GE Logiq 3 Instruction Manual
Have a look at the manual GE Logiq 3 Instruction Manual online for free. It’s possible to download the document as PDF or print. UserManuals.tech offer 45 GE manuals and user’s guides for free. Share the user manual or guide on Facebook, Twitter or Google+.
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-5 Direction 5122542-100 Rev. 2 Mechanical Bioeffect In a similar manner ,the interact ion of ultrasound energy with tissue can produce a number of non-thermal or mechanical effects.The most significant is cavitation which results from the action of the oscillating ultrasou nd pressure on tiny gas bubbles with in the tissues. Cavitation has caused mechanical damage on a cellular level such as microscopic tears and hemorrhage in laboratory tests with small animals.t he major contributing factors can again be categorised as either tissue characteristics or control parameters: • The physical characterist ics of tissue such as the presence and size of micr oscopic gas bubbles and the sensitivity of the tissue to th e effects of cavitation will influence the potential for and magnitude of cavitation. • Acoustic field parameters like output frequency,peak pulse amplitude s and perhaps pulse length are the primary parameters effecting the onset of cavitation.these are controlled by the operator through appropriate equipment selections. Althogh it is generally accepted that no harmful biological effects have been demonstrated at the frequency ,intensity,and exposure times used in diognostic examinations,research in to the potential for harmful effects continues.The operator is encouraged to survey the literature for future developments on bioeffects and to become familiar with the references at the end of this section. ARM_Rev_2.book Page 5 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-6LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Operator Awareness and Actions to Minimize bioeffect The operator must be aware of the particular conditions that exist during the examination to recognize the potential for bioeffect and then take appropriate action to reduce the risk.The recognization of potential harm comes from an understanding of tissue characteristics and a re al-time knowledge of acoustic output.Taking appropriate action requires familiarity with equipment operation and examination skills like implementing alternative techniques for obtaining the same diognostic information. Tissue Characteristics Tissue Characteristics vary considerably throughout the body.they influence the acoustic field and determine the heating / cooling rates and cavitation potential.Ultrasound energy dissipates as it passes throug h the tissue causing the deeper tissue to encounter much lowe r levels.Some tissues like bone readiy convert ultrasound energy to heat,while others like blood and amniotic fluid pass the energy on to adjucent tissue relatively unattenuated. A perticular situation that represents a tissue combination requiring extra precaution is a third trimester transabdominal fetal examination where there is a very thin abdominal wall and a long fluid path.The relative lack of attenuating tissue along the acoustic path will significantly increase the available energy in the fetal tissue.Additionally,fetal tissues are more susceptible to long term injury due to nature of developing tissue.focussing the ultra sound beam on or near fetal bone further increases the risk. Otherthan fetal tissue,there is increased susceptability for heating in any tissue that cannot easily conduct or distribute heat due to low blood perfusion.As the examination progresses,the operator must be aware of changing tissue conditions. ARM_Rev_2.book Page 6 Monday, March 15, 2010 2:19 PM
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-7 Direction 5122542-100 Rev. 2 Acoustic Output awareness of the acoustic output level can be a difficult task for the operator,espcially when the objective is to obtaine a quality image.Older ultrasound equipment had limited means,if any,for indicating the acoustic output le vel.In most cases ,the operator had to be familiar with the output intensities as described in the operator manual.To improve op erator awareness of acoustic output,this system incorporates an output display that directly indicates the potential for mechanical and thermal bioeffects as equipment controls are adjusted. The output display consists of four numeric index values that indicate the potential for producing bioeffects(three indices are for heating effect and one for cavitation).As the user changes equipment settings that alter the acoustic output,the output display indices are immediately updated to reflect the changes in potential for producing .The indices are based on mathematical models and each is normalized so that the potential for bioeffect becomes more significant as the indices reach a value 1.0 or larger. A mechanical index (MI) provides an indication of the potential for the possible onset of transient cavitation