interface. 
    Taken from worklist if it is there (from Scheduled 
    Performing Physician’s Name) Operator’s Name (0008,1070) 3 May be entered from User Interface. Default is login 
    id.  Referenced Performed Procedure Step 
    Sequence (0008,1111) 3 Used if Modality Performed Procedure Step is 
    enabled. >Referenced SOP Class UID (0008,1150) 1C Used if Modality Performed Procedure Step is 
    enabled. >Referenced SOP Instance UID (0008,1155) 1C Used if Modality Performed Procedure Step is 
    enabled. Body Part Examined (0018,0015) 3 Not used Patient Position (0018,5100) 2C Not used Smallest Pixel Value in Series (0028,0108) 3 Not used Largest Pixel Value in Series (0028,0109) 3 Not used Request Attributes Sequence  (0040,0275) 3 Used if Modality Worklist and/or Modality Performed 
    Procedure Step is enabled. >Requested Procedure ID  (0040,1001) 1C Taken from worklist if it is there. >Scheduled Procedure Step ID  (0040,0009) 1C Taken from worklist if it is there. >Scheduled Procedure Step Description (0040,0007) 3 Taken from worklist if it is there. >Scheduled Protocol Code Sequence (0040,0008) 3 Taken from worklist if it is there. >>Include ‘Code Sequence Macro’    Performed Procedure Step ID  (0040,0253) 3 Used if Modality Performed Procedure Step is 
    enabled. Performed Procedure Step Start Date  (0040,0244) 3 Used if Modality Performed Procedure Step is 
    enabled. Performed Procedure Step Start Time  (0040,0245) 3 Used if Modality Performed Procedure Step is 
    enabled.  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     37Performed Procedure Step Description  (0040,0254) 3 Used if Modality Performed Procedure Step is 
    enabled. Performed Protocol Code Sequence (0040,0260) 3 Taken from worklist if it is there (from Scheduled 
    Protocol Code Sequence) >Include ‘Code Sequence Macro’    4.5.4 Common Equipment Entity Modules 
    The following Equipment IE Module is common to all Composite Image IODs, which reference the 
    Equipment IE. 
    4.5.4.1 General Equipment Module 
    This section specifies the attributes that identify and describe the piece of equipment, which produced a 
    Series of Images. 
    TABLE 4.5-5  
    GENERAL EQUIPMENT MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Manufacturer (0008,0070) 2 Is set to “GE Healthcare” Institution Name (0008,0080) 3 Is set to configured Institution Name. Institution Address (0008,0081) 3 Not used Station Name (0008,1010) 3 Is set to configured Station Name. Institutional Department Name (0008,1040) 3 May be entered from User Interface. 
    Default is configured Department name. Manufacturers Model Name (0008,1090) 3 Is set to “LOGIQ 7”. Device Serial Number (0018,1000) 3 Not used Software Versions (0018,1020) 3 Is set to LOGIQ 7 software version Spatial Resolution (0018,1050) 3 Not used Date of Last Calibration (0018,1200) 3 Not used Time of Last Calibration (0018,1201) 3 Not used Pixel Padding Value (0028,0120) 3 Not used  
    4.5.5 Common Image Entity Modules 
    The following Image IE Modules are common to all Composite Image IODs, which reference the Image 
    IE. 
    4.5.5.1 General Image Module 
    This section specifies the attributes that identify and describe an image within a particular series. 
    TABLE 4.5-6  
    GENERAL IMAGE MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Instance Number (0020,0013) 2 Internal value which is incremented for each 
    captured imagewithin a series(within a study), Patient Orientation (0020,0020) 2C Sent with empty value. Content Date (0008,0023) 2C Set from Image date Content Time (0008,0033) 2C Set from Image time  
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     38Attribute Name Tag Type Attribute Description Image Type (0008,0008) 3 The first two values contain 
    “ORIGINAL/PRIMARY” or 
    “DERIVED/ PRIMARY”. 
    Value is the description of the type of application 
    per the DICOM PS3.3-2001, section C.8.5.6.1.1 
    Value 4 is a description of the mode. 
