Page 51
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 40Attribute Name Tag Type Attribute Description Contrast/Bolus Volume (0018,1041) 3 Not used Contrast/Bolus Start Time (0018,1042) 3 Not used Contrast/Bolus Stop Time (0018,1043) 3 Not used Contrast/Bolus Total Dose (0018,1044) 3 Not used Contrast Flow Rate(s) (0018,1046) 3 Not used Contrast Flow Duration(s) (0018,1047) 3 Not used Contrast/Bolus Ingredient (0018,1048) 3 Not used Contrast/Bolus Ingredient...
Page 52
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 41TABLE 4.5-10 VOI LUT MODULE ATTRIBUTES Attribute Name Tag Type Attribute Description VOI LUT Sequence (0028,3010) 3 Not used >LUT Descriptor (0028,3002) 3 Not used >LUT Explanation (0028,3003) 3 Not used >LUT Data (0028,3006) 3 Not used Window Center (0028,1050) 3 Value set to 127 if Photometric Interpretation has value MONOCHROME2. Window Width (0028,1051) 3 Value set to 256 if Photometric Interpretation...
Page 53
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 42TABLE 4.5-12 US REGION CALIBRATION MODULE ATTRIBUTES Attribute Name Tag Type Attribute Description Sequence of Ultrasound Regions (0018,6011) 1 Object contains 0 or 1 regions >Region Location Min x 0 (0018,6018) 1 Value is 0 >Region Location Min y 0 (0018,601A) 1 Value is 0 >Region Location Max x 1 (0018,601C) 1 Value is image width-1. >Region Location Max y 1 (0018,601E) 1 Value is image height-1 >Physical...
Page 54
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 43Attribute Name Tag Type Attribute Description >Doppler Sample Volume Y Position (0018,603A) 3 Not used >TM-Line Position x 0 (0018,603C) 3 Not used >TM-Line Position y 0 (0018,603E) 3 Not used >TM-Line Position x 1 (0018,6040) 3 Not used >TM-Line Position y 1 (0018,6042) 3 Not used
Page 55
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 444.5.7.2 US Image Module This section specifies the attributes that describe ultrasound images. TABLE 4.5-13 US IMAGE MODULE ATTRIBUTES Attribute Name Tag Type Attribute Description Samples Per Pixel (0028,0002) 1 Value of ‘1’ if Photometric Interpretation element value has value ‘MONOCHROME2’ Value of ‘3’ when Photometric Interpretation element value has value ‘RGB’ or YBR_FULL or YBR_FULL_422...
Page 56
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 45Attribute Name Tag Type Attribute Description Stage Number (0008,2122) 3 Number of stage, starting at one. Sent if image is acquired in a stress test. View Name (0008,2127) 3 Name of view of stress test. Sent if image is acquired in a stress test. The name is defined in the User Interface. View Number (0008,2128) 3 Number of view, starting at one. Sent if image is acquired in a stress test. Number of Event...
Page 57
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 46Attribute Name Tag Type Attribute Description Trigger Time (0018,1060) 3 Not used Nominal Interval (0018,1062) 3 Not used Beat Rejection Flag (0018,1080) 3 Not used Low R-R Value (0018,1081) 3 Not used High R-R Value (0018,1082) 3 Not used Heart Rate (0018,1088) 3 Set to heart rate Output Power (0018,5000) 3 Not used Transducer Data (0018,5010) 3 Not used Transducer Type (0018,6031) 3 Not used Focus Depth...
Page 58
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 47 5. ULTRASOUND MULTIFRAME (US MF) INFORMATION OBJECT IMPLEMENTATION 5.1 INTRODUCTION This section specifies the use of the DICOM US Multi-frame Image IOD to represent the information included in US images produced by this implementation. Corresponding attributes are conveyed using the module construct. The contents of this section are: 5.2 - IOD Implementation 5.3 - IOD Entity-Relationship Model 5.4 - IOD...
Page 59
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 48 ILLUSTRATION 5.3-1 US MULTIFRAME IMAGE ENTITY RELATIONSHIP DIAGRAM Patient Study is the subject of contains creates Equipment US MF Image contains Series 0,n 1 1 1,n 1,n 1 1,n 1 5.3.1 Entity Descriptions Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the entities contained within the US Multi-Frame Information Object. 5.3.2 LOGIQ 7 Mapping of DICOM...
Page 60
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 49 5.4 IOD MODULE TABLE Within an entity of the DICOM US Multi-Frame IOD, attributes are grouped into related set of attributes. A set of related attributes is termed a module. A module facilitates the understanding of the semantics concerning the attributes and how the attributes are related with each other. A module grouping does not infer any encoding of information into datasets. Table 5.4-1 identifies the...