Philips Hd 7 User Manual
Here you can view all the pages of manual Philips Hd 7 User Manual. The Philips manuals for Ultrasound System are available online for free. You can easily download all the documents as PDF.
Page 21
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 21 4.2.1.4.1.3 SOP Specific Conformance for Storage Commitment Push Model SOP Class 4.2.1.4.1.3.1 Storage Commitment Notifications (N-EVENT-REPORT) Upon receipt of an N-EVENT-REPORT the timer associated with the Transaction UID will be canceled. Table 14 summarizes the behavior of Storage AE when receiving Event Types within the N-EVENT-REPORT. Table 15 summarizes the reasons for returning specif ic status codes in an...
Page 22
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 22 Table 22 DICOM IMPLEMENTATION CLASS AND VERSION FOR AE WORKFLOW Implementation Class UID 1.3.46.670589.14.3000.300 Implementation Version Name HD7_300 4.2.2.3 Association Initiation Policy 4.2.2.3.1 Activity – Worklist Update 4.2.2.3.1.1 Description and Sequencing of Activities Worklist queries for Modality (U S) or modalities may be initiated by the user or will occur at a preset interval set as one of the...
Page 23
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 23 4.2.2.3.1.2 Proposed Presentation Contexts HD7/HD7XE 3.0 will propose Presentation Contexts as shown in the following table: Table 23 PROPOSED PRESENTATION CONTEXTS FOR ACTIVITY WORKLIST UPDATE Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Modality Worklist Information Model – FIND 1.2.840.10008.5.1. 4.31 Implicit VR Little Endian Explicit VR Little...
Page 24
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 24 Table 25 summarizes the behavior of HD7/HD7XE 3.0 during communication failure. Table 25 MODALITY WORKLIST COMMUNICATION FAILURE BEHAVIOR Exception Behavior Timeout Same as Service Status “Refused” in the table above. Association aborted by the SCP or network layersSame as Service Status “Refused” in the table above. Table 26 describes the HD7/HD7XE 3.0 Worklist Matching Keys and requested attributes. Unexpected...
Page 25
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 25 Module Name Attribute Name Tag VR M RD IOD Visit Admission Current Patient Location10 (0038,0300) LO x x Patient Identification Patient’s Name Patient ID Other Patient IDs11 (0010,0010) (0010,0020) (0010,1000) PN LO LO x x x x x x x x Patient Demographic Patient’s Birth Date12 Patient’s Birth Time12 Patient’s Sex13 Patient’s Age14 Patient Size15 Ethnic...
Page 26
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 26 4 Scheduled Procedure Step Location sets the “Loc ation: field in the Patient Selection Screen. 5 Requested Procedure Description value is se t in the “Description” field of the Patient Selection screen and “Study Description” of the Patient ID screen. Manual entry to Study Description field is also sent in Image and MPPS messages. 6 May be used to set “Indication” field on Patient Selection screen. 1 st choice,...
Page 27
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 27 Workflow AEDepartmentScheduler 1 Open Association 2 N-CREATE (MPPS) - IN PROGRESS 3 Close Association 7 Open Association 8 N-SET (MPPS) - COMPLETED 9 Close Association start exam, acquire image end exam 4 Open Association 5 N-SET (MPPS) - DISCONTINUED 6 Close Associationcancel exam Figure 6 SEQUENCING OF ACTIVITY – ACQUIRE IMAGES A possible sequence of interactions between the Workflow AE and a Departmental Scheduler (e.g....
Page 28
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 28 Table 27 PROPOSED PRESENTATION CONTEXTS FOR REAL-WORLD ACTIVITY ACQUIRE IMAGES Presentation Context Table Abstract Syntax Transfer Syntax Role Ext. Neg. Name UID Name List UID List Modality Performed Procedure Step 1.2.840.10008.3.1. 2.3.3 Implicit VR Little Endian Explicit VR Little Endian 1.2.840.10008.1.2 1.2.840.10008.1.2.1 SCU None 4.2.2.3.2.3 SOP Specific Conformance for MPPS Table 28 summarizes...
Page 29
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 29 Table 29 MPPS COMMUNICATION FAILURE BEHAVIOR Exception Behavior Timeout Same as “Failure” above. Association aborted by the SCP or network layersSame as “Failure” above. Table 30 provides a description of the MPPS N-CREATE and N-SET request identifiers. Empty cells in the N- CREATE and N-SET columns indicate that the attribute is not sent. Table 30 MPPS N-CREATE / N-SET REQUEST IDENTIFIER Attribute Name Tag VR...
Page 30
HD7/HD7XE 3.0 DICOM Conformance Statement 000245000000008 Rev. A Page 30 Attribute Name Tag VRN-CREATE N-SET (0032,1064) from MWL No value sent for unscheduled study. unscheduled study. >Code Value (0008,0100)SHAs received from MWL No value sent for unscheduled study. As received from MWL No value sent for unscheduled study. >Coding Scheme Designator (0008,0102)SHAs received from MWL No value sent for unscheduled study. As received from MWL No value sent for unscheduled study....