Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-65 Direction 5122542-100 Rev. 2 Probe Name:7S Table 1-51: Transducer Mode 7S Operating Mode: B Mode Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 1.061.3 IECFDAUnits pra pr.3 (MPa) 2.76 PWo (mW)89.9 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 2.52 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.993.99 Dim of A aprtX(cm)1.0 Y(cm) 0.7 t dPD (µsec) 0.6 prr PRF (Hz) 4192.9 pr at max. Ipi pr@PIImax (MPa) 3.35 deq at max. Ipi deq@PIImax (cm) Focal Length FLX(cm)4.8 FL Y(cm)8.0 I pa, αat max. MI IPA . 3@MImax (W/cm2) 344.35 Frequency (MHz) T4T4 Image Depth (cm) 14.0 10.0 Other Conditions ARM_Rev_2.book Page 65 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-66LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Table 1-52: Transducer Mode 7S Operating Mode: M Mode (Inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 1.481.1 IECFDAUnits pra pr.3 (MPa) 2.32 PWo (mW)79.4 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 2.33 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.973.97 Dim of A aprtX(cm)1 Y(cm) 0.7 t dPD (µsec) 0.61 prr PRF (Hz) 4192.8 7 pr at max. Ipi pr@PIImax (MPa) 3.35 deq at max. Ipi deq@PIImax (cm) Focal Length FLX(cm) FLY(cm) Ipa,αat max. MI IPA . 3@MIma x (W/ cm2)362.72 Frequency (MHz) T4T4 Image Depth (cm) 14.0 14 ARM_Rev_2.book Page 66 Monday, March 15, 2010 2:19 PM
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-67 Direction 5122542-100 Rev. 2 Table 1-53: Transducer Mode 7S Operating Mode: Color Flow Mode (Inc. M-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 1.42.1 1.7 IECFDAUnits pra pr. 3 (MPa) 3.4 PWo (mW)44.9 105.0 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 2.262.26 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)2.84 ƒawffc(MHz) 3.463.46 3.46 Dim of A aprtX(cm)1.0 1 Y(cm) 0.7 0.7 t dPD (µsec) 095 prr PRF (Hz) 4251.7 pr at max. Ipi pr@PIImax (MPa) 2.5 deq at max. Ipi deq@PIImax (cm)2.84 Focal Length FLX(cm)3.5 FL Y(cm)8.0 I pa, αat max. MI IPA . 3@MImax (W/cm2) 171.65 Frequency (MHz) 3.3 Image Depth (cm)2.4 Vel Scale (kHz)4.3 Line Density Penet OFF ROI6010 PS 10 10 10 ARM_Rev_2.book Page 67 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-68LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Table 1-54: Transducer Mode: 7S Operating Mode: Pulsed Doppler (inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 1.32.1 IECFDAUnits pra pr.3 (MPa) 2.4 PWo (mW)8.244.9 min of [Pα(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 2.32.33 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.24 ƒawffc(MHz) 3.4 5.13.4 Dim of AaprtX(cm)0.31.8 Y(cm)0.30.7 tdPD (µsec) 1.17 prr PRF (Hz) 1736.1 1 pr at max. Ipi pr@PIImax (MPa) 3.13 deq at max. Ipi deq@PIImax (cm)0.24 Focal Length FLX(cm)0.8 FLY(cm)1.5 Ipa,αat max. MI IPA . 3@MImax (W/cm2) 384.41 Frequency (MHz) 3.3 5.03.3 Image Depth (cm) 3.5 4.0 Vel Scale (kHz)1.71.7 B Pause(on/off) SV 22 PenetOFFOFF ARM_Rev_2.book Page 68 Monday, March 15, 2010 2:19 PM
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-69 Direction 5122542-100 Rev. 2 Table 1-55: Transducer Mode:7S Operating Mode :CW Doppler Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.66 3.034.29 0.0 IECFDAUnits pra pr.3 (MPa) 3.1 PWo (mW)158.73158.7 3158.7 3 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 2.842.84 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.70 ƒawffc(MHz) 4.014.014.01 4.01 Dim of A aprtX(cm)0.930.93 0.93 Y(cm) 0.70.7 0.7 t dPD (µsec) prr PRF (Hz) pr at max. Ipi pr@PIImax (MPa) 1.16 deq at max. Ipi deq@PIImax (cm)0.70 Focal Length FLX(cm)5.765.76 FL Y(cm)3.503.50 I pa, αat max. MI IPA . 3@MImax (W/cm2) Frequency (MHz) D4.0D4.0D4.0 Image Depth (cm) Other conditions ARM_Rev_2.book Page 69 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-70LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 ARM_Rev_2.book Page 70 Monday, March 15, 2010 2:19 PM
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-1 Direction 5122542-100 Rev. 2 Chapter 2 Measurement Formulas Reference for measurement formula tables for the various applications. ARM_Rev_2.book Page 1 Monday, March 15, 2010 2:19 PM
