Page 111
System Care and Maintenance LOGIQ E9 User Guide 3-33 Direction 5180302-100 Rev. 1 System Care and Maintenance Inspecting the System Examine the following on a monthly basis: • Connectors on cables for any mechanical defects. • Entire length of electrical and power cables for cuts or abrasions. • Equipment for loose or missing hardware. • Control panel and keyboard for defects. • Casters for proper locking operation. CAUTIONTo avoid electrical shock hazard, do not remove panels or covers from...
Page 112
After the Exam is Over 3-34 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Weekly Maintenance The system requires weekly care and maintenance to function safely and properly. Clean the following: • LCD • Operator control panel • B/W Printer • CD/DVD-R Multi Drive Failure to perform required maintenance may result in unnecessary service calls. System Cabinet To clean the system cabinet: 1. Moisten a soft, non-abrasive folded cloth with a mild, general purpose, non-abrasive soap and water...
Page 113
System Care and Maintenance LOGIQ E9 User Guide 3-35 Direction 5180302-100 Rev. 1 Operator Controls ONLY use the following cleaners on the LOGIQ E9 Operator Panel: • Palmolive Dishwashing Liquid (manufactured by Colgate- Palmolive) • Sani Wipes Alcohol-free (manufactured by Micorgen Inc.) • T-Spray II (manufactured by Pharmaceutical Innovations, Inc.) DO NOT USE: • Original T-Spray (T-Spray I) • SaniCloth Plus (red cap); Use Sani-Cloth HB (green cap) instead. • Any cleaning/disinfecting...
Page 114
After the Exam is Over 3-36 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Other Maintenance Manually Moving the Operator Panel Up/Down The LOGIQ E9 is equipped with an automatic Operator Panel up/down, left/right, forward/backward control. But if the system is powered down when you need to lower the Operator Panel, you can use the manual release handle located at the back of the unit to manually move the Operator Panel up/down. To move the Operator Panel up/down using the manual control, 1....
Page 115
System Care and Maintenance LOGIQ E9 User Guide 3-37 Direction 5180302-100 Rev. 1 Cleaning the Air Filters Clean the systems air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability. It is recommended the filters be cleaned quarterly (once every three months). LocatingThe LOGIQ E9 has two air filters which need to be cleaned. The top air filter is located on the back of the LOGIQ E9 below the power cord and the bottom air filter...
Page 116
After the Exam is Over 3-38 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Cleaning the Air Filters (continued) CleaningTo clean the top air filter, 1. Gently remove the filter cover. Pull the cover out and away from the LOGIQ E9. Set it down. 2. Remove the foam filter. 3. Clean the foam filter with soap and water. Let dry. 4. Replace the filter. 5. Replace the filter cover. Position it back into place on the back of the LOGIQ E9. The bottom filter is held in place with magnets. To clean...
Page 117
System Care and Maintenance LOGIQ E9 User Guide 3-39 Direction 5180302-100 Rev. 1 Supplies/Accessories Not all features or products described in this document may be available or cleared for sale in all markets. Contact the distributor, GE affiliate or sales representative for approved peripherals. For HCATs, contact your sales person. For 2million/5million number part numbers, these are service replacement part numbers that may be either new or refurbished. To order these, contact CARES in the US,...
Page 118
After the Exam is Over 3-40 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Console Table 3-4: Console Accessories AccessoryPart Number / HCAT Protective Cover H4908DC Volume Navigation Stand 5268740 / H4908NS Volume Navigation Bracket Starter Kit H4908NK LOGIQ E9 Technical Course H4908TC Swedish Keyboard H4908SY Norwegian Keyboard H4908NY Greek Keyboard H4908GY Russian Keyboard H4908RY Power Cord - North American 5148381 / H4000ZA Power Cord - Argentina 5148381-9 / H4000ZH Power Cord - Europe...
Page 119
System Care and Maintenance LOGIQ E9 User Guide 3-41 Direction 5180302-100 Rev. 1 Probes Table 3-5: Probes and Accessories Probe HCAT Global P/N Japan P/N Biopsy Guide Biopsy Guide HCAT V Nav Capable V Nav Bracket HCAT 3CRF Micro ConvexH40442LP 5196216 5214820Single Angle, Disposable with a Reusable BracketH40442LR No C1-5 ConvexH40452LE 5304539 5314506Multi Angle, Disposable with a Reusable BracketH40432LE Yes H40432LF M6C-D Matrix Array ConvexH40432LM 5182312 5189066Multi Angle, Disposable...
Page 120
After the Exam is Over 3-42 LOGIQ E9 User Guide Direction 5180302-100 Rev. 1 Options RIC5-9-D 4D Convex Vo l u m e IntracavitaryH48651MS KTZ157043 5189803Single Angle, Reusable H46721R No RNA5-9-D 4D Convex Vo l u m eH48651MY KTZ156994 5260751Single Angle, Disposable with a Reusable Bracket Single Angle, ReusableH46701AF H468651DGNo RSP6-16-D 4D Linear Vo l u m eH48651MR KTZ157046 5189162Single Angle, Disposable with a Reusable Bracket Single Angle, ReusableH47001AD H46721WNo Table 3-6:...