Page 21

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 10Application Context Name 1.2.840.10008.3.1.1.1  
The Maximum Length PDU negotiation is included in all association establishment requests.  The 
maximum length PDU for an association initiated by LOGIQ 7 is: 
Maximum Length PDU 32768  
The SOP Class Extended Negotiation is not supported. 
The user information Items sent by this product are: 
· Maximum PDU Length 
· Implementation UID 
· Implementation Version Name...

Page 22

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 112.3.1.2.1 Real-World Activity A (‘Exam save’ Operation) 
2.3.1.2.1.1 Associated Real-World Activity 
Upon a request by the operator (manual or automatic), images will be sent to a DICOM Storage SCP. 
2.3.1.2.1.2 Proposed Presentation Context Tables  
The Proposed Presentation Context Table depends on compression according to the following table: 
Presentation Context Table – Proposed Abstract Syntax Transfer Syntax...

Page 23

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 12Storage (retired) RLE  
Explicit VR Little 
Endian Explicit VR Big 
Endian Implicit VR 
Little Endian  
1.2.840.10008.1.2.1 
1.2.840.10008.1.2.2 
1.2.840.10008.1.2  Presentation Context Table: Compression set to JPEG Secondary Capture Image 
Storage 1.2.840.10008.5.1.4.1.1.7 JPEG Baseline  
JPEG Lossless Non 
Hierarchical (Process 
14) 1.2.840.10008.1.2.4.50 
1.2.840.10008.1.2.4.70 SCU None Ultrasound Image Storage...

Page 24

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 13Associations will be released upon the receipt of each C-ECHO confirmation.   
In the event that the SCP does not respond for some reason, the operation will time out and LOGIQ 7 will close the 
association. 
2.3.1.2.2.2 Proposed Presentation Context Table 
Presentation Context Table - Proposed Abstract Syntax Transfer Syntax 
Role Name  UID Name List UID List  Extended 
Negotiation Verification SOP Class...

Page 25

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 14Failed A900 Identifier does not 
match SOP Class Terminate the 
association and 
operation (0000, 0901) 
(0000, 0902)  Cxxx Unable to process Terminate the 
association and 
operation (0000,0901) 
(0000,0902) Success 0000 Matching is 
complete – No 
final identifier is 
supplied  None Pending FF00 Matches are 
continuing – 
Current Match is 
supplied and any 
Optional Keys 
were supported in 
the same manner...

Page 26

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 15LOGIQ 7 includes attributes in the Modality Performed Procedure StepN-SET as described in Section 
10.2.1 
The mapping from Worklist attributes is described in Section 9.5.LOGIQ 7 sends N-SET after the exam 
is ended. The N-SET will include all acquired images’ UIDs and the status of COMPLETED or 
DISCONTINUED. 
For this SOP class, all status codes with status Refused or Error are treated as failures and terminate...

Page 27

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 16Name  UID Name List UID List  Negotiation Study Root Query/Retrieve 
Information Model - FIND 1.2.840.10008.5.1.4.1.2.2.1 Explicit VR Little 
Endian Explicit VR 
Big Endian Implicit 
VR Little Endian 1.2.840.10008.1.2.1 
1.2.840.10008.1.2.2 
1.2.840.10008.1.2 SCU None Study Root Query/Retrieve 
Information Model - MOVE 1.2.840.10008.5.1.4.1.2.2.2 Explicit VR Little 
Endian Explicit VR 
Big Endian Implicit 
VR...

Page 28

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 17Patient Comments (0010,4000) S,*, U - Other Study Numbers (0020,1070) S,*, U - Number of Patient Related 
Studies (0020,1200) S,*,U - Number of Patient Related Series (0020,1202) S,*,U - Number of Patient Related 
Instances (0020,1204) S,*,U - Number of Study Related Series (0020,1206) S,*,U - Number of Study Related 
Instances (0020,1208) S,*,U - Interpretation Author (4008,010C) S,*,U - Series Level    Modality...

Page 29

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 182.3.1.3 Association Acceptance Policy 
The AE accepts an association when it receives a Verification Request from another network device or a 
store request from a Q/R SCP or a Storage Commitment result from a Storage Commitment SCP. 
 
 2.3.1.3.1 Real-World Activity B – (‘Echo’ operation) 
2.3.1.3.1.1 Associated Real-World Activity 
An incoming Verification Request will cause the AE to accept the association and...

Page 30

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 19Ultrasound Image Storage 1.2.840.10008.5.1.4.1.1.6.1 JPEG Baseline  
JPEG Lossless Non 
Hierarchical (Process 
14) 
Explicit VR Little 
Endian Explicit VR Big 
Endian Implicit VR 
Little Endian 
Run Length Encoding, 
RLE 1.2.840.10008.1.2.4.50 
1.2.840.10008.1.2.4.70 
1.2.840.10008.1.2.1 
1.2.840.10008.1.2.2 
1.2.840.10008.1.2 
1.2.840.10008.1.2.5 
 SCP None Ultrasound Multi-frame 
Image Storage...