Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-15 Direction 5122542-100 Rev. 2 Suggested Guiedlines(cont’d) Methodology: Doppler studies on the umbilical arteries are carried out after manual palpation of the uterus and fetus. In pregnancies before 28 weeks, the fetus cannot be easily felt hencethe probe is placed on the upper third of the ut erus at the midline. The probe is gently rotated until the umbilic al signal is heard and clearly identified on the screen. Ther e should be 3-4 clear dense waveforms will not be equ al in intensity or frequency;hence,measurements must be postponeduntil the fetus is quiet. If there is ta chycardia or bradycardia, the measurements have no clear meaning because now the waveform represents a complex mixure of resistance, cardiac output, and time needed to empty a vascular region. The image is frozen and the measurement are taken.We have found it useful to continuouslyrecord the examination on audiotape because the signal is immediat ely captured and not lost when the fetus or cord moves. This process is repeated at least twice more and the results from the three studies are averaged. Although the S/D ratio is ,in theory, an angle-independent measurement, there may be differences in the ratio when the artery is studied near the abdomen or near the insertion in thye placenta.When three different angle measurements are taken, we found an average experimental error of 6% and a maximum of 16%.the uterine artery signals may be more difficult to obtain and require more training to develop expertise. The probe is directed in the lower quadrants at the paracervical area where the uterine artery enters the uterus. Three potent ial signals may be encountered wave form is identi fied. The mature waveform is generally present by 20 weeks,but should be achived no later than 26 weeks.The pulse wave frequency should coincide with the meternal pulse and 4 equla and dense images should be seen on each side of the uterus. The results from both sides are averaged to give a single number.the S/D ratio appears to provide a simple and reproducible measurement, but a number of other calculations have also been proposed.Interoserver error with this technique is 4% with a maximum error of 10%. ARM_Rev_2.book Page 15 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-16LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 FDA Acoustic Output Data Maximum Output Summary The following tables list the typical maximum acoustic output levels achievable with LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 for all probes and operational modes. It is intended that this is information be useful in making ALARA decisions and selecting the most appropriate probe for the application. In accordance with US FDA Guidelines, the ov erall maximum acoustic SPTA intensity for LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 is limited to 720 mW/cm 2 and MI is limited to 1.9. Modes for which TI does not exceed 0.1 are indicated by
FDA Acoustic Output Data LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-17 Direction 5122542-100 Rev. 2 Maximum Probe Temperature Table 1-3: Summary Table (Continued) Operating Mode12L3S7S B-Mode XXX M-Mode (including B-Mode) XXX Color Flow XXX M Color Flow XXX Pulsed Doppler (including B-Mode) XXX Continuous Wave (CW) Doppler XX Table 1-4: Maximum Probe Temperature Probe Max Te m p Probe Max Te m p Probe Max Te m p Probe Max Te m p 3C 29.64 E8C 28.11 3S 25.59 3.5C 30.70 8L 37.95 7S 26.20 5C 26.50 10L 40.89 8C 41.17 10LB 28.64 BE9C 40.97 12L 39.90 Notes: Lens temperature measured under foll owing conditions per IEC 60601-2-37: 1. Thermocouple was placed at the geometric center of the lens. 2. Probe placed upright in contac t with above thermal phantom. 3. Autofreeze capability disabled. 4. Lens temperature monitored for 30 min. 5. Measurement uncertainty: 2 degC. *Thermal phantom made with tissue-mimicking mate rial as referenced in IEC60601-2-37: Appendix J. ARM_Rev_2.book Page 17 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-18LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Acoustic Output Ta b l e K e y Table 1-5: Table Key FDAIECMeaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3 aαAcoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz) A aprtAarpt-12db Output Beam Area / Active aperture area C MINormalizing Coefficient D eqDeqEquivalent Aperture Diameter / (same) d -6d-6Pulse Beam Width / Beam diameter at –6 dB d eqdeqEquivalent Beam Diameter f cƒawfAcoustic Working Frequency / Center frequency I paIpaPulse-Average Intensity I pa.3Ipa, αAttenuated Pulse-Average Intensity PII I piPulse-Intensity Integral PII .3Ipi, αAttenuated Pulse-Intensity Integral I TAIta( z ) Temporal-Average Intensity I TA . 3(Z) Ita,α(z ) Attenuated Temporal-Average Intensity / (at depth z) I SPTA(Z) Izpta(z ) Spatial-Peak Temporal-Average Intensity I SPTA.3(Z) Izpta,α(z ) Attenuated Spatial-Peak Temporal-Average Intensity MI MIMechanical Index W oP Output Power / Time average acoustic power at the source W .3(Z) PαAttenuated Output Power / Time average acoustic power derated to depth z W o1P1Bounded Output Power / Power emitted from the central 1cm of aperture PII p iPulse Pressure Squared Integral / Pulse intensity integral p rprPeak-Rarefactional Acoustic Pressure / (same) ARM_Rev_2.book Page 18 Monday, March 15, 2010 2:19 PM
