    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-15
    Direction 5122542-100 Rev. 2
    Suggested Guiedlines(cont’d)
    Methodology:
    Doppler studies on the umbilical  arteries are carried out after 
    manual palpation of the uterus  and fetus. In pregnancies before 
    28 weeks, the fetus cannot be  easily felt hencethe probe is 
    placed on the upper third of the ut erus at the midline. The probe 
    is gently rotated until the umbilic al signal is heard and clearly 
    identified on the screen. Ther e should be 3-4 clear dense 
    waveforms will not be equ al in intensity or 
    frequency;hence,measurements must be postponeduntil the 
    fetus is quiet. If there is ta chycardia or bradycardia, the 
    measurements have no clear meaning because now the 
    waveform represents a complex  mixure of resistance, cardiac 
    output, and time needed to empty a vascular region. The image 
    is frozen and the measurement are taken.We have found it 
    useful to continuouslyrecord the examination on audiotape 
    because the signal is immediat ely captured and not lost when 
    the fetus or cord moves.
    This process is repeated at least twice more and the results from 
    the three studies are averaged. Although the S/D ratio is ,in 
    theory, an angle-independent measurement, there may be 
    differences in the ratio when the artery is studied near the 
    abdomen or near the insertion in thye placenta.When three 
    different angle measurements are taken, we found an average 
    experimental error of 6% and a maximum of 16%.the uterine 
    artery signals may be more difficult to obtain and require more 
    training to develop expertise. The probe is directed in the lower 
    quadrants at the paracervical area where the uterine artery 
    enters the uterus. Three potent ial signals may be encountered 
    wave form is identi fied. The mature waveform is generally 
    present by 20 weeks,but should be achived no later than 26 
    weeks.The pulse wave frequency should coincide with the 
    meternal pulse and 4 equla and dense images should be seen 
    on each side of the uterus. The results from both sides are 
    averaged to give a single number.the S/D ratio appears to 
    provide a simple and reproducible measurement, but a number 
    of other calculations have also been proposed.Interoserver error 
    with this technique is 4% with a maximum error of 10%.
    ARM_Rev_2.book  Page 15  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-16LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    FDA Acoustic Output Data
    Maximum Output Summary
    The following tables list the typical maximum acoustic output 
    levels achievable with LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 
    for all probes and operational modes. It is intended that this is 
    information be useful in making ALARA decisions and selecting 
    the most appropriate probe for the application. In accordance 
    with US FDA Guidelines, the ov erall maximum acoustic SPTA 
    intensity for LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 is limited to 
    720 mW/cm
    2 and MI is limited to 1.9. Modes for which TI does 
    not exceed 0.1 are indicated by 
    						
    							FDA Acoustic Output Data
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-17
    Direction 5122542-100 Rev. 2
    Maximum Probe Temperature
    Table 1-3: Summary Table  (Continued)
    Operating Mode12L3S7S
    B-Mode XXX
    M-Mode (including
    B-Mode) XXX
    Color Flow XXX
    M Color Flow XXX
    Pulsed Doppler 
    (including
    B-Mode) XXX
    Continuous Wave (CW) 
    Doppler XX
    Table 1-4: Maximum Probe Temperature
    Probe
    Max 
    Te m p
    Probe
    Max 
    Te m p
    Probe
    Max 
    Te m p
    Probe
    Max 
    Te m p
    3C 29.64 E8C 28.11 3S 25.59
    3.5C 30.70 8L 37.95 7S 26.20
    5C 26.50 10L 40.89
    8C 41.17 10LB 28.64
    BE9C 40.97 12L 39.90
    Notes: 
    Lens temperature measured under foll owing conditions per IEC 60601-2-37:
    1. Thermocouple was placed at the geometric center of the lens.
    2. Probe placed upright in contac t with above thermal phantom.
    3. Autofreeze capability disabled.
    4. Lens temperature monitored for 30 min.
    5. Measurement uncertainty: 2 degC.
    *Thermal phantom made with tissue-mimicking mate rial as referenced in IEC60601-2-37: Appendix J.
