     130 Attribute Name Tag Req. Type 
    N-CREATE   Req. Type 
    N-SET   >>Referenced SOP Instance UID (0008,1155) 1C, supported Not allowed Performed Procedure Step Information Performed Procedure Step ID (0040,0253) 1 Not allowed Performed Station AE Title (0040,0241) 1 Not allowed Performed Station Name (0040,0242) 2, supported Not allowed Performed Location (0040,0243) 2, supported 
    (Institution 
    Name, 
    truncated if 
    necessary to 16 
    characters) Not allowed Performed Procedure Step Start Date (0040,0244) 1 Not allowed Performed Procedure Step Start Time (0040,0245) 1 Not allowed Performed Procedure Step Status (0040,0252) 1 3, supported 
     
     Performed Procedure Step Description (0040,0254) 2, supported 3, supported Performed Procedure Type Description (0040,0255) 2, always 
    empty 2, always empty Procedure Code Sequence  (0008,1032) 2, supported 3, supported >Include ‘Code Sequence Macro’    Performed Procedure Step End Date (0040,0250) 2, always 
    empty 3, supported Performed Procedure Step End Time (0040,0251) 2, always 
    empty 3, supported Image Acquisition Results Modality (0008,0060) 1 Not allowed Study ID (0020,0010) 2, supported Not allowed Performed Protocol Code Sequence (0040,0260) 2, supported 3, supported >Include ‘Code Sequence Macro’    Performed Series Sequence (0040,0340) 2, always 
    empty 3, supported >Performing Physicians Name (0008,1050) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present) >Protocol Name (0018,1030) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >Operators Name (0008,1070) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present)  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     131 Attribute Name Tag Req. Type 
    N-CREATE   Req. Type 
    N-SET   >Series Instance UID (0020,000E) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >Series Description (0008,103E) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present) >Retrieve AE Title (0008,0054) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present) >Referenced Image Sequence (0008,1140) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present) >>Referenced SOP Class UID (0008,1150) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >>Referenced SOP Instance UID (0008,1155) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >Referenced Non-Image Composite SOP 
    Instance Sequence (0040,0220) 2C 
    (Required if 
    Sequence Item 
    is present) 2C 
    (Required if Sequence Item 
    is present) >>Referenced SOP Class UID (0008,1150) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >>Referenced SOP Instance UID (0008,1155) 1C 
    (Required if 
    Sequence Item 
    is present) 1C 
    (Required if Sequence Item 
    is present) >Referenced Frame Number (0008,1160) 3, not supported 3, not supported  
     
    10.2.2 Operations 
    Action Information 
    Covered under IOD Description in Section 10.2.1 
    10.2.2.1 Service Class User Behavior 
    LOGIQ 7 sends N-CREATE when first image in examination is acquired or when the 
    exam is ended for the case where there are no images  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     132 LOGIQ 7 sends N-SET after the exam is ended. The N-SET will include all acquired 
    images’ UIDs and the status of COMPLETED or DISCONTINUED. 
    10.2.2.2 Status Codes 
    No Service Class specific status values are defined for the N-ACTION Service.  See PS 
    3.7 for general response status codes.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     133 11. STORAGE COMMITMENT PUSH MODEL SOP CLASS 
    DEFINITION 
    11.1 INTRODUCTION 
    This section of the DICOM Conformance Statement specifies the Storage Commitment 
    Push Model SOP Class, the optional attributes and service elements supported, the valid 
    range of values for mandatory and optional attributes, and the status code behavior.  
    11.2 STORAGE COMMITMENT PUSH MODEL SOP CLASS DEFINITION 
    11.2.1 IOD Description 
    11.2.1.1 STORAGE COMMITMENT MODULE 
    TABLE 11.2-1 
    STORAGE COMMITMENT MODULE 
    Attribute Name Tag Attribute Description Transaction UID (0008,1195) Uniquely generated by the equipment  Retrieve AE Title (0008,0054) Not used Storage Media File-Set ID (0088,0130) Not used Storage Media File-Set UID (0088,0140) Not used Referenced SOP Sequence (0008,1199) Supported >Referenced SOP Class UID (0008,1150) Supported >Referenced SOP Instance 
    UID (0008,1155) Supported >Retrieve AE Title (0008,0054) Not used >Storage Media File-Set ID (0088,0130) Not used >Storage Media File-Set UID (0088,0140) Not used Failed SOP Sequence (0008,1198) Supported >Referenced SOP Class UID (0008,1150) Supported >Referenced SOP Instance 
    UID (0008,1155) Supported >Failure Reason (0008,1197) Supported 11.2.2 DIMSE Service Group 
    DIMSE Service Element Usage SCU/SCP N-EVENT-REPORT M/M N-ACTION M/M  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     134 11.2.3 Operations 
    11.2.3.1 Action Information 
    Covered under IOD Description in Table 11.2.1 
     
    11.2.3.2 Service Class User Behavior 
    LOGIQ 7 sends the N-ACTION primitive (Storage Commitment Request) after 
    successful exam save to a DICOM Storage SCP.  
