    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     140 12.3.3.2 Basic Color Image Box SOP Class 
    The Basic Color Image Box IOD is an abstraction of the presentation of an image and 
    image related data in the image area of a film.  The DIMSE services that are applicable to 
    the IOD are shown in Table 12.3.3-2. 
    TABLE 12.3.3-2 DIMSE SERVICE GROUP 
    DIMSE Service Element Usage 
    SCU Reference N-SET      M see 12.3.3.2.1 12.3.3.2.1 DIMSE Service Group (N-SET) 
    The N-SET DIMSE Service is used by LOGIQ 7 to update the Basic Color Image Box 
    SOP Instance. Table 12.5.2-5 defines the Basic Image Box Presentation Module 
    attributes used.  
    12.3.4 Printer SOP Class 
    The Printer IOD is an abstraction of the hard copy printer and is the basic Information 
    Entity to monitor the status of the printer. The DIMSE services that are applicable to the 
    IOD are shown in table 11.3.4-1. 
    12.3.4.1 DIMSE Service Group 
    TABLE 12.3.4-1 DIMSE SERVICE GROUP 
    DIMSE Service Element Usage 
    SCU Reference N-EVENT-REPORT      M see 11.3.4.1.1 N-GET      U see 12.3.4.1.2 12.3.4.1.1 N-EVENT_REPORT 
    LOGIQ 7 confirms the N-EVENT-REPORT initiated by the SCP (printer). 
    12.3.4.1.2 N-GET 
    Used by LOGIQ 7 to request the SCP to get a Printer SOP Instance.  Table 12.5.2-6 
    defines the Printer Module attributes. 
    12.4 PRINT MANAGEMENT IODS  
    Within an entity of a DICOM Print Management, attributes are grouped into a related set 
    of attributes.  A set of related attributes is termed a module.  A module facilitates the 
    understanding of the semantics concerning the attributes and how the attributes are 
    related with each other.  A module grouping does not infer any encoding of information 
    into datasets. 
    Table 12.4.1-1, Table 12.4.2-1, Table 12.4.3-1, and Table 12.4.4-1 identify the defined 
    modules within the entities which comprise the DICOM Print Management Service 
    IODs.  Modules are identified by Module Name. 
    See DICOM for a complete definition of the entities, modules and attributes.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     141 12.4.1 Film Session IOD Module  
    TABLE 12.4.1-1 FILM SESSION IOD MODULES 
    Module Name Reference Module Description SOP Common Module 12.5.2.1 Contains SOP Common information Basic Film Session Presentation 
    Module 12.5.2.1 Contains Film Session presentation 
    information Basic Film Session Relationship 
    Module 12.5.2.2 References to related SOPs  
    12.4.2 Basic Film Box IOD Module Table 
    TABLE 12.4.2-1 BASIC FILM BOX IOD MODULES 
    Module Name Reference SOP Common Module 12.5.1.1 Basic Film Box Presentation Module 12.5.2.3 Basic Film Box Relationship Module 12.5.2.2  
    12.4.3 Basic Image Box IOD Module Table 
    TABLE 12.4.3-1 BASIC IMAGE BOX IOD MODULES 
    Module Name Reference SOP Common Module 12.5.1.1 Image Box Pixel Presentation Module 12.5.2.5  
    12.4.4 Printer IOD Module Table 
    TABLE 12.4.4-1 PRINTER IOD MODULES 
    Module Name Reference SOP Common Module 12.5.1.1 Printer Module  12.5.2.6 12.5 INFORMATION MODULE DEFINITIONS 
    Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description 
    of each of the entities and modules that comprise the Print Management. 
    The following modules are included to convey Enumerated Values, Defined Terms, and 
    Optional Attributes supported.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     142 12.5.1 General Modules 
    12.5.1.1 SOP Common Module 
    This section defines the attributes that are required for proper functioning and 
    identification of the associated SOP Instances. They do not specify any semantics about 
    the Real-World Object represented by the IOD. 
    TABLE 12.5.1-1 SOP COMMON MODULE ATTRIBUTES 
    Attribute Name Tag Type Attribute Description SOP Class UID (0008,0016) 1 Varies with Module Instance and 
    DIMSE Service being used. 
    1.2.840.100011.5.1.1.1 (Film 
    Session) 
    1.2.840.100011.5.1.1.2 (Film 
    Box) 
    1.2.840.100011.5.1.1.4 (Image 
    Box) SOP Instance UID (0008,0018) 1 Provided by SCP (printer). Specific Character Set (0008,0005) 1C Not used as expanded or 
    replacement character sets not 
    used. Instance Creation Date (0008,0012) 3 Not used. Instance Creation Time (0008,0013) 3 Not used. Instance Creator UID (0008,0014) 3 Not used.  
    12.5.2 Print Management Modules 
    For all user configurable tags with no default, no value will be sent if the tag is not 
    configured. 