Measurement Formulas 2-2LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Measurement Acronyms Cardiac Measurements The following table shows tha cardiac measurements available on the LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 When you make a measurement,you select the abbreviationfor the measurement on the Touch Panel. Most abbreviations are made using acronyms. The following table lists acronyms used for naming cardiac measurements. Table 2-1: Cardiac Measurements AbbreviationsDefinitionUnit %FS LV Fractional Shortening, 2D % %FS LV Fractional Shortening, M-mode % %IVS Thck IVS Fractional Shortening, 2D % %IVS Thck IVS Fractional Shortening, M-mode % %LVPW Thck LV Posterior Wall Fractional Shortening, 2D % %LVPW Thck LV Posterior Wall Fractional Shortening, M-mode % Ao Arch Diam Aortic Arch Diameter cm Ao asc Ascending Arch Diameter cm Ao Desc Diam Descending Aortic Diameter cm Ao Isthmus Aortic Isthmus cm Ao Root Diam Aortic Root Diameter cm Ao Root Diam Aortic Root Diameter, M-mode cm AR ERO PISA:Regurgitant Orifice Area cm 2 AR FlowPISA:Regurgitant Flow ml/s AR PHT AV Insuf. Pressure Half Time ms AP Rad PISA:Radius of Aliased Point cm AR RV PISA:Regurgitant Volume Flow ml AR Vel PISA:Aliased Velocity m/s ARM_Rev_2.book Page 2 Monday, March 15, 2010 2:19 PM
Measurement Acronyms LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-3 Direction 5122542-100 Rev. 2 AR Vmax Aortic Insuf. Peak Velocity m/s AR VTI Aortic Insuf. Velocity Time Integral cm ARed max PC Aortic Insuf. End-Diastole Pressure Gradient mm Hg ARed Vmax Aortic Insuf. End-Diastole Velocity m/s AV Acc Slope Aortic Valve Flow Acceleration m/s 2 AV Acc TimeAortic Valve Acceration Time ms AV AccT/ET AV Acceleration to Ejection Time Ratio AV CO Cardiac Output by AOrtic Flow Vmin AV Cusp Aortic Valve Cusp Separation, 2D cm AV Cusp Aortic Valve Cusp Separation, M-mode cm AV Dec Time Aortic Valve Decelation Time ms AV Diam Aortic Diameter, 2D cm AV max PG Aortic Valve Peak Pressure Gradient mm Hg AV mean PG Aortic Valve Mean Pressure Gradient mm Hg AV SV Stroke Volume by Aortic Flow ml AV Vmax Aortic Valve Peak Velocity m/s AV VTI Aortic Valve Velocity Time Integral cm AVA (Vmax) AV Area by Continuity Equation by Peak V cm 2 AVA (VTY)AV Area by Continuity Equation VTI cm2 AVA PlanimetryAortic Valve Area cm2 AVETAortic Valve Ejection Time ms AVET Aortic Valve Ejection Time, M-mode ms CO (A-L A2C) CO 2CH, Single Plane, Area-Length l/min CO (A-L A4C) CO 4CH, Single Plane, Area-Length l/min CO (Biplane) CO, Bi-Plane, MOD l/min CO (Bullet) CO, Bi-Plane, Bullet l/min CO (MOD A2C) CO 2CH, Single Plane, MOD (Simpson) l/min CO (MOD A4C) CO 4CH, Single Plane, 4CH, MOD (Simpson) l/min CO (Cube) Cardiac Output, 2D, Cubic l/min Table 2-1: Cardiac Measurements AbbreviationsDefinitionUnit ARM_Rev_2.book Page 3 Monday, March 15, 2010 2:19 PM
Measurement Formulas 2-4LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 CO (Cube)Cardiac Output, M-mode, Cubic l/min CO (Tech) Cardiac Output, 2D, Teicholtz l/min CO (Tech) Cardiac Outpu, M-mode, Teicholtz l/min D-E Excursion MV Anterior Leaflet Excursion cm D-E Excursion Mitral Valve D-E Slope cm EDV (Bullet) LV Volume, Diastolic, Bi-Plane, Bullet ml EDV (Cube) Left Ventricle Volume, Diastolic, 2D, Cubic ml EDV (Cube) Left Ventricle Volume, Diastolic, M-mode, Cubic ml EDV (Teich) Left Ventricle Volume, Diastolic, 2D, Teicholtz ml EDV (Teich) Left Ventricle Volume, Diastolic, M-mode, Teicholtz ml E-F (A-L A2C) Ejection Fraction, 2CH, Sigle Plane, Area-Length % EF (A-L A4C) Ejection Fraction, 4CH, Sigle Plane, Area-Length % EF (Biplane) Ejection Fraction, Bi-Plane, MOD % EF (Bullet) Ejection Fraction, 2CH, Bi-Plane, Bullet % EF (MOD A2C) Ejection Fraction, 2CH, Single Plane, MOD % EF (MOD A4C) Ejection Fraction, 4CH, Single Plane, 4CH, MOD (Simpson) % E-F Slope Mitral Valve E-F Slope m/s EF (Cube) Ejection Fraction, 2D, Cubic % EF (Cube) Ejection Fraction, M-mode, Cubic % EF (Teich) Ejection Fraction, 2D, Teicholtz % EF (Teich) Ejection Fraction, M-mode, Teicholtz % EPSS E-Point-toSeptum Separation, M-mode cm EPSS 2D E-Point-to-Septum Separation, 2D cm ERC Effective Regurgitant Orifice cm 2 ESV (Bullet)LV Volume, Systolic, Bi-Plane, Bullet ml ESV (Cube) Left Ventricle Volume, Systolic, 2D, Cubic ml ESV (Cube) Left Ventircle Volume, Systolic, M-mode, Cubic ml ESV (Teich) Left Ventricle Volume, Systolic, 2D, Teicholtz ml ESV (Teich) Left Ventricle Volume, Systolic, M-mode, Teicholtz ml HR AV Heart Rate, Dop bpm Table 2-1: Cardiac Measurements AbbreviationsDefinitionUnit ARM_Rev_2.book Page 4 Monday, March 15, 2010 2:19 PM