Page 151

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 140 12.3.3.2 Basic Color Image Box SOP Class 
The Basic Color Image Box IOD is an abstraction of the presentation of an image and 
image related data in the image area of a film.  The DIMSE services that are applicable to 
the IOD are shown in Table 12.3.3-2. 
TABLE 12.3.3-2 DIMSE SERVICE GROUP 
DIMSE Service Element Usage 
SCU Reference N-SET      M see 12.3.3.2.1 12.3.3.2.1 DIMSE Service Group (N-SET) 
The N-SET DIMSE...

Page 152

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 141 12.4.1 Film Session IOD Module  
TABLE 12.4.1-1 FILM SESSION IOD MODULES 
Module Name Reference Module Description SOP Common Module 12.5.2.1 Contains SOP Common information Basic Film Session Presentation 
Module 12.5.2.1 Contains Film Session presentation 
information Basic Film Session Relationship 
Module 12.5.2.2 References to related SOPs  
12.4.2 Basic Film Box IOD Module Table 
TABLE 12.4.2-1 BASIC FILM BOX...

Page 153

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 142 12.5.1 General Modules 
12.5.1.1 SOP Common Module 
This section defines the attributes that are required for proper functioning and 
identification of the associated SOP Instances. They do not specify any semantics about 
the Real-World Object represented by the IOD. 
TABLE 12.5.1-1 SOP COMMON MODULE ATTRIBUTES 
Attribute Name Tag Type Attribute Description SOP Class UID (0008,0016) 1 Varies with Module Instance and...

Page 154

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 143 TABLE 12.5.2-1 BASIC FILM SESSION PRESENTATION MODULE ATTRIBUTES  
Attribute Name Tag USAGE 
(SCU) Attribute Description Number of Copies (2000,0010) U Defined Terms used (user 
configurable): 
Default is 1. Max is 99. Print Priority (2000,0020) U Defined Terms used (user 
configurable): 
HIGH, MED, LOW. Default 
is HIGH. Medium Type (2000,0030) U  Defined Terms used (user 
configurable): 
      PAPER 
      BLUE FILM...

Page 155

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 144 TABLE 12.5.2-3 BASIC FILM BOX PRESENTATION MODULE ATTRIBUTES  
Attribute Name Tag USAGE 
(SCU) Attribute Description Image Display Format (2010,0010) M Enumerated values used (user 
configurable): 
      STANDARD\X,Y,  
      where X and Y can take 
       values from 1 to 5. 
Default is STANDARD\1,1. Annotation Display Format ID (2010,0030) U Not used. Film Orientation (2010,0040) U Defined Terms used (user...

Page 156

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 145 Trim (2010,0140) U Enumerated Values Used (user 
configurable): 
     YES 
      NO 
Default is NO. Configuration Information (2010,0150) U User configurable.   
No default.  
12.5.2.4 Basic Film Box Relationship Module 
This section defines the attributes that describe the common parameters, which apply for 
all images on a given sheet of film.  
TABLE 12.5.2-4 BASIC FILM BOX RELATIONSHIP MODULE ATTRIBUTES 
Attribute...

Page 157

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 146 TABLE 12.5.2-5 IMAGE BOX PIXEL PRESENTATION MODULE ATTRIBUTES 
Attribute Name Tag USAGE 
(SCU) Attribute Description Image Position (2020,0010) M Based on the image display 
format. Polarity (2020,0020) U Defined term, NORMAL Requested Image Size (2020,0030) U Not sent Basic Grayscale Image Sequence (2020,0110) M  >Samples Per Pixel (0028,0002) M Value = ‘1’  >Photometric Interpretation (0028,0004) M Defined Term...

Page 158

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 147 TABLE 12.5.2-6 PRINTER MODULE ATTRIBUTES 
Attribute Name Tag USAGE 
(SCU) Attribute Description Printer Status (2110,0010) U Used to check the status of the 
printer before a print operation is 
started.  
If the status is different from 0x0, 
Success the print operation is 
aborted, a message is displayed 
and the print files reside in the 
print buffer. Printer Status Info (2110,0020) U If return status is FAILURE...

Page 159

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 148 13. STUDY ROOT QUERY/RETRIEVE 
INFORMATION MODEL DEFINITION 
13.1 INTRODUCTION 
This section specifies the use of the DICOM Study Root Query/Retrieve Model used to 
organize data and against which a Query/Retrieve will be performed. The contents of this 
section are: 
13.2 - Information Model Description 
13.3 - Information Model Entity-Relationship Model 
13.4 - Information Model Keys 
13.2 STUDY ROOT INFORMATION...

Page 160

 LOGIQ7 
GE HEALTHCARE DICOM CONFORMANCE STATEMENT 
DIRECTION 2316173-100REV 7.03 
 149 ILLUSTRATION 13.3-1 
STUDY ROOT QUERY/RETRIEVE INFORMATION MODEL E/R DIAGRAM Study contains Image or Report contains  
Series 
0,n 1 1,n 1  
13.3.1 Entity Descriptions 
Please refer to DICOM Standard PS 3.4 (Service Class Specifications) for a description 
of each of the levels contained within the Study Root Query/Retrieve Information Model. 
13.3.2 LOGIQ 7 Mapping of DICOM entities 
TABLE 13.3-1  
MAPPING OF DICOM...