Home > GE > Ultrasound System > GE Logiq 3 Instruction Manual

GE Logiq 3 Instruction Manual

    Download as PDF Print this page Share this page

    Have a look at the manual GE Logiq 3 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.

    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-5
    Direction 5122542-100 Rev. 2
    Mechanical Bioeffect
    In a similar manner ,the interact ion of ultrasound energy with 
    tissue can produce a number of non-thermal or mechanical 
    effects.The most significant is cavitation which results from the 
    action of the oscillating ultrasou nd pressure on tiny gas bubbles 
    with in the tissues. Cavitation has caused mechanical damage 
    on a cellular level such as microscopic tears and hemorrhage in 
    laboratory tests with small animals.t he major contributing factors 
    can again be categorised as either tissue characteristics or 
    control parameters:
    • The physical characterist ics of tissue such as the 
    presence and size of micr oscopic gas bubbles and the 
    sensitivity of the tissue to th e effects of cavitation will 
    influence the potential for and magnitude of cavitation.
    • Acoustic field parameters like output frequency,peak  pulse amplitude s and perhaps pulse length are the 
    primary parameters effecting the onset of 
    cavitation.these are controlled by the operator through 
    appropriate equipment selections.
    Althogh it is generally accepted that no harmful biological effects 
    have been demonstrated at the frequency ,intensity,and 
    exposure times used in diognostic examinations,research in to 
    the potential for harmful effects continues.The operator is 
    encouraged to survey the literature for future developments on 
    bioeffects and to become familiar with the references at the end 
    of this section.
    ARM_Rev_2.book  Page 5  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-6LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Operator Awareness and Actions to Minimize bioeffect 
    The operator must be aware of the particular conditions that 
    exist during the examination to recognize the potential for 
    bioeffect and then take appropriate action to reduce the risk.The 
    recognization of potential harm comes from an understanding of 
    tissue characteristics and a re al-time knowledge of acoustic 
    output.Taking appropriate action requires familiarity with 
    equipment operation  and examination skills  like implementing 
    alternative techniques for obtaining the same diognostic 
    information.
    Tissue Characteristics
    Tissue Characteristics vary  considerably throughout the 
    body.they influence the acoustic  field and determine the heating 
    / cooling rates and cavitation potential.Ultrasound energy 
    dissipates as it passes throug h the tissue causing the deeper 
    tissue to encounter much lowe r levels.Some tissues like bone 
    readiy convert ultrasound energy to heat,while others like blood 
    and amniotic fluid pass the  energy on to adjucent tissue 
    relatively unattenuated.
    A perticular situation that represents a tissue combination 
    requiring extra precaution is a third trimester transabdominal 
    fetal examination where there is a very thin abdominal wall and 
    a long fluid path.The relative lack  of attenuating tissue along the 
    acoustic path will significantly  increase the available energy in 
    the fetal tissue.Additionally,fetal tissues are more susceptible to 
    long term injury due to nature of developing tissue.focussing the 
    ultra sound beam on or near fetal bone further increases the 
    risk.
    Otherthan fetal tissue,there is increased susceptability for 
    heating in any tissue that cannot easily conduct or distribute 
    heat due to low blood perfusion.As the examination 
    progresses,the operator must be aware of changing tissue 
    conditions.
    ARM_Rev_2.book  Page 6  Monday, March 15, 2010  2:19 PM 
    						