    12.5.2.1 Basic Film Session Presentation Module 
    This section defines the attributes that are common for all films of a film session. The 
    attributes described in table 12.5.2-1 apply when the N-CREATE DIMSE service is used.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     143 TABLE 12.5.2-1 BASIC FILM SESSION PRESENTATION MODULE ATTRIBUTES  
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Number of Copies (2000,0010) U Defined Terms used (user 
    configurable): 
    Default is 1. Max is 99. Print Priority (2000,0020) U Defined Terms used (user 
    configurable): 
    HIGH, MED, LOW. Default 
    is HIGH. Medium Type (2000,0030) U  Defined Terms used (user 
    configurable): 
          PAPER 
          BLUE FILM 
          CLEAR FILM 
    Default is CLEAR FILM. Film Destination (2000,0040) U  Defined Terms used (user 
    configurable): 
          MAGAZINE - default 
          PROCESSOR  Film Session Label (2000,0050) U User configurable.  No default. Memory Allocation (2000,0060) U Not Used  
    12.5.2.2 Basic Film Session Relationship Module 
    TABLE 12.5.2-2 BASIC FILM SESSION RELATIONSHIP MODULE ATTRIBUTES 
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Referenced Film Box Sequence (2000,0500) U Not used >Referenced SOP Class UID (0008,1150) U  >Referenced SOP Instance UID (0008,1155) U   
    12.5.2.3 Basic Film Box Presentation Module 
    The attributes described in table 12.5.2-3 apply when the N-CREATE DIMSE service is 
    used.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     144 TABLE 12.5.2-3 BASIC FILM BOX PRESENTATION MODULE ATTRIBUTES  
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Image Display Format (2010,0010) M Enumerated values used (user 
    configurable): 
          STANDARD\X,Y,  
          where X and Y can take 
           values from 1 to 5. 
    Default is STANDARD\1,1. Annotation Display Format ID (2010,0030) U Not used. Film Orientation (2010,0040) U Defined Terms used (user 
    configurable): 
          PORTRAIT - default 
          LANDSCAPE Film Size ID (2000,0050) U  Defined Terms used (user 
    configurable): 
         8INX10IN - default 
         10INX12IN 
         10INX14IN 
         11INX14IN 
         14INX14IN 
         14INX17IN 
         24CMX24CM 
         24CMX30CM Magnification Type (2010,0060) U Defined Terms Used (user 
    configurable): 
         REPLICATE  
         BILINEAR 
         CUBIC 
         NONE Smoothing Type (2010,0080) U Free form text entry field (user 
    configurable) and only sent if 
    Magnification Type is CUBIC. 
    No default Border Density (2010,0100) U Defined Terms Used (user 
    configurable): 
         BLACK 
         WHITE 
    Default is BLACK. Empty Image Density (2010,0110) U Defined Terms Used (user 
    configurable): 
         BLACK 
         WHITE 
    Default is WHITE. Min Density (2010,0120) U User configurable.  Defaults to 
    0. Max is 999. Max Density (2010,0130) U User configurable.  Defaults to 
    300. Max is 999.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     145 Trim (2010,0140) U Enumerated Values Used (user 
    configurable): 
         YES 
          NO 
    Default is NO. Configuration Information (2010,0150) U User configurable.   
    No default.  
    12.5.2.4 Basic Film Box Relationship Module 
    This section defines the attributes that describe the common parameters, which apply for 
    all images on a given sheet of film.  
    TABLE 12.5.2-4 BASIC FILM BOX RELATIONSHIP MODULE ATTRIBUTES 
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Referenced Film Session Sequence (2010,0500) M  >Referenced SOP Class UID (0008,1150) M 1.2.840.10008.5.1.1.1 >Referenced SOP Instance UID (0008,1155) M Provided by SCP (printer) Referenced Image Box Sequence (2010,0510) U Not used >Referenced SOP Class UID (0008,1150) U  >Referenced SOP Instance UID (0008,1155) U  Referenced Basic Annotation Sequence (2010,0520) U Not used >Referenced SOP Class UID (0008,1150) U  >Referenced SOP Instance UID (0008,1155) U   
    12.5.2.5 Image Box Pixel Presentation Module 
    The attributes described in table 12.5.2-5 apply when the DIMSE Service N-SET is used.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     146 TABLE 12.5.2-5 IMAGE BOX PIXEL PRESENTATION MODULE ATTRIBUTES 
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Image Position (2020,0010) M Based on the image display 
    format. Polarity (2020,0020) U Defined term, NORMAL Requested Image Size (2020,0030) U Not sent Basic Grayscale Image Sequence (2020,0110) M  >Samples Per Pixel (0028,0002) M Value = ‘1’  >Photometric Interpretation (0028,0004) M Defined Term 
    MONOCHROME2 used >Rows (0028,0010) M Value depends on scanning 
    mode and configuration setup.  >Columns (0028,0011) M Value depends on scanning 