    LOGIQ 7 may request storage commitment for all generated SOP Class UIDs: 
    Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3 Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Ultrasound Image Storage (Retired)  1.2.840.10008.5.1.4.1.1.6 Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 The association for the N-ACTION is disconnected after processing the response. Thus, 
    the N-EVENT-REPORT must be sent on a separate association. 
    The Referenced Study Component Sequence Attribute is not supported. 
    The Transaction UID is valid for two days. If no answer is received, the request will be 
    removed without warning the user. 
    The optional Storage Media File-Set ID & UID Attributes in the N-ACTION are not 
    supported. 
    On receipt of an unsuccessful N-ACTION Response Status Code from the SCP, the 
    request will be put in a holding queue for the user to manually retry the request. 
    11.2.3.3 Status Codes 
    11.2.4 No Service Class specific status values are defined for the N-ACTION Service.  See PS 3.7 
    for general response status codes.Notifications 
    LOGIQ 7 will only listen for an N-EVENT-REPORT from the SCP in a new association 
    on the listen port for Verification and Storage Commitment.   
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     135 11.2.4.1 Event Information 
    TABLE 11.2-2 
    STORAGE COMMITMENT RESULT - EVENT INFORMATION 
    Event Type 
    Name Event 
    Type 
    ID Attribute Tag Requirement Type SCU/SCP Storage 
    Commitment 
    Request 
    Successful 1 Transaction UID (0008,1195) -/1   Retrieve AE Title (0008,0054) Not used   Storage Media File-Set 
    ID (0088,0130) Not used   Storage Media File-Set 
    UID (0088,0140) Not used   Referenced SOP 
    Sequence (0008,1199) -/1   >Referenced SOP Class 
    UID (0008,1150) -/1   >Referenced SOP 
    Instance UID (0008,1155) -/1   >Retrieve AE Title (0008,0054) Not used   >Storage Media File-Set 
    ID (0088,0130) Not used   >Storage Media File-Set 
    UID (0088,0140) Not used Storage 
    Commitment 
    Request 
    Complete - 
    Failures Exist 2 Transaction UID (0008,1195) -/1   Retrieve AE Title (0008,0054) Not used   Storage Media File-Set 
    ID (0088,0130) Not used   Storage Media File-Set 
    UID (0088,0140) Not used   Referenced SOP 
    Sequence (0008,1199) -/1C   >Referenced SOP Class 
    UID (0008,1150) -/1   >Referenced SOP 
    Instance UID (0008,1155) -/1   >Retrieve AE Title (0008,0054) Not used   >Storage Media File-Set 
    ID (0088,0130) Not used   >Storage Media File-Set 
    UID (0088,0140) Not used  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     136   Failed SOP Sequence (0008,1198) -/1   >Referenced SOP Class 
    UID (0008,1150) -/1   >Referenced SOP 
    Instance UID (0008,1155) -/1   >Failure Reason (0008,1197) -/1 11.2.4.2 Service Class User Behavior 
    If a successful answer is received, the request will be removed without warning the user. 
    If a non-successful answer is received, the request will be left in the holding queue.  
    If no answer is received, the request will be removed without warning the user after two 
    days.  
    11.2.4.3 Status Codes 
    No Service Class specific status values are defined for the N-EVENT-REPORT Service.  
    See PS 3.7 for general response status code.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     137 12. PRINT MANAGEMENT SOP CLASS DEFINITION 
    12.1 INTRODUCTION 
    This section of the DICOM Conformance Statement specifies the supported Print 
    Management SOP and Meta SOP Classes, the optional attributes and service elements 
    supported, the valid range of values for mandatory and optional attributes, and the status 
    code behavior.  
    12.2 - Basic Print Management Meta SOP Classes 
    12.3 - Print Management SOP Class Definitions 
    12.4 - Print Management IODs 
    12.5 - IOD Module Definition 
    12.2 BASIC PRINT MANAGEMENT META SOP CLASSES 
    The Basic Print Management Meta SOP Classes correspond with the minimum 
    functionality that an implementation of the Print Management Service Class shall 
    support. 
    LOGIQ 7 supports the Basic Grayscale Print Management Meta SOP Class and the Basic 
    Color Print Management Meta SOP Class.  These are defined in Table 12.2.1-1 and Table 
    12.2.2-1. 