    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-7
    Direction 5122542-100 Rev. 2
    Acoustic Output
    awareness of the acoustic output level can be a difficult task for 
    the operator,espcially when the objective is to obtaine a quality 
    image.Older ultrasound equipment had limited means,if any,for 
    indicating the acoustic output le vel.In most cases ,the operator 
    had to be familiar with the output intensities as described in the 
    operator manual.To improve op erator awareness of acoustic 
    output,this system incorporates an  output display that directly 
    indicates the potential for mechanical and thermal bioeffects as 
    equipment controls are adjusted.
    The output display consists of four numeric index values that 
    indicate the potential for producing bioeffects(three indices are 
    for heating effect and one for cavitation).As the user changes 
    equipment settings that alter the acoustic output,the output 
    display indices are immediately updated to reflect the changes 
    in potential for producing .The indices are based on 
    mathematical models and each is normalized so that the 
    potential for bioeffect becomes more significant as the indices 
    reach a value 1.0 or larger.
    A mechanical index (MI) provides an indication of the potential 
    for the possible onset of transient cavitation within tissue while 
    the three thermal indices provide an indication of the potential 
    for heat generation within tissue.The different thermal indices 
    may be used depending on the type of tissue being examined:
    • Soft Tissue Thermal Index (TIS) is used as an indicator ofthe potential to generate heat within soft tissues.this is 
    the most used thermal indicator.
    • bone Thermal Index (TIB) is used as an indicator of the  potential to generate heat at the beam focus when 
    focussing on or near bone that is adjucent to very 
    sensitive tissue.this index  is intended as a thermal 
    indicator for second and third trimester fetal examination 
    or transfontanelle neonatal cephalic exams.
    • Cranial Bone Thermal Index (TIC) is used as an  indicator of the potential to generate heat in the near-
    field when the beam passes through bone at the surface 
    as with adult or pediatric  cranial applications.
    ARM_Rev_2.book  Page 7  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-8LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Acoustic Output(cont’d)
    Tissue heating is more of a concern when the acoustic beam is 
    stationary, so the thermal index is likely to increase when 
    Doppler or M-Modes are selected.The influence of specific 
    operator controls on acoustic ou tput is described along with the 
    functional purpose of the control through out the user manual 
    and a summary is provided in safety.
    The operator now has easy access to the status of acoustic 
    output and , when combined with the knowledge of tissue 
    characteristics and beam location, the risk of potential bioeffect 
    can be readily assessed.this di splay conforms to the AIUM/
    NEMA Output Display Standard[1] for Ultrasound imaging 
    equipment.
    Operator Intervention
    When conditions indicate a potential for harmful bioeffect, the 
    operator should take action promptly to reduce the risk by 
    changing equipment settings or altering procedural techniques:
    • Output display index values much greater than 1.0 represents an increased risk for tissues in perticular 
    beam locations.the potenti al for heating will normally 
    only occur near the surface or at the focus,while the 
    potential for cavitation is reduced away from the 
    focus.selecting non scan operating modes such as PW 
    or CW Doppler and M-Mode will significantly increase 
    the thermal index because the beam is stationary.
    • Optimise gain and other im age enhancement features 
    before increasing the acoustic output control or other 
    equipment controls that sign ificantly affect the output 
    level.Become thoroughly fam iliar with all controls that 
    affect output and observe the output display for 
    results.controls affecting output are described 
    throughout the user manual.
    ARM_Rev_2.book  Page 8  Monday, March 15, 2010  2:19 PM 
    						
    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-9
    Direction 5122542-100 Rev. 2
    Operator Intervention(Cont’d)
    • Development and practice sk ills to localize anatomy and 
    optimize image quality rapidly, then freeze the image as 
    soon as the necessary diognostic information is 
    obtained.It takes time for tissue temparatures to 
    increase ,so reducing exposue r time can significantly 
    reduce the potential for injury.
    • Avoid susceptible tissues  , if possible ,by changing 
    probe position ,entrace angles or probe type. Higher 
    frequency probes will not penetrate as deep while linear 
    probes have lower near field energy density.Avoid 
    focussing on bone or poorly perfused tissue.do not 
    allow the acoustic beam to penetrate or focus on or near 
    the eye.
    Although choices like probe selection , mode of operation and 
    other control adjustments have a significant affect on output 
    levels, the ability to ch nage these selections is often restricted 
    by the type of examination or  clinical objectives.Therefore , 
    some examinations may require re latively high output levels to 
    achieve success.
    The decission to raise acoustic output to potentially harmful 
    levels must include an assessment of the risk/benefit potential. 
    Such decessions are routine with imaging modalities 
    incorporating ionizing radiation such as Nuclear Medicine, X-
    Ray and CT.The principle of ALARA is widely used in these 
    modalities for minimizing the exposure risk and is now a 
    recommended practice with highlevel diagnostic ultrasound.
    CAUTION:
    During each ultrasound examination , the clinical user is 
    expected to weigh the medical benefit of the diognostic 
    information obtained against the risk of harmful effects. Once an 
    optimal image is achived the need for incresing acoustic output 
    or prolonging the exposure can no t be justified. It is important, 
    therefore , for the user to be  familiar with system controls that 
    affect image quality as well as acoustic output.Complete 
    descriptions of image optimization  and acoustic output controls 
    are provided in the user interactions.
    ARM_Rev_2.book  Page 9  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-10LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Implementing the ALARA Methods
    The Primary objective for any ultrasound examination is to 
    obtaine diagnostic information of  sufficient quality to benefit the 
    patient. Image Quality can usually be improved by increasing 
    the acoustic output or taking  more time to refine the 
    image.These same actions, however, will also  increase the risk 
    of harmful bioeffects when imaging sensitive tissues or when 
    high output levels are used.the o perator is therefore encouraged 
    to use the lowest acoustic output setting necessary to produce 
    clinically acceptable data.
    The principle of ALARA , which stands for As low as reasonably 
    achievable, is to keep the radi ation exposure at the minimum 
    level necessary to obtain the diagnostic information.This 
    principle is widely practiced in medical x-ray protection where 
    exposure at any level is potentially harmful.Historically,ALARA 
    was initiated as a cautions approachfor dealing with uncertain 
    hazards but has since become the principle method for reducing 
    the risk of injury from hazard s that do not have safe minimum 
    threshold.
    While no minimum thresolds for harmful bioeffects have been 
    established with the use of diagn ostic ultrasound, the principle 
    of ALARA can be readily implemented on equipment 
    incorporating an output display. As the operator adjusts the 
    equipment to optimize the image quality, the display interactively 
    updates to indicate the effect on output.
    Controls that have no noticeable impact on image quality should 
    be set to minimize the output while controls that improve the 
    image quality and also increase acoustic output should be set 
    no higher than needed to achive a diagnostic quality image. If 
    the output display indicates values much greater than 1.0 , the 
    operator should reduce the exposure time and freeze the image 
    as soon as possible.
    At very low levels(
    						