    mode and configuration setup.  >Pixel Aspect Ratio (0028,0034) MC Not used >Bits Allocated (0028,0100) M Value always = 0008H >Bits Stored (0028,0101) M Value always = 0008H >High Bit (0028,0102) M Value always = 0007H >Pixel Representation (0028,0103) M Defined Value ‘0’ - unsigned 
    integer >Pixel Data (7FE0,0010) M  Basic Color Image Sequence (2020,0111) M  >Samples Per Pixel (0028,0002) M Value = ‘3’  >Photometric Interpretation (0028,0004) M Defined Term RGB used >Rows (0028,0010) M Value depends on scanning 
    mode and configuration setup.  >Columns (0028,0011) M Value depends on scanning 
    mode and configuration setup.  >Pixel Aspect Ratio (0028,0034) MC Not used >Bits Allocated (0028,0100) M Value always = 0008H >Bits Stored (0028,0101) M Value always = 0008H >High Bit (0028,0102) M Value always = 0007H >Pixel Representation (0028,0103) M Defined Value ‘0’ - unsigned 
    integer >Pixel Data (7FE0,0010) M  Planar Configuration (0028, 0006) M 0001H, color-by-plane, when 
    Basic Color Image Sequence is 
    set  
    12.5.2.6 Printer Module 
    This section defines the attributes that are used to monitor the status of the printer. The 
    attributes described in Table 12.5.2-6 apply when the DIMSE Service N-GET is used.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     147 TABLE 12.5.2-6 PRINTER MODULE ATTRIBUTES 
    Attribute Name Tag USAGE 
    (SCU) Attribute Description Printer Status (2110,0010) U Used to check the status of the 
    printer before a print operation is 
    started.  
    If the status is different from 0x0, 
    Success the print operation is 
    aborted, a message is displayed 
    and the print files reside in the 
    print buffer. Printer Status Info (2110,0020) U If return status is FAILURE an 
    error message is displayed, and 
    the print files resides in the print 
    buffer. Printer Name (2110,0030) U Requested, but not used Manufacturer (0008,0070) U Requested, but not used Manufacturer Model Name (0008,1090) U Requested, but not used Device Serial Number (0018,1000) U Requested, but not used Software Versions (0018,1020) U Requested, but not used Date Last Calibration (0018,1200) U Requested, but not used Last Calibration (0018,1201) U Requested, but not used  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     148 13. STUDY ROOT QUERY/RETRIEVE 
    INFORMATION MODEL DEFINITION 
    13.1 INTRODUCTION 
    This section specifies the use of the DICOM Study Root Query/Retrieve Model used to 
    organize data and against which a Query/Retrieve will be performed. The contents of this 
    section are: 
    13.2 - Information Model Description 
    13.3 - Information Model Entity-Relationship Model 
    13.4 - Information Model Keys 
    13.2 STUDY ROOT INFORMATION MODEL DESCRIPTION 
    This section defines the implementation of Study Root Query/Retrieve Information 
    Model. 
    13.3 STUDY ROOT INFORMATION MODEL ENTITY-RELATIONSHIP MODEL 
    The Entity-Relationship diagram for the Study Root Information Model schema is shown 
    in Illustration 13.3-1. In this figure, the following diagrammatic convention is established 
    to represent the information organization : 
    · each entity is represented by a rectangular box 
    · each relationship is represented by a diamond shaped box. 
    · the fact that a relationship exists between two entities is depicted 
    by lines connecting the corresponding entity boxes to the 
    relationship boxes. 
    The relationships are fully defined with the maximum number of possible entities in the 
    relationship shown. In other words, the relationship between Series and Image can have 
    up to n Images per Series.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     149 ILLUSTRATION 13.3-1 
    STUDY ROOT QUERY/RETRIEVE INFORMATION MODEL E/R DIAGRAM Study contains Image or Report contains  
    Series 
    0,n 1 1,n 1  
    13.3.1 Entity Descriptions 
    Please refer to DICOM Standard PS 3.4 (Service Class Specifications) for a description 
    of each of the levels contained within the Study Root Query/Retrieve Information Model. 
    13.3.2 LOGIQ 7 Mapping of DICOM entities 
    TABLE 13.3-1  
    MAPPING OF DICOM ENTITIES TO LOGIQ 7 ENTITIES 
    DICOM LOGIQ 7 Entity Study Exam Series Exam Image Image  
    13.4 INFORMATION MODEL KEYS 
    Please refer to DICOM Standard PS 3.4 (Service Class Specifications) for a description 
    of each of the levels contained within the Study Root Query/Retrieve Information Model. 
    The following Level descriptions are included to specify what data elements are 
    supported and what type of matching can be applied. It should be noted that they are the 
    same ones as defined in the DICOM Standard PS 3.4 (Service Class Specifications). 
    13.4.1 Supported Matching 
    Following are the types of matching that can be request by the implementation :  
    						