GE Logiq 7 User Manual
Here you can view all the pages of manual GE Logiq 7 User Manual. The GE manuals for Medical Equipment are available online for free. You can easily download all the documents as PDF.
Page 21
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 10Application Context Name 1.2.840.10008.3.1.1.1 The Maximum Length PDU negotiation is included in all association establishment requests. The maximum length PDU for an association initiated by LOGIQ 7 is: Maximum Length PDU 32768 The SOP Class Extended Negotiation is not supported. The user information Items sent by this product are: · Maximum PDU Length · Implementation UID · Implementation Version Name...
Page 22
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 112.3.1.2.1 Real-World Activity A (‘Exam save’ Operation) 2.3.1.2.1.1 Associated Real-World Activity Upon a request by the operator (manual or automatic), images will be sent to a DICOM Storage SCP. 2.3.1.2.1.2 Proposed Presentation Context Tables The Proposed Presentation Context Table depends on compression according to the following table: Presentation Context Table – Proposed Abstract Syntax Transfer Syntax...
Page 23
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 12Storage (retired) RLE Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 Presentation Context Table: Compression set to JPEG Secondary Capture Image Storage 1.2.840.10008.5.1.4.1.1.7 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 SCU None Ultrasound Image Storage...
Page 24
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 13Associations will be released upon the receipt of each C-ECHO confirmation. In the event that the SCP does not respond for some reason, the operation will time out and LOGIQ 7 will close the association. 2.3.1.2.2.2 Proposed Presentation Context Table Presentation Context Table - Proposed Abstract Syntax Transfer Syntax Role Name UID Name List UID List Extended Negotiation Verification SOP Class...
Page 25
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 14Failed A900 Identifier does not match SOP Class Terminate the association and operation (0000, 0901) (0000, 0902) Cxxx Unable to process Terminate the association and operation (0000,0901) (0000,0902) Success 0000 Matching is complete – No final identifier is supplied None Pending FF00 Matches are continuing – Current Match is supplied and any Optional Keys were supported in the same manner...
Page 26
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 15LOGIQ 7 includes attributes in the Modality Performed Procedure StepN-SET as described in Section 10.2.1 The mapping from Worklist attributes is described in Section 9.5.LOGIQ 7 sends N-SET after the exam is ended. The N-SET will include all acquired images’ UIDs and the status of COMPLETED or DISCONTINUED. For this SOP class, all status codes with status Refused or Error are treated as failures and terminate...
Page 27
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 16Name UID Name List UID List Negotiation Study Root Query/Retrieve Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 SCU None Study Root Query/Retrieve Information Model - MOVE 1.2.840.10008.5.1.4.1.2.2.2 Explicit VR Little Endian Explicit VR Big Endian Implicit VR...
Page 28
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 17Patient Comments (0010,4000) S,*, U - Other Study Numbers (0020,1070) S,*, U - Number of Patient Related Studies (0020,1200) S,*,U - Number of Patient Related Series (0020,1202) S,*,U - Number of Patient Related Instances (0020,1204) S,*,U - Number of Study Related Series (0020,1206) S,*,U - Number of Study Related Instances (0020,1208) S,*,U - Interpretation Author (4008,010C) S,*,U - Series Level Modality...
Page 29
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 182.3.1.3 Association Acceptance Policy The AE accepts an association when it receives a Verification Request from another network device or a store request from a Q/R SCP or a Storage Commitment result from a Storage Commitment SCP. 2.3.1.3.1 Real-World Activity B – (‘Echo’ operation) 2.3.1.3.1.1 Associated Real-World Activity An incoming Verification Request will cause the AE to accept the association and...
Page 30
LOGIQ 7 GE HEALTHCARE DICOM CONFORMANCE STATEMENT DIRECTION 2316173-100REV 7.03 19Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline JPEG Lossless Non Hierarchical (Process 14) Explicit VR Little Endian Explicit VR Big Endian Implicit VR Little Endian Run Length Encoding, RLE 1.2.840.10008.1.2.4.50 1.2.840.10008.1.2.4.70 1.2.840.10008.1.2.1 1.2.840.10008.1.2.2 1.2.840.10008.1.2 1.2.840.10008.1.2.5 SCP None Ultrasound Multi-frame Image Storage...