within tissue while the three thermal indices provide an indication of the potential for heat generation within tissue.The different thermal indices may be used depending on the type of tissue being examined: • Soft Tissue Thermal Index (TIS) is used as an indicator ofthe potential to generate heat within soft tissues.this is the most used thermal indicator. • bone Thermal Index (TIB) is used as an indicator of the potential to generate heat at the beam focus when focussing on or near bone that is adjucent to very sensitive tissue.this index is intended as a thermal indicator for second and third trimester fetal examination or transfontanelle neonatal cephalic exams. • Cranial Bone Thermal Index (TIC) is used as an indicator of the potential to generate heat in the near- field when the beam passes through bone at the surface as with adult or pediatric cranial applications. ARM_Rev_2.book Page 7 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-8LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Acoustic Output(cont’d) Tissue heating is more of a concern when the acoustic beam is stationary, so the thermal index is likely to increase when Doppler or M-Modes are selected.The influence of specific operator controls on acoustic ou tput is described along with the functional purpose of the control through out the user manual and a summary is provided in safety. The operator now has easy access to the status of acoustic output and , when combined with the knowledge of tissue characteristics and beam location, the risk of potential bioeffect can be readily assessed.this di splay conforms to the AIUM/ NEMA Output Display Standard[1] for Ultrasound imaging equipment. Operator Intervention When conditions indicate a potential for harmful bioeffect, the operator should take action promptly to reduce the risk by changing equipment settings or altering procedural techniques: • Output display index values much greater than 1.0 represents an increased risk for tissues in perticular beam locations.the potenti al for heating will normally only occur near the surface or at the focus,while the potential for cavitation is reduced away from the focus.selecting non scan operating modes such as PW or CW Doppler and M-Mode will significantly increase the thermal index because the beam is stationary. • Optimise gain and other im age enhancement features before increasing the acoustic output control or other equipment controls that sign ificantly affect the output level.Become thoroughly fam iliar with all controls that affect output and observe the output display for results.controls affecting output are described throughout the user manual. ARM_Rev_2.book Page 8 Monday, March 15, 2010 2:19 PM
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-9 Direction 5122542-100 Rev. 2 Operator Intervention(Cont’d) • Development and practice sk ills to localize anatomy and optimize image quality rapidly, then freeze the image as soon as the necessary diognostic information is obtained.It takes time for tissue temparatures to increase ,so reducing exposue r time can significantly reduce the potential for injury. • Avoid susceptible tissues , if possible ,by changing probe position ,entrace angles or probe type. Higher frequency probes will not penetrate as deep while linear probes have lower near field energy density.Avoid focussing on bone or poorly perfused tissue.do not allow the acoustic beam to penetrate or focus on or near the eye. Although choices like probe selection , mode of operation and other control adjustments have a significant affect on output levels, the ability to ch nage these selections is often restricted by the type of examination or clinical objectives.Therefore , some examinations may require re latively high output levels to achieve success. The decission to raise acoustic output to potentially harmful levels must include an assessment of the risk/benefit potential. Such decessions are routine with imaging modalities incorporating ionizing radiation such as Nuclear Medicine, X- Ray and CT.The principle of ALARA is widely used in these modalities for minimizing the exposure risk and is now a recommended practice with highlevel diagnostic ultrasound. CAUTION: During each ultrasound examination , the clinical user is expected to weigh the medical benefit of the diognostic information obtained against the risk of harmful effects. Once an optimal image is achived the need for incresing acoustic output or prolonging the exposure can no t be justified. It is important, therefore , for the user to be familiar with system controls that affect image quality as well as acoustic output.Complete descriptions of image optimization and acoustic output controls are provided in the user interactions. ARM_Rev_2.book Page 9 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-10LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Implementing the ALARA Methods The Primary objective for any ultrasound examination is to obtaine diagnostic information of sufficient