Measurement Formulas
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-25
Direction 5122542-100 Rev. 2
AV SV [Aortic]
Mode: CW:PW:VRCW:VRPW
Formula: {AV Diam}^2*0.785*{AV VTI}
Needs measurement: AV Diam [Aortic], AV VTI [Aortic]
Measured by: AV Trace [SDMANTRACE]
AVA (VTI) [Aortic]
Mode: 2D:CW:PW:VRCW:VRPW
Formula: 3.14/4*{LVOT Diam }^2*{LVOT VTI}/{AV VTI}
Needs measurement: LVOT Diam [Aortic], LVOT VTI [Aortic], AV VTI 
[Aortic]
Measured by: AV Trace [AUTOCALC]
AVA Vmax [Aortic]
Mode:...

Page 102

Measurement Formulas
2-26LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
CI A-L A4C [Single Plane A4C, AutoBiplane]
Mode: 2D:CF:VR2D:Trace
Formula: (({LVEDV A-L A4C}-{LVESV A-L A4C})*{HR}/60)/{BSA}
Needs measurement: LVEDV A-L A4C [Single Plane A4C, AutoBiplane], 
LVESV A-L A4C [Single Plane A4C, AutoBiplane], HR [Single Plane A4C, 
AutoBiplane]
Measured by: R-R [2DCALIPER], A4C [2DAUTOVOLUME]
CI A-L A4C [Single Plane A4C]
Mode: 2D:CF:VR2D
Formula: (({LVEDV A-L...

Page 103

Measurement Formulas
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-27
Direction 5122542-100 Rev. 2
CI MOD A2C [Single Plane A2C]
Mode: 2D:CF:VR2D
Formula: (({LVEDV MOD A2C}-{ LVESV MOD A2C})*{HR}/60)/{BSA}
Needs measurement: LVEDV MOD A2C [Single Plane A2C], LVESV MOD 
A2C [Single Plane A2C], 
HR [Single Plane A2C]
Measured by: LVESV A2C [2DVOLUMETRACE]
CI MOD A4C [Single Plane A4C, AutoBiplane]
Mode: 2D:CF:VR2D:Trace
Formula: (({LVEDV MOD A4C}-{ LVESV MOD A4C})*{HR}/60)/{BSA}
Needs...

Page 104

Measurement Formulas
2-28LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
CI(Cube) [Generic, Dimension]
Mode: MM:CM:AMM:CAMM:VRMM
Formula: ((dv-sv)*{MM/HeartRate}/60)/{BSA} where: sv = {MM/LVIDs}^3 
dv = {LVIDd}^3
Needs measurement: LVIDd [Generic , Dimension], LVIDs [Generic, 
Dimension], HR [Generic, Dimension]
Measured by: Heartrate [MMTIMECALIPER]
CI(Teich) [Dimension, Cube/Teicholz]
Mode: 2D:CF:VR2D:Trace
Formula: ((d-s)*{ECG/HeartRate}/60)/{BSA} where: s =...

Page 105

Measurement Formulas
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-29
Direction 5122542-100 Rev. 2
CO A-L LAX [Single Plane LAX, AutoBiplane]
Mode: 2D:CF:VR2D:Trace
Formula: ({LVEDV A-L LAX}-{LVESV A-L LAX})*{HR}/60
Needs measurement: LVEDV A-L LAX [Single Plane LAX, AutoBiplane], 
LVESV A-L LAX [Single Plane LAX, AutoBiplane], HR [Single Plane LAX, 
AutoBiplane]
Measured by: R-R [2DCALIPER], AutoVolume [2DAUTOVOLUME]
CO A-L LAX [Single Plane LAX]
Mode: 2D:CF:VR2D
Formula: ({LVEDV A-L...

Page 106

Measurement Formulas
2-30LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
CO MOD A4C [Single Plane A4C, AutoBiplane]
Mode: 2D:CF:VR2D:Trace
Formula: ({LVEDV MOD A4C}-{LVESV MOD A4C})*{HR}/60
Needs measurement: LVEDV MOD A4C [Single Plane A4C, AutoBiplane], 
LVESV MOD A4C [Single Plane, A4C, AutoBiplane], HR [Single Plane 
A4C, AutoBiplane]
Measured by: R-R [2DCALIPER], A4C [2DAUTOVOLUME]
CO MOD A4C [Single Plane A4C]
Mode: 2D:CF:VR2D
Formula: ({LVEDV MOD...

Page 107

Measurement Formulas
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-31
Direction 5122542-100 Rev. 2
CO(Cube) [Generic, Dimension]
Mode: MM:CM:AMM:CAMM:VRMM
Formula: (dv-sv)*{MM/HeartRate}/ 60 where: sv = {MM/LVIDs}^3 dv = 
{LVIDd}^3
Needs measurement: LVIDd [Generic , Dimension], LVIDs [Generic, 
Dimension], HR [Generic, Dimension]
Measured by: Heartrate [MMTIMECALIPER]
CO(Teich) [Dimension, Cube/Teicholz]
Mode: 2D:CF:VR2D:Trace
Formula: (d-s)*{ECG/HeartRate}/60 where: s =...

Page 108

Measurement Formulas
2-32LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
EDV(Cube) [Generic, Dimension]
Mode: MM:CM:AMM:CAMM:VRMM
Formula: {LVIDd}^3
Needs measurement: LVIDd [Generic, Dimension]
Measured by: LV Study [MMLV], LVIDd [MMDISCALIPER]
EDV(Teich) [Dimension, Cube/Teicholz]
Mode: 2D:CF:VR2D
Formula: 7/(2.4+{LVIDd})*{LVIDd}^3
Needs measurement: LVIDd [D imension, Cube/Teicholz]
Measured by: LVd [2DLV], LVIDd [2DCALIPER], EF(Cube) [AUTOCALC]
EDV(Teich)...

Page 109

Measurement Formulas
LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 2-33
Direction 5122542-100 Rev. 2
EF mod sim [Modified Simpson]
Mode: 2D:CF:VR2D
Formula: ({LVLd apical}/9)*((4*{L VAd (sax MV)})+(2*{LVAd sax 
PM})+sqrt({LVAd (sax MV)}*{LVAd sax PM}))
Needs measurement: LVLd apical [Mod ified Simpson], LVAd (sax MV) 
[Modified Simpson], LVAd sax PM [Modified Simpson], LVLs apical 
[Modified Simpson], LVAs sax MV [Modified Simpson], LVAs sax PM 
[Modified Simpson]
Measured by: EF mod sim...

Page 110

Measurement Formulas
2-34LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual
Direction 5122542-100 Rev. 2
ESV bp el [Biplane Ellipse]
Mode: 2D:CF:VR2D
Formula: (8/(3*3.14159))*{LVAs A4C}*{LVAs sax MV}/{LVIDs}
Needs measurement: LVAs A4C [Biplane Ellipse], LVAs sax MV [Biplane 
Ellipse], LVIDs [Biplane Ellipse]
Measured by: LVEF BP-EL [AUTOCALC]
ESV bullet [Bullet]
Mode: 2D:CF:VR2D
Formula: 5/6*{LVAs sax)}*{LVLs apical}
Needs measurement: LVAs sax)  [Bullet], LVLs apical [Bullet]
Measured by:...