    ARM_Rev_2.book  Page 17  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-18LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Acoustic Output
    Ta b l e  K e y
    Table 1-5:  Table Key
    FDAIECMeaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3
    aαAcoustic Attenuation Coefficient  / Derating factor (usually 0.3 dB/cm-MHz)
    A
    aprtAarpt-12db Output Beam Area  / Active aperture area
    C
    MINormalizing Coefficient
    D
    eqDeqEquivalent Aperture Diameter  / (same)
    d
    -6d-6Pulse Beam Width  / Beam diameter at –6 dB
    d
    eqdeqEquivalent Beam Diameter
    f
    cƒawfAcoustic Working Frequency  / Center frequency
    I
    paIpaPulse-Average Intensity
    I
    pa.3Ipa, αAttenuated Pulse-Average Intensity
    PII I
    piPulse-Intensity Integral
    PII
    .3Ipi, αAttenuated Pulse-Intensity Integral
    I
    TAIta( z ) Temporal-Average Intensity
    I
    TA . 3(Z) Ita,α(z ) Attenuated Temporal-Average Intensity  / (at depth z)
    I
    SPTA(Z) Izpta(z ) Spatial-Peak Temporal-Average Intensity
    I
    SPTA.3(Z) Izpta,α(z ) Attenuated Spatial-Peak Temporal-Average Intensity
    MI MIMechanical Index
    W
    oP Output Power / Time average acoustic power at the source
    W
    .3(Z) PαAttenuated Output Power  / Time average acoustic power derated to depth z
    W
    o1P1Bounded Output Power / Power emitted from the central 1cm of aperture
    PII p
    iPulse Pressure Squared Integral / Pulse intensity integral
    p
    rprPeak-Rarefactional Acoustic Pressure  / (same)
    ARM_Rev_2.book  Page 18  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-19
    Direction 5122542-100 Rev. 2
    pr.3pr αAttenuated Peak-Rarefactional Acoustic Pressure  / (same)
    PRF prrPulse Repetition Rate  / Pulse repetition frequency
    TI TIThermal Index / (same)
    TIB TIBBone Thermal Index  / (same)
    TIC TICCranial-Bone Thermal Index  / (same)
    TIS TISSoft-Tissue Thermal Index  / (same)
    PD t
    dPulse Duration / (same)
    x
    -12,y-12X, Y -12 dB Output Beam Dimensions / (same)
    Z ZDistance from the Source to a Specified Point  / (same)
    Z
    spZbDepth for TIB  / Depth at which the relevant index is maximum
    Z
    bpZbpBreak-Point Depth  / (same)
    Z
    spZsDepth for TIS  / Depth at which the relevant index is maximum
    Table 1-5:  Table Key
    FDAIECMeaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3
    ARM_Rev_2.book  Page 19  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-20LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Probe Name:3C
    Table 1-6:  Transducer Mode 3C Operating Mode: B Mode
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.9
    IECFDAUnits
    pra pr.3 (MPa) 1.5
    PWo (mW)
    min of [Pα(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 1.1
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)
    ƒawffc(MHz) 3.4 3.43.4
    Dim of A
    aprtX(cm)3.63.6
    Y(cm)
    1.031.03
    t
    dPD (µsec) 0.4
    prr PRF (Hz) 2245.7
    pr at max. Ipi   pr@PIImax (MPa) 1.7
    deq at max. Ipi deq@PIImax (cm)
    Focal Length FLX(cm)0.220.22
    FL
    Y(cm)0.90.9
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 62.62
    Frequency (MHz)
    3.0 3.03.0
    Image Depth (cm) 30.0 20.0
    20.0
    Other Conditions
    ARM_Rev_2.book  Page 20  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-21
    Direction 5122542-100 Rev. 2 Table 1-7:  Transducer Mode 3C
     Operating Mode: M Mode (Inc. B-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.850.7 0.3
    IECFDAUnits
    pra pr.3 (MPa) 1.51
    PWo (mW)31.0 21.0
