Page 241
A - 23 (2) Operating Procedures (a) Set PMT Voltage to 400 V and Data Mode to “Fluorescence”. Perform “Zero Adjust” to calibrate the zero point. (b) Set the triangular cell filled with Rhodamine B into the cell holder in the sample compartment. (See Fig. F-2). The triangular cell must be set in the direction shown in Fig. F-3. Fig. F-2 Setting of Rhodamine B Fig. F-3 Setting Direction of Triangular Cell (Top view)...
Page 242
A - 24 (c) Select “EX” of “Instrumental Response: Utility”. EM wavelength is set at 640 nm and the instrument starts EX wavelength scan (from 198.6 nm to 601.4 nm). At this stage, the instrument is automatically set in the following conditions. Slit (EX/EM) ........... 5.0/20.0 nm Scan speed ............ 60 nm/min Scan mode............. Fluorescence PMT Voltage .......... 400 V Response............... Auto (b) Upon completing the measurement, all of measuring conditions are...
Page 243
A - 25 F.2 Measurement of Instrumental Response on Emission Side This is the function to obtain the instrumental response on the emission side such as wavelength characteristics of the emission monochromator and detector (photomultiplier). The instrumental response on the emission side is determined by measuring a combination of instrumental response on both the excitation and emission sides by synchronous wavelength scan, and dividing it by the instrumental response on the excitation...
Page 244
A - 26 (c) Select “EM (200 to 600 nm)” of “Instrumental Response: Utility”. EM wavelength is set at 198.6 nm and the instrument starts synchronous wavelength scan (from 198.6 nm to 601.4 nm). At this stage, the instrument is automatically set in the following conditions. Slit (EX/EM) ......... 5.0/20.0 nm Scan speed .......... 60 nm/min Scan mode .......... Fluorescence PMT Voltage ........ 400 V Response ............ Auto (d) Upon completing the measurement, all of measuring...
Page 245
A - 27 F.3 Measurement of Instrumental Response on the Long Wavelength Range of Emission Side (only with Model F-4500) When a corrected spectrum in the long wavelength area of 600 nm is required on the emission side, the beam of a lamp for which the wavelength characteristics are known is allowed to be irradiated to the emission optics, and the instrumental response in the long wavelength range on the emission side is read for the correction. As a light source, a sub standard light source...
Page 246
A - 28 (e) Upon completing the measurement, all of measuring conditions are set to those set before. Now, instrumental response on the emission side has been measured and stored. Remove the sub standard light source.
Page 247
INDEX - 1 INDEX A ASCII text file ............................................................................................................. 1-3, 6-25 Activity ....................................................................................................................... 3-16, A-6 Analog output gain ................................................................................................... 1-41, 8-10 Auto pre-scan...
Page 248
INDEX - 2 F Factor input ......................................................................................................................... 4-7 Fonts tab ........................................................................................................................... 1-32 G General tab .................................................................................................... 2-3, 3-2, 4-2, 5-3 Gradient...
Page 249
INDEX - 3 N Number of wavelengths ...................................................................................................... 4-10 P Peak area ........................................................................................................................... 4-5 Peak detection ......................................................................................... 1-3, 2-12, 3-10, 5-10 Peak height...
Page 250
INDEX - 4 Spectral calculation ........................................................................................................... 6-16 Spectrophotometer menu ................................................................................................. 1-40 Spectrum toolbar ..................................................................................................... 1-20, 1-22 Stack...