    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-11
    Direction 5122542-100 Rev. 2
    Clinical Instruction for Fetal Use
    Following are illustrative examples  of clinical instructions for 
    fetal Doppler use which were prepared by Harold 
    Schulman,M.D. and Jhon Hobbins,M.D.
    Umbilical artery
    To Obtain a signal:
    For continuous wave doppler instrument , Place the pencil probe 
    with appropriate conducting gel on the meternal abdomen. The 
    volume should be at a comportable level and the power output 
    and gain at a mid-setting. Slowly move the pencil probe by 
    changing the angle or location on the meternal abdomen until 
    the characteristic sound of the fetal umbilical flow can be 
    heard.This is a swishing sound witha rate usually between 120 
    and 160 BPm. It should be lack ing in discernible clicks or 
    sounds similar to valve moveme nt on cordiac auscultation. 
    When the image is obtained , adjust the controls in order to 
    optimize the imageby reducing the power output to the lowest 
    setting at which a good qualitysignal is obtained.The gain setting 
    should also be reduced to make the signal appear crisp with 
    minimal background noise. If diagnostic flow signals are not 
    obtained, the angle of the beam  incident to the ubbilical cord 
    should be examined to determine whether diognostic flow is 
    present. When a good quality signal is obtained all the way 
    across the display screen , the image amy be frozen.
    Prior to taking measurements, the signals should be examined 
    to exposure that the variations  in wave form size that may be 
    caused by fetal breathing movements are not present.Fetal 
    breathing movements will invalidat e any measurements due to 
    the variations they cause in umbilical flow.
    If a duplex doppler system is being  used, then the umbilical cord 
    should be visualized and attempted to be seen in as much of its 
    length as is feasible, Considering its usual cooling.The sample 
    volume size(“gate”) should be adjusted to encompass the entire 
    vessel. If this vessel is being  seen only in crossection , and 
    diastollic flow is not seen, this ma y be an artifact caused by the 
    angle between the beam and the vessel and anothersampling 
    site should be examined.when this crosssectional view is 
    obtained, rotating the transducer 90 degrees may bring a 
    greater length of the cord into view and permit examination of an 
    area where the artery can be seen at a more advantageous 
    angle.Indices of pulsality , if calculated, should be determined 
    for each of several heartbeats and averaged.
    ARM_Rev_2.book  Page 11  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-12LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Uterine artery 
    For Continuous wave doppler systems,place the pencil probe in 
    the lower lateral portion of the mother’s abdomen.generally just 
    above the groin and directed toward the cervix.Adjust the angle 
    or position until the characterist ic wave form of the uterine/ 
    arcuate artery is obtained.Without visualization, it is important to 
    search carefully for the wave form with the greatest amount of 
    diastolic flow compared to the systolic flow in a way that might 
    appear abnormal if obtained from the uterine artery itself. The 
    power and gain settings should be adjusted to the lowest levels 
    at which adequate signals are obtained with a minimul amount 
    of noise relative to the signal. The image may be frozen when 
    the desired number of waveforms are present on the screen.
    For pulsed doppler systems, Place the transducer in lower other 
    portion of the meternal abdomen. The orientation should be 
    towards the parametria area can then be examinedby 
    maneuvering the sample volume until the characteristics wave 
    forms are obtained.examinations should be confined to the area 
    that is composedof the myometrium amd may be most fruitful 
    close to the placenta. Indices fo r pulasatility , if calculated, 
    should be determined for each of several heartbeats and 
    averaged.
    Fetal Heart
    Fetal cardiac doppler studies can be performed only with duplex 
    ultrasound systems. the fetal heart should be carefully examined 
    in all standard planes to determine whether or not the anatomy 
    is normal. These planes should inclued :four-chamber view, long 
    axis left ventricles, short axis of the great vessels, aortic arch, 
    and pulmonary artery ductus views. The area for sampling 
    should be visualized as clearly as possible to the ultrasound 
    beam. The sample volume should be placed in this area with an 
    appropriate size selected to samp le as desired. Power output 
    control should be at the lowest settingscompatible with obtaining 
    an adequate image and the gain setting should be maintained to 
    keep an adequate signal without ex cessive noise. Excluding the 
    actual valves from the sample  volume will help to minimize 
    unnecessary noise from the signals obtained.
    ARM_Rev_2.book  Page 12  Monday, March 15, 2010  2:19 PM 
    						