    12.2.1 Basic Grayscale Print Management Meta SOP Class 
    The Basic Grayscale Print Management Meta SOP Class is defined by the following set 
    of supported SOP Classes. 
    TABLE 12.2.1-1 BASIC GRAYSCALE PRINT MANAGEMENT META SOP CLASS 
    SOP Class Name Usage SCU Reference Basic Film Session SOP Class M 
    see 12.3.1 Basic Film Box SOP Class M see 12.3.2 Basic Grayscale Image Box SOP 
    Class M see 12.3.3.1 Printer SOP Class M see 12.3.4 12.2.2 Basic Color Print Management Meta SOP Class 
    The Basic Color Print Management Meta SOP Class is defined by the following set of 
    supported SOP Classes  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     138 TABLE 12.2.2-1 BASIC COLOR PRINT MANAGEMENT META SOP CLASS 
    SOP Class Name Usage 
    SCU Reference Basic Film Session SOP Class      M see 12.3.1 Basic Film Box SOP Class      M see  12.3.2 Basic Color Image Box SOP Class      M see 12.3.3.2 Printer SOP Class      M see 12.3.4 12.3 PRINT MANAGEMENT SOP CLASS DEFINITIONS 
    12.3.1 Basic Film Session SOP Class 
    The Basic Film Session IOD describes the presentation parameters, which are common 
    for all the films of a film session.  The DIMSE services that are applicable to the IOD are 
    shown in Table 12.3.1-1. 
    TABLE 12.3.1-1 DIMSE SERVICE GROUP 
    DIMSE Service Element Usage 
    SCU Reference N-CREATE      M see 12.3.1.1.1 N-SET      U see 12.3.1.1.2 N-DELETE      U see 12.3.1.1.3 N-ACTION      U see 12.3.1.1.4 12.3.1.1 DIMSE Service Group 
    12.3.1.1.1 N-CREATE  
    The N-CREATE DIMSE Service is used by LOGIQ 7 to request that the SCP (printer) 
    create a Film Session SOP Instance.  Table 12.4.2-1 defines the Basic Film Session 
    Presentation Module attributes used in this request.  
    12.3.1.1.2 N-SET 
    Not used in this implementation. 
    12.3.1.1.3 N-DELETE 
    Not used in this implementation. 
    12.3.1.1.4 N-ACTION 
    Not used in this implementation 
    12.3.2 Basic Film Box SOP Class 
    The Basic Film Box IOD is an abstraction of the presentation of one film of the film 
    session.  The DIMSE services that are applicable to the IOD are shown in Table 12.3.2-1.  
    						
    							 LOGIQ7 
    GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
    DIRECTION 2316173-100REV 7.03 
     139 TABLE 12.3.2-1 DIMSE SERVICE GROUP 
    DIMSE Service Element Usage 
    SCU Reference N-CREATE      M see 12.3.2.1.1 N-ACTION      M see 12.3.2.1.2 N-DELETE      U see 12.3.2.1.3 N-SET      U see 12.3.2.1.4 12.3.2.1 DIMSE Service Group 
    12.3.2.1.1 N-CREATE 
    The N-CREATE DIMSE Service is used by LOGIQ 7 to request that the SCP create a 
    Film Box SOP Instance.  Table 12.4.2-1 defines the Basic Film Box Presentation Module 
    attributes used in this request.  
    12.3.2.1.2 N-ACTION 
    The N-ACTION DIMSE Service is used by LOGIQ 7 to request the SCP (printer) to 
    print the number of copies configured by the user to a film of the film session. 
    12.3.2.1.3 N-DELETE 
    The N-DELETE DIMSE Service is used by LOGIQ 7 to request the SCP (printer) to 
    delete the complete Film Box.  The root Film Box Instance UID is sent to the SCP to 
    accomplish this. 
    12.3.2.1.4 N-SET 
    Not used in this implementation. 
    12.3.3 Image Box SOP Class 
    12.3.3.1 Basic Grayscale Image Box  SOP Class 
    The Basic Grayscale Image Box IOD is an abstraction of the presentation of an image 
    and image related data in the image area of a film.  The DIMSE services that are 
    applicable to the IOD are shown in Table 12.3.3-1. 
    TABLE 12.3.3-1 DIMSE SERVICE GROUP 
    DIMSE Service Element Usage 
    SCU Reference N-SET      M see 12.3.3.1.1 12.3.3.1.1 DIMSE Service Group (N-SET) 
    The N-SET DIMSE Service is used by LOGIQ 7 to update the Basic Grayscale Image 
    Box SOP Instance. Table 12.5.2-5 defines the Basic Image Box Presentation Module 