Page 511
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 23 10-6-8-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less th an the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-20 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage...
Page 512
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 24 Section 10-6 - Electrical Safety Tests 10-6-9 Probe Leakage Current Test 10-6-9-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-6-9-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with the unit...
Page 513
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 25 10-6-9-3Meter Procedure Using Probe Adapter (cont’d) 6.) Set the meters “FUNCTION” switch to EXTERNAL position. 7.) Add the saline probe and the imaging area of the probe into the saline bath. 8.) Have unit power ON for the first par t; turn it OFF for the second half. 9.) Depress the ISO TEST rocker switch and record the highest current reading. 10.) Follow the test co nditions...
Page 514
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 26 Section 10-6 - Electrical Safety Tests 10-6-9-4 No Meter Probe Adapter Procedure Follow these steps to test each transducer for leakage current. 1.) Turn the LOGIQ™ P5 unit OFF. 2.) Plug the unit into the test meter, and the meter into the tested AC wall outlet. 3.) Plug the external probe into the meters (Dale 600) “EXTERNAL” connector. 4.) Set the meters “FUNCTION” switch to EXTERNAL position. 5.) Connect the probe for test...
Page 515
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 27 10-6-9-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-15 and Table 10- 16 . Record all data on the PM Inspection Certificate. . CAUTIONEquipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON. Be sure to turn the unit power OFF before switching them...
Page 516
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 28 Section 10-6 - Electrical Safety Tests Section 10-7 When Theres Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for cont inuity. Ensure the ground is not broken, frayed, or intermittent. Replac e any defective part. Tighten all grounds. Ensure star washers are under all ground studs. Inspect wiring for bad crimps, poor connections, or damage. Test the wall outlet; verify it is grounded...
Page 517
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL Section 10-6 - Electrical Safety Tests 10 - 29 ULTRASOUND INSPECTION CERTIFICATE * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other FUNCTIONAL CHECKSPHYSICAL INSPECTION AND CLEANING COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________...
Page 518
GE HEALTHCARE DIRECTION 5394141, REVISION 1LOGIQ™ P5 SERVICE MANUAL 10 - 30 Section 10-6 - Electrical Safety Tests ELECTRICAL SAFETY Final Check. All system covers are in plac e. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Electrical Test Performed Max Value Allowed Value Measured OK? Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage...
Page 520