 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 50 5.5.1 Common Image Modules 
The following Image IE Modules are common to all Composite Image IODs which reference the Image 
IE. 
5.5.1.1 Cine Module 
TABLE 5.5-2  
 CINE MODULE ATTRIBUTES 
Attribute Name Tag Type Attribute Description Frame Time (0018,1063) 1C Is set to the interframe time 
 Frame Time Vector (0018,1065) 1C Supported. 
Reading: Average value is set to 
interframe time Start Trim (0008,2142) 3...

Page 62

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 51   

Page 63

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 526. SC INFORMATION OBJECT IMPLEMENTATION 
6.1 INTRODUCTION 
This section specifies the use of the DICOM SC Image IOD to represent the information 
included in SC images produced by this implementation. Corresponding attributes are 
conveyed using the module construct. The contents of this section are: 
6.2 - IOD Implementation 
6.3 - IOD Entity-Relationship Model 
6.4 - IOD Module Table 
6.5 - IOD Module Definition 
6.2...

Page 64

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 53ILLUSTRATION 6.3-1  
SC IMAGE ENTITY RELATIONSHIP DIAGRAM Patient Study is 
the subject of contains creates Equipment SC Image contains  
Series 
0,n 1 1 1,n 1,n 1 1,n 1  
6.3.1 Entity Descriptions 
Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description 
of each of the entities contained within the SC Information Object. 
6.3.2 LOGIQ 7 Mapping of DICOM Entities 
TABLE 6.3-1  
MAPPING OF...

Page 65

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 546.4 IOD MODULE TABLE 
Within an entity of the DICOM SC IOD, attributes are grouped into related set of 
attributes. A set of related attributes is termed a module. A module facilitates the 
understanding of the semantics concerning the attributes and how the attributes are 
related with each other. A module grouping does not infer any encoding of information 
into datasets. 
Table 6.4-1 identifies the defined modules...

Page 66

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 55TABLE 6.5-2  
SC EQUIPMENT MODULE ATTRIBUTES 
Attribute Name Tag Type Attribute Description Conversion Type (0008,0064) 1 Set to WSD Modality (0008,0060) 3 Defined Value “US” used. 
When reading SC all modalities 
are accepted Secondary Capture Device ID (0018,1010) 3 Defined Value “LOGIQ 7” Secondary Capture Device Manufacturer (0018,1016) 3 Implementation defined string 
“GE Healthcare” Secondary Capture Device...

Page 67

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 567. SR INFORMATION OBJECT IMPLEMENTATION 
7.1 INTRODUCTION 
This section specifies the use of the DICOM Comprehensive SR IOD to represent results 
produced by this implementation. Corresponding attributes are conveyed using the 
module construct. The contents of this section are: 
7.2- IOD Implementation 
7.3 - IOD Entity-Relationship Model 
7.4 - IOD Module Table 
7.5 - IOD Module Definition 
In this section, supported...

Page 68

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 57ILLUSTRATION 7.3-1  
COMPREHENSIVE SR ENTITY RELATIONSHIP DIAGRAM Patient Study is 
the subject of contains creates Equipment SR Document contains  
Series 
0,n 1 1 1,n 1,n 1 1,n 1  
7.3.1 Entity Descriptions 
Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description 
of each of the entities contained within the Comprehensive SR Information Object. 
7.3.2 LOGIQ 7 Mapping of DICOM Entities...

Page 69

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 587.4 IOD MODULE TABLE 
Within an entity of the DICOM Comprehensive SR IOD, attributes are grouped into 
related set of attributes. A set of related attributes is termed a module. A module 
facilitates the understanding of the semantics concerning the attributes and how the 
attributes are related with each other. A module grouping does not infer any encoding of 
information into data sets. 
Table 7.4-1 identifies the...

Page 70

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 597.5.1 SR Document Series Module 
TABLE 7.5-1 
SR DOCUMENT SERIES MODULE 
Attribute Name Tag Type  Attribute Description Modality (0008,0060) 1 Defined Term  “SR” used. Series Instance UID 
 (0020,000E) 1 Uniquely generated by the 
equipment Series Number (0020,0011) 1 Internal number which is 
incremented for each new exam 
within a study. Referenced Performed Procedure Step Sequence 
 (0008,1111) 2 Not used....