GE Logiq 7 User Manual
Have a look at the manual GE Logiq 7 User Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 130 Attribute Name Tag Req. Type N-CREATE Req. Type N-SET >>Referenced SOP Instance UID (0008,1155) 1C, supported Not allowed Performed Procedure Step Information Performed Procedure Step ID (0040,0253) 1 Not allowed Performed Station AE Title (0040,0241) 1 Not allowed Performed Station Name (0040,0242) 2, supported Not allowed Performed Location (0040,0243) 2, supported (Institution Name, truncated if necessary to 16 characters) Not allowed Performed Procedure Step Start Date (0040,0244) 1 Not allowed Performed Procedure Step Start Time (0040,0245) 1 Not allowed Performed Procedure Step Status (0040,0252) 1 3, supported Performed Procedure Step Description (0040,0254) 2, supported 3, supported Performed Procedure Type Description (0040,0255) 2, always empty 2, always empty Procedure Code Sequence (0008,1032) 2, supported 3, supported >Include ‘Code Sequence Macro’ Performed Procedure Step End Date (0040,0250) 2, always empty 3, supported Performed Procedure Step End Time (0040,0251) 2, always empty 3, supported Image Acquisition Results Modality (0008,0060) 1 Not allowed Study ID (0020,0010) 2, supported Not allowed Performed Protocol Code Sequence (0040,0260) 2, supported 3, supported >Include ‘Code Sequence Macro’ Performed Series Sequence (0040,0340) 2, always empty 3, supported >Performing Physicians Name (0008,1050) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present) >Protocol Name (0018,1030) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >Operators Name (0008,1070) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present)
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 131 Attribute Name Tag Req. Type N-CREATE Req. Type N-SET >Series Instance UID (0020,000E) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >Series Description (0008,103E) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present) >Retrieve AE Title (0008,0054) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present) >Referenced Image Sequence (0008,1140) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present) >>Referenced SOP Class UID (0008,1150) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >>Referenced SOP Instance UID (0008,1155) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >Referenced Non-Image Composite SOP Instance Sequence (0040,0220) 2C (Required if Sequence Item is present) 2C (Required if Sequence Item is present) >>Referenced SOP Class UID (0008,1150) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >>Referenced SOP Instance UID (0008,1155) 1C (Required if Sequence Item is present) 1C (Required if Sequence Item is present) >Referenced Frame Number (0008,1160) 3, not supported 3, not supported 10.2.2 Operations Action Information Covered under IOD Description in Section 10.2.1 10.2.2.1 Service Class User Behavior LOGIQ 7 sends N-CREATE when first image in examination is acquired or when the exam is ended for the case where there are no images
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 132 LOGIQ 7 sends N-SET after the exam is ended. The N-SET will include all acquired images’ UIDs and the status of COMPLETED or DISCONTINUED. 10.2.2.2 Status Codes No Service Class specific status values are defined for the N-ACTION Service. See PS 3.7 for general response status codes.
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 133 11. STORAGE COMMITMENT PUSH MODEL SOP CLASS DEFINITION 11.1 INTRODUCTION This section of the DICOM Conformance Statement specifies the Storage Commitment Push Model SOP Class, the optional attributes and service elements supported, the valid range of values for mandatory and optional attributes, and the status code behavior. 