GE Logiq 7 User Manual
Here you can view all the pages of manual GE Logiq 7 User Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 31
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 202.3.1.3.2.4 Transfer Syntax Selection Policies The selected transfer syntax is based on the proposed transfer syntax list. The priority order is JPEG Baseline Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian and Run Length Encoding, RLE. 2.3.1.3.3 Real-World Activity A (‘ Exam save’ Operation) 2.3.1.3.3.1 Associated Real-World Activity LOGIQ 7 will only listen for an...
Page 32
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 212.4.2.1 API Not applicable to this product. 2.5 EXTENSIONS / SPECIALIZATIONS / PRIVATIZATIONS If so configured, the product will send ultrasound raw data information in private data elements designated by the Private Creator element: Element Name Tag VR VM Description Private Creator 7FE1,00xx LO 1 GEMS_Ultrasound_MovieGroup_001 This means that all private tags starting with 7FE1,xx will belong to the...
Page 33
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 22· Disabling/enabling and setting constant values for query fields – This is a way to disable search criteria in the worklist query or setting them to a fixed value. · Maximum number of downloaded entries Storage: · The AE Title, IP address and port number of the SCP · Max retries, Retry interval, Timeout · Enable/disable raw data · Frame rate reduction · Enable/disable multi-frame · Compression...
Page 34
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 232.8 CODES AND CONTROLLED TERMINOLOGY 2.8.1 Fixed Coded Terminology The product uses the fixed (non-configurable, non-extensible) coded terminology in SR Document attributes, as described in Section 7 SR INFORMATION OBJECT IMPLEMENTATION. 2.9 SECURITY PROFILES The product does not conform to any defined DICOM Security Profiles. It is assumed that the product is used within a secured environment. It is...
Page 35
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 243. MEDIA STORAGE CONFORMANCE STATEMENT 3.1 INTRODUCTION This section of the conformance statement (CS) specifies the LOGIQ 7compliance to DICOM Media Interchange. It details the DICOM Media Storage Application Profiles and roles, which are supported by this product. LOGIQ 7 is able to export images to DICOM media, browse DICOM media or read images from DICOM media. 3.2 IMPLEMENTATION MODEL 3.2.1 Application...
Page 36
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 253.2.3 Sequencing Requirements None applicable 3.2.4 File Meta Information Options (See PS3.10) The File Meta-Information for this implementation is: File Meta-Information Version 1 LOGIQ 7 Implementation UID 1.2.840.113619.6.115 Implementation Version Name LOGIQ 7_4 Note: The Implementation Version Name and may change in the future without modification of this document. 3.3 AE SPECIFICATIONS 3.3.1 LOGIQ 7...
Page 37
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 26Following are the default value set in the File Meta Information for this AE Title: Source Application Entity Title LOGIQ 7-0 (User Configurable) 3.3.1.2 Real-World Activities for the LOGIQ 7 Application Entity 3.3.1.2.1 Real-World Activity “Exam save” “Exam save” saves a DICOM SOP instance to media and updates DICOM File Set. 3.3.1.2.1.1 Media Storage Application Profile for the Real-World Activity “Exam...
Page 38
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 273.3.1.2.2 Real-World Activity “DICOM Browse” DICOM Browse is activated when the user searches for an exam in Search screen. 3.3.1.2.3 Media Storage Application Profile for the Real-World Activity “DICOM Browse” For the list of Application Profiles that invoke this AE for Image Read Real-World Activity, see the Table in Section 3.2.1 “LOGIQ 7 AE Specification”. 3.3.1.2.3.1.1 Options Following are the SOP...
Page 39
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 28Explicit VR Little Endian 1.2.840.10008.1.2.1 Explicit VR Big Endian 1.2.840.10008.1.2.2 Implicit VR Little Endian 1.2.840.10008.1.2 Run Length Encoding, RLE 1.2.840.10008.1.2.5 JPEG Baseline 1.2.840.10008.1.2.4.50 Ultrasound Image Storage (retired) 1.2.840.10008.5.1.4.1.1.6 JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.70 Explicit VR Little Endian 1.2.840.10008.1.2.1 Explicit VR Big...
Page 40
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 29Explicit VR Little Endian 1.2.840.10008.1.2.1 Explicit VR Big Endian 1.2.840.10008.1.2.2 Implicit VR Little Endian 1.2.840.10008.1.2 Run Length Encoding, RLE 1.2.840.10008.1.2.5 JPEG Baseline 1.2.840.10008.1.2.4.50 Ultrasound Multi-frame Image Storage(retired) 1.2.840.10008.5.1.4.1.1.3 JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.70 Explicit VR Little Endian 1.2.840.10008.1.2.1...