Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-5 Direction 5122542-100 Rev. 2 Mechanical Bioeffect In a similar manner ,the interact ion of ultrasound energy with tissue can produce a number of non-thermal or mechanical effects.The most significant is cavitation which results from the action of the oscillating ultrasou nd pressure on tiny gas bubbles with in the tissues. Cavitation has caused mechanical damage on a cellular level such as microscopic tears and hemorrhage in...
Page 12
Acoustic Output 1-6LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Operator Awareness and Actions to Minimize bioeffect The operator must be aware of the particular conditions that exist during the examination to recognize the potential for bioeffect and then take appropriate action to reduce the risk.The recognization of potential harm comes from an understanding of tissue characteristics and a re al-time knowledge of acoustic output.Taking appropriate...
Page 13
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-7 Direction 5122542-100 Rev. 2 Acoustic Output awareness of the acoustic output level can be a difficult task for the operator,espcially when the objective is to obtaine a quality image.Older ultrasound equipment had limited means,if any,for indicating the acoustic output le vel.In most cases ,the operator had to be familiar with the output intensities as described in the operator manual.To improve op erator awareness of...
Page 14
Acoustic Output 1-8LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Acoustic Output(cont’d) Tissue heating is more of a concern when the acoustic beam is stationary, so the thermal index is likely to increase when Doppler or M-Modes are selected.The influence of specific operator controls on acoustic ou tput is described along with the functional purpose of the control through out the user manual and a summary is provided in safety. The operator now has easy...
Page 15
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-9 Direction 5122542-100 Rev. 2 Operator Intervention(Cont’d) • Development and practice sk ills to localize anatomy and optimize image quality rapidly, then freeze the image as soon as the necessary diognostic information is obtained.It takes time for tissue temparatures to increase ,so reducing exposue r time can significantly reduce the potential for injury. • Avoid susceptible tissues , if possible ,by changing probe...
Page 16
Acoustic Output 1-10LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Implementing the ALARA Methods The Primary objective for any ultrasound examination is to obtaine diagnostic information of sufficient quality to benefit the patient. Image Quality can usually be improved by increasing the acoustic output or taking more time to refine the image.These same actions, however, will also increase the risk of harmful bioeffects when imaging sensitive tissues or...
Page 17
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-11 Direction 5122542-100 Rev. 2 Clinical Instruction for Fetal Use Following are illustrative examples of clinical instructions for fetal Doppler use which were prepared by Harold Schulman,M.D. and Jhon Hobbins,M.D. Umbilical artery To Obtain a signal: For continuous wave doppler instrument , Place the pencil probe with appropriate conducting gel on the meternal abdomen. The volume should be at a comportable level and the...
Page 18
Acoustic Output 1-12LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Uterine artery For Continuous wave doppler systems,place the pencil probe in the lower lateral portion of the mother’s abdomen.generally just above the groin and directed toward the cervix.Adjust the angle or position until the characterist ic wave form of the uterine/ arcuate artery is obtained.Without visualization, it is important to search carefully for the wave form with the greatest...
Page 19
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-13 Direction 5122542-100 Rev. 2 Efficacy Of Fetal Doppler The following are clinical obstritrical where ther are experimental data that demonstrate the effica cy of Doppler.Provided by Jhon C.Hobbins,M.D. and Peter Burns,Ph.D.4 IUGR Many studies have shown a good correlation between abnorma waveform( or decreased blood flow) and increased perinatal mortality5, Fetal hypoxia 6,7,8 ,and neonatal eveents such as necrotizing...
Page 20
Acoustic Output 1-14LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Efficacy Of Fetal Doppler(cont’d) Summary It is important to realize that current investigation DOES NOT support the concept that Doppler waveform analysis in the above obestritrical conditions prov ides information that is meant to replace other conventional tests, or the biophysical profile, also, it should not replace classical non-doppler ultrasound scanning methods. Rather , it a ppears...