GE Logiq 3 Instruction Manual
Here you can view all the pages of manual GE Logiq 3 Instruction Manual. The GE manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 21
Bioeffects LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-15 Direction 5122542-100 Rev. 2 Suggested Guiedlines(cont’d) Methodology: Doppler studies on the umbilical arteries are carried out after manual palpation of the uterus and fetus. In pregnancies before 28 weeks, the fetus cannot be easily felt hencethe probe is placed on the upper third of the ut erus at the midline. The probe is gently rotated until the umbilic al signal is heard and clearly identified on the screen. Ther...
Page 22
Acoustic Output 1-16LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 FDA Acoustic Output Data Maximum Output Summary The following tables list the typical maximum acoustic output levels achievable with LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 for all probes and operational modes. It is intended that this is information be useful in making ALARA decisions and selecting the most appropriate probe for the application. In accordance with US FDA Guidelines, the ov erall...
Page 23
FDA Acoustic Output Data LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-17 Direction 5122542-100 Rev. 2 Maximum Probe Temperature Table 1-3: Summary Table (Continued) Operating Mode12L3S7S B-Mode XXX M-Mode (including B-Mode) XXX Color Flow XXX M Color Flow XXX Pulsed Doppler (including B-Mode) XXX Continuous Wave (CW) Doppler XX Table 1-4: Maximum Probe Temperature Probe Max Te m p Probe Max Te m p Probe Max Te m p Probe Max Te m p 3C 29.64 E8C 28.11 3S 25.59 3.5C 30.70 8L 37.95...
Page 24
Acoustic Output 1-18LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Acoustic Output Ta b l e K e y Table 1-5: Table Key FDAIECMeaning—IEC 60601-2-37 / FDA & NEMA UD2, UD3 aαAcoustic Attenuation Coefficient / Derating factor (usually 0.3 dB/cm-MHz) A aprtAarpt-12db Output Beam Area / Active aperture area C MINormalizing Coefficient D eqDeqEquivalent Aperture Diameter / (same) d -6d-6Pulse Beam Width / Beam diameter at –6 dB d eqdeqEquivalent Beam Diameter f...
Page 25
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-19 Direction 5122542-100 Rev. 2 pr.3pr αAttenuated Peak-Rarefactional Acoustic Pressure / (same) PRF prrPulse Repetition Rate / Pulse repetition frequency TI TIThermal Index / (same) TIB TIBBone Thermal Index / (same) TIC TICCranial-Bone Thermal Index / (same) TIS TISSoft-Tissue Thermal Index / (same) PD t dPulse Duration / (same) x -12,y-12X, Y -12 dB Output Beam Dimensions / (same) Z ZDistance from the Source to a...
Page 26
Acoustic Output 1-20LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Probe Name:3C Table 1-6: Transducer Mode 3C Operating Mode: B Mode Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.9 IECFDAUnits pra pr.3 (MPa) 1.5 PWo (mW) min of [Pα(zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.1 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.4 3.43.4 Dim of A aprtX(cm)3.63.6 Y(cm) 1.031.03 t dPD...
Page 27
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-21 Direction 5122542-100 Rev. 2 Table 1-7: Transducer Mode 3C Operating Mode: M Mode (Inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.850.7 0.3 IECFDAUnits pra pr.3 (MPa) 1.51 PWo (mW)31.0 21.0 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.43.1 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.3 ƒawffc(MHz) 3.4 3.43.4 3.4 Dim of A...
Page 28
Acoustic Output 1-22LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Table 1-8: Transducer Mode 3C Operating Mode: Color Flow Mode (Inc. M-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.84 3.22.9 4.6 IECFDAUnits pra pr. 3 (MPa) 1.55 PWo (mW)270.1179.7 270.1 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 3.963.3 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)0.4 ƒawffc(MHz) 2.5...
Page 29
Acoustic Output LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual 1-23 Direction 5122542-100 Rev. 2 Table 1-9: Transducer Mode: 3C Operating Mode: Pulsed Doppler (inc. B-Mode) Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.7 0.50.9 0.6 IECFDAUnits pra pr.3 (MPa) 1.0 PWo (mW)42.679.7 42.6 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 5.087.7 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm)1.0 ƒawffc(MHz) 3.3 3.33.5...
Page 30
Acoustic Output 1-24LOGIQ 3 Expert/LOGIQ 3 Pro/LOGIQ 3 Advanced Reference Manual Direction 5122542-100 Rev. 2 Probe Name:3.5C Table 1-10: Transducer Mode __3.5C____ Operating Mode: B Mode Index LabelMI TISTIB TIC scannon-scannon- scan Aaprt1 Global Maximum: Index Value 0.52.1 IECFDAUnits pra pr.3 (MPa) 1.54 PWo (mW)147.0 min of [P α( zs, Ita, α(zs)] [(W.3(Z1),ITA . 3(z1)]) zs z1(cm) zbp zbp(cm) zb zsp(cm) 1.0 z at max. I pi,α zsp(cm) deq(zb) deq(zsp) (cm) ƒawffc(MHz) 3.042.8 Dim of A aprtX(cm)1.8...