quality to benefit the patient. Image Quality can usually be improved by increasing the acoustic output or taking more time to refine the image.These same actions, however, will also increase the risk of harmful bioeffects when imaging sensitive tissues or when high output levels are used.the o perator is therefore encouraged to use the lowest acoustic output setting necessary to produce clinically acceptable data. The principle of ALARA , which stands for As low as reasonably achievable, is to keep the radi ation exposure at the minimum level necessary to obtain the diagnostic information.This principle is widely practiced in medical x-ray protection where exposure at any level is potentially harmful.Historically,ALARA was initiated as a cautions approachfor dealing with uncertain hazards but has since become the principle method for reducing the risk of injury from hazard s that do not have safe minimum threshold. While no minimum thresolds for harmful bioeffects have been established with the use of diagn ostic ultrasound, the principle of ALARA can be readily implemented on equipment incorporating an output display. As the operator adjusts the equipment to optimize the image quality, the display interactively updates to indicate the effect on output. Controls that have no noticeable impact on image quality should be set to minimize the output while controls that improve the image quality and also increase acoustic output should be set no higher than needed to achive a diagnostic quality image. If the output display indicates values much greater than 1.0 , the operator should reduce the exposure time and freeze the image as soon as possible. At very low levels(
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-11 Direction 5122542-100 Rev. 2 Clinical Instruction for Fetal Use Following are illustrative examples of clinical instructions for fetal Doppler use which were prepared by Harold Schulman,M.D. and Jhon Hobbins,M.D. Umbilical artery To Obtain a signal: For continuous wave doppler instrument , Place the pencil probe with appropriate conducting gel on the meternal abdomen. The volume should be at a comportable level and the power output and gain at a mid-setting. Slowly move the pencil probe by changing the angle or location on the meternal abdomen until the characteristic sound of the fetal umbilical flow can be heard.This is a swishing sound witha rate usually between 120 and 160 BPm. It should be lack ing in discernible clicks or sounds similar to valve moveme nt on cordiac auscultation. When the image is obtained , adjust the controls in order to optimize the imageby reducing the power output to the lowest setting at which a good qualitysignal is obtained.The gain setting should also be reduced to make the signal appear crisp with minimal background noise. If diagnostic flow signals are not obtained, the angle of the beam incident to the ubbilical cord should be examined to determine whether diognostic flow is present. When a good quality signal is obtained all the way across the display screen , the image amy be frozen. Prior to taking measurements, the signals should be examined to exposure that the variations in wave form size that may be caused by fetal breathing movements are not present.Fetal breathing movements will invalidat e any measurements due to the variations they cause in umbilical flow. If a duplex doppler system is being used, then the umbilical cord should be visualized and attempted to be seen in as much of its length as is feasible, Considering its usual cooling.The sample volume size(“gate”) should be adjusted to encompass the entire vessel. If this vessel is being seen only in crossection , and diastollic flow is not seen, this ma y be an artifact caused by the angle between the beam and the vessel and anothersampling site should be examined.when this crosssectional view is obtained, rotating the transducer 90 degrees may bring a greater length of the cord into view and permit examination of an area where the artery can be seen at a more advantageous angle.Indices of pulsality , if calculated, should be determined for each of several heartbeats and averaged. ARM_Rev_2.book Page 11 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-12LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Uterine artery For Continuous wave doppler systems,place the pencil probe in the lower lateral portion of the mother’s abdomen.generally just above the groin and directed toward the cervix.Adjust the angle or position until the characterist ic wave form of the uterine/ arcuate artery is obtained.Without visualization, it is important to search carefully for the wave form with the greatest amount of diastolic flow compared to the systolic flow in a way that might appear abnormal if obtained from the uterine artery itself. The power and gain settings should be adjusted to the lowest levels at which adequate signals are obtained with a minimul amount of noise relative to the signal. The image may be frozen when the desired