    Values 5 and 6 may be used for private data. Acquisition Number (0020,0012) 3 Not used Acquisition Date (0008,0022) 3 Not used Acquisition Time (0008,0032) 3 Not used Referenced Image Sequence (0008,1140) 3 Not used >Referenced SOP Class UID (0008,1150) 1C Not used >Referenced SOP Instance UID (0008,1155) 1C Not used >Referenced Frame Number (0008,1160) 3 Not used Derivation Description (0008,2111) 3 May contain additional derivation information if 
    Image Type is DERIVED. Source Image Sequence (0008,2112) 3 Not used >Referenced SOP Class UID (0008,1150) 1C Not used >Referenced SOP Instance UID (0008,1155) 1C Not used >Referenced Frame Number (0008,1160) 3 Not used Images in Acquisition (0020,1002) 3 Not used Image Comments (0020,4000) 3 May be used for description of the image Quality Control Image (0028,0300) 3 Not used Burned In Annotation (0028,0301) 3 Not used Lossy Image Compression (0028,2110) 3 Set to 01 if images is lossy compressed. Lossy Image Compression Ratio (0028,2112) 3 Used if lossy compressed. 4.5.5.2 Image Pixel Module 
    This section specified the attributes that describe the pixel data of the image. 
    TABLE 4.5-7  
    IMAGE PIXEL MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Samples per Pixel (0028,0002) 1 Value of ‘1’ if Photometric Interpretation element value 
    has value 
     ‘MONOCHROME2’  
    Value of ‘3’ when Photometric Interpretation element 
    value has value  
     ‘RGB’, 
     YBR_FULL or  
      YBR_FULL_422 Photometric Interpretation (0028,0004) 1 Defined Values used: 
     “MONOCHROME2” , 
     “RGB”, 
     “YBR_FULL” or 
     “YBR_FULL_422”   
    						
    							 LOGIQ 7   
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     
     39Attribute Name Tag Type Attribute Description Rows (0028,0010) 1 Value depends on scanning mode and configuration 
    setup Columns (0028,0011) 1 Value depends on scanning mode and configuration 
    setup. Bits Allocated (0028,0100) 1 Value always = 0008H. Bits Stored (0028,0101) 1 Value always = 0008H. High Bit (0028,0102) 1 Value always = 0007H. Pixel Representation (0028,0103) 1 Defined Value ‘0’ - unsigned integer. Pixel Data (7FE0,0010) 1 Pixel Data of image. Planar Configuration (0028,0006) 1C Enumerated value 0000H, color-by-pixel, if 
    Photometric Interpretation element value has value 
    ‘RGB’ or if image is JPEG compressed. 
    Enumerated value 0001H, color-by-plane if image is 
    RLE compressed. Pixel Aspect Ratio (0028,0034) 1C Not used Smallest Image Pixel Value (0028,0106) 3 Not used Largest Image Pixel Value (0028,0107) 3 Not used Red Palette Color Lookup Table Descriptor  (0028,1101) 1C Only used when reading Palette images. Green Palette Color Lookup Table Descriptor  (0028,1102) 1C Only used when reading Palette images. Blue Palette Color Lookup Table Descriptor  (0028,1103) 1C Only used when reading Palette images. Red Palette Color Lookup Table Data (0028,1201) 1C Only used when reading Palette images. Green Palette Color Lookup Table Data (0028,1202) 1C Only used when reading Palette images. Blue Palette Color Lookup Table Data (0028,1203) 1C Only used when reading Palette images. 4.5.5.3 Contrast/Bolus Module  
    This section specifies the attributes that describe the contrast /bolus used in the acquisition of the Image. 
    TABLE 4.5-8  
    CONTRAST/BOLUS MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description Contrast/Bolus Agent (0018,0010) 2 May be entered from User Interface. Contrast/Bolus Agent Sequence (0018,0012) 3 Not used >Include ‘Code Sequence Macro’ 
        Contrast/Bolus Route (0018,1040) 3 Not used Contrast/Bolus Administration Route Sequence (0018,0014) 3 Not used >Include ‘Code Sequence Macro’    >Additional Drug Sequence (0018,002A) 3 Not used >>Include ‘Code Sequence Macro’     
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 7 User Manual