Page 51

Acoustic Output
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-45
Direction 5122542-100 Rev. 2 Table 1-31:  Transducer Mode 
8L Operating Mode: M Mode (Inc. B-Mode)
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 0.80.20.7 0.3
IECFDAUnits
pra pr.3 (MPa) 1.7
PWo (mW)8.28.2 8.2
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. I pi,α zsp(cm) 1.1
deq(zb) deq(zsp) (cm)0.2
ƒawffc(MHz) 5.15.15.1 5.1
Dim of A...

Page 52

Acoustic Output
1-46LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
Table 1-32:  Transducer Mode 
8L Operating Mode: Color Flow Mode (Inc. M-Mode)
Index Label
MITIS
TIB
TIC
scan non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value 0.9 2.24.8
IECFDAUnits
pra pr.3 (MPa) 1.9
PWo (mW)134.2134.2
min of [P
α( zs,  Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. Ipi, αzsp(cm) 0.9
deq(zb) deq(zsp) (cm)
ƒawffc(MHz) 4.5 4.54.5
Dim...

Page 53

Acoustic Output
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-47
Direction 5122542-100 Rev. 2 Table 1-33:  Transducer Mode: 
8L Operating Mode: Pulsed Doppler (inc. B-Mode)
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 0.51.23.1 2.1
IECFDAUnits
pra pr.3 (MPa) 1.1
PWo (mW)58.458.4 58.4
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. I pi,α zsp(cm) 0.8
deq(zb) deq(zsp) (cm)0.3
ƒawffc(MHz) 4.24.54.5 4.5
Dim...

Page 54

Acoustic Output
1-48LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
Probe Name:10L
Table 1-34:  Transducer Mode 10L Operating Mode: B Mode
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 1.4 0.20.3
IECFDAUnits
pra pr.3 (MPa) 3.2
PWo (mW)8.28.2
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. I pi,α zsp(cm) 1.9
deq(zb) deq(zsp) (cm)
ƒawffc(MHz) 5.3 5.35.3
Dim of A
aprtX(cm)1.01.0...

Page 55

Acoustic Output
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-49
Direction 5122542-100 Rev. 2 Table 1-35:  Transducer Mode 
10LOperating Mode: M Mode (Inc. B-Mode)
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 1.40.20.8 0.3
IECFDAUnits
pra pr.3 (MPa) 3.2
PWo (mW)8.28.2 8.2
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. I pi,α zsp(cm) 1.9
deq(zb) deq(zsp) (cm)0.4
ƒawffc(MHz) 5.35.35.3 5.3
Dim of A...

Page 56

Acoustic Output
1-50LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
Table 1-36:  Transducer Mode 
10L Operating Mode: Color Flow Mode (Inc. M-Mode)
Index Label
MITIS
TIB
TIC
scan non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value 1.8 0.230.4
IECFDAUnits
pra pr. 3 (MPa) 3.9
PWo (mW)9.99.9
min of [ P
α(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. Ipi, αzsp(cm) 0.9
deq(zb) deq(zsp) (cm)
ƒawffc(MHz) 4.9 4.94.9
Dim of...

Page 57

Acoustic Output
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-51
Direction 5122542-100 Rev. 2 Table 1-37:  Transducer Mode: 
10L Operating Mode: Pulsed Doppler (inc. B-Mode)
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 1.50.31.9 0.5
IECFDAUnits
pra pr.3 (MPa) 3.4
PWo (mW)13.213.2 13.2
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm)
z  at max. I pi,α zsp(cm) 1.4
deq(zb) deq(zsp) (cm)0.3
ƒawffc(MHz) 5.05.05.0 5.0...

Page 58

Acoustic Output
1-52LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
Probe Name:10LB
Table 1-38:  Transducer Mode 10LB Operating Mode: B Mode
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 1.6 2.12.0
IECFDAUnits
pra pr.3 (MPa) 3.8
PWo (mW)73.973.9
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.2
z  at max. I pi,α zsp(cm)
deq(zb) deq(zsp) (cm)
ƒawffc(MHz) 5.8 5.85.8
Dim of A...

Page 59

Acoustic Output
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-53
Direction 5122542-100 Rev. 2 Table 1-39:  Transducer Mode 
10LB Operating Mode: M Mode (Inc. B-Mode)
Index LabelMI
TISTIB
TIC
scannon-scannon-
scan
Aaprt1
Global Maximum: Index Value 1.45 2.12.1
IECFDAUnits
pra pr.3 (MPa) 3.52
PWo (mW)73.969.96
min of [P
α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.32
z  at max. I pi,α zsp(cm)
deq(zb) deq(zsp) (cm)
ƒawffc(MHz) 5.91 5.95.4
Dim of A...

Page 60

Acoustic Output
1-54LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
Table 1-40:  Transducer Mode 
10LB Operating Mode: Color Flow Mode (Inc. M-Mode)
Index Label
MITIS
TIB
TIC
scan non-scan
non-
scan
Aaprt<
Aaprt>1
Global Maximum: Index Value 1.45 3.02.0 5.3
IECFDAUnits
pra pr. 3 (MPa) 3.92
PWo (mW)127.0139.0 127.0
min of [ P
α(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)])
zs z1(cm)
zbp zbp(cm)
zb zsp(cm) 1.22.9
z  at max. Ipi, αzsp(cm)
deq(zb) deq(zsp) (cm)0.4...
Start reading GE Logiq 3 Instruction Manual