Philips Hd 7 User Manual
Here you can view all the pages of manual Philips Hd 7 User Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 41
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 41 Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Verification 1.2.840.10008.1.1 Explicit VR Little Endian Implicit VR Little Endian 1.2.840.10008.1.2.1 1.2.840.10008.1.2 SCU /SCP None US Image Storage 1.2.840.10008.5.1. 4.1.1.6.1 Implicit VR Little Endian Explicit VR Little Endian JPEG Lossy Baseline RLE Lossless 1.2.840.10008.1.2 1.2.840.10008.1.2.1...
Page 42
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 42 1. Open Association 2. C-ECHO Request 4. Close Association Verification SCUVerification SCP 3. C-ECHO Response Figure 8a SEQUENCING OF ACTIVITY – ISSUE VERIFY SCP: The system listens on the port configured on the “This System” configuration screen for Verification requests initiated by other remote devices. The calling device AE must already be configured as a remote device in HD7/HD7XE 3.0 or the association is...
Page 43
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 43 4.2.4.3.3 Proposed Presentation Contexts HD7/HD7XE 3.0 will propose Presentation Contexts as shown in the following table: Table 51.6 PROPOSED PRESENTATION CONTEX TS FOR ACTIVITY VERIFICATION Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Verification 1.2.840.10008.1.1 Implicit VR Little Endian Explicit VR Little Endian 1.2.840.10008.1.2...
Page 44
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 44 Entry screen. DICOM Structured Report Server Comprehensive Structured Report Storage SR Storage Commit SCP Storage Commitment DICOM BW Printer Basic Grayscale META Print All bw printers configure this entry. If the printer supports both BW and Color, then this must be configured to allow BW on that printer. DICOM Color Printer Basic Color META Print May be the same printer if color is also supported. 4.3...
Page 45
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 45 4.4.1.2 Remote AE Title/Presentation Address Mapping The AE Titles, IP Addresses and Port numbers of remote applications are manually confi gured using the DICOM setup screen. The remote system’s IP Address may be entered ma nually if known or the Host Name of the remote device may be entered and resolved by the DNS if the network includes this service. 4.4.1.2.2 Workflow Setup is used to set the AE Title, port-number and...
Page 46
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 46 patients, studies, series or instances (images, or Structured Reports). Throughout this section, the term “Media” refers to any of the media listed below which is in use. HD7/HD7XE 3.0 will support the use of most writable media including CD-R, CD-RW, DVD-R, DVD+R, DVD-RW and DVD+RW. DICOM structure will be t he same regardless of media used. 5.1.2 Functional Definition of AEs 5.1.2.1 Functional Definition of...
Page 47
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 47 The contents of the export job will be written together with a corresponding DICOMDIR to media. The user can cancel an export job in the job queue. Writing in multi-session format to CDs and DVDs is supported. Each export job is written as one session. 5.2.1.2.2 Activity – Import from Media The Media Application Entity acts as an FSR using the interchange option when requested to import SOP Instances from media to the...
Page 48
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 48 6 SUPPORT OF CHARACTER SETS All HD7/HD7XE 3.0 DICOM applications support the ISO_IR 100 Latin Alphabet No. 1 ISO-IR 87 Japanese Kanji (ideographic), Hi ragana (phonetic) and Katakana (phonetic ISO-IR 13 Japanese Katakana (phonetic) ISO-IR 159 Supplementary Kanji (ideographic) ISO-IR 144 Russian Cyrillic 6.1 SUPPORT FOR RUSSIAN AND JAPANESE MARKETS HD7/HD7XE 3.0 use Code-extension tech niques to encode Japanese stroke...
Page 49
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 49 Phonetic-last^Phonetic-first^Phonetic-middle^Phonetic-prefix^Phonetic-suffix” In the above string the five components are separated with the ‘^’ Ascii character and the three language variants are separated by the ‘=’ Ascii character. The only required co mponent is the Roman Last name. All other components are optional. Trailing ‘^’ and ‘=’ characters can be excluded. When this string is encoded in a DICOM image file or...
Page 50
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 50 8 ANNEXES 8.1 CREATED IOD INSTANCES Table 69 specifies the attributes of an Ultrasound Ima ge transmitted by the HD7/HD7XE 3.0 storage application. Table 70 specifies the attributes of a Comprehensive St ructured Reports transmitted by the HD7/HD7XE 3.0 storage application. The following tables use a number of abbreviations. The abbreviations used in the “Presence of …” column are: VNAP Value Not Always Present...