    							Bioeffects
    LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-13
    Direction 5122542-100 Rev. 2
    Efficacy Of Fetal Doppler
    The following are clinical  obstritrical where ther are experimental 
    data that demonstrate the effica cy of Doppler.Provided by Jhon 
    C.Hobbins,M.D. and Peter Burns,Ph.D.4
    IUGR
    Many studies have shown a good correlation between abnorma 
    waveform( or decreased blood flow) and increased perinatal 
    mortality5, Fetal hypoxia 6,7,8 ,and neonatal eveents such as 
    necrotizing entero colitics(NEC), and interventricular 
    hemorrhage(IVH)8. It must be pointed out that the doppler 
    studies initially designed to identity altered fetal growth had a 
    predictably low sensitivity beca use the category of small for 
    dates frequently includes genetically small but completely 
    normal fetuses and neonates.Thus far, all studies have shown 
    that the rare doppler  pattern of reverse diastollic flow in the 
    umbilical artery has been very  highly correlated with adverse 
    outcome and may warrent immediate intervention. conversely, 
    anormal waveform in the umbilic al artery has been rarely 
    associated with stillbirth in a high risk pregnancy.
    Cardiac Anomalies
    Doppler has become an integral part of fetal schocardiaographic 
    studies. Its use has been directly derived from well-documented 
    pediatric and adult Doppler research.The etiology and 
    seriousness of fetal arrhythmias  is relatively easy to determine 
    with pulsed doppler interrogation of ventricular diastollic filling 
    patterns.Benign patterns such as  premature atrial extrasystoles 
    can be more easily diognised with Doppler than with older 
    approach of M-Mode echocardiography.Doppler has proven 
    useful as well in the understanding and interpretation of 
    structural heart disease and the appearence of non-immune 
    hydrops. Improved diognostic  and prognostic ability has 
    permitted more accurate counseling of parents whose fetuses 
    have structural ca rdiac abnormalities.
    ARM_Rev_2.book  Page 13  Monday, March 15, 2010  2:19 PM 
    						
    							Acoustic Output
    1-14LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
    Direction 5122542-100 Rev. 2
    Efficacy Of Fetal Doppler(cont’d)
    Summary
    It is important to realize that current investigation DOES NOT 
    support the concept that Doppler waveform analysis in the 
    above obestritrical conditions prov ides information that is meant 
    to replace other conventional tests, or the biophysical profile, 
    also, it should not replace classical non-doppler ultrasound 
    scanning methods. Rather , it a ppears that Doppler is of value 
    when used in conjuction with thes e other tests. Lastly, there is 
    little data to suggest that an is olated Doppler examination of the 
    fetus can be used as a screening tool in a low-risk population to 
    identify the compromised fetus.
    As with the evoluti on of any new diognostic technique, the 
    indications for use may change when new experimental data 
    become avilable.
    Suggested Guiedlines
    Following are suggested guidelines prepared by Harold 
    Schulman,M.D
    Overview
    There are two types of Doppler instruments, continuous 
    Wave(CW) and pulsed .Modern instruments use directional 
    Dopplers, that is they portray forward and reverse flow on a split 
    screen.The CW Doppler traverses the entire vessel Diameter 
    and summerises a veriety of refl ections, including those from the 
    vessel wall and neighboring vess els. The pulsed Doppler is 
    focused and may sample the red cell flow in different loci within 
    the radius, but has the advantage of rejecting extraneous 
    current instrumentation utilize  output energies which exceede 
    guidelines for safety  obestrtric ultrasound.
    ARM_Rev_2.book  Page 14  Monday, March 15, 2010  2:19 PM 
    						
    All GE manuals Comments (0)

    Related Manuals for GE Logiq 3 Instruction Manual