11.2 STORAGE COMMITMENT PUSH MODEL SOP CLASS DEFINITION 11.2.1 IOD Description 11.2.1.1 STORAGE COMMITMENT MODULE TABLE 11.2-1 STORAGE COMMITMENT MODULE Attribute Name Tag Attribute Description Transaction UID (0008,1195) Uniquely generated by the equipment Retrieve AE Title (0008,0054) Not used Storage Media File-Set ID (0088,0130) Not used Storage Media File-Set UID (0088,0140) Not used Referenced SOP Sequence (0008,1199) Supported >Referenced SOP Class UID (0008,1150) Supported >Referenced SOP Instance UID (0008,1155) Supported >Retrieve AE Title (0008,0054) Not used >Storage Media File-Set ID (0088,0130) Not used >Storage Media File-Set UID (0088,0140) Not used Failed SOP Sequence (0008,1198) Supported >Referenced SOP Class UID (0008,1150) Supported >Referenced SOP Instance UID (0008,1155) Supported >Failure Reason (0008,1197) Supported 11.2.2 DIMSE Service Group DIMSE Service Element Usage SCU/SCP N-EVENT-REPORT M/M N-ACTION M/M
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 134 11.2.3 Operations 11.2.3.1 Action Information Covered under IOD Description in Table 11.2.1 11.2.3.2 Service Class User Behavior LOGIQ 7 sends the N-ACTION primitive (Storage Commitment Request) after successful exam save to a DICOM Storage SCP. LOGIQ 7 may request storage commitment for all generated SOP Class UIDs: Ultrasound Multi-frame Image Storage 1.2.840.10008.5.1.4.1.1.3.1 Ultrasound Multi-frame Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.3 Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 Ultrasound Image Storage (Retired) 1.2.840.10008.5.1.4.1.1.6 Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 The association for the N-ACTION is disconnected after processing the response. Thus, the N-EVENT-REPORT must be sent on a separate association. The Referenced Study Component Sequence Attribute is not supported. The Transaction UID is valid for two days. If no answer is received, the request will be removed without warning the user. The optional Storage Media File-Set ID & UID Attributes in the N-ACTION are not supported. On receipt of an unsuccessful N-ACTION Response Status Code from the SCP, the request will be put in a holding queue for the user to manually retry the request. 11.2.3.3 Status Codes 11.2.4 No Service Class specific status values are defined for the N-ACTION Service. See PS 3.7 for general response status codes.Notifications LOGIQ 7 will only listen for an N-EVENT-REPORT from the SCP in a new association on the listen port for Verification and Storage Commitment.
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 135 11.2.4.1 Event Information TABLE 11.2-2 STORAGE COMMITMENT RESULT - EVENT INFORMATION Event Type Name Event Type ID Attribute Tag Requirement Type SCU/SCP Storage Commitment Request Successful 1 Transaction UID (0008,1195) -/1 Retrieve AE Title (0008,0054) Not used Storage Media File-Set ID (0088,0130) Not used Storage Media File-Set UID (0088,0140) Not used Referenced SOP Sequence (0008,1199) -/1 >Referenced SOP Class UID (0008,1150) -/1 >Referenced SOP Instance UID (0008,1155) -/1 >Retrieve AE Title (0008,0054) Not used >Storage Media File-Set ID (0088,0130) Not used >Storage Media File-Set UID (0088,0140) Not used Storage Commitment Request Complete - Failures Exist 2 Transaction UID (0008,1195) -/1 Retrieve AE Title (0008,0054) Not used Storage Media File-Set ID (0088,0130) Not used Storage Media File-Set UID (0088,0140) Not used Referenced SOP Sequence (0008,1199) -/1C >Referenced SOP Class UID (0008,1150) -/1 >Referenced SOP Instance UID (0008,1155) -/1 >Retrieve AE Title (0008,0054) Not used >Storage Media File-Set ID (0088,0130) Not used >Storage Media File-Set UID (0088,0140) Not used
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 136 Failed SOP Sequence (0008,1198) -/1 >Referenced SOP Class UID (0008,1150) -/1 >Referenced SOP Instance UID (0008,1155) -/1 >Failure Reason (0008,1197) -/1 11.2.4.2 Service Class User Behavior If a successful answer is received, the request will be removed without warning the user. If a non-successful answer is received, the request will be left in the holding queue. If no answer is received, the request will be removed without warning the user after two days. 11.2.4.3 Status Codes No Service Class specific status values are defined for the N-EVENT-REPORT Service. See PS 3.7 for general response status code.