number of waveforms are present on the screen. For pulsed doppler systems, Place the transducer in lower other portion of the meternal abdomen. The orientation should be towards the parametria area can then be examinedby maneuvering the sample volume until the characteristics wave forms are obtained.examinations should be confined to the area that is composedof the myometrium amd may be most fruitful close to the placenta. Indices fo r pulasatility , if calculated, should be determined for each of several heartbeats and averaged. Fetal Heart Fetal cardiac doppler studies can be performed only with duplex ultrasound systems. the fetal heart should be carefully examined in all standard planes to determine whether or not the anatomy is normal. These planes should inclued :four-chamber view, long axis left ventricles, short axis of the great vessels, aortic arch, and pulmonary artery ductus views. The area for sampling should be visualized as clearly as possible to the ultrasound beam. The sample volume should be placed in this area with an appropriate size selected to samp le as desired. Power output control should be at the lowest settingscompatible with obtaining an adequate image and the gain setting should be maintained to keep an adequate signal without ex cessive noise. Excluding the actual valves from the sample volume will help to minimize unnecessary noise from the signals obtained. ARM_Rev_2.book Page 12 Monday, March 15, 2010 2:19 PM
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-13 Direction 5122542-100 Rev. 2 Efficacy Of Fetal Doppler The following are clinical obstritrical where ther are experimental data that demonstrate the effica cy of Doppler.Provided by Jhon C.Hobbins,M.D. and Peter Burns,Ph.D.4 IUGR Many studies have shown a good correlation between abnorma waveform( or decreased blood flow) and increased perinatal mortality5, Fetal hypoxia 6,7,8 ,and neonatal eveents such as necrotizing entero colitics(NEC), and interventricular hemorrhage(IVH)8. It must be pointed out that the doppler studies initially designed to identity altered fetal growth had a predictably low sensitivity beca use the category of small for dates frequently includes genetically small but completely normal fetuses and neonates.Thus far, all studies have shown that the rare doppler pattern of reverse diastollic flow in the umbilical artery has been very highly correlated with adverse outcome and may warrent immediate intervention. conversely, anormal waveform in the umbilic al artery has been rarely associated with stillbirth in a high risk pregnancy. Cardiac Anomalies Doppler has become an integral part of fetal schocardiaographic studies. Its use has been directly derived from well-documented pediatric and adult Doppler research.The etiology and seriousness of fetal arrhythmias is relatively easy to determine with pulsed doppler interrogation of ventricular diastollic filling patterns.Benign patterns such as premature atrial extrasystoles can be more easily diognised with Doppler than with older approach of M-Mode echocardiography.Doppler has proven useful as well in the understanding and interpretation of structural heart disease and the appearence of non-immune hydrops. Improved diognostic and prognostic ability has permitted more accurate counseling of parents whose fetuses have structural ca rdiac abnormalities. ARM_Rev_2.book Page 13 Monday, March 15, 2010 2:19 PM
Acoustic Output 1-14LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Efficacy Of Fetal Doppler(cont’d) Summary It is important to realize that current investigation DOES NOT support the concept that Doppler waveform analysis in the above obestritrical conditions prov ides information that is meant to replace other conventional tests, or the biophysical profile, also, it should not replace classical non-doppler ultrasound scanning methods. Rather , it a ppears that Doppler is of value when used in conjuction with thes e other tests. Lastly, there is little data to suggest that an is olated Doppler examination of the fetus can be used as a screening tool in a low-risk population to identify the compromised fetus. As with the evoluti on of any new diognostic technique, the indications for use may change when new experimental data become avilable. Suggested Guiedlines Following are suggested guidelines prepared by Harold Schulman,M.D Overview There are two types of Doppler instruments, continuous Wave(CW) and pulsed .Modern instruments use directional Dopplers, that is they portray forward and reverse flow on a split screen.The CW Doppler traverses the entire vessel Diameter and summerises a veriety of refl ections, including those from the vessel wall and neighboring vess els. The pulsed Doppler is focused and may sample the red cell flow in different loci within the radius, but has the advantage of rejecting extraneous current instrumentation utilize output energies which exceede guidelines for safety obestrtric ultrasound. ARM_Rev_2.book Page 14 Monday, March 15, 2010 2:19 PM