Page 91
Measurement Formulas LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-15 Direction 5122542-100 Rev. 2 . Table 2-3: Volume Calculation Formulas Calc NameInput MeasurementsFormula Volume (spherical) one distance Vo l [ m l ] = (π/6)xd^3 Volume (prolate spheroidal) two distances, d1>d2 Vol[ml]= ( π/6)xd1xd2^2 Volume (prolate spheroidal) one ellipse, d1 major axis, d2 minor axis Vol[ml]= ( π/6)xd1xd2^2 Volume (spheroidal) three distances Vol[ml]= ( π/6)xd1xd2xd3 Volume (spheroidal) one...
Page 92
Measurement Formulas 2-16LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Formulas–OB Table 2-5: OB Calculation Formulas Calc Mnemonic Calc Name Input Measurements Formula Author Reference AC Abdominal Circumference circumference by trace, ellipse, circle or two distances AC=13.3+1.61 (GA) - 0.00998 (GA)2Hadlock et al, Radiology, 152:497-501, 1984 BPD Biparietal Diameter one distance BPD = -3.08+0.41 (GA) - 0.000061 (GA) 3 CRLCrown Rump Length one...
Page 93
Measurement Formulas LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-17 Direction 5122542-100 Rev. 2 Table 2-6: CUA Hadlock Formulas Calc MnemonicCalc Name Formula Author Reference CUAaComposite Ultrasound Age 1. CUA (BPD) = 9.54 + 1.482 * BPD + 0.1676 * BPD2 2. CUA (HC) = 8.96 + 0.540 * HC + 0.0003 * HC3 3. CUA (AC) = 8.14 + 0.753 * AC + 0.0036 * AC2 4. CUA (FL) = 10.35 + 2.460 * FL + 0.170 * FL2 5. CUA (BPD, HC) = 10.32 + 0.009 * HC2 + 1.3200 * BPD + 0.00012 * HC3 6. CUA (BPD, AC)...
Page 94
Measurement Formulas 2-18LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 EFWEstimated Fetal Weight AC, HC, and FL EFW [g]=10^(1.326- (0.00326 * AC [cm] * FL [cm]) + (0.0107 * HC [cm]) + (0.0438 * AC [cm]) + (0.158 * FL [cm])) Hadlock, AJOG, 151:333, 1985 EFW Estimated Fetal Weight AC, HC, BPD, FL EFW [g] = 10^(1.3596- (0.00386 * AC [cm] * FL [cm]) + (0.0064 * HC [cm]) + (0.00061 * BPD [cm] * AC [cm]) + (0.0424 * AC [cm]) + (0.174 * FL [cm])) Hadlock,...
Page 95
Measurement Formulas LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-19 Direction 5122542-100 Rev. 2 Amniotic Fluid Index (AFI) The normal values are considered to be: 36-40 weeks 0-5 cm = very low 5.1-8.0 cm = low 8.1 - 18.0 cm = normal >18.0 = high Dr. Rutherford/Dr. Phelan, Obstetrics and Gynecology , Volume 70, No. 3, Part 1, p.353-6, Sept. 1987. 28-40 weeks 15.0 cm = average >20.0 - 24.0 = hydramnios
Page 96
Measurement Formulas 2-20LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Formulas–GYN Table 2-8: GYN Calculation Formulas Calc Mnemonic Calc Name Input Measurements Formula Ut-L Uterine Length one distanceUt-L[cm or mm]=d1 Ut-H Uterine Height one distanceUt-H[cm or mm]=d1 Ut-W Uterine Width one distanceUt-W[cm or mm]=d1 Ut-Volume Uterine Volume Ut-L, Ut-H, Ut-W Ut-Volume=0.5236 x Ut-L[cm or mm] x Ut-H[cm or mm] x Ut-W[cm or mm] UtPFD Uterus Portio-Fundus...
Page 97
Measurement Formulas LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-21 Direction 5122542-100 Rev. 2 LtOvFo[ml] Left Ovary Follicles One distancebD1[cm]3 x π/6 Two distances bD1[cm]2 x D2x[cm] π/6 : (D1
Page 98
Measurement Formulas 2-22LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Formulas–Vascular Table 2-9: Vascular Calculation Formulas Calc Mnemonic Calc NameInput MeasurementsFormula RT ECA Right External Carotid Artery Ve l o c i t y one Doppler blood flow peak velocityRT ECA=v1 [cm/s or m/s] RT CCA Right Common Carotid Artery Ve l o c i t y one Doppler blood flow peak velocityRT CCA=v1 [cm/s or m/s] RT BIFURC Right Carotid Bifurcation Ve l o c i t y one...
Page 99
Measurement Formulas LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-23 Direction 5122542-100 Rev. 2 Formulas-Cardiac The Following table lists the card iac calculations. the folders where to find tha calculations and related measurements are indicated in brackets “[ ]”. Table 2-10: Cardiac Calculation Formulas %FS [Dimension, Cube/Teicholz] Mode : 2D, CF, VR2D Formula : (({LFIDd}-[LVIDs])/{LVIDd]) Needs measurement : LVIDd [Dimension, Cube/Teichloz], LVIDs [Dimension, Cube/Teichloz]...
Page 100
Measurement Formulas 2-24LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Ao/LA [Generic, Dimension] Mode: MM:CM:AMM:CAMM:VRMM Formula: {Ao Diam}/{LA Diam} Needs measurement: Ao Diam [Generic , Dimension], LA Diam [Generic, Dimension] Measured by: LA/Ao [MMLAAO] AP Area [Aortic] Mode: CW:PW:VRCW:VRPW Formula: {LVOT Diam}^2*0.785*({LVOT VTI}/{AP VTI}) Needs measurement: LVOT Diam [Aortic], LVOT VTI [Aortic], AP VTI [Aortic] Measured by: AP Area [SDMANTRACE] AR...