Home > GE > Medical Equipment > GE Logiq 7 User Manual

GE Logiq 7 User Manual

Here you can view all the pages of manual GE Logiq 7 User Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.

Page 31

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 202.3.1.3.2.4 Transfer Syntax Selection Policies 
The selected transfer syntax is based on the proposed transfer syntax list. The priority order is JPEG 
Baseline  
Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian and Run Length Encoding, RLE. 
 
 
2.3.1.3.3 Real-World Activity A (‘ Exam save’ Operation) 
 
2.3.1.3.3.1 Associated Real-World Activity 
 
LOGIQ 7 will only listen for an...

Page 32

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 212.4.2.1 API 
Not applicable to this product. 
2.5 EXTENSIONS / SPECIALIZATIONS / PRIVATIZATIONS 
If so configured, the product will send ultrasound raw data information in private data elements 
designated by the Private Creator element: 
Element Name  Tag VR VM Description Private Creator 7FE1,00xx LO 1 
GEMS_Ultrasound_MovieGroup_001  
This means that all private tags starting with 7FE1,xx will belong to the...

Page 33

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 22· Disabling/enabling and setting constant values for query fields – This is a way to disable search 
criteria in the worklist query or setting them to a fixed value. 
· Maximum number of downloaded entries 
 
Storage: 
· The AE Title, IP address and port number of the SCP 
· Max retries, Retry interval, Timeout 
· Enable/disable raw data 
· Frame rate reduction 
· Enable/disable multi-frame 
· Compression...

Page 34

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 232.8 CODES AND CONTROLLED TERMINOLOGY  
2.8.1 Fixed Coded Terminology 
The product uses the fixed (non-configurable, non-extensible) coded terminology in SR 
Document attributes, as described in Section 7 SR INFORMATION OBJECT 
IMPLEMENTATION. 
2.9 SECURITY PROFILES 
The product does not conform to any defined DICOM Security Profiles.   
It is assumed that the product is used within a secured environment.  It is...

Page 35

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 243. MEDIA STORAGE CONFORMANCE STATEMENT 
3.1 INTRODUCTION 
This section of the conformance statement (CS) specifies the LOGIQ 7compliance to DICOM Media 
Interchange. It details the DICOM Media Storage Application Profiles and roles, which are supported by 
this product. 
LOGIQ 7 is able to export images to DICOM media, browse DICOM media or read images from 
DICOM media. 
3.2 IMPLEMENTATION MODEL 
3.2.1 Application...

Page 36

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 253.2.3 Sequencing Requirements 
None applicable 
3.2.4 File Meta Information Options (See PS3.10) 
The File Meta-Information for this implementation is: 
File Meta-Information Version 1 LOGIQ 7 Implementation UID 
1.2.840.113619.6.115 Implementation Version Name 
LOGIQ 7_4 Note: The Implementation Version Name and may change in the future without modification of this 
document. 
3.3 AE SPECIFICATIONS 
3.3.1 LOGIQ 7...

Page 37

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 26Following are the default value set in the File Meta Information for this AE Title: 
Source Application Entity Title 
LOGIQ 7-0 (User Configurable)  
3.3.1.2 Real-World Activities for the LOGIQ 7 Application Entity 
3.3.1.2.1 Real-World Activity “Exam save” 
“Exam save” saves a DICOM SOP instance to media and updates DICOM File Set.  
3.3.1.2.1.1 Media Storage Application Profile for the Real-World Activity “Exam...

Page 38

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 273.3.1.2.2 Real-World Activity “DICOM Browse” 
DICOM Browse is activated when the user searches for an exam in Search screen.  
3.3.1.2.3 Media Storage Application Profile for the Real-World Activity “DICOM Browse” 
For the list of Application Profiles that invoke this AE for Image Read Real-World Activity, see the Table in 
Section 3.2.1 “LOGIQ 7 AE Specification”. 
3.3.1.2.3.1.1 Options 
Following are the SOP...

Page 39

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 28Explicit VR Little 
Endian 1.2.840.10008.1.2.1 Explicit VR Big Endian  1.2.840.10008.1.2.2  Implicit VR Little 
Endian 1.2.840.10008.1.2 Run Length Encoding, 
RLE 1.2.840.10008.1.2.5 JPEG Baseline 1.2.840.10008.1.2.4.50 Ultrasound Image Storage 
(retired) 1.2.840.10008.5.1.4.1.1.6 
JPEG Lossless Non 
Hierarchical (Process 
14) 1.2.840.10008.1.2.4.70 Explicit VR Little 
Endian 1.2.840.10008.1.2.1 Explicit VR Big...

Page 40

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 29Explicit VR 
Little Endian 1.2.840.10008.1.2.1 Explicit VR Big 
Endian  1.2.840.10008.1.2.2  Implicit VR 
Little Endian 1.2.840.10008.1.2 Run Length 
Encoding, RLE 1.2.840.10008.1.2.5 JPEG Baseline 1.2.840.10008.1.2.4.50 Ultrasound Multi-frame 
Image Storage(retired) 1.2.840.10008.5.1.4.1.1.3 
JPEG Lossless 
Non 
Hierarchical 
(Process 14) 1.2.840.10008.1.2.4.70 Explicit VR 
Little Endian 1.2.840.10008.1.2.1...