Miele G 7881 Service Manual
Here you can view all the pages of manual Miele G 7881 Service Manual. The Miele manuals for Washer Disinfector are available online for free. You can easily download all the documents as PDF.
Page 101
2.5.3 General Information - Service Mode 1 Service mode 1 can be used to advance through the individual program steps of a selected program. Steps should be advanced rapidly with a maximum of 2 seconds of pause. If a step is not advanced, the selected program is continued normally from the step reached. The parameters affecting the fault-monitoring system also apply during the service mode. If, e.g., a heating step is accessed, then a fault is registered if sufficient water has not previously entered the...
Page 102
printer interface is present and activated. When stepping back and forth through the various programming levels the active programming level is onlydisplayed, e.g.,S 23, when the program selector switch is turned to the 12 o'clock setting. Settings in service mode 2 always apply for all cleaning programs, except when impermissible settings have been made in individual cleaning programs. Service mode 2 can be quit without saving any parameter changes that may have been made by interrupting the power...
Page 103
x Program BlockG 7882 Wash Programs1) CHEM 60°C-5 mincombi CHEM 60°C-5 minA (customisable) vario TD NR UNIVERSAL VR 1— — —10 L cold water 10 L cold water 1 min 1 min 60 s drainage 60 s drainage VR 2[10 L cold water] [10 L cold water] — — — [1 min] [1 min] [60 s drainage] [60 s drainage] HR 1[10.5 L cold water] [10.5 L cold water] — —10.5 L cold water [DOS 1 / 40°C] [DOS 1 / 40°C]DOS 1 / 40°C [60°C / 3 min] [55°C / 3 min]60°C / 3 min [60 s drainage] [—]60 s drainage HR 2[—] [—]—10.5 L cold water — DOS 1 /...
Page 104
Program Block G 7883 Wash Programs1) SPECIAL 93°C-10 min vario TD Customer Anorganica VR 1[—][10 L cold water]10 L cold water — [1 min]1 min [60 s drainage]60 s drainage VR 2[—]— —10 L cold water DOS 3 immediately 50°C / 1 min 60 s drainage HR 1[10.5 L cold water]10.5 L cold water — — [DOS 1 / 40°C]DOS 1 / 40°C [[93°C / 10 min]]55°C / 5 min [60 s drainage]60 s drainage HR 2[—]—10.5 L hot water 10.5 L hot water DOS 1 / 40°C DOS 1 / 40°C 80°C / 3 min 75°C / 3 min 60 s drainage 60 s drainage CH-DESIN[—]...
Page 105
Program BlockG 7883 Wash Programs1) Organica Standard Universal Intensive Plastic HR 210.5 L hot water[10.5 L hot water] [10.5 L hot water]10.5 L hot water 10.5 L hot water DOS 1 / 40°C[DOS 1 / 40°C] [DOS 1 / 40°C]DOS 1 / 40°C DOS 1 / 40°C 85°C / 3 min[75°C / 3 min] [80°C / 3 min]85°C / 3 min 55°C / 5 min 60 s drainage[60 s drainage] [60 s drainage]60 s drainage 60 s drainage CH-DESIN—[—] [—] [—] [—] SP 1— — — — — SP 29 L hot water 9 L hot water 9 L hot water 9 L hot water 9 L hot water DOS 3 immediately...
Page 106
2.7 Software Index Versions, Program Sequence Plans 2.7.1 List of Software Index Versions and Modifications –Software index P 00 (start of pilot series) –Software index P 01 (ID 488) –Software index P 02 (ID 524) –Software index P 03 (ID 560) Introduction of service mode function S 27. –Software index P 04 (ID 585) 1. Parameter adjustment of all "vario TD”programs. The washing parameters of programs "vario TD", "vario TD AN" and "B" (vario TD NR of G 7882/CD) increased from...
Page 107
1. If an aerosol-injection steam condenser is installed, the steam condenser (DHK) pause can also be activated without drying in order to, e.g., reduce the level of steam escaping with high-temperature programs. 2. Depending on model (with/without steam condenser (DHK) and with/ without drying unit (TA)), service mode 2 (level S 26 - Function 3) offers various options as follows: 1. S 26 - "-": DHK (aerosol-injection steam condenser) not installed 2. S 26 - 30: Without additional steam condenser...
Page 108
7. If a reduced water quantity is programmed in program blocks VR2, SP1 or SP2, then heating should not take place in these blocks because the wash pressure is not sufficiently constant and this can lead to the registration of a fault (F 3EorF 27). The interdependence has been adjusted. 8. With the electronic module EGPL 081 P 05 (ID 860), a diagnostic support file (DUD) for the Miele diagnostic utility (MDU) has been created for the first time. 9. In the programming mode, the range of dispensing options...
Page 109
this fault is registered when an interface is active, the fault codeF 13is always correctly generated. The modified software ensures that this communication problem is correctly identified. 6. A new diagnostic support file (DUD) for the Miele diagnostic utility (MDU) has been created for electronic module EGPL 081 P 06 (ID 1180). 7. With the introduction of the new control, the machine technical documentation is no longer located behind the cover plate under the door, but is now included with the machine...
Page 110
F 32: Faulty activation, dispenser relay on EGPL or EZL stuck (dispensing relay). F 33: Maximum permitted activation time exceeded (dispensing time). 6. Control modifications, corrections and extensions. The automatic ventilation times stored in the electronic module have been adjusted for the internal dispenser pump DOS 3. When an electronic module is reset, the dispenser control settings (E 02 - Functions 9 - 11) are no longer overwritten. This means that now when an old electronic module is replaced...