Hitachi F 2500 Manual
Here you can view all the pages of manual Hitachi F 2500 Manual. The Hitachi manuals for Spectrophotometer are available online for free. You can easily download all the documents as PDF.
Page 231
A - 13 And, eventually the molecule returns to the ground state while emitting fluorescent radiation. Also, if radiationless transition to the triplet state takes place, then phosphorescence is emitted during triplet-to-singlet transition (from the excited triplet state to the ground singlet state). Generally phosphorescence persists for 10 -4 sec or longer due to the selection rule imposed on the triplet-to-singlet transition. In contrast, fluorescence persists for a period of 10 -8 to...
Page 232
A - 14 E.2 Advantages of Fluorometry As contrasted with fluorometry, absorptiometry for a low-concentration sample is explained in the following : A sample having 99% transmittance to blank is taken as an example. In the absorbance measurement of such a substance, inaccuracies must always be taken into consideration. Here, the inaccuracy is assumed to be 0.1%. Since it has an effect on both the blank and sample, Percent transmittance of blank 100.0 ± 0.1% Percent transmittance of...
Page 233
A - 15 Amplified Fig. E-2 Comparison between Absorptiometry and Fluorometry In addition to high sensitivity, fluorometry is advantageous in that more information is attainable. A fluorescence spectrum is also available besides an excitation spectrum which corresponds to an absorption spectrum in absorptiometry. The two kinds of wavelengths can be selected as desired, and a fluorescence spectrum can be recorded using a properly selected excitation wavelength (or vice versa). Thus,...
Page 234
A - 16 Table E-1 compares information attainable in absorptiometry and that in fluorometry. Table E-1 Comparison of Information Attainable in Absorptiometry and Fluorometry Absorptiometry Fluorometry Absorption spectrum only (Corresponding to excitation spectrum in fluorometry) Excitation spectrum Fluorescence spectrum E.3 Notes of Fluorescence Analysis Measurement For most kinds of samples, an increase of 1°C in the temperature of sample causes the intensity of fluorescence to...
Page 235
A - 17 In fluorescence measurement, spectra having different natures from that of fluorescence may be observed. These are called Rayleigh scattering spectrum and Raman scattering spectra; the former appearing at the same wavelength position as the excitation spectrum, and the latter appearing at the longer-wavelength side near Rayleigh scattering. In a fluorescence spectrum, when the excitation wavelength is shifted, only the peak height is changed while the peak wavelength position remains...
Page 236
A - 18 Table E-3 Raman Peak Positions at Respective Excitation Wavelengths (Excitation wavelength) WaterEthanolCyclohexane Carbon Tetrachloride Chloroform 248 271 267 267 313 350 344 344 320 346 365 416 405 408 375 410 405 469 459 458 418 461 Excitation wavelength and Raman peak position (nm) 436 511 500 499 450 502 In measurement of a high-concentration sample, a variety of error factors may be involved. The most significant error factor...
Page 237
A - 19 If only the incident point of excitation beam is bright, it is necessary to dilute the sample properly for measurement. The second significant error factor consists in extinction due to concentration. This condition is caused by preventing activation through interaction of molecules. The third significant error factor consists in re-absorption of fluorescence. As shown in Figure E-6, this condition occurs due to overlapping between the shot-wavelength side of fluorescence spectrum...
Page 238
A - 20 Where the excitation and emission wavelengths are plotted near each other, care should be exercised not to mistake the Raman and Rayleigh scattering for the fluorescence spectrum as mentioned in E.3.3. Where the excitation and emission wavelengths are plotted apart from each other, care should be exercised not to mistake the second-order and third-order scattered radiations for the fluorescence spectrum. The second-order scattered radiation appears at a wavelength two times longer than...
Page 239
A - 21 Figure E-7 shows a measurement example of fluorescence spectrum. (1) Scattering of exciting radiation (2) Raman spectrum of solvent (3) Fluorescence of impurities, solvent, etc. (4) Fluorescence of sample (5) Second-order spectrum of exciting radiation Fig. E-7 Measurement Example of Fluorescence Spectrum As shown in Figure E-7, other peaks than a fluorescence peak of sample appear in fluorescence spectral measurement. With reference to this example, it is...
Page 240
A - 22 APPENDIX F MEASUREMENT OF INSTRUMENTAL RESPONSE (CORRECTED SPECTRA) Spectrum correction is performed to enable measuring a true spectrum by eliminating instrumental response such as wavelength characteristics of the monochromator or detectors. The measurement of instrumental response is needed to perform spectrum correction. “Instrumental Response” is the function to measure and save the instrumental response. F.1 Measurement of Instrumental Response on Excitation...