Page 41

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 30 
3.4 AUGMENTED AND PRIVATE APPLICATION PROFILES 
LOGIQ 7 creates Secondary Capture Image Objects in addition to the objects defined in the application 
profiles. 
3.5 EXTENSIONS, SPECIALIZATIONS, PRIVATIZATIONS OF SOP CLASSES AND TRANSFER 
SYNTAXES 
If so configured, the product will send ultrasound raw data information in private data elements 
designated by the Private Creator element: 
Element Name  Tag VR VM...

Page 42

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 314. ULTRASOUND (US) INFORMATION OBJECT IMPLEMENTATION 
4.1 INTRODUCTION 
This section specifies the use of the DICOM US Image IOD to represent the information included in US 
images produced by this implementation. Corresponding attributes are conveyed using the module 
construct. The contents of this section are: 
4.2 - IOD Implementation 
4.3 - IOD Entity-Relationship Model 
4.4 - IOD Module Table 
4.5 - IOD...

Page 43

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 32ILLUSTRATION 4.3-1  
US IMAGE ENTITY RELATIONSHIP DIAGRAM Patient Study is 
the subject of contains creates Equipment US Image contains  
Series 
0,n 1 1 1,n 1,n 1 1,n 1  
4.3.1 Entity Descriptions 
Please refer to DICOM Standard Part 3 (Information Object Definitions) for a description of each of the 
entities contained within the US Information Object. 
4.3.2 LOGIQ 7 Mapping of DICOM Entities 
TABLE 4.3-1...

Page 44

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 334.4 IOD MODULE TABLE 
Within an entity of the DICOM US IOD, attributes are grouped into related set of attributes. A set of 
related attributes is termed a module. A module facilitates the understanding of the semantics 
concerning the attributes and how the attributes are related with each other. A module grouping does not 
infer any encoding of information into data sets. 
Table 4.4-1 identifies the defined...

Page 45

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 34TABLE 4.5-1 
 PATIENT MODULE ATTRIBUTES 
Attribute Name Tag Type Attribute Description Patients Name (0010,0010) 2 May be entered from User Interface.  
Taken from worklist if it is there. Patient ID (0010,0020) 2 May be entered from User Interface.  
Taken from worklist if it is there. Patients Birth Date (0010,0030) 2 May be entered from User Interface. 
Taken from worklist if it is there. Patients Sex...

Page 46

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 35Name of Physician(s) Reading Study (0008,1060) 3 Not used Referenced Study Sequence (0008,1110) 3 Taken from worklist if it is there. (Not used in SR 
Documents.) >Referenced SOP Class UID (0008,1150) 1C Taken from worklist if it is there. (Not used in SR 
Documents.) >Referenced SOP Instance UID (0008,1155) 1C Taken from worklist if it is there. (Not used in SR 
Documents.)  
4.5.2.2 Patient Study Module 
This...

Page 47

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 364.5.3  Common Series Entity Modules 
The following Series IE Modules are common to all Composite Image IODs, which reference the Series 
IE. 
4.5.3.1 General Series Module 
This section specifies the attributes that identify and describe general information about the Series 
within a Study. 
TABLE 4.5-4  
GENERAL SERIES MODULE ATTRIBUTES 
Attribute Name Tag Type  Attribute Description Modality (0008,0060) 1 Defined...

Page 48

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 37Performed Procedure Step Description  (0040,0254) 3 Used if Modality Performed Procedure Step is 
enabled. Performed Protocol Code Sequence (0040,0260) 3 Taken from worklist if it is there (from Scheduled 
Protocol Code Sequence) >Include ‘Code Sequence Macro’    4.5.4 Common Equipment Entity Modules 
The following Equipment IE Module is common to all Composite Image IODs, which reference the 
Equipment IE....

Page 49

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 38Attribute Name Tag Type Attribute Description Image Type (0008,0008) 3 The first two values contain 
“ORIGINAL/PRIMARY” or 
“DERIVED/ PRIMARY”. 
Value is the description of the type of application 
per the DICOM PS3.3-2001, section C.8.5.6.1.1 
Value 4 is a description of the mode. 
Values 5 and 6 may be used for private data. Acquisition Number (0020,0012) 3 Not used Acquisition Date (0008,0022) 3 Not used...

Page 50

 LOGIQ 7   
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 
 39Attribute Name Tag Type Attribute Description Rows (0028,0010) 1 Value depends on scanning mode and configuration 
setup Columns (0028,0011) 1 Value depends on scanning mode and configuration 
setup. Bits Allocated (0028,0100) 1 Value always = 0008H. Bits Stored (0028,0101) 1 Value always = 0008H. High Bit (0028,0102) 1 Value always = 0007H. Pixel Representation (0028,0103) 1 Defined Value ‘0’ - unsigned integer....