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-19 Direction 5122542-100 Rev. 2 pr.3pr αAttenuated Peak-Rarefactional Acoustic Pressure / (same) PRF prrPulse Repetition Rate / Pulse repetition frequency TI TIThermal Index / (same) TIB TIBBone Thermal Index / (same) TIC TICCranial-Bone Thermal Index / (same) TIS TISSoft-Tissue Thermal Index / (same) PD t dPulse Duration / (same) x -12,y-12X, Y -12 dB Output Beam Dimensions / (same) Z ZDistance from the Source to a Specified Point / (same) Z spZbDepth for TIB / Depth at which the relevant index is maximum Z bpZbpBreak-Point Depth / (same) Z spZsDepth for TIS / Depth at which the relevant index is maximum Table 1-5: Table Key FDAIECMeaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3 ARM_Rev_2.book Page 19 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-20LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Probe Name:3C Table 1-6: Transducer Mode 3C Operating Mode: B Mode Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.9 IECFDAUnits pra pr.3 (MPa) 1.5 PWo (mW) min of [Pα(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.1 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.4 3.43.4 Dim of A aprtX(cm)3.63.6 Y(cm) 1.031.03 t dPD (µsec) 0.4 prr PRF (Hz) 2245.7 pr at max. Ipi pr@PIImax (MPa) 1.7 deq at max. Ipi deq@PIImax (cm) Focal Length FLX(cm)0.220.22 FL Y(cm)0.90.9 I pa, αat max. MI IPA . 3@MImax (W/cm2) 62.62 Frequency (MHz) 3.0 3.03.0 Image Depth (cm) 30.0 20.0 20.0 Other Conditions ARM_Rev_2.book Page 20 Monday, March 15, 2010 2:19 PM
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-21 Direction 5122542-100 Rev. 2 Table 1-7: Transducer Mode 3C Operating Mode: M Mode (Inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.850.7 0.3 IECFDAUnits pra pr.3 (MPa) 1.51 PWo (mW)31.0 21.0 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.43.1 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.3 ƒawffc(MHz) 3.4 3.43.4 3.4 Dim of A aprtX(cm)3.61.4 3.6 Y(cm) 1.031.03 1.03 t dPD (µsec) 0.38 prr PRF (Hz) 2245.7 pr at max. Ipi pr@PIImax (MPa) 1.8 deq at max. Ipi deq@PIImax (cm)0.3 Focal Length FLX(cm)1.70.17 FL Y(cm)0.890.89 I pa, αat max. MI IPA . 3@MImax (W/cm2) 80 Frequency (MHz) 3.0 3.03.0 3.0 Image Depth (cm) 30.0 20.0 24.0 20.0 ARM_Rev_2.book Page 21 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-22LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Table 1-8: Transducer Mode 3C Operating Mode: Color Flow Mode (Inc. M-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.84 3.22.9 4.6 IECFDAUnits pra pr. 3 (MPa) 1.55 PWo (mW)270.1179.7 270.1 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 3.963.3 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.4 ƒawffc(MHz) 2.5 3.253.2 2.5 Dim of A aprtX(cm)3.61.4 3.6 Y(cm) 1.031.03 1.03 t dPD (µsec) 0.84 prr PRF (Hz) 1698.7 pr at max. Ipi pr@PIImax (MPa) 2.36 deq at max. Ipi deq@PIImax (cm)0.4 Focal Length FLX(cm)0.170.17 FL Y(cm)0.890.89 I pa, αat max. MI IPA . 3@MImax (W/cm2) 207.8 Frequency (MHz) 3.3 2.53.3 2.5 Image Depth (cm) 3.5 4.8 3.5 3.5 Vel Scale (kHz) 1.9 2.45 1.07 1.07 Line Density Penet OFF ON OFF OFF ROI 60 10 PS 14 1414 14 ARM_Rev_2.book Page 22 Monday, March 15, 2010 2:19 PM
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-23 Direction 5122542-100 Rev. 2 Table 1-9: Transducer Mode: 3C Operating Mode: Pulsed Doppler (inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.7 0.50.9 0.6 IECFDAUnits pra pr.3 (MPa) 1.0 PWo (mW)42.679.7 42.6 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 5.087.7 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)1.0 ƒawffc(MHz) 3.3 3.33.5 3.3 Dim of A aprtX(cm)3.63.5 3.6 Y(cm) 1.031.03 1.03 t dPD (µsec) 1.13 prr PRF (Hz) 1736.1 1 pr at max. Ipi pr@PIImax (MPa) 1.75 deq at max. Ipi deq@PIImax (cm)1.0 Focal Length FLX(cm)0.170.17 FL Y(cm)0.890.89 I pa, αat max. MI IPA . 3@MImax (W/cm2) 93.5 Frequency (MHz) 3.3 3.33.3 3.3 Image Depth (cm) 3.2 21.0 7.7 7.7 Vel Scale (kHz) 1.7 5.84 5.84 SV 2 5.84 5.84 B Pause (on/off) Penet OFF-OFF OFF ARM_Rev_2.book Page 23 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-24LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Probe Name:3.5C Table 1-10: Transducer Mode __3.5C____ Operating Mode: B Mode Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.52.1 IECFDAUnits pra pr.3 (MPa) 1.54 PWo (mW)147.0 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.0 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.042.8 Dim of A aprtX(cm)1.8 Y(cm) 1.3 t dPD (µsec) 0.41 prr PRF (Hz) 2240.6 5 pr at max. Ipi pr@PIImax (MPa) 31.71 deq at max. Ipi deq@PIImax (cm) Focal Length FLX(cm)6.4 FL Y(cm)6.6 I pa, αat max. MI IPA . 3@MImax (W/cm2) 70.55 Frequency (MHz) 3.33.3 Image Depth (cm) 30.0 21.0 other conditions ARM_Rev_2.book Page 24 Monday, March 15, 2010 2:19 PM