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 137 12. PRINT MANAGEMENT SOP CLASS DEFINITION 12.1 INTRODUCTION This section of the DICOM Conformance Statement specifies the supported Print Management SOP and Meta SOP Classes, the optional attributes and service elements supported, the valid range of values for mandatory and optional attributes, and the status code behavior. 12.2 - Basic Print Management Meta SOP Classes 12.3 - Print Management SOP Class Definitions 12.4 - Print Management IODs 12.5 - IOD Module Definition 12.2 BASIC PRINT MANAGEMENT META SOP CLASSES The Basic Print Management Meta SOP Classes correspond with the minimum functionality that an implementation of the Print Management Service Class shall support. LOGIQ 7 supports the Basic Grayscale Print Management Meta SOP Class and the Basic Color Print Management Meta SOP Class. These are defined in Table 12.2.1-1 and Table 12.2.2-1. 12.2.1 Basic Grayscale Print Management Meta SOP Class The Basic Grayscale Print Management Meta SOP Class is defined by the following set of supported SOP Classes. TABLE 12.2.1-1 BASIC GRAYSCALE PRINT MANAGEMENT META SOP CLASS SOP Class Name Usage SCU Reference Basic Film Session SOP Class M see 12.3.1 Basic Film Box SOP Class M see 12.3.2 Basic Grayscale Image Box SOP Class M see 12.3.3.1 Printer SOP Class M see 12.3.4 12.2.2 Basic Color Print Management Meta SOP Class The Basic Color Print Management Meta SOP Class is defined by the following set of supported SOP Classes
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 138 TABLE 12.2.2-1 BASIC COLOR PRINT MANAGEMENT META SOP CLASS SOP Class Name Usage SCU Reference Basic Film Session SOP Class M see 12.3.1 Basic Film Box SOP Class M see 12.3.2 Basic Color Image Box SOP Class M see 12.3.3.2 Printer SOP Class M see 12.3.4 12.3 PRINT MANAGEMENT SOP CLASS DEFINITIONS 12.3.1 Basic Film Session SOP Class The Basic Film Session IOD describes the presentation parameters, which are common for all the films of a film session. The DIMSE services that are applicable to the IOD are shown in Table 12.3.1-1. TABLE 12.3.1-1 DIMSE SERVICE GROUP DIMSE Service Element Usage SCU Reference N-CREATE M see 12.3.1.1.1 N-SET U see 12.3.1.1.2 N-DELETE U see 12.3.1.1.3 N-ACTION U see 12.3.1.1.4 12.3.1.1 DIMSE Service Group 12.3.1.1.1 N-CREATE The N-CREATE DIMSE Service is used by LOGIQ 7 to request that the SCP (printer) create a Film Session SOP Instance. Table 12.4.2-1 defines the Basic Film Session Presentation Module attributes used in this request. 12.3.1.1.2 N-SET Not used in this implementation. 12.3.1.1.3 N-DELETE Not used in this implementation. 12.3.1.1.4 N-ACTION Not used in this implementation 12.3.2 Basic Film Box SOP Class The Basic Film Box IOD is an abstraction of the presentation of one film of the film session. The DIMSE services that are applicable to the IOD are shown in Table 12.3.2-1.
LOGIQ7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 139 TABLE 12.3.2-1 DIMSE SERVICE GROUP DIMSE Service Element Usage SCU Reference N-CREATE M see 12.3.2.1.1 N-ACTION M see 12.3.2.1.2 N-DELETE U see 12.3.2.1.3 N-SET U see 12.3.2.1.4 12.3.2.1 DIMSE Service Group 12.3.2.1.1 N-CREATE The N-CREATE DIMSE Service is used by LOGIQ 7 to request that the SCP create a Film Box SOP Instance. Table 12.4.2-1 defines the Basic Film Box Presentation Module attributes used in this request. 12.3.2.1.2 N-ACTION The N-ACTION DIMSE Service is used by LOGIQ 7 to request the SCP (printer) to print the number of copies configured by the user to a film of the film session. 12.3.2.1.3 N-DELETE The N-DELETE DIMSE Service is used by LOGIQ 7 to request the SCP (printer) to delete the complete Film Box. The root Film Box Instance UID is sent to the SCP to accomplish this. 12.3.2.1.4 N-SET Not used in this implementation. 12.3.3 Image Box SOP Class 12.3.3.1 Basic Grayscale Image Box SOP Class The Basic Grayscale Image Box IOD is an abstraction of the presentation of an image and image related data in the image area of a film. The DIMSE services that are applicable to the IOD are shown in Table 12.3.3-1. TABLE 12.3.3-1 DIMSE SERVICE GROUP DIMSE Service Element Usage SCU Reference N-SET M see 12.3.3.1.1 12.3.3.1.1 DIMSE Service Group (N-SET) The N-SET DIMSE Service is used by LOGIQ 7 to update the Basic Grayscale Image Box SOP Instance. Table 12.5.2-5 defines the Basic Image Box Presentation Module attributes used.