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 1.43.1
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)0.3
    ƒawffc(MHz) 3.4 3.43.4 3.4
    Dim of A
    aprtX(cm)3.61.4 3.6
    Y(cm)
    1.031.03 1.03
    t
    dPD (µsec) 0.38
    prr PRF (Hz) 2245.7
    pr at max. Ipi   pr@PIImax (MPa) 1.8
    deq at max. Ipi deq@PIImax (cm)0.3
    Focal Length FLX(cm)1.70.17
    FL
    Y(cm)0.890.89
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 80
    Frequency (MHz)
    3.0 3.03.0 3.0
    Image Depth (cm) 30.0 20.0
    24.0 20.0
    ARM_Rev_2.book  Page 21  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-22LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Table 1-8:  Transducer Mode 3C Operating Mode: Color Flow Mode (Inc. M-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.84 3.22.9 4.6
    IECFDAUnits
    pra pr. 3 (MPa) 1.55
    PWo (mW)270.1179.7 270.1
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 3.963.3
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)0.4
    ƒawffc(MHz) 2.5 3.253.2 2.5
    Dim of A
    aprtX(cm)3.61.4 3.6
    Y(cm)
    1.031.03 1.03
    t
    dPD (µsec) 0.84
    prr PRF (Hz) 1698.7
    pr at max. Ipi   pr@PIImax (MPa) 2.36
    deq at max. Ipi deq@PIImax (cm)0.4
    Focal Length FLX(cm)0.170.17
    FL
    Y(cm)0.890.89
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 207.8
    Frequency (MHz)
    3.3 2.53.3 2.5
    Image Depth (cm) 3.5 4.8
    3.5 3.5
    Vel Scale (kHz) 1.9 2.45
    1.07 1.07
    Line Density
    Penet OFF ON
    OFF OFF
    ROI 60 10
    PS 14 1414 14
    ARM_Rev_2.book  Page 22  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-23
    Direction 5122542-100 Rev. 2 Table 1-9:  Transducer Mode: 
    3C Operating Mode: Pulsed Doppler (inc. B-Mode)
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.7 0.50.9 0.6
    IECFDAUnits
    pra pr.3 (MPa) 1.0
    PWo (mW)42.679.7 42.6
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 5.087.7
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)1.0
    ƒawffc(MHz) 3.3 3.33.5 3.3
    Dim of A
    aprtX(cm)3.63.5 3.6
    Y(cm)
    1.031.03 1.03
    t
    dPD (µsec) 1.13
    prr PRF (Hz) 1736.1
    1
    pr at max. Ipi   pr@PIImax (MPa) 1.75
    deq at max. Ipi deq@PIImax (cm)1.0
    Focal Length FLX(cm)0.170.17
    FL
    Y(cm)0.890.89
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 93.5
    Frequency (MHz)
    3.3 3.33.3 3.3
    Image Depth (cm) 3.2 21.0
    7.7 7.7
    Vel Scale (kHz) 1.7
    5.84 5.84
    SV 2
    5.84 5.84
    B Pause (on/off)
    Penet OFF-OFF OFF
    ARM_Rev_2.book  Page 23  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-24LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Probe Name:3.5C
    Table 1-10:  Transducer Mode __3.5C____ Operating Mode: B Mode
    Index LabelMI
    TISTIB
    TIC
    scannon-scannon-
    scan
    Aaprt1
    Global Maximum: Index Value 0.52.1
    IECFDAUnits
    pra pr.3 (MPa) 1.54
    PWo (mW)147.0
    min of [P
    α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
    zs z1(cm)
    zbp zbp(cm)
    zb zsp(cm) 1.0
    z  at max. I pi,α zsp(cm)
    deq(zb) deq(zsp) (cm)
    ƒawffc(MHz) 3.042.8
    Dim of A
    aprtX(cm)1.8
    Y(cm)
    1.3
    t
    dPD (µsec) 0.41
    prr PRF (Hz) 2240.6
    5
    pr at max. Ipi   pr@PIImax (MPa) 31.71
    deq at max. Ipi deq@PIImax (cm)
    Focal Length FLX(cm)6.4
    FL
    Y(cm)6.6
    I
    pa, αat max. MI IPA . 3@MImax 
    (W/cm2) 70.55
    Frequency (MHz)
    3.33.3
    Image Depth (cm) 30.0
    21.0
    other conditions
    ARM_Rev_2.book  Page 24  Monday, March 15, 2010  